FLEX Microwave Ablation System and Accessories

K163118 · Neuwave Medical, Inc. · NEY · Mar 9, 2017 · General, Plastic Surgery

Device Facts

Record IDK163118
Device NameFLEX Microwave Ablation System and Accessories
ApplicantNeuwave Medical, Inc.
Product CodeNEY · General, Plastic Surgery
Decision DateMar 9, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The NEUWAVE™ Flex Microwave Ablation System and Accessories are indicated for the ablation (coagulation) of soft tissue. The NEUWAVE™ Flex Microwave Ablation System is not indicated for use in cardiac procedures. The system is designed for facility use and should only be used under the orders of a physician.

Device Story

System performs soft tissue ablation via microwave energy; utilizes 2.45 GHz frequency-modulated microwave source. Input: user-selected power settings (40W–100W) and probe placement. Transformation: microwave energy delivered through flexible ablation probes (FLEX4/FLEX6) to target tissue, causing necrosis; CO2-based cooling system (Joule-Thompson effect) maintains probe temperature. Output: thermal ablation zone in target tissue. Used in clinical facilities (OR/clinic) by physicians. System includes Power Distribution Module (PDM) and support arm for probe stabilization. Healthcare providers monitor parameters (time, temperature, coolant, power) via interface; system automatically stops power if safety thresholds (e.g., reflected power, probe communication) are exceeded. Benefits: enables precise soft tissue ablation in varied anatomical access points (percutaneous, endoscopic) with reduced risk of probe movement.

Clinical Evidence

Bench testing only. Ex-vivo studies conducted in bovine liver, bovine lung, and porcine kidney tissue to characterize ablation zones and verify performance against predicate devices. Results confirmed that FLEX system ablation sizes are substantially equivalent to the predicate. Device complies with IEC 60601-1, IEC 60601-2-2, IEC 60601-2-6, IEC 60601-1-2, EN ISO 11607-1, and ISO 10993-1.

Technological Characteristics

2.45 GHz microwave energy source; CO2-based cooling (Joule-Thompson effect). Probes: French 6 diameter, 150 cm length, flexible cable with conical tip. PDM with magnetic retention and support arm. Materials: thermoformed plastic tray/Tyvek lid packaging. Sterilization: Ethylene Oxide. Connectivity: PDM-based monitoring of temperature, coolant, and power. Software: embedded control system for power delivery and safety monitoring (alarms for temperature, reflected power, probe authentication).

Indications for Use

Indicated for the ablation (coagulation) of soft tissue in percutaneous, open surgical, or lumen/scope-accessed (endoscope/laparoscopic) procedures. Not indicated for cardiac procedures. For use by physicians in facility settings.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is simple and recognizable, representing the department's mission to protect the health of all Americans. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 9, 2017 NeuWave Medical, Inc. Mr. Dan Kosednar Director of Regulatory Affairs 3529 Anderson Street Madison, Wisconsin 53704 Re: K163118 Trade/Device Name: NEUWAVE Flex Microwave Ablation System and Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: NEY Dated: February 17, 2017 Received: February 21, 2017 Dear Mr. Kosednar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K163118 Device Name NEUWAVE Flex Microwave Ablation System and Accessories Indications for Use (Describe) The NEUWAVE™ Flex Microwave Ablation System and Accessories are indicated for the ablation (coagulation) of soft tissue. The NEUWAVE™ Flex Microwave Ablation System is not indicated for use in cardiac procedures. The system is designed for facility use and should only be used under the orders of a physician. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------|------------------------------------------------------------------------------| | <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | Date: | 03/06/2017 | |--------------------|--------------------------------------------------------------------------------------------------------------------------| | Subject: | 510(k) Summary of Safety and Effectiveness Information for the<br>NEUWAVE Flex Microwave Ablation System and Accessories | | Company: | NeuWave Medical, Inc.<br>3529 Anderson Street<br>Madison, WI 53704 | | FDA Establishment# | 3008769756 | | Contact: | Dan Kosednar, Director of Regulatory Affairs<br>P - 608-512-1592<br>F - 608-512-1509 | | Proprietary: | NEUWAVE Flex Microwave Ablation System and Accessories | | Common: | System, Ablation, Microwave and Accessories | | Classification: | Surgical Devices, 73 NEY, 21 CFR 878.4440 | The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992. #### Predicate Devices The NEUWAVE Flex Microwave Ablation System and Accessories is substantially equivalent to the following currently marketed device: - . Certus 140 2.45 GHz Ablation System and Accessories – Class II – 21CFG878.4400 which has been the subject of a cleared 510(k) with the FDA log number K100744, K113237, K122217, K130399 and K160936. #### Indications For Use The NEUWAVE™ Flex Microwave Ablation System and Accessories are indicated for the ablation (coagulation) of soft tissue. The NEUWAVE™ Flex Microwave Ablation System is not indicated for use in cardiac procedures. The system is designed for facility use and should only be used under the orders of a physician. #### Device Description The design of the NEUWAVE Flex Microwave Ablation System (FLEX) is an evolution of the design of the legally marketed Certus 140 2.45GHz Ablation System (K160936). The FLEX is a fully featured soft tissue ablation system that uses small diameter flexible ablation probes, a single microwave source operating at 2.45 GHz, a CO2-based cooling system, a Power Distribution Module (PDM), and a support arm designed to hold the PDM in a user-selected position. Microwave energy is applied to the target tissue via a single flexible ablation probe, heating the tissue to the point of necrosis. The FLEX is designed to be used in Target Ablation involves placing a probe into a substantial target and then ablating for up to several minutes until the target tissue is necrotic. The FLEX FLEX 510(k) Submission Safety Summary Page 1 of 6 {4}------------------------------------------------ is designed for ablations in soft tissue in percutaneous (via an introducer) procedures, open surgical procedures, and procedures in which the target tissue is accessed via a lumen or scope such as an endoscope or laparoscopic port. Two microwave ablation probes are available for use with the NEUWAVE Flex Microwave Ablation System: the FLEX4 and the FLEX6. They are comprised of a conical tip on the end of a flexible cable and a connector assembly. The probe with the sharper tip (FLEX6) is designed for easier tissue penetration while the probe with the less sharp (FLEX4) tip designed for improved navigation. Both FLEX probe models are French gauge 6 (outer diameter of less than 2 mm) and 150 cm long. Both probe models have the same antenna design and ablation performance characteristics. Each probe contains three (3) temperature measurement sensors that help monitor performance and ensure patient and operator safety. The FLEX probe antenna was designed to produce an ablation zone substantially equivalent to the predicate Certus® probe but within a flexible probe shaft. Like the predicate Certus® probe, the FLEX probes are designed to produce ablations that encompass the tip of the probe while limiting the overall length of the ablation. Testing in ex-vivo liver, lung and kidney tissue confirm that the FLEX probes produce ablations that are substantially equivalent to the predicate probes. A PDM and Accessories Support Arm (The Arm) holds the PDM in a user-selected position. The Arm attaches to an imaging/procedure table via a rail mount. This allows the PDM and probe to move with the patient, greatly reducing the potential for patient injury due to accidental probe movement. The Arm is comprised of three jointed lengths with two tension knobs that allow the user to maneuver and fix the Arm at the user-selected position. The PDM mount on the ARM has a dovetail design with magnetic retention for quick and easy mounting. Additionally, the Arm has a grasping mechanism that can be used to hold a scope such as an endoscope (if used) in a fixed position for an extended period of time. A CO2 based cooling system ensures the non-active portion of the probe does not exceed temperature requirements. The system uses two (2) customer supplied E-sized CO2 cylinders. The system monitors the pressure of the tanks and heats the tanks to maintain the desired tank pressure. The FLEX System will select which cylinder to initially use based upon tank pressures. #### Comparison to Predicate The FLEX Microwave Ablation System is an evolution of the predicate Certus 140 2.45GHz Ablation System and Accessories. Below is a tabular comparison of the similarities and differences: {5}------------------------------------------------ K163118 | Feature/Specification | NEUWAVE Flex Microwave<br>Ablation System | Predicate Certus 140 Ablation<br>System and Accessories (K160936) | Comments/impact on<br>safety and<br>effectiveness | | |------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Indication for Use | The NEUWAVE Flex Microwave<br>Ablation System and Accessories<br>is indicated for the ablation<br>(coagulation) of soft tissue.<br>The NEUWAVE Flex is not<br>indicated for use in cardiac<br>procedures. | The NeuWave Medical Certus 140™<br>2.45 GHz Ablation System and<br>Accessories are indicated for the<br>ablation (coagulation) of soft tissue in<br>percutaneous, open surgical and in<br>conjunction with laparoscopic surgical<br>settings.<br>The Certus 140 is not indicated for use<br>in cardiac procedures. | The FLEX indications for<br>use are a subset of the<br>predicate indications. | | | Energy Source | 2.45 GHz Frequency Modulated<br>Source Microwave | 2.45 GHz Frequency Modulated Source<br>Microwave | No change from<br>predicate | | | Power Distribution<br>Module (PDM) | PDM mounts to the PDM and<br>Accessories Support Arm and<br>allows for the connection of a<br>single (1) FLEX ablation probe. | PDM mounts to the imaging table rail<br>and allows for the connection of up to<br>three (3) ablation probes | The PDM concept and<br>construct is similar to<br>the predicate. However,<br>the FLEX system only<br>supports the use of a<br>single ablation probe at<br>a time. | | | Probe applications | The FLEX ablation probes can be<br>used in multiple applications,<br>including percutaneous (via an<br>introducer), open surgical and<br>procedures in which the target<br>tissue is accessed via a lumen or<br>scope such as an endoscope or<br>coaxial introducer. | Percutaneous, open surgical and in<br>conjunction with laparoscopic. | The nature of the<br>flexible probe enables<br>the FLEX system to be<br>used in additional<br>applications in which the<br>target tissue is accessed<br>via a lumen or scope<br>such as an endoscope or<br>coaxial introducer. | | | Probe dimensions | French 6 diameter and 150cm in<br>length. | CertusLK, CertusLN and CertusPR probes<br>are 17 gauge and available in 15 and<br>20cm lengths.<br><br>CertusSR probe is 13 gauge and<br>available only in a 25cm length.<br><br>CertuSurgGT has a 17 gauge cannula<br>and is only available in one length. | The FLEX system uses a<br>flexible ablation probe<br>instead of the metal<br>cannula-based<br>percutaneous probes of<br>the predicate. | | | Feature/Specification | NEUWAVE Flex Microwave<br>Ablation System | Predicate Certus 140 Ablation<br>System and Accessories (K160936) | Comments/impact on<br>safety and<br>effectiveness | K163118 | | Generator Output<br>Power | The FLEX system allows an output<br>of 40W – 100W. The user controls<br>the power setting on the user<br>interface and can adjust in 5<br>Watts increments. | CertusLK, CertusLN and CertusSR Probes<br>are limited to 140W for a single probe,<br>95W if 2 probes are selected and 65W<br>if 3 probes are selected.<br><br>CertusPR Probes Maximum of 65W per<br>probe regardless of the number of<br>probes when used in Ablation Mode.<br>In Surgical Mode, the maximum power<br>for CertusPR probes is 95W if one or<br>two probes are used and 65W if three<br>probes are used. | The power levels<br>available on the FLEX are<br>a subset of the power<br>levels on the predicate. | | | | | For the CertuSurgGT, the maximum<br>power is 110W for a single probe, 95W<br>if two probes are used and 65W if 3<br>probes are used. | | | | Antenna Design | Coaxial Antenna Flex probe | Triaxial Antenna for CertusLK, CertusLN<br>and Certus SR<br><br>Modified Triaxial Antenna for CertusPR<br>and CertuSurgGT | The FLEX probe antenna<br>was designed to produce<br>an ablation zone<br>substantially equivalent<br>to the predicate CertusPR<br>probe but within a<br>flexible probe shaft.<br>Testing in ex-vivo liver,<br>lung and kidney tissue<br>confirm that the FLEX<br>probes produce<br>ablations that are<br>substantially equivalent<br>to the predicate probes. | | | Cooling Mechanism | CO2 cooling using the Joule-<br>Thompson effect of high pressure<br>CO2 gas | CO2 cooling using the Joule-Thompson<br>effect of high pressure CO2 gas | The same CO2 cooling<br>method is employed to<br>keep the non-active<br>portions of the probe at<br>acceptable<br>temperatures. | | | Tissu-Loc Feature | No Tissu-Loc feature | Tissu-Loc feature | The predicate employs a<br>Tissu-Loc feature to<br>adhere the probe to<br>tissue prior to delivering<br>microwave energy. This<br>feature is not available<br>on the FLEX system. | | | Feature/Specification | NEUWAVE Flex Microwave<br>Ablation System | Predicate Certus 140 Ablation<br>System and Accessories (K160936) | Comments/impact on<br>safety and<br>effectiveness | | | Sterilization/Packaging | | | | | | Probe Sterilization Method | Ethylene Oxide | Ethylene Oxide | This aspect of<br>system/probe design has<br>not been modified from<br>the predicate device. | | | Probe Packaging | Thermoformed plastic tray with<br>Tyvek Lid, with E-flute outer box. | Thermoformed plastic tray with Tyvek<br>Lid, with E-flute outer box. | The packaging used on<br>the FLEX probes uses the<br>same tray and lid as the<br>predicate. | | | Monitored<br>Parameters | | | | | | Elapsed Time<br>Probe Temperature<br>Coolant Level<br>Power in Watts<br>Coolant Pressure | Yes | Yes | This aspect of<br>system/probe design has<br>not been modified from<br>the predicate device. | | | Alarms | The following parameters are<br>monitored and will cause the FLEX<br>system to stop delivering power if<br>values are unacceptable:<br><br>Gas Coolant Level<br>Generator Temperature<br>Probe Temperature<br>Reflected Power Level<br>Probe Communication<br>Probe Authentication | The following parameters are<br>monitored and will cause the Certus<br>140 to stop delivering power if values<br>are unacceptable:<br><br>Gas Coolant Level<br>Generator Temperature<br>Probe Temperature<br>Reflected Power Level<br>Probe Communication<br>Probe Authentication | This aspect of<br>system/probe design has<br>not been modified from<br>the predicate device. | | {6}------------------------------------------------ K163118 {7}------------------------------------------------ K163118 ### Performance Data The NEUWAVE Flex Microwave Ablation System and Accessories has been designed to comply with the applicable portions of various International Standards, including: - IEC 60601-1:2005 + CORR 1:2006, CORR 2:2007 + Amendment 1:2012 ● - IEC 60601-2-2:2009 - IEC 60601-2-6:2012 - . IEC 60601-1-2:2007/AC:2010 - EN ISO 11607-1:2009 - ISO 10993-1: 2009 The NEUWAVE Flex Microwave Ablation System and Accessories and the predicate devices are substantially equivalent in design concepts, technologies and materials. The NEUWAVE Flex Microwave Ablation System and Accessories has been verified through rigorous testing that, in part, supports the compliance of NEUWAVE Flex Microwave Ablation System and Accessories to the standards listed above. The system passed all pre-determined acceptance criteria identified in the test plan. Ex-vivo testing in bovine liver and lung and porcine kidney was conducted to produce data for the Instructions for Use and to compare to the predicate devices. Ex-vivo ablation sizes for the FLEX system were deemed to be substantially equivalent to the predicate devices. FLEX 510(k) Submission Safety Summary Page 5 of 6 {8}------------------------------------------------ Verification and validation testing were completed in accordance with the company's Design Control process in compliance with 21 CFR Part 820.30, which included testing that fulfills the requirements of FDA "Guidance on Software Contained in Medical Devices". Potential risks were analyzed and satisfactorily mitigated in the device design.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...