# Youha electric breast pump (K163136)

_Ningbo Youhe Electrical Appliance Technology Co., Ltd. · HGX · Jul 31, 2017 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K163136

## Device Facts

- **Applicant:** Ningbo Youhe Electrical Appliance Technology Co., Ltd.
- **Product Code:** [HGX](/productcode/HGX.md)
- **Decision Date:** Jul 31, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The Youha electric breast pump is intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.

## Device Story

Youha electric breast pump is a single-user, powered device for lactating women to express and collect breast milk. It utilizes a microprocessor-controlled reciprocating diaphragm vacuum pump to generate suction up to 280mmHg. The device features three operational modes: massage (fast cycles, low vacuum), low expression (slow cycles, high vacuum), and high expression (slower cycles, higher vacuum). Users interact with the device via a front panel keypad and LCD display to select modes and adjust vacuum pressure. It is powered by an internal rechargeable battery or an external AC/DC adapter. The device is intended for home use by the patient. Output consists of expressed milk collected into a container. The device includes backflow protection to prevent milk contamination.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), home-use safety (IEC 60601-1-11), and biocompatibility (ISO 10993). Performance testing verified vacuum levels, cycle rates, and backflow protection across all power sources and pumping configurations.

## Technological Characteristics

Powered by internal rechargeable battery or AC/DC adapter. Diaphragm vacuum pump controlled by a microprocessor. User interface includes LCD and keypad. Biocompatibility per ISO 10993. Electrical safety per IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11. Software developed per IEC 62304.

## Regulatory Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

## Predicate Devices

- Spectra 9 Plus powered breast pump ([K162415](/device/K162415.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, composed of three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The seal is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 31, 2017

Ningbo Youhe Electrical Appliance Technology Co., Ltd. % Daniel Qiu Project Manager Shanghai Qisheng Business Consulting Co.,ltd. Room 1301, Bld 46, Jing Gu Zhong Rd., No.58, Min Hang District, Shanghai, CN Shanghai, 200240 China

Re: K163136 Trade/Device Name: Youha electric breast pump, Models: YH-8004, YH-8016, YH-8006IV, YH-8015 Regulation Number: 21 CFR& 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: June 22, 2017 Received: June 22, 2017

Dear Daniel Qiu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Joyce M. Whang -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K163136

### Device Name

Youha electric breast pump, Models: YH-8004, YH-8016, YH-8006IV, YH-8015

Indications for Use (Describe)

The Youha electric breast pump is intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# K163136

#### 1. Submitter Information

| Name:      | Ningbo Youhe Electrical Appliance Technology Co., Ltd                  |
|------------|------------------------------------------------------------------------|
| Address:   | No.226-228 San Jiao Tian Road, Fu Hai Town, CiXi City, Zhejiang, China |
| Telephone: | +86-574-63860315                                                       |
| Fax:       | +86-574-23660066                                                       |

Contact Person: Huang Yun Jun

- 2. Date Prepared: June 30, 2017
#### 3. Device Information

| Trade Name:        | Youha electric breast pump           |
|--------------------|--------------------------------------|
| Common Name:       | Powered breast pump                  |
| Models:            | YH-8004, YH-8016, YH-8006IV, YH-8015 |
| Regulation Class:  | Class II                             |
| Regulation Number: | 21CFR 884.5160, Powered breast pump  |
| Review Panel:      | Obstetrics/Gynecology                |
| Product Code:      | HGX (pump, breast, powered)          |

#### 4. Predicate Device

K162415 510(k): Device Name: Spectra 9 Plus powered breast pump Manufacturer: Uzinmedicare Company

#### 5. Device Description

The Youha electric breast pump is a single-user, powered breast pump intended to be used by lactating women to express and collect milk from their breasts. YH-8004 and YH-8016 have both single (pumping from one breast) and double pumping (pumping from both breasts) configurations. YH-8006IV and YH-8015 have a single pumping configuration only.

A reciprocating diaphragm vacuum pump, driven by a microprocessor, generates the suction to extract the milk at vacuum levels up to 280mmHg. The user interface consists of a front panel keypad and LCD display in which the user switches between modes and controls the vacuum pressure.

The device has three modes of operation:

- . Massage mode: Suction patterns with fast cycles and low vacuum to start milk flowing

{4}------------------------------------------------

- Low expression mode: Suction patterns with slow cycles and high vacuum to . express more milk gently and efficiently.
- High expression mode: Suction patterns with slower cycles and higher vacuum to . express more milk efficiently.

YH-8004, YH-8015 and YH-8016 have 3 modes including massage mode, low expression mode and high expression mode. YH-8006IV has 2 modes including massage mode and high expression mode.

The device is electrically powered from either an internal rechargeable battery or an external AC power adapter. The external adapter also charges the battery.

#### 6. Indications for Use

The Youha electric breast pump is intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.

#### 7. Predicate Comparison

The following table compares the Youha Electric breast pump to the predicate device with respect to indications for use and technological characteristics:

| Device & Predicate Device(s):                | Subject Device:                                                                                                                                                     | Predicate Device:                                                                                                                          |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
|                                              | Youha electric breast pump<br>K163136                                                                                                                               | Spectra 9plus<br>K162415                                                                                                                   |
| Indications for Use                          | The Youha electric breast pump is<br>intended to be used by lactating women<br>to express and collect milk from their<br>breasts. It is intended for a single user. | The Spectra 9Plus is a single-user,<br>powered breast pump intended to<br>express and collect milk from the<br>breasts of lactating women. |
| Single user                                  | Yes                                                                                                                                                                 | Yes                                                                                                                                        |
| Provided Non-sterile                         | Yes                                                                                                                                                                 | Yes                                                                                                                                        |
| Re-usable                                    | Yes                                                                                                                                                                 | Yes                                                                                                                                        |
| Direct user contact                          | Yes                                                                                                                                                                 | Yes                                                                                                                                        |
| Power source                                 | AC/DC wall converter and<br>Rechargeable                                                                                                                            | Power source                                                                                                                               |
| Suction Strength (Expression<br>Mode) (mmHg) | Low<br>YH-8004:75-250<br>YH-8016:75-250<br>YH-8015:115-250<br>YH-8006IV:N/A                                                                                         | 50-270                                                                                                                                     |
|                                              | High<br>YH-8004:120-280<br>YH-8016:120-280<br>YH-8015: 120-280<br>YH-8006IV:125-<br>280                                                                             |                                                                                                                                            |
| Suction Strength (Massage<br>Mode) (mmHg)    | YH-8004:34-190<br>YH-8016:34-190<br>YH-8015: 50-190<br>YH-8006IV: 50-190                                                                                            | 50-150                                                                                                                                     |
| Adjustable Suction levels                    | Yes                                                                                                                                                                 | Yes                                                                                                                                        |
| Suction Settings (High/low                   | YH-8004:6                                                                                                                                                           | 10                                                                                                                                         |

{5}------------------------------------------------

| Expression Mode)                           | YH-8016:6    |                                                |                           |
|--------------------------------------------|--------------|------------------------------------------------|---------------------------|
|                                            | YH-8015:9    |                                                |                           |
|                                            | YH-8006IV:10 |                                                |                           |
| Suction Settings (Massage Mode)            | YH-8004:6    | 5                                              |                           |
|                                            | YH-8016:6    |                                                |                           |
|                                            | YH-8015:9    |                                                |                           |
|                                            | YH-8006IV:10 |                                                |                           |
| Cycle Speed (Expression Mode) (cycles/min) | Low          | YH-8004:41-69                                  | 24-60                     |
|                                            |              | YH-8016:41-69                                  |                           |
|                                            |              | YH-8015:36-60                                  |                           |
|                                            |              | YH-8006IV:N/A                                  |                           |
|                                            | High         | YH-8004:16-33                                  |                           |
|                                            |              | YH-8016:16-33                                  |                           |
|                                            |              | YH-8015:16-34                                  |                           |
|                                            |              | YH-8006IV:32-57                                |                           |
| Cycle Speed (Massage Mode)                 |              | YH-8004:94-113                                 | 70                        |
|                                            |              | YH-8016:94-113                                 |                           |
|                                            |              | YH-8015:80-122                                 |                           |
|                                            |              | YH-8006IV:71-100                               |                           |
| Backflow Protection                        |              | Yes                                            | Yes                       |
| Single or Double Pumping                   |              | YH-8004/ YH-8016: Single and Double<br>Pumping | Single and Double Pumping |
|                                            |              | YH-8006IV/YH-8015:Single Pumping<br>Only       |                           |
| Visual Indicator                           |              | LCD                                            | LCD                       |
| Pump Type                                  |              | Diaphragm                                      | Diaphragm                 |

The Youha electric breast pump has the same intended use but different technological characteristics compared to the predicate device. The differences in technological characteristics do not raise different questions of safety and effectiveness.

#### 8. Non-Clinical Performance Testing

Non-clinical tests were conducted to verify that the Youha electric breast pump met all design specifications and is substantially equivalent to the predicate. The test results demonstrated that the proposed device complies with the following standards:

- Risk Analysis developed in accordance with ISO 14971:2007.
- Electrical Safety Testing in accordance with IEC 60601-1:2005 (3m Edition) with US ● deviations per AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012.
- Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2014.
- . Safety Testing for use in the home in accordance with IEC 60601-1-11:2015.
- Biocompatibility Tests in accordance with ISO-10993, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2010) and irritation (ISO 10993-10:2010).
- Software Validation Software Life Cycle Processes in accordance with IEC 62304. ●

{6}------------------------------------------------

- . FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005).
Performance testing was conducted at minimum and maximum vacuum settings to determine the minimum and maximum vacuum levels of the pump as compared to its specifications. The specifications were met for vacuum level, cycle rate, and backflow protection. These results held under conditions of single and double pumping mode with varying power sources (e.g., AC/DC power vs. battery power).

#### Clinical Performance Data 9.

Not Applicable

#### 10. Conclusion

Based on the comparison and analysis above, the Youha electric breast pump is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K163136](https://fda-staging.innolitics.com/device/K163136)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).