K163163 · Natus Medical Incorporated Dba Excel-Tech Ltd. (Xltek) · GWQ · Dec 14, 2016 · Neurology
Device Facts
Record ID
K163163
Device Name
XLTEK EMU40EX EEG Headbox
Applicant
Natus Medical Incorporated Dba Excel-Tech Ltd. (Xltek)
Product Code
GWQ · Neurology
Decision Date
Dec 14, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The XLTEK EMU40EX EEG Headbox is an electroencephalograph that works in conjunction with XLTEK NeuroWorks software. The XLTEK EMU40EX EEG Headbox is used to acquire, digitize, store and transmit physiological signals (such as EEG, pulse and oximetry signals) for EEG studies in research and clinical environments. The XLTEK EMU40EX EEG Headbox requires competent user input, and its output must be reviewed and interpreted by trained medical professionals who will exercise professional judgement in using this information.
Device Story
The XLTEK EMU40EX EEG Headbox is an electroencephalograph system used to acquire, digitize, store, and transmit physiological signals, including EEG, pulse, and oximetry. The system comprises a patient electrode set, the headbox, and a Natus base unit. It connects to a host computer running Natus NeuroWorks software for data review and analysis. The headbox connects to the base unit via cable or secure Bluetooth. It features a touchscreen LCD display for control. The device is operated by trained medical professionals in clinical or research settings. Output is reviewed by clinicians to support diagnostic decision-making for EEG studies.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was verified through bench testing for signal quality, functionality, and user interface, alongside compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-26, IEC 62366, and ISO 80601-2-61 standards.
Technological Characteristics
Electroencephalograph with 40 referential channels and 4 isolated DC channels. Features touch-proof electrode connections, oximeter/photic stimulator ports, and patient event button. Powered by two AAA rechargeable batteries or external battery pack. Connectivity via cable or Bluetooth. Interface includes touchscreen LCD. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-2-26 (EEG safety), and ISO 80601-2-61 (pulse oximetry).
Indications for Use
Indicated for use in research and clinical environments to acquire, digitize, store, and transmit physiological signals, including EEG, pulse, and oximetry, for EEG studies. Requires operation by trained medical professionals.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The image is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 14, 2016
Natus Medical Incorporated DBA Excel-tech Ltd. (XLTEK) Mr. Sanjay Mehta Senior Manager, Quality Assurance & Regulatory Affairs 2568 Bristol Circle Oakville, Ontario, L6H 5S1 Canada
Re: K163163
Trade/Device Name: XLTEK EMU40EX EEG Headbox Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWQ Dated: November 3, 2016 Received: November 10, 2016
Dear Mr. Mehta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K163163
Device Name XLTEK EMU40EX EEG Headbox
#### Indications for Use (Describe)
The XLTEK EMU40EX EEG Headbox is an electroencephalograph that works in conjunction with XLTEK NeuroWorks software.
The XLTEK EMU40EX EEG Headbox is used to acquire, digitize, store and transmit physiological signals (such as EEG, pulse and oximetry signals) for EEG studies in research and clinical environments.
The XLTEK EMU40EX EEG Headbox requires competent user input, and its output must be reviewed and interpreted by trained medical professionals who will exercise professional judgment in using this information.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submission Date: | 03 November 2016 | | |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|------------------------------------|
| Submitter: | Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK)<br>2568 Bristol Circle<br>Oakville, Ontario, L6H 5S1<br>Canada | | |
| Submitter and<br>Application<br>Correspondent | Mr. Sanjay Mehta<br>Phone: +1 (905) 287-5055<br>Fax: +1 (905) 829-5304<br>Email: sanjay.mehta@natus.com | | |
| Manufacturing Site: | Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK)<br>2568 Bristol Circle<br>Oakville, Ontario, L6H 5S1<br>Canada | | |
| Trade Name: | XLTEK EMU40EX EEG Headbox | | |
| Common and<br>Classification<br>Name: | Electroencephalograph | | |
| Classification<br>Regulation: | 21 CFR §882.1400 | | |
| Product Code: | GWQ | | |
| Classification<br>Panel: | Neurology | | |
| Substantially<br>Equivalent Devices: | New Model | Predicate 510(k)<br>Number | Predicate<br>Manufacturer / Model |
| | XLTEK EMU40EX EEG<br>Headbox | K053386 | XLTEK / XLTEK<br>EMU40 EEG Headbox |
| Device Description: | The Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK)<br>EMU40EX System is a powerful and flexible EEG recording system.<br>The EMU40EX System consists of a patient electrode set, EMU40EX<br>EEG Headbox, and the Natus Base Unit. | | |
| | The EMU40EX System is designed to work with an XLTEK host<br>computer system running the Natus Database and Natus NeuroWorks®<br>data review and analysis software. | | |
| | The EMU40EX EEG Headbox can be connected to the Natus Base Unit<br>with a cable, or can connect wirelessly through a secure BlueTooth link | | |
{4}------------------------------------------------
Intended Use: The XLTEK EMU40EX EEG Headbox is an electroencephalograph that works in conjunction with XLTEK NeuroWorks® software.
> The XLTEK EMU40EX EEG Headbox is used to acquire, digitize, store and transmit physiological signals (such as EEG, pulse and oximetry signals) for EEG studies in research and clinical environments.
The XLTEK EMU40EX EEG Headbox requires competent user input, and its output must be reviewed and interpreted by trained medical professionals who will exercise professional judgement in using this information.
Technology Comparison:
The XLTEK EMU40EX EEG Headbox employs the same technological characteristics as the predicate device.
| ECG System<br>Characteristic | XLTEK EMU40<br>EEG Headbox<br>(K053386) | XLTEK EMU40EX<br>EEG Headbox<br>(Proposed Device) |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Intended Use | The XLTEK EMU40 EEG<br>Headbox is an<br>electroencephalograph that works<br>in conjunction with XLTEK<br>NeuroWorks software.<br><br>The XLTEK EMU40EX EEG<br>Headbox is used to acquire,<br>digitize, store and transmit<br>physiological signals (such as<br>EEG, pulse and oximetry signals)<br>for EEG studies in research and<br>clinical environments.<br><br>The XLTEK EMU40EX EEG<br>Headbox requires competent user<br>input, and its output must be<br>reviewed and interpreted by<br>trained medical professionals whojudgement in using this<br>information. | Same |
| Referential Channels | 40 | Same |
| DC Channels | Four (4) isolated | Same |
| Electrode Connections<br>(including common<br>input) | Touch-proof | Same |
| Oximeter Connection | Yes | Same |
| Photic Stimulator<br>Connection | Yes | Same |
| Patient Event Button | Yes | Same |
{5}------------------------------------------------
### Technology Comparison (continued):
| ECG System<br>Characteristic | XLTEK EMU40<br>EEG Headbox<br>(K053386) | XLTEK EMU40EX<br>EEG Headbox<br>(Proposed Device) |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Batteries | Two (2) AAA Rechargeable | Same |
| Compatible with<br>Rechargeable External<br>Battery Pack | Yes | Same |
| Front Panel Base Unit<br>Controls | LCD Display<br>Four (4) buttons | Touchscreen LCD display. |
| Electromagnetic<br>Compatibility | The XLTEK EMU40EX EEG Headbox was verified for performance in<br>accordance with the following standard:<br>IEC 60601-1-2: 2007, Medical electrical equipment – Part 1-2:<br>General requirements for basic safety and essential performance –<br>Collateral standard: Electromagnetic compatibility – Requirements<br>and tests. | |
| | Results indicate that the XLTEK EMU40EX EEG Headbox passed<br>verification, and complies with the applicable standards. | |
| Performance<br>Testing – Bench | The XLTEK EMU40EX EEG Headbox was verified and validated for<br>performance in accordance with product requirements including:<br>Signal quality. Functionality. User interface. | |
| | and the applicable clauses of the following standards:<br>IEC 60601-1-6: 2010, Medical electrical equipment – Part 1-6:<br>General requirements for basic safety and essential performance –<br>Collateral standard: Usability. IEC 60601-2-26: 2012, Medical electrical equipment – Part 2-26:<br>Particular requirements for the basic safety and essential<br>performance of electroencephalographs. IEC 62366: 2007, Medical devices – Application of usability<br>engineering to medical devices. ISO 80601-2-61: 2011, Medical electrical equipment – Part 2-61:<br>Particular requirements for basic safety and essential performance of<br>pulse oximeter equipment. | |
| | Results indicate that the XLTEK EMU40EX EEG Headbox passed<br>verification and validation, and complies with the product requirements<br>and the applicable standards. | |
| Conclusion | Verification and validation activities were conducted to establish the<br>performance and safety characteristics of the device modifications made<br>to the XLTEK EMU40EX EEG Headbox. The results of these activities<br>demonstrate that the XLTEK EMU40EX EEG Headbox is as safe, as<br>effective, and performs as well as or better than the predicate device.<br>Therefore, the XLTEK EMU40EX EEG Headbox is considered<br>substantially equivalent to the predicate device | |
### Summary of Performance Testing:
Software
The XLTEK EMU40EX EEG Headbox firmware was designed and developed according to a robust software development process, and was rigorously verified and validated. Firmware information is provided in accordance with internal firmware requirements and the following FDA guidance documents and standards:
- The content of premarket submissions for software contained in medical devices, 11 May 05.
- Off-the-shelf software use in medical devices, 09 Sep 99.
- General principles of software validation; Final guidance for industry and FDA staff, 11 Jan 02.
- . Content of premarket submissions for management of cybersecurity in medical devices, 02 Oct 14.
- . IEC 62304: 2006, Medical device software – Software life cycle processes
Results indicate that the XLTEK EMU40EX EEG Headbox firmware passed verification and validation, complies with its predetermined firmware specifications, the applicable guidance documents, and the applicable standards.
#### The XLTEK EMU40EX EEG Headbox was verified for performance in Electrical Safety accordance with the following standard:
- IEC 60601-1: 2005, Am1: 2012, Medical electrical equipment Part . 1: General requirements for basic safety and essential performance.
Results indicate that the XLTEK EMU40EX EEG Headbox passed verification, and complies with the applicable standards.
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.