K163211 · Siemens Medical Solutions USA, Inc. · LNH · Jan 27, 2017 · Radiology
Device Facts
Record ID
K163211
Device Name
MAGNETOM Sempra with syngo MR E11S
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Jan 27, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The MAGNETOM Sempra is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis. The MAGNETOM Sempra may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-Safe biopsy needles.
Device Story
MAGNETOM Sempra is a 1.5T MRI system; inputs include magnetic resonance signals from patient anatomy; transforms signals into transverse, sagittal, coronal, oblique cross-sectional images, and spectroscopic data. Features Tim4G + Dot technology, Quiet Suite for noise reduction, CAIPIRINHA for breath-hold acceleration, and Advanced WARP for metallic implant distortion correction. Operated by healthcare professionals in clinical settings. Output displayed for physician interpretation to assist in diagnosis. Includes Eco-Power and Zero Helium Boil-off for operational efficiency. Hardware includes modified gradient amplifier and fixed patient table. Benefits include consistent image quality via Dot engines, reduced scan times, and improved soft tissue evaluation near implants.
Clinical Evidence
Clinical study of 34 individuals conducted per IEC 60601-2-33 (Edition 3.1: 2013) to determine peripheral nerve stimulation (PNS) threshold levels for gradient system output. Additional non-clinical bench testing performed for SNR, image uniformity, and heating.
Technological Characteristics
1.5T MRI system; Tim4G + Dot technology; modified gradient amplifier; fixed patient table. Software platform syngo MR E11S. Conforms to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, IEC 60601-1-6, and ISO 14971. Connectivity includes standard MR imaging interfaces.
Indications for Use
Indicated for patients requiring diagnostic MR imaging of the head, body, or extremities. Used for cross-sectional imaging, spectroscopic imaging, and interventional procedures with MR-compatible devices. Contraindications include patients with non-MR-safe implants or devices.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
MAGNETOM Amira with software version syngo MR E11N (K152283)
Reference Devices
Software syngo MR E11C for the MAGNETOM System Aera (K153343)
Software syngo MR D14 for the MAGNETOM ESSENZA (1.5T) (K130262)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 27, 2017
Siemens Medical Solutions USA, Inc. % Cordell L. Fields, Esq. Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 65 Valley Stream Pkwy. MALVERN PA 19355
Re: K163211
Trade/Device Name: MAGNETOM Sempra with syngo MR E11S Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH, LNI, MOS Dated: November 15, 2016 Received: November 16, 2016
Dear Mr. Fields:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163211
Device Name MAGNETOM Sempra with syngo MR E11S
#### Indications for Use (Describe)
The MAGNETOM Sempra is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.
The MAGNETOM Sempra may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-Safe biopsy needles.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/>Prescription Use (Part 21 CFR 201, Subpart D)</div> | <div style="display:flex; align-items:center;"><input type="checkbox"/>Over-The-Counter Use (21 CFR 201, Subpart C)</div> |
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
# I. General Information
| Establishment | Siemens Medical Solutions USA. Inc.<br>40 Liberty Boulevard<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Registration Number: 2240869 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | November 15, 2016 |
| Manufacturer | Siemens Shenzhen Magnetic Resonance Ltd.<br>Siemens MRI Center, Gaoxin C. Ave., 2nd<br>Hi-Tech Industrial Park<br>518057 Shenzhen<br>China<br>Registration Number: 3004754211 |
| Contact Person | Mr. Cordell L. Fields, Esq.<br>Regulatory Affairs Specialist<br>Siemens Healthcare<br>Siemens Medical Solutions USA, Inc.<br>Customer Solutions Group<br>40 Liberty Boulevard<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Phone: (610) 448-6469<br>Fax: (610) 640-4481 |
| Device Name: | MAGNETOM Sempra with syngo MR E11S |
| Trade name: | MAGNETOM Sempra |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD)<br>Emission Computed Tomography System |
MAGNETOM Sempra with syngo MR E11S
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a cluster of orange dots arranged in a pattern.
| Classification Panel: | Radiology |
|--------------------------|----------------------------------------|
| Regulation Number: | 21 CFR § 892.1200<br>21 CFR § 892.1000 |
| Device Class: | II |
| Primary Product Code: | LNH |
| Secondary Product Codes: | LNI, MOS |
# II. Safety and Effectiveness Information Supporting Substantial Equivalence
The MAGNETOM Sempra has the same intended use as the legally marketed predicate device, MAGNETOM Amira (K152283).
## Intended Use
The MAGNETOM Sempra is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.
The MAGNETOM Sempra may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-Safe biopsy needles.
## Device Description
The subject device, MAGNETOM Sempra (1.5T) with syngo MR E11S, is an MRI system that is substantially equivalent to the previously cleared predicate device MAGNETOM Amira with software version syngo MR E11N; K152283).
MAGNETOM Sempra is equipped with Tim4G + Dot technology and Siemens latest software platform syngo MR E11 which includes latest applications like Quiet Suite for quiet brain, spine and musculoskeletal exams, CAIPIRINHA for shorter breathhold times and Advanced WARP which enables correction of in- through-plane distortions for better evaluation of soft tissue around metallic implants.
MAGNETOM Sempra with syngo MR E11S
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
From the operational side, Brain, Spine and Large Joint Dot engines included in standard configuration of MAGNETOM Sempra deliver consistency in imaging of core body regions. The same software platform E11 would help users to shorten learning curve, plus the consistent result from Dot engines, MAGNETOM Sempra helps the user to get uniform quality. The Eco-Power technology is included and works by automatically switching off the cold head compressor during system standby or power off at night, and intelligently switching on/off of components between patients together with Zero Helium Boil-off technology to help customers to save operating cost.
MAGNETOM Sempra utilizes the similar design of hardware as the predicate device MAGNETOM Amira with syngo E11N (K152283), the main hardware changes are the modified Gradient Amplifier and introducing a new fixed patient table without vertical movement.
# Technological Characteristics
The subject device, MAGNETOM Sempra with syngo MR E11S and the predicate device, MAGNETOM Amira with syngo MR E11N, are substantially equivalent with reqard to acquiring MR images steps/features and with regard to the operational environment, programming language, operating system and performance.
The subject device, MAGNETOM Sempra with syngo MR E11S conforms to the standard for software medical devices (IEC 62304:2006) and IEC as well as NEMA standards.
## Nonclinical Tests
The following performance testing was conducted on the subject device:
- The coils were tested for SNR, image uniformity, and heating. ●
- . All other software features were verified and validated.
The results from each set of tests demonstrate that the device performs as intended and is thus substantially equivalent to the predicate devices to which it has been compared.
## Clinical Tests
A clinical study of 34 individuals was conducted in accordance with IEC 60601-2-33 (Edition 3.1: 2013) to determine the nerve stimulation thresholds used to limit the gradient system output. The observed parameters were the PNS (Peripheral Nerve Stimulation) THRESHOLD LEVEL which are required in IEC 60601-2-33 (Edition 3.1:2013)
#### Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards in a risk analysis beginning early in the design phase and continuing throughout the development of the product. These risks are controlled via measures realized in software development, SW testing and product labeling. To minimize risks, Siemens adheres to recognized and established industry practices and standards, such as the IEC 60601-1 series, to minimize electrical and mechanical risk. Furthermore, the operators are healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.
MAGNETOM Sempra with Software syngo MR E11S, conforms to the applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as recommended by the respective MR FDA Guidance Document. The standards conformed to are the following:
| Recognition<br>Number | Product Area | Title of Standard | Reference<br>Number and<br>date | Standards<br>Development<br>Organization |
|-----------------------|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|------------------------------------------|
| 19-4 | General II<br>(ES/EMC) | C1:2009/(R)2012 and<br>A2:2010/(R)2012 (Consolidated<br>Text) Medical electrical equipment<br>- Part 1: General requirements for<br>basic safety and essential<br>performance (IEC 60601-1:2005,<br>MOD) | ES60601-<br>1:2005/(R)2012<br>and A1:2012, | AAMI<br>ANSI |
| 19-1 | General II<br>(ES/EMC) | Medical electrical equipment - Part<br>1-2: General requirements for<br>basic safety and essential<br>performance - Collateral standard:<br>Electromagnetic compatibility -<br>Requirements and tests | 60601-1-2 Edition<br>3: 2007-03 | IEC |
| 12-271 | Radiology | Medical electrical equipment - Part<br>2-33: Particular requirements for<br>the basic safety and essential<br>performance of magnetic<br>resonance equipment for medical<br>diagnosis | 60601-2-33 Ed.<br>3.1:2013 | IEC |
| 5-40 | General I | Medical devices - Application of | 14971 Second | ISO |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a pattern.
| | (QS/RM) | risk management to medical<br>devices | Edition 2007-03-<br>01 | |
|-------|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|---------------------|
| 5-89 | General I<br>(QS/RM) | Medical electrical equipment - Part<br>1-6: General requirements for<br>basic safety and essential<br>performance - Collateral standard:<br>Usability | 60601-1-6 Edition<br>3.1 2013-10 | IEC |
| 13-32 | Software/I<br>nformatics | Medical device software -<br>Software life cycle processes | 62304: 2006 | AAMI<br>ANSI<br>IEC |
# Substantial Equivalence
The MAGNETOM Sempra with software syngo MR E11S is substantially equivalent to the following devices:
| Predicate Device | FDA Clearance<br>Number | FDA<br>Clearance<br>Date | Product<br>Code |
|-------------------------------------------------------|-------------------------|--------------------------|------------------|
| MAGNETOM Amira with software<br>version syngo MR E11N | K152283 | December<br>24, 2015 | LNH, LNI,<br>MOS |
| Reference Devices | FDA Clearance<br>Number | FDA<br>Clearance<br>Date | Product<br>Code |
|-------------------------------------------------------------|-------------------------|--------------------------|------------------|
| Software syngo MR E11C for the<br>MAGNETOM System Aera | K153343 | April 15,<br>2016 | LNH, LNI,<br>MOS |
| Software syngo MR D14 for the<br>MAGNETOM ESSENZA<br>(1.5T) | K130262 | March 1,<br>2013 | LNH |
## Conclusion as to Substantial Equivalence
MAGNETOM Sempra with Software syngo MR E11S has the same intended use as the predicate device MAGNETOM Amira (K152283, cleared on December 24,2015), with respect to the magnetic resonance features and functionalities.
MAGNETOM Sempra with Software syngo MR E11S will be used for acquiring MR images (transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra). The predicate device MAGNETOM Amira (K152283), is also capable of acquiring MR images (transverse, sagittal, coronal
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots arranged in a circular pattern.
and oblique cross sectional images, spectroscopic images and/or spectra). Though there are differences between the subject device and the predicate device, which include the new and modified software and hardware features, MAGNETOM Sempra with software syngo MR E11S has similar functionality as the predicate device, and does not introduce new issues of safety or effectiveness. Therefore, Siemens is of the opinion that MAGNETOM Sempra with Software syngo MR E11S does not raise new questions of safety or effectiveness and is substantially equivalent to the predicate device MAGNETOM Amira (K152283).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.