Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series
Applicant
Grandway Technology (Shenzhen) Limited
Product Code
DXN · Cardiovascular
Decision Date
Jan 19, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure of an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the wrist.
Device Story
Digital Automatic Wrist Blood Pressure Monitor (SWBPM22 Series) uses oscillometric methodology to measure systolic/diastolic blood pressure and pulse rate. Device consists of inflatable wrist cuff, pressure sensor, electric pump, electric valve, and electronic control module. User wraps cuff around wrist; device automatically inflates/deflates cuff; sensor detects brachial artery blood movement; CPU converts signals into digital readings displayed on LCD panel. Used in clinical or home settings by medical professionals or patients. Provides objective blood pressure/pulse data to assist in monitoring cardiovascular health. Includes Irregular Heart Beat (IHB) detection and WHO classification indicators.
Clinical Evidence
Clinical study conducted per ISO 81060-2 with 100 patients (49 male, 51 female). Compared device measurements against standard auscultation method in the arm. Results demonstrated clinical accuracy equivalent to predicate device.
Technological Characteristics
Oscillometric non-invasive blood pressure measurement. Components: pressure sensor, electric valve, electric pump, CPU, PCB, ABS cabinet. Power: 2x1.5V AAA batteries. Standards: ISO 81060-1:2007, EN 1060-3:1997+A2:2009, IEC 60601-1:2012, IEC 60601-1-2:2014, FCC Part 15 Subpart B, IEC 81060-2:2013.
Indications for Use
Indicated for adult individuals for non-invasive measurement of systolic and diastolic blood pressure and pulse rate. Suitable for use by medical professionals or home users.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series (K142088)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. Inside the circle is an abstract image of an eagle with three faces in profile.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 19, 2017
Grandway Technology (Shenzhen) Limited Patrick Chow General Manager Block 7, Zhu Keng Industrial Zone, Ping Shan District, Shenzhen, 518118 CHINA
Re: K163113
Trade/Device Name: Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series, Models MD2200 and MD2210 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: December 16, 2016 Received: December 23, 2016
Dear Patrick Chow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M. A. Hillemann
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K163113
Device Name
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series, models MD2200 and MD2210
Indications for Use (Describe)
This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure of an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the wrist.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Re: K
### CHECK ONLY ONE
ত্রি 510(k) Summary. Attached is a summary of safety and effectiveness information upon which an equivalence determination could be based.
510(k) Statement. I certify that, in my capacity as (the position held in company by person required to submit the premarket notification, preferably the official correspondent) of (company name), I will make available all information included in this premarket notification on safety and effectiveness within 30 days of request by any person if the device described in the premarket notification submission is determined to be substantially equivalent. The information I agree to make available will be a duplicate of the premarket notification submission, including any adverse safety and effectiveness information, but excluding all patient identifiers, and trade secret and confidential commercial information, as defined in 21 CFR 20.61.
Oh ha-by
Patrick Chow, General Manager (00852)-2851-6789 02-Nov-2016
{4}------------------------------------------------
### 1. Submitter identification
| 510(k) Submitter | GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED |
|------------------|-------------------------------------------------------------------------------------------------------------------|
| Address | Block 7, Zhu Keng Industrial Zone, Ping Shan, Long Gang District, 518118,<br>Shenzhen, People's Republic of China |
| Phone Number | (00852)-2851-6789 |
| Fax Number | (00852)-2851-6278 |
| Contact Person | Mr. Patrick Chow |
| 01 | 02-Nov-2016 |
# 2. Device identification
| Trade Name | Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series<br>[Model No.: MD2200 and MD2210] |
|---------------------|-------------------------------------------------------------------------------------------------|
| Common Name | Non-invasive Blood Pressure Measurement System |
| Classification Name | Non-invasive Blood Pressure Measurement System<br>(CFR 870.1130, Class II, Product Code DXN) |
### 3. Predicate device
| Predicate Device | Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series |
|------------------|---------------------------------------------------------------|
| Manufacturer | GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED |
| 510(k) Number | K142088 |
#### Device Description 4.
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is a non-invasive blood pressure measurement system for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate of an individual in each measurement and then display the readings on a digital panel.
The device utilizes the oscillometric methodology, in which an inflatable cuff is wrapped around the wrist of an individual, for blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movement into a digital reading.
The table below illustrate the feature presence in Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series.
| Model | BP<br>Measurement | Pulse Rate<br>Measurement | WHO<br>Classification | IHB | LCD Type | LCD Size<br>(mm) | User x<br>Memory |
|--------|-------------------|---------------------------|-----------------------|-----|---------------------|------------------|------------------|
| MD2200 | ✓ | ✓ | ✓ | ✓ | Positive reflective | 42 x 48 | 2 x 120 |
| MD2210 | ✓ | ✓ | ✓ | ✓ | Positive reflective | 62 x 59 | 2 x 120 |
{5}------------------------------------------------
| GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED | | 510(k) Summary |
|----------------------------------------|--|----------------|
| | | |
# 5. Indication for use
This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure on an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the wrist.
# 6. Comparison of technological characteristics between new Device and predicate Devices
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is compared to the predicate device, Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series (K142088) in the device comparison table below.
| Item | Predicate Device | New device | Comment |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Indication for<br>Use | This device is for use by medical<br>professional or home users. It is<br>intended to measure the systolic and<br>diastolic blood pressure on an adult<br>individual by using a non-invasive<br>technique, in which an inflatable cuff<br>is wrapped around the wrist. | This device is for use by medical<br>professional or home users. It is<br>intended to measure the systolic and<br>diastolic blood pressure on an adult<br>individual by using a non-invasive<br>technique, in which an inflatable cuff<br>is wrapped around the wrist. | Identical |
| Measurement<br>Method | Non-invasive, Oscillometric | Non-invasive, Oscillometric | Identical |
| IHB Detection | Yes | Yes | Identical |
| Patient<br>Population | Adult | Adult | Identical |
| BP<br>Measurement<br>Range | Cuff Pressure: 0 - 300 mmHg<br>Systolic Pressure: 50 - 250 mmHg<br>Diastolic Pressure: 30 - 200 mmHg | Cuff Pressure: 0 - 300 mmHg<br>Systolic Pressure: 50 - 250 mmHg<br>Diastolic Pressure: 30 - 200 mmHg | Identical |
| Number of User | 2 independent users | 2 independent users | Identical |
| Memory Space | 2 users × 120 memory space | 2 users × 120 memory space | Identical |
| Resolution of<br>Measurement | Blood Pressure: 1 mmHg or 0.1kPa<br>Pulse Rate: 1 beat/ min | Blood Pressure: 1 mmHg or 0.1kPa<br>Pulse Rate: 1 beat/ min | Identical |
| Blood Pressure<br>Measurement<br>Accuracy | ± 3 mmHg | ± 3 mmHg | Identical |
| Pulse Rate<br>Measurement<br>Range | 30 - 180 beats/min | 40 - 180 beats/min | Similar |
| Pulse Rate<br>Measurement<br>Accuracy | ± 5 % of the reading | ± 5 % of the reading | Identical |
| Display Type | LCD | LCD | Identical |
| Power Source | 2 × 1.5 V AAA-batteries | 2 × 1.5 V AAA-batteries | Identical |
| Pressurization<br>Mode | Automatic Inflation | Automatic Inflation | Identical |
| Deflation Mode | Automatic Exhaust/ Deflation | Automatic Exhaust/ Deflation | Identical |
| Operating<br>Condition | Temperature: +5 to +40 °C<br>Humidity: 15 to 93 % R.H. max<br>Atmospheric Pressure: 700-1060 hPa | Temperature: +5 to +40 °C<br>Humidity: 15 to 93 % R.H. max<br>Atmospheric Pressure: 700-1060 hPa | Identical |
{6}------------------------------------------------
### GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED
| Item | Predicate Device | New device | Comment |
|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Storage and<br>Transportation<br>Condition | Temperature: -25 to +70 °C<br>Humidity: up to 93% R.H. max<br>Atmospheric Pressure: 700-1060 hPa | Temperature: -25 to +70 °C<br>Humidity: up to 93% R.H. max<br>Atmospheric Pressure: 700-1060 hPa | Identical |
| Material | Resistances, capacitance, transistors,<br>amplifiers, pressure sensor, CPU,<br>PCB, cuff ABS button, ABS cabinet,<br>batteries and packaging | Resistances, capacitance, transistors,<br>amplifiers, pressure sensor, CPU,<br>PCB, cuff ABS button, ABS cabinet,<br>batteries and packaging | Identical |
| Compatibility<br>with<br>Environment<br>and Other<br>Devices | No influence with environment and<br>other device | No influence with environment and<br>other device | Identical |
| Applicable<br>Standard | - EN 1060-1:1995+A2:2009<br>- EN 1060-3:1997+A2:2009<br>- IEC 60601-1:2012<br>- IEC 60601-1-2:2007<br>- FCC Part 15 Subpart B<br>- IEC 81060-2:2009 | - ISO 81060-1:2007<br>- EN 1060-3:1997+A2:2009<br>- IEC 60601-1:2012<br>- IEC 60601-1-2:2014<br>- FCC Part 15 Subpart B<br>- IEC 81060-2:2013 | Equivalent |
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is a non-invasive measuring device and utilizes the oscillometric methodology to measure the blood pressure reading. The key components of device are: a pressure sensor, an electric valve and an electronic control module together with an electric pump. The electric pump inflate (and deflate) the inflatable cuff automatically according to our designed architecture. The predicate device adopts exactly same methodology and key components for measuring blood pressure.
## 7. Clinical and Non-clinical Tests
#### Clinical Test Summary
Testing to insure clinical accuracy of the device in accordance with ISO 81060-2 as documented in Clinical Test report.
One hundred patients (49 males and 51 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left arm. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in . 209008 051
#### Non-Clinical Test Summary
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series has performed several non-clinical tests to show that all requirement specifications and standard requirements are met. The tests includes the follows:
- 4 ISO 81060-1:2007
- EN 1060-3:1997+A2:2009 や
- や IEC 60601-1:2012
- � IEC 60601-1-2:2014
{7}------------------------------------------------
### GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED
#### ◆ FCC Part 15 Subpart B
All of the clinical and non-clinical testing performed on Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series are same as the predicate device. Therefore, no bench test is conducted to show the performance of Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is equivalent to the predicate device.
### 8. Conclusion
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series has the same intended use and same technological characteristics as the predicate device, Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series (K142088). Moreover both clinical and non-clinical testing has demonstrated that no differences in the technological characteristics and questioning on safety or effectiveness to be raised. Thus, Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.