K163205 · Ge Healthcare Coils (Usa Instruments, Inc.) · MOS · Jan 12, 2017 · Radiology
Device Facts
Record ID
K163205
Device Name
48CH Head Coil
Applicant
Ge Healthcare Coils (Usa Instruments, Inc.)
Product Code
MOS · Radiology
Decision Date
Jan 12, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The 3.0T 48CH Head Coil is a receive-only RF coil designed for use with select 3.0T MRI systems manufactured by GE Healthcare. The coil is indicated for use for head imaging. The nucleus excited is hydrogen.
Device Story
48CH Head Coil is a receive-only phased-array RF coil for 3.0T GE MRI systems; captures hydrogen proton signals for head/brain imaging. Features 48 elements with integrated preamplifiers and MEMS switches for decoupling from system body coil. Used in clinical settings by MRI technicians/radiologists. Provides high signal-to-noise ratio, uniform coverage, and supports multiband imaging. Output is raw MR signal processed by the host MRI system into diagnostic images; assists clinicians in visualizing head anatomy for diagnostic decision-making. Optimized pads enhance patient comfort.
Clinical Evidence
No clinical data required. Substantial equivalence supported by bench testing, including AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-33, MS 6-2008, blocking network analysis, and heat testing. Biocompatibility demonstrated via ISO 10993. Sample clinical images provided.
Technological Characteristics
Receive-only phased-array RF coil; 48 elements; MEMS decoupling switches; integrated preamplifiers. Materials: flame retardant. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-33, MS 6-2008, ISO 10993. Connectivity: P-connectors for GE 3.0T MRI systems.
Indications for Use
Indicated for head imaging in patients undergoing MRI procedures using GE 3.0T MRI systems. No specific age or gender restrictions; no specific contraindications listed.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 13, 2017
GE Healthcare Coils (USA Instruments, Inc.) % Mary A. Mayka, Ph.D. Regulatory Affairs Manager 1515 Danner Drive AURORA OH 44202
Re: K163205
Trade/Device Name: 48CH Head Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: November 14, 2016 Received: November 15, 2016
Dear Dr. Mayka:
This letter corrects our substantially equivalent letter of January 12, 2016
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K163205
Device Name 48CH Head Coil
Indications for Use (Describe)
The 3.0T 48CH Head Coil is a receive-only RF coil designed for use with select 3.0T MRI systems manufactured by GE Healthcare. The coil is indicated for use for head imaging. The nucleus excited is hydrogen.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle has a wavy or swirling pattern around the letters, giving it a dynamic and recognizable appearance.
GE Healthcare 510(k) Premarket Notification Submission
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | January 5, 2017 |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare Coils (USA Instruments, Inc.)<br>1515 Danner Drive<br>Aurora, OH 44202<br>USA |
| Primary Contact Person: | Mary A. Mayka, Ph.D.<br>Regulatory Affairs Manager<br>GE Healthcare<br>Phone: 262-527-3148<br>Fax: 262-364-2785 |
| Secondary Contact Person: | Glen Sabin<br>Regulatory Affairs Director<br>GE Healthcare<br>Phone: 262-521-6848<br>Fax: 262-364-2785 |
| Device Trade Name: | 48CH Head Coil |
| Common/Usual Name: | Coil, Magnetic Resonance, Specialty |
| Classification Names: | Magnetic Resonance Diagnostic Device per 21 CFR<br>892.1000 |
| Product Code: | MOS |
| Predicate Device(s): | TDI Head Neck Unit (K143345) |
| Device Description: | The 48CH Head Coil is a phased-array receive-only coil<br>designed to provide optimum signal-to noise, uniform<br>coverage and high acceleration including multiband<br>imaging of the head and brain. It is a 48-element coil<br>tuned to image proton nuclei and designed for use with<br>GE 3.0T MRI Systems. Each coil element has an<br>integrated preamplifier to improve image quality and a<br>custom MEMS (Micro Electro Mechanical System)<br>switch to decouple the receive coil from the MR System<br>body coil. The 48CH Head Coil has an anterior coil and a<br>posterior coil with P-connectors. The coil has optimized<br>pads to maximize patient comfort and image uniformity. |
| Intended Use: | The 3.0T 48CH Head Coil is a receive-only RF coil<br>designed for use with select 3.0T MRI systems<br>manufactured by GE Healthcare. The coil is indicated for<br>use for head imaging. The nucleus excited is hydrogen. |
| Comparison of<br>Intended Use: | Both the 48CH Head Coil and the predicate device are<br>classified as coils for magnetic resonance imaging devices<br>and are intended for diagnostic use. Both indications for<br>use statements are functional in nature, and do not list<br>specific diseases or conditions. The 48CH Head Coil and<br>the predicate device are indicated for the same patient<br>population, and for the same clinical setting. The<br>predicate device, the TDI Head Neck Unit is one coil as<br>part of the TDI Coil Suite, and covers head, neck,<br>brachial-plexus and vessel anatomies. The 48CH Head<br>Coil also references head anatomy. Therefore, GE<br>Healthcare believes that the 48CH Head Coil has the same<br>intended use as the predicate device in accordance with<br>the FDA's guidance document "The 510(k) Program:<br>Evaluating Substantial Equivalence in Premarket<br>Notifications [510(k)]", dated 28 July 2014. |
| Comparison of<br>Technological<br>Characteristics: | Overall, the 48CH Head Coil employs the same<br>fundamental scientific technology as its predicate device.<br>Coil Design: The most notable technological difference<br>between the 48CH Head Coil and the predicate device is<br>that the 48CH Head Coil contains 48 elements and utilizes<br>MEMS technology for decoupling as compared to the<br>predicate's 21 elements and hybrid decoupling.<br>Operating Principles: The 48CH Head Coil operates on the<br>same principles as the predicate device.<br>Materials: The 48CH Head Coil and the predicate device<br>both use flame retardant materials.<br>Safety and Performance Testing: Both the 48CH Head<br>Coil and the predicate device complies with the same<br>safety and performance testing (see Determination of<br>Substantial Equivalence, below).<br>These technological differences do not raise any different<br>questions of safety and effectiveness. |
| Determination of | Summary of Non-Clinical Tests: |
| Substantial Equivalence: | The predicate and modified devices have been subject to<br>similar risk management testing to demonstrate<br>substantial equivalence of safety and performance.<br>Testing included: |
| | AAMI/ANSI ES60601-1 IEC 60601-1-2 IEC 60601-2-33 MS 6-2008 Maximum B1 peak Blocking Network Analysis Heat Testing |
| | Additionally, both predicate and modified devices have a<br>successful biocompatibility track record, as demonstrated<br>by ISO 10993 testing and by their history of use in<br>previously cleared devices. |
| | The following quality assurance measures were applied to<br>the development of the device: Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation) |
| | Summary of Clinical Tests:<br>The subject of this premarket submission, the 48CH Head<br>Coil, did not require clinical studies to support substantial<br>equivalence. Sample clinical images have been included<br>in this submission. |
| | Substantial Equivalence Conclusion: |
| | The indications for use of the proposed device are<br>comparable to the claimed predicate device. The 48CH<br>Head Coil employs equivalent technology to the claimed<br>predicate device. Additionally, the results from the above<br>non-clinical tests demonstrate that the device performs as<br>intended. Thus, the 48CH Head Coil is substantially<br>equivalent to the predicate device to which it has been<br>compared. |
| Conclusion: | GE Healthcare considers the 48CH Head Coil to be as<br>safe, as effective, and performance is substantially<br>equivalent to the predicate devices. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font inside. There are also some white swirls around the letters. The logo is simple and recognizable.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font inside. There are also some curved lines around the letters, which give the logo a dynamic and modern look. The blue color is vibrant and eye-catching.
# GE Healthcare 510(k) Premarket Notification Submission
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by swirling, wave-like elements, also in blue, which give the impression of motion or energy. The overall design is simple, clean, and recognizable, representing the brand identity of General Electric.
# GE Healthcare
510(k) Premarket Notification Submission
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.