← Product Code [NXC](/productcode/NXC) · K163155

# MTM Clear Aligner (K163155)

_Dentsply Sirona · NXC · Feb 3, 2017 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K163155

## Device Facts

- **Applicant:** Dentsply Sirona
- **Product Code:** [NXC](/productcode/NXC.md)
- **Decision Date:** Feb 3, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K163155 · Feb 3, 2017 | MTM Clear Aligner | Dentsply Sirona | Scientific literature (published clinical studies) | A literature review was conducted to provide valid scientific evidence for intrusive and extrusive orthodontic tooth movements achieved with clear aligner therapy to support the expansion of the device's indications for use. | Literature review; Orthodontic tooth movement; Clear aligner therapy |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Literature review of clear aligner therapy; Literature review | Patients undergoing clear aligner therapy | Not applicable for this study | Intrusive and extrusive orthodontic tooth movements |

## Indications for Use

MTM® Clear Aligner is indicated for the treatment of anterior tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental minor tooth movements, MTM® Clear Aligner sequentially positions teeth by way of continuous gentle force.

## Device Story

MTM Clear Aligner consists of custom-made, removable, thermoformed plastic orthodontic appliances; used to treat anterior tooth malocclusions. Device applies continuous gentle force to teeth to achieve incremental movement. Patient wears aligners 20-22 hours per day; removes them for eating and cleaning. Device is used in clinical settings under professional supervision. Modification from predicate allows for all types of tooth movement, including intrusion and extrusion, by removing specific movement restrictions from the indications for use statement.

## Clinical Evidence

No new human clinical data included. Relies on clinical data previously reviewed for K132145 and K123925. A literature review was conducted to provide scientific evidence supporting the use of clear aligner therapy for intrusive and extrusive orthodontic tooth movements.

## Technological Characteristics

Sequential thermoformed plastic aligners made of co-polyester or co-polymer. Operates via continuous gentle force application. Removable by patient. No electronic components or software algorithms described.

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- MTM® Clear Aligner ([K132145](/device/K132145.md))
- ClearPath Aligner ([K123514](/device/K123514.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 3, 2017

Dentsply Sirona Ms. Helen Lewis Director Corporate Regulatory Affairs 221 West Philadelphia Street Suite 60 York, Pennsylvania 17408

Re: K163155

Trade/Device Name: MTM Clear Aligner Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: November 9, 2016 Received: November 10, 2016

Dear Helen Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K163155

## Device Name

MTM® Clear Aligner

## Indications for Use (Describe)

MTM® Clear Aligner is indicated for the treatment of anterior tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental minor tooth movements, MTM® Clear Aligner sequentially positions teeth by way of continuous gentle force.

| Type of Use (Select one or both, as applicable) | <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)   </span> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)   </span> |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time
to review instructions, search existing data sources, gather and maintain the data needed and complete and review
the collection of information. Send comments regarding this burden estimate or any other aspect of this
information collection, including suggestions for reducing this burden, to:

 Department of Health and Human  

 Services Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

 PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Dentsply Sirona 221 West Philadelphia Street Suite 60 York, PA 17404

Image /page/3/Picture/1 description: The image contains the logo for Dentsply Sirona. On the left side of the logo is a gray abstract leaf-like shape. To the right of the leaf shape is the text "Dentsply" on the top line and "Sirona" on the bottom line. The text is in a simple sans-serif font.

# 510(k) SUMMARY for MTM® Clear Aligner (K163155)

Submitter Information: 1.

> Dentsply Sirona 221 West Philadelphia Street Suite 60 York, PA 17404

| Contact Person:   | Helen Lewis  |
|-------------------|--------------|
| Telephone Number: | 717-487-1332 |
| Fax Number:       | 717-849-4343 |

Date Prepared: 19 January 2017

#### 2. Device Name:

- MTM® Clear Aligner Proprietary Name: ●
- Classification Name: Aligner, Sequential ●
- CFR Number: 872.5470 ●
- Device Class: . II
- Product Code: NXC ●

#### 3. Predicate Devices:

| Predicate Device Name | 510(k)  | Company Name           |
|-----------------------|---------|------------------------|
| MTM® Clear Aligner    | K132145 | Raintree Essix Inc.    |
| ClearPath Aligner     | K123514 | ClearPath Orthodontics |

#### 4. Description of Device:

MTM® Clear Aligner consists of a series of custom made removable plastic orthodontic appliances that sequentially positions teeth by way of continuous gentle force.

#### న. Indications for Use:

MTM® Clear Aligner is indicated for the treatment of anterior tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental minor tooth movements, MTM® Clear Aligner sequentially positions teeth by way of continuous gentle force.

# Indications for Use Comparison:

In this submission we focus on removing mention of specific movement types from the indications for use statement of the primary predicate. The indications for use are similar between the primary predicate MTM® Clear Aligner and the proposed MTM® Clear Aligner with expanded indications. Both devices are indicated for the correction of anterior tooth misalignments (malocclusions) in 

{4}------------------------------------------------

patients with permanent teeth, using a series of aligners. The primary predicate specifies that four types of tooth movement are used, whereas the proposed MTM® Clear Aligner removes the specific tooth movement types from the indications for use statement, to allow all types of tooth movement to be used.

The indications for use are fundamentally the same between the reference predicate, ClearPath Aligner and the proposed MTM® Clear Aligner. Both devices are indicated for the correction of anterior tooth misalignments (malocclusions) in patients with permanent teeth, using a series of aligners. Neither of them specify the types of tooth movement to be used.

The change in the indications for use to remove specific tooth movement types for the proposed MTM® Clear Aligner does not result in a new intended use for the current, primary predicate MTM® Clear Aligner. This change does not raise new concerns as evidenced in the scientific literature reviewed.

{5}------------------------------------------------

#### 6. Substantial Equivalence:

#### Table 5.1 Device Comparison Table

| Element                    | Proposed Device                                                                                                                                                                                                                                                                                                       | Primary Predicate                                                                                                                                                                                                                                                                                                                                                        | Reference Predicate                                                                                                                                                                                                                                                                                                           | Comparison                                                                                                                                                           |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)                     | K163155                                                                                                                                                                                                                                                                                                               | K132145                                                                                                                                                                                                                                                                                                                                                                  | K123514                                                                                                                                                                                                                                                                                                                       | N/A                                                                                                                                                                  |
| Device Name                | MTM® Clear Aligner                                                                                                                                                                                                                                                                                                    | MTM® Clear Aligner                                                                                                                                                                                                                                                                                                                                                       | ClearPath Aligner                                                                                                                                                                                                                                                                                                             | N/A                                                                                                                                                                  |
| Manufacturer               | Raintree Essix Inc.                                                                                                                                                                                                                                                                                                   | Raintree Essix Inc.                                                                                                                                                                                                                                                                                                                                                      | ClearPath Orthodontics                                                                                                                                                                                                                                                                                                        | N/A                                                                                                                                                                  |
| Intended Use               | Treatment of anterior tooth<br>malocclusions in patients with<br>permanent dentition                                                                                                                                                                                                                                  | Treatment of anterior tooth<br>malocclusions in patients with<br>permanent dentition                                                                                                                                                                                                                                                                                     | Correction of dental malocclusion<br>in patients with permanent<br>dentition                                                                                                                                                                                                                                                  | Same                                                                                                                                                                 |
| Indications for<br>Use     | MTM® Clear Aligner is indicated<br>for the treatment of anterior tooth<br>malocclusions in patients with<br>permanent dentition (i.e. all second<br>molars).  Utilizing a series of<br>incremental minor tooth<br>movements, MTM® Clear Aligner<br>sequentially positions teeth by way<br>of continuous gentle force. | MTM® Clear Aligner is indicated for<br>the treatment of anterior tooth<br>malocclusions in patients with<br>permanent dentition (i.e. all second<br>molars).  Utilizing a series of<br>incremental minor tooth movements<br>(torque, tipping, rotation and bodily<br>movement), MTM® Clear Aligner<br>sequentially positions teeth by way of<br>continuous gentle force. | The ClearPath Aligner is a series<br>of clear, lightweight, plastic<br>appliances indicated for the<br>correction of dental malocclusion<br>in patients with permanent<br>dentition (i.e. all second molars).<br>The ClearPath Aligner is intended<br>for minor anterior tooth movement<br>by way of continuous gentle force. | Same as reference<br>predicate. Removing<br>specific movement types<br>from primary predicate<br>IFU statement.                                                      |
| Tooth<br>Movement<br>Types | All                                                                                                                                                                                                                                                                                                                   | All except intrusion and extrusion.                                                                                                                                                                                                                                                                                                                                      | All                                                                                                                                                                                                                                                                                                                           | Same as reference<br>predicate. Removing<br>specific movement types<br>from primary predicate<br>IFU statement to allow<br>for intrusive and extrusive<br>movements. |
| Material                   | Co-polyester or Co-polymer                                                                                                                                                                                                                                                                                            | Co-polyester or Co-polymer                                                                                                                                                                                                                                                                                                                                               | Co-Polyester                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                 |
| Mode of<br>Action          | Continuous gentle force applied to<br>teeth to achieve movement.                                                                                                                                                                                                                                                      | Continuous gentle force applied to teeth<br>to achieve movement.                                                                                                                                                                                                                                                                                                         | Continuous gentle force applied to<br>teeth to achieve movement.                                                                                                                                                                                                                                                              | Same                                                                                                                                                                 |
| Features                   | Sequential thermoformed plastic<br>aligners used in the alignment of<br>teeth by application of continuous<br>gentle force. Aligners can be<br>removed by the patient for eating<br>and cleaning. Aligners are most<br>effective when worn 20-22 hours<br>per day.                                                    | Sequential thermoformed plastic<br>aligners used in the alignment of teeth<br>by application of continuous gentle<br>force. Aligners can be removed by the<br>patient for eating and cleaning. Aligners<br>are most effective when worn 20-22<br>hours per day.                                                                                                          | Sequential thermoformed plastic<br>aligners used in the alignment of<br>teeth by application of continuous<br>gentle force. Aligners can be<br>removed by the patient for eating<br>and cleaning. Aligners are most<br>effective when worn 20-22 hours<br>per day.                                                            | Same                                                                                                                                                                 |

The technological characteristics of the proposed MTM® Clear Aligner are unchanged from those of the primary predicate MTM® Clear Aligner device cleared under premarket notification K132145.

{6}------------------------------------------------

#### 7. Non-Clinical Performance Data

No additional non-clinical bench testing has been included in this submission. Nonclinical data was previously submitted and reviewed to support clearance of the primary predicate MTM® Clear Aligner under premarket notification K132145.

#### 8. Clinical Performance Data

No new human clinical data has been included in this submission: however a literature review was conducted. Clinical data was previously submitted and reviewed to support clearance of the primary predicate MTM® Clear Aligner under premarket notification K132145, as well as in a prior submission for the MTM® Clear Aligner In-Office system under premarket notification K123925. This data demonstrated performance of the predicate MTM® Clear Aligner for sequential minor incremental tooth movements using Raintree Essix plastics with force point technology. A literature review was conducted in support of the proposed device modification. The literature review is included in this submission to provide valid scientific evidence for intrusive and extrusive orthodontic tooth movements achieved with clear aligner therapy.

#### 9. Conclusion Regarding Substantial Equivalence

The data submitted for non-clinical and clinical performance testing, as well as the technological characteristics of the proposed MTM® Clear Aligner device support substantial equivalence when compared to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K163155](https://fda-staging.innolitics.com/device/K163155)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).