← Product Code [LOX](/productcode/LOX) · K163200

# Maverick XL Percutaneous Transluminal Coronary Angioplasty Monorail Dilatation Cathete (K163200)

_Boston Scientific Corporation · LOX · Dec 14, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K163200

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [LOX](/productcode/LOX.md)
- **Decision Date:** Dec 14, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.5100
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Maverick XL balloon catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft for the purpose of improving myocardial perfusion. The Maverick XL balloon catheter is also indicated for the post delivery expansion of balloon expandable stents.

## Device Story

The Maverick XL is a Monorail percutaneous transluminal coronary angioplasty (PTCA) dilatation catheter featuring a distal balloon. Used by physicians in clinical settings, the device is inserted into the coronary vasculature to dilate stenotic lesions or expand stents. The catheter utilizes a dual-lumen, coaxial design; the balloon is inflated to a specific diameter and length at controlled pressures to restore blood flow. The device includes a modified corewire compared to the predicate. It is a mechanical tool; no electronic or software components are involved.

## Clinical Evidence

Bench testing and first article testing were performed. No clinical data was required or provided.

## Technological Characteristics

Monorail PTCA dilatation catheter; dual-lumen coaxial distal segment; balloon at distal tip; modified corewire design. Sterilization method is consistent with the predicate device.

## Regulatory Identification

Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.

## Predicate Devices

- Maverick XL Percutaneous Transluminal Coronary Angioplasty Monorail Dilatation Catheter ([P860019](/device/P860019.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 14, 2016

Boston Scientific Corporation Ka Zoua Xiong Regulatory Affairs Specialist One Scimed Place Maple Grove, Minnesota 55311-1566

Re: K163200

Trade/Device Name: Maverick XL Percutaneous Transluminal Coronary Angioplasty Monorail Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II (Special Controls) Product Code: LOX Dated: November 14, 2016 Received: November 15, 2016

Dear Ka Zoua Xiong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Brian D. Pullin -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K163200

Device Name

Maverick™ XL Percutaneous Transluminal Coronary Angioplasty Monorail™ Dilatation Catheter

Indications for Use (Describe)

The Maverick XL balloon catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft for the purpose of improving myocardial perfusion. The Maverick XL balloon catheter is also indicated for the post delivery expansion of balloon expandable stents.

Type of Use (Select one or both, as applicable)

| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|--------------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>                    |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

# Per 21 CFR §807.92

| Common or Usual<br>Name                            | PTCA Dilatation Catheter                                                                                                                                          |                                                                                                     |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Trade Name(s)                                      | Maverick™ XL Percutaneous Transluminal Coronary Angioplasty<br>Monorail™ Dilatation Catheter                                                                      |                                                                                                     |
| Product Code                                       | LOX - Catheters, Transluminal Coronary Angioplasty,<br>Percutaneous                                                                                               |                                                                                                     |
| Classification of<br>Device                        | Class II (special controls) - 21 CFR 870.5100                                                                                                                     |                                                                                                     |
| Submitter's Name<br>and Address                    | Boston Scientific Corporation<br>One Scimed Place<br>Maple Grove, MN 55311-1566                                                                                   |                                                                                                     |
| Contact Name and<br>Information                    | Ka Zoua Xiong<br>Regulatory Affairs Specialist<br>Phone: 763-494-2970<br>Fax: 763-494-2222<br>Email: Kazoua.Xiong@bsci.com                                        |                                                                                                     |
| Date Prepared                                      | 11 November 2016                                                                                                                                                  |                                                                                                     |
| Section 514 of the<br>Act Performance<br>Standards | The special control for this device is FDA's "Class II Special<br>Controls Guidance Document: Percutaneous Transluminal<br>Coronary Angioplasty (PTCA) Catheters" |                                                                                                     |
| Establishment<br>Registration<br>Numbers           | Owner /Operator:                                                                                                                                                  | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, MA 01752<br>ERN: 9912058 |
|                                                    | Manufacturing<br>Facility:                                                                                                                                        | Boston Scientific Corporation<br>Two Scimed Place<br>Maple Grove, MN 55311<br>ERN: 2134265          |
|                                                    | Sterilization<br>Facilities:                                                                                                                                      | BSC Coventry<br>8 Industrial Drive<br>Coventry, RI 02816<br>USA                                     |
| Predicate Devices                                  | P860019 - Maverick™ XL Percutaneous Transluminal Coronary<br>Angioplasty Monorail™ Dilatation Catheter, date reclassified                                         | 08 October 2010                                                                                     |

{4}------------------------------------------------

| Device Description                                            | The Maverick XL Percutaneous Transluminal Coronary Angioplasty<br>PTCA (Maverick XL) dilatation catheter is a Monorail catheter with a<br>balloon near the distal tip. The distal catheter segment is dual<br>lumen and coaxial, characteristic of all monorail style catheters. The<br>balloon is located at the distal end of the catheter and is designed<br>to provide an inflatable segment of known diameter and length at<br>recommended pressures. |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/<br>Indications for Use                          | The Maverick XL balloon catheter is indicated for balloon dilatation<br>of the stenotic portion of a coronary artery or bypass graft for the<br>purpose of improving myocardial perfusion. The Maverick XL<br>balloon catheter is also indicated for the post delivery expansion of<br>balloon expandable stents.                                                                                                                                          |
| Comparison of<br>Required<br>Technological<br>Characteristics | The proposed Maverick XL is substantially equivalent to the existing<br>Maverick XL approved under premarket approval P860019, which<br>have since been down-classified by FDA on 08Oct2010. Maverick<br>XL has the same intended use, scientific technology, design (with<br>the exception of the corewire design), materials, sterilization<br>method, and packaging materials as the applicable predicate<br>device.                                    |
| Summary of Non-<br>Clinical Test<br>Summary                   | Bench testing and first article testing were performed to support a<br>determination of substantial equivalence. The results of these tests<br>provide reasonable assurance that the proposed device with the<br>modified corewire has been designed and tested to assure<br>conformance to the requirements for its intended use. No new<br>safety or performance issues were raised during the device testing.                                           |
| Conclusion                                                    | Based on the indications for use, technological characteristics, and<br>safety and performance testing, the proposed Maverick XL with the<br>modified corewire has been shown to be appropriate for its<br>intended use and is considered to be substantially equivalent to its<br>predicate (P860019).                                                                                                                                                    |

---

**Source:** [https://fda-staging.innolitics.com/device/K163200](https://fda-staging.innolitics.com/device/K163200)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
