ACTION II Laser System

K163196 · Lutronic Corporation · GEX · May 24, 2017 · General, Plastic Surgery

Device Facts

Record IDK163196
Device NameACTION II Laser System
ApplicantLutronic Corporation
Product CodeGEX · General, Plastic Surgery
Decision DateMay 24, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ACTION II Laser System is indicated for: coagulation, vaporization, ablation or cutting of soft tissue (skin) in Dermatology, Plastic Surgery, Oral Surgery, ENT, Gynecology, General Surgery, Podiatry and Ophthalmology (skin around the eyes). The ACTION II Laser Module Fractional Handpiece is indicated for: use in soft tissue (skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes; organs, and glands) such as, but not limited to: Dermatology and Plastic Surgery: skin resurfacing treatment of wrinkles; epidermal nevi; telangiectasia; spider veins; actinic chelitis; keloids; verrucae; skin tags; anal tags; keratoses; scar revision (including acne scars).

Device Story

ACTION II Laser System is a pulsed Er:YAG laser (2940 nm) for soft tissue surgery. Device consists of laser console, articulated arm, and handpieces (including fractional module). User operates via footswitch and LCD touch screen to control parameters; aiming beam assists targeting. Used in clinical settings (dermatology, plastic surgery, etc.) by trained healthcare professionals. Laser energy causes coagulation, vaporization, or cutting of tissue. Fractional handpiece creates micro-ablative zones for skin resurfacing and scar revision. Benefits include precise tissue ablation with controlled thermal effects. No clinical data provided; substantial equivalence based on technological comparison to predicate Er:YAG laser systems.

Clinical Evidence

No clinical data. Bench testing only, including compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-22 standards.

Technological Characteristics

Er:YAG laser source; 2940 nm wavelength; pulsed beam. Articulated arm delivery with fractional handpiece. Integrated LCD touch screen interface. Max power 22W; max energy 2.2 J; pulse width 200-300 µs. Compliance with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 60601-2-22 (laser equipment).

Indications for Use

Indicated for coagulation, vaporization, ablation, or cutting of soft tissue (skin) in dermatology, plastic surgery, oral surgery, ENT, gynecology, general surgery, podiatry, and ophthalmology (skin around eyes). Fractional handpiece indicated for soft tissue (skin, muscle, cartilage, organs, glands) for resurfacing, scar revision, and treatment of various dermatological lesions (e.g., wrinkles, nevi, telangiectasia, keloids, verrucae, keratoses).

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a family of three figures, representing health and human services. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 24, 2017 Lutronic Corporation Jhung Vojir VP. Ouality And Regulatory Affairs 6 Neshaminy Interplex, Suite 100 Trevose, Pennsylvania 19053 Re: K163196 Trade/Device Name: ACTION II Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX. ONG Dated: May 12, 2016 Received: May 15, 2016 Dear Jhung Vojir: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K163196 Device Name ACTION II Laser System #### Indications for Use (Describe) The ACTION II Laser System is indicated for: coagulation, vaporization or cutting of soft tissue (skin) in Dermatology, Plastic Surgery, Oral Surgery, ENT, Gynecology, General Surgery, Podiatry and Ophthalmology (skin around the eyes). The ACTION II Laser Module Fractional Handpiece is indicated for: use in soft tissue (skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage membrane, lymph vessels and nodes; organs, and glands) such as, but not limited to: Dermatology and Plastic Surgery: skin resurfacing treatment of wrinkles; epidermal nevi; telangiectasia; spider veins; actinic chelitis; keloids; verrucae; skin tags; keratoses; scar revision (including acne scars). | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------|-----------------------------------------------| | <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary Lutronic ACTION II Laser System This 510(k) Summary is being submitted in accordance with 21 CFR § 807.92. | 1. Submitter's Information | | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant: | Lutronic Corporation | | | 219 Sowon-ro<br>Haengsin-dong, Deogyang-gu<br>Goyang-si, Gyeonggi-do 410-220<br>Republic of Korea<br>Tel: (82) 31-908-3440<br>Fax: (82) 31-907-3440 | | Contact Person: | Jhung Won Vojir, Ph.D. | | | VP of Quality and Regulatory Affairs | | | Lutronic Corporation<br>Six Neshaminy Interplex, Suite 100<br>Trevose, PA 19053<br>Telephone: 215-205-2219<br>Fax: 609-488-6958<br>Email: jvojir@lutronic.com | | Date Prepared: | May 22, 2017 | | 2. Device Information | | | Trade Name: | ACTION II Laser System | | Common Name: | Medical Laser System | | Classification Name: | Powered Laser Surgical Instrument | | Product Code: | GEX, ONG | | Panel: | General and Plastic Surgery | # 3. Predicate Devices | 510(K) Number | K153229 | K150140 | |---------------|-------------|---------------------------------------| | Company Name | Bios S.R.L. | Asclepion Laser<br>Technologies, GmbH | {4}------------------------------------------------ | Device Name | Superbium | MCL 31 Dermablate | |-------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | | Regulation Name | Laser surgical instrument for use in general and plastic surgery and in dermatology | Laser surgical instrument for use in general and plastic surgery and in dermatology | | Regulatory Class | II | II | | Product Code | GEX, ONG, ONF | GEX | ## 4. Device Description The ACTION II Laser System produces a pulsed beam of coherent near-infrared (2940 nm) light upon activation by a footswitch. The beam is then directed to the treatment zone by means of an articulated arm coupled to a handpiece. An integrated LCD touch screen gives the user control over the necessary laser system parameters. An aiming beam is coincident with the 2940 nm main laser beam. ## 5. Indications for Use The ACTION II Laser System is indicated for: coagulation, vaporization, ablation or cutting of soft tissue (skin) in Dermatology, Plastic Surgery, Oral Surgery, ENT, Gynecology, General Surgery, Podiatry and Ophthalmology (skin around the eyes). The ACTION II Laser Module Fractional Handpiece is indicated for: use in soft tissue (skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes; organs, and glands) such as, but not limited to: Dermatology and Plastic Surgery: skin resurfacing treatment of wrinkles; epidermal nevi; telangiectasia; spider veins; actinic chelitis; keloids; verrucae; skin tags; anal tags; keratoses; scar revision (including acne scars). ## 6. Substantial Equivalence The ACTION II Laser System is substantially equivalent to the legally marketed Superbium and MCL 31 Dermablate. | Manufacturer | Lutronic Corporation | Bios S.R.L. | Asclepion Laser<br>Technologies, GmbH | |-----------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------| | Device Name | ACTION II Laser System | Superbium<br>(Primary Predicate) | MCL 31 Dermablate | | 510(k) # | K163196 | K153229 | K150140 | | Indications for Use | The ACTION II Laser<br>System is indicated for:<br>coagulation, vaporization,<br>ablation or cutting of soft | The 2940 nm Er:YAG Laser<br>Module handpiece is indicated<br>for: Coagulation, vaporization,<br>ablation or cutting of soft tissue | The MCL 31 Dermablate<br>laser system is intended for<br>coagulation, vaporization,<br>ablation or cutting of soft | | | tissue (skin) in Dermatology,<br>Plastic Surgery, Oral<br>Surgery, ENT, Gynecology, | (skin) in Dermatology, Plastic<br>Surgery, Oral Surgery, ENT,<br>Gynecology, General Surgery, | tissue (skin) in<br>Dermatology, Plastic<br>Surgery, Oral Surgery, | | | General Surgery, Podiatry | Podiatry and Ophthalmology | ENT, Gynecology, General | | | and Ophthalmology (skin | (skin around the eyes). | Surgery, Podiatry and | | | around the eyes). | Laser skin resurfacing | Ophthalmology (skin | | | | procedures for the treatment of: | around the eyes). | | | The ACTION II Laser | • Acne scars | | | | Module Fractional | • Wrinkles | | | | Handpieces are indicated for: | | | | | use in soft tissue (skin, | The 2940 nm Er:YAG | | | | cutaneous tissue,<br>subcutaneous tissue, striated | Fractional tips are indicated<br>for: use in soft tissue (skin, | | | | and smooth tissue, muscle, | cutaneous tissue, subcutaneous | | | | cartilage meniscus, mucous | tissue, striated and smooth | | | | membrane, lymph vessels | tissue, muscle, cartilage | | | | and nodes; organs, and | meniscus, mucous membrane, | | | | glands) such as, but not | lymph vessels and nodes; | | | | limited to: Dermatology and | organs, and glands) such as, but | | | | Plastic Surgery: skin | not limited to: Dermatology | | | | resurfacing treatment of | and Plastic Surgery: skin | | | | wrinkles; epidermal nevi; | resurfacing treatment of | | | | telangiectasia; spider veins; | wrinkles; epidermal nevi; | | | | actinic chelitis; keloids; | telangiectasia; spider veins;<br>actinic chelitis; keloids; | | | | verrucae; skin tags; anal<br>tags; keratoses; scar revision | verrucae; skin tags; anal tags; | | | | (including acne scars). | keratoses; scar revision | | | | | (including acne scars). | | | Classification | GEX 21 CFR 878.4810, | GEX 21 CFR 878.4810, ONF | GEX 21 CFR 878.4810 | | | ONF 21 CFR 878.4810 | 21 CFR 878.4810 | | | Laser Source | Er: YAG | Er: YAG | Er: YAG | | Wavelength | 2940 nm | 2940 nm | 2940 nm | | Weight of Laser | 86.5 kg | 35 kg | 75 kg | | Power max | 22W | 12 W | 20W | | Energy max | 2.2 J | 2.5 J | 2.5 J | | Pulse Width | 200 – 300 µs | 200-2200 µs | 100-1000 µs | | Spot Size | 1-14 mm | 1 - 9 mm | 1-12 mm | | Frequency max | 20 Hz | 5 Hz | 20 Hz | | Fluence max | 280 J/cm² (spot 1 mm) | 316 J/cm² (spot 1 mm) | 318 J/cm² (spot 1 mm) | | | | | | | Module Fractional Handpiece | | | | | Beam Diameter (dots) | 9x9 | 7x7, 9x9 | | | Fluence | 13 mJ/dot | 13 mJ/dot | | | Number of Dots | 81 dots/cm² | 81 dots/cm² (9x9) | | | Max Frequency | 5 Hz | 5 Hz | | | Max Power | 12 W | 12 W | | {5}------------------------------------------------ {6}------------------------------------------------ #### 7. Non-Clinical Performance Data The ACTION II Laser System complies with the following standards: IEC 60601-1 2005 + 1 (2006) + CORR 2 (2007) + AM1 (2012) or IEC 60601-1 2012. Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance. IEC 60601-1-2 (Third Edition, 2007), Medical Electrical Equipment-Part 1-2: General Requirements for Basic Safety and Essential Performance Collateral Standard: Electromagnetic Compatibility. IEC 60601-2-22 Third Edition 2007-05 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment. #### 8. Clinical Performance Data None. #### 9. Conclusion The ACTION II Laser System and the legally marketed Superbium cleared under 510(k) number K153229 have the same intended use and Indications for Use statement. Any technological differences between the ACTION II Laser System and its predicate devices are considered minor and do not raise any different types of questions regarding safety or effectiveness when used according to its intended use. The ACTION II Laser System is substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...