K163138 · Clarius Mobile Health Corp. · IYN · Nov 30, 2016 · Radiology
Device Facts
Record ID
K163138
Device Name
Clarius Ultrasound System
Applicant
Clarius Mobile Health Corp.
Product Code
IYN · Radiology
Decision Date
Nov 30, 2016
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
AI/ML, Pediatric, 3rd-Party Reviewed
Indications for Use
The Clarius Ultrasound System is a software-based ultrasound imaging system and accessories intended for use in Point-of-Care Imaging of Medical Conditions on the general public. Point-of-Care clinical applications include: - Emergency triage exam to look at trauma conditions - Procedure guidance to guide needles into the body; and - Other targeted diagnostic and measurement applications: fetal echo, abdominal, small organ, musculo-skeletal (conventional), musculo-skeletal (superficial), urology, gynecology, cardiac adult, cardiac pediatric, peripheral vessel, pediatric, carotid - The Clarius Ultrasound System is intended for use in environments where healthcare is provided by trained medical professionals. The device is not intended for use in emergency medical service, ambulance, or aircraft. Users will be trained medical professionals (e.g., doctors, nurses, technicians).
Device Story
Portable, wireless ultrasound system; consists of handheld C3 convex or L7 linear transducers and Clarius App for iOS/Android. Piezoelectric transducers transmit high-frequency sound waves; detect echoes; convert to electronic signals via internal beamformer/pre-scan converter. Data transmitted via Wi-Fi Direct to COTS tablet/smartphone for real-time display. Used by trained medical professionals (doctors, nurses, technicians) in clinical settings for point-of-care imaging, trauma triage, and procedure guidance. Provides high-resolution visualization of internal structures; assists in needle/catheter placement and nerve blocks. Benefits include portability and ease of use for rapid clinical assessment.
Clinical Evidence
Bench testing only. Evidence includes verification and validation of system design, risk analysis, and compliance with international standards for medical electrical equipment safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37) and biological evaluation (ISO 10993-1).
Technological Characteristics
Handheld wireless transducers (convex/linear) with piezoelectric elements. Features internal battery, power generator, multichannel beamformer, pre-scan converter, and Wi-Fi Direct connectivity. Operates with COTS iOS/Android devices. Software-controlled imaging. Complies with IEC 60601-1 (Ed 3.1), IEC 60601-1-2 (4th Ed), IEC 60601-2-37 (3rd/4th Ed), and ISO 10993-1.
Indications for Use
Indicated for point-of-care diagnostic ultrasound imaging and fluid flow analysis in adults and pediatrics. Applications include emergency triage (trauma), procedure guidance (needle/catheter placement, nerve blocks), and diagnostic imaging of fetal, abdominal, small organ, musculoskeletal, urological, gynecological, cardiac, and vascular structures. Contraindicated for ophthalmic use, intra-operative use, endocavitary use, imaging open wounds, or use in emergency medical services/transport.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
SONON Ultrasound Imaging System (Model: SONON 300C) (K151339)
Philips Lumify Diagnostic Ultrasound System (K152899)
FUJIFILM SonoSite Edge II Ultrasound System (K153626)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 30, 2016
Clarius Mobile Health Corp. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street. NW BUFFALO MN 55313
Re: K163138
Trade/Device Name: Clarius Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: November 7, 2016 Received: November 8, 2016
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
For
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K163138
Device Name Clarius Ultrasound System
#### Indications for Use (Describe)
The Clarus Ultrasound System is a software-based ultrasound imaging system and accessories intended for use in Pointof-Care Imaging of Medical Conditions on the general public.
Point-of-Care clinical applications include:
- · Emergency triage exam to look at trauma conditions
- · Procedure guidance to guide needles into the body; and
· Other targeted diagnostic and measurement applications: fetal, fetal echo, abdominal, small organ, musculo-sketal (conventional), musculo-skeletal (superficial), urology, cardiac adult, cardiac pediatric, peripheral vessel, pediatric, carotid
· The Clarius Ultrasound System is intended for use in environments where is provided by trained medical professionals. The device is not intended for use in emergency medical service, ambulance, or aircraft.
Users will be trained medical professionals (e.g., doctors, nurses, technicians).
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Diagnostic Ultrasound Indication for Use Form
510(k) Number: NA
Device Name: Clarius Ultrasound System
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|--------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------|
| | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(specify) | Other* |
| General<br>(Track 1 Only) | Ophthalmic | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | N | | | | | | |
| | Abdominal | N | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | N | | | | | | |
| | Small organ (thyroid,<br>scrotum, prostate,<br>breast) | N | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.<br>(non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | N | | | | | | |
| | Musculo-skeletal<br>(Superficial) | N | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Urology,<br>Gynecology) | N | | | | | | |
| Cardiac | Cardiac Adult | N | | | | | | |
| | Cardiac Pediatric | N | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Fetal Echo) | N | | | | | | |
| Peripheral<br>Vessel | Peripheral Vessel | N | | | | | | |
| | Other (Carotid) | N | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging.
Additional Comments: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for nerve block procedures.
{4}------------------------------------------------
### Diagnostic Ultrasound Indication for Use Form
510(k) Number: NA
Device Name: C3: convex scanner
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|--------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------|
| | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(specify) | Other* |
| General<br>(Track 1 Only) | Ophthalmic | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | N | | | | | | |
| | Abdominal | N | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | N | | | | | | |
| | Small organ (thyroid,<br>scrotum, prostate,<br>breast) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.<br>(non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | N | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Urology,<br>Gynecology) | N | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Fetal Echo) | N | | | | | | |
| Peripheral<br>Vessel | Peripheral Vessel | N | | | | | | |
| | Other (Carotid) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging.
Additional Comments: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for nerve block procedures.
{5}------------------------------------------------
### Diagnostic Ultrasound Indication for Use Form
510(k) Number: NA
Device Name: L7: linear scanner
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Specific | Mode of Operation | | | | | | |
|---------------------------|--------------------------------------------------------|-------------------|---|-----|-----|---------------|--------------------|--------|
| | (Tracks 1 & 3) | B | M | PWD | CWD | Color Doppler | Combined (specify) | Other* |
| General<br>(Track 1 Only) | | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | | |
| | Abdominal | N | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | N | | | | | | |
| | Small organ (thyroid,<br>scrotum, prostate,<br>breast) | N | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.<br>(non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | N | | | | | | |
| | Musculo-skeletal<br>(Superficial) | N | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Urology,<br>Gynecology) | | | | | | | |
| Cardiac | Cardiac Adult | N | | | | | | |
| | Cardiac Pediatric | N | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Fetal Echo) | | | | | | | |
| Peripheral<br>Vessel | Peripheral Vessel | N | | | | | | |
| | Other (Carotid) | N | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging.
Additional Comments: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for nerve block procedures.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange icon on the left and the word "clarius" in dark blue on the right. The icon is made up of three curved lines that form a stylized "C" shape.
## Summary
1. Submitter's name, address, telephone number; Contact person
Submitter: Abhijit Ahir, Director of QA/RA Clarius Mobile Health Corp. #350 – 3605 Gilmore Way Burnaby, B.C. V5G 4X5 Canada Email: abhijit.ahir@clarius.me Tel: (+1) 778-800-9975 Fax: (+1) 778-800-9974
Contact Person: Emergo Global Representative, LLC 816 Congress Avenue, Suite 1400 Austin, TX 78701 USA Tel: (+1) 512-327-9997
Date Prepared: November 28, 2016
2. Name of the device, including the trade or proprietary name, if applicable, the common or usual name, and the classification, if known
Clarius Ultrasound System Device Name:
Common Name: Diagnostic Ultrasound System and Accessories
Classification: Class II
| 21 CFR<br>Section | Classification Name | Product Code |
|-------------------|------------------------------------------|--------------|
| 892.1550 | Ultrasonic Pulsed Doppler Imaging System | 90 IYN |
| 892.1560 | Ultrasonic Pulsed Echo Imaging System | 90 IYO |
| 892.1570 | Diagnostic Ultrasound Transducer | 90 ITX |
- 3. Substantially Equivalent Devices:
| Device Name | 510(k) Number |
|-----------------------------------------------------|---------------|
| SONON Ultrasound Imaging System (Model: SONON 300C) | K151339 |
| Philips Lumify Diagnostic Ultrasound System | K152899 |
| FUJIFILM SonoSite Edge II Ultrasound System. | K153626 |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange symbol that looks like three curved lines forming a stylized "C", followed by the word "clarius" in a dark blue, sans-serif font. The logo is simple and modern.
# 1. Device Description
The Clarius Ultrasound System is a portable, general-purpose, software-controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through a commercial off-theshelf (COTS) iOS or Android device. The Clarius Ultrasound System comprises a series of wireless transducers employing Bluetooth and Wi-Fi-based technology to communicate with traditional tablet/smartphone devices via direct Wi-Fi. This allows the user to export ultrasound images and display them across a range of portable personal devices.
Image /page/7/Picture/3 description: The image shows a Clarius Ultrasound device wirelessly connected to a compatible viewing device, which appears to be a tablet. The ultrasound device is a handheld scanner, and the tablet displays the ultrasound image. The text labels "Clarius Ultrasound" and "Compatible Viewing Device" are present in the image. The wireless connection is indicated by a blue wifi symbol.
Piezoelectric material in the systems transmits high frequency, non-ionizing sound waves to the designated region of the body and converts the subsequent echoes detected to electronic signals in order to construct an image of the internal structures of an anatomical field. This image is sent wirelessly to an external (COTS) iOS or Android viewing device on which the image can be displayed. The transducer houses a battery and power generator, multichannel beam former, pre-scan converter, and Wi-Fi components. The battery is removable and comes with a separate charger. Communication between the transducer and the compatible viewing device will be via Wi-Fi Direct® (Ad-Hoc mode with security) for easy pairing.
The Clarius Ultrasound System includes:
Product/Package Components:
- 1. Software:
- The Ultrasound App (Clarius App) for iOS; OR
- The Ultrasound App (Clarius App) for Android ●
- 2. Transducers/Scanners:
- Clarius Scanner C3 (C3 Convex Transducer); OR ●
- . Clarius Scanner L7 (L7 Linear Transducer)
- 3. Accessories:
- a. Clarius-Built:
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange icon on the left and the word "clarius" in dark blue on the right. The icon is made up of three curved lines that form a stylized "C" shape.
- Battery Pack (Li-ion) O
- Battery Charger O
OEM/Off-The-Shelf Product(s):
Medical Power Supply (Off-the-shelf power adaptor from SL Power Electronics, USA; Model Number ME20A1203B02; Approved in the US.
Aquasonic 100 Ultrasound Transmission Gel (Off-the-shelf ultrasound gel from Parker Laboratories Inc., USA; Approved in the US.
Image /page/8/Figure/6 description: The image shows a diagram of an ultrasound system. On the left side of the image, there is a diagram of an ultrasound probe, which includes components such as a battery, Wi-Fi, power circuitry, ultrasound system-on-chip, transmit/receive analog interface, and transducer stack. On the right side of the image, there is a diagram of a Wi-Fi enabled smartphone or tablet, which displays an ultrasound image and a user interface.
Ultrasound Probe
User Interface / Display
The concept of the Clarius Ultrasound System transducers and software is primarily to provide an easy to use, high-performance, low-cost, ultrasound platform for teaching and clinical applications. The Clarius Ultrasound System is intended for use in professional healthcare facilities where healthcare is provided by trained medical professionals. The device is not intended for use in Emergency Medical Service, ambulance, or aircraft environments.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange icon on the left and the word "clarius" in dark blue on the right. The icon is made up of three curved lines that form a stylized "C" shape.
## 2. Indication for Use
The Clarius Ultrasound System is a software-based ultrasound imaging system and accessories intended for use in Point-of-Care Imaging of Medical Conditions on the general public.
Point-of-Care clinical applications include:
- Emergency triage exam to look at trauma conditions ●
- Procedure guidance to guide needles into the body; and
- Other targeted diagnostic and measurement applications: fetal echo, abdominal, small organ, musculo-skeletal (conventional), musculo-skeletal (superficial), urology, gynecology, cardiac adult, cardiac pediatric, peripheral vessel, pediatric, carotid
- The Clarius Ultrasound System is intended for use in environments where healthcare is provided by trained medical professionals. The device is not intended for use in emergency medical service, ambulance, or aircraft.
Users will be trained medical professionals (e.g., doctors, nurses, technicians).
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange symbol that looks like three curved lines forming a "C" shape, followed by the word "clarius" in a dark blue, sans-serif font. The logo is clean and modern, with a focus on simplicity and readability.
## 2.1 Contraindications
Do not use the Clarius Ultrasound System in the following situations. Doing so may produce images with inaccurate results:
- Patients who have had surgery, which may have changed the composition of the examining tissue (for example, a mastectomy), as this could skew or alter the measured density.
- Patients whose bodies contain foreign artifacts (for example, implants). ●
- Ophthalmic use or any use causing the acoustic beam to pass through the eye. ●
- Intra-operative use (e.g., defined as introducing a scanner into a surgical incision or burr hole). ●
- Endocavitary use (i.e., defined as introducing a scanner within a (body) cavity or organ. E.g. an atrium, esophagus, rectum, or vagina).
- . Imaging an open wound.
- During transportation of a patient to a professional healthcare facility, or between professional healthcare facilities.
- At the scene of an emergency outside of a professional healthcare facility.
In terms of equivalent devices, we will refer to FDA's standard terminology for comparable devices: "predicate devices". The predicate devices selected to demonstrate equivalence are;
| 1. Device Name<br>SONON Ultrasound Imaging System. (Model 300C) | FDA 510(k) Number<br>(K151339) |
|-----------------------------------------------------------------|--------------------------------|
| 2. Device Name<br>Lumify Diagnostic Ultrasound System. | FDA 510(k) Number<br>(K152899) |
| 3. Device Name<br>FUJIFILM SonoSite Edge II Ultrasound System. | FDA 510(k) Number<br>(K153626) |
## 3. Determination of Substantial Equivalence
They Clarius Ultrasound system is a Track 3 system that employs the same fundamental scientific technology as that cleared with K151339, K153626. All indications for use introduced by Clarius are indications used by one of the predicate devices. The K152899) and (K153626) do not connect with the display wirelessly. The Clarius ultrasound system is equivalent to the K151339in this respect.
### Non-Clinical Performance Data
Non-clinical performance tests relied on this premarket notification for a determination of substantial equivalence include rests which show compliance to the following standards:
| Reference No. | Year | Title |
|---------------|------|--------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 | 2012 | Medical electrical equipment – Part 1: General requirements for basic safety<br>and essential performance. (Edition 3.1) |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange symbol that looks like three concentric arcs, with a small gap in each arc. To the right of the symbol is the word "clarius" in a dark blue, sans-serif font. The logo is clean and modern, and the colors are bright and eye-catching.
| IEC 60601-1-2 | 2014 | Medical electrical equipment – Part 1-2: General requirements for basic<br>safety and essential performance – Collateral Standard: Electromagnetic<br>Capability – Requirements and tests. (4th Edition) |
|--------------------|------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-2-<br>37 | 2004 | Medical electrical equipment – Part 2-37: Particular requirements for the<br>basic safety and essential performance of ultrasonic medical diagnostic and<br>monitoring equipment. (3rd and 4th Edition) |
| ISO 10993-1 | 2009 | Biological evaluation of medical devices - Part 1: Evaluation and testing<br>within a risk management process |
## 4. Quality Assurance measures
Quality assurance measures applied to the system design and development include, but were not limited to: risk analysis, product specifications, design reviews, and verification and validation.
## 5. Conclusion
The Clarius Ultrasound System is essentially the same as it's predicate devices. These are prescription devices. This statement appears in the labeling. Sterilization is not applicable.
This is a Class II Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.