K163107 · Ansell Healthcare Products, LLC · HIS · Jan 23, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K163107
Device Name
Lifestyles Zero Lubricated Latex Male Condom
Applicant
Ansell Healthcare Products, LLC
Product Code
HIS · Obstetrics/Gynecology
Decision Date
Jan 23, 2017
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
The Lifestyles Zero Lubricated Latex Male Condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).
Device Story
Single-use, non-sterile, lubricated natural rubber latex male condom; straight-walled with reservoir tip. Intended for home use by consumers. Provides physical barrier to fluids to reduce risk of pregnancy and STIs. Dimensions: 180+/-10 mm length, 52+/-2 mm width, 0.045+/-0.005 mm thickness. Conforms to ASTM D3492:2015. No electronic or software components.
Clinical Evidence
Bench testing only. Device evaluated for biocompatibility per ISO 10993-5, ISO 10993-10, and ISO 10993-11. Results: non-toxic, non-sensitizing, and non-irritating (vaginal and penile). Mechanical performance met all requirements of ASTM D3492-15 (air burst pressure/volume, leakage AQL, package integrity).
Technological Characteristics
Material: Natural rubber latex. Design: Straight-walled, lubricated, reservoir tip. Dimensions: 180+/-10 mm length, 52+/-2 mm width, 0.045+/-0.005 mm thickness. Standards: ASTM D3492:2015, ISO 4074:2015. Lubricant: Silicone gel (400-600 mg). Sterilization: Non-sterile. Shelf life: 5 years.
Indications for Use
Indicated for contraception and reduction of risk of pregnancy and STIs in individuals using male condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
Latex Condom with Silicone Oil Lubricant (K941258)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, composed of three overlapping profiles facing to the right. The profiles are rendered in black. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 23, 2017
Ansell Healthcare Products, LLC Robert Mahler Director, Regulatory Affairs 111 Wood Avenue South, Suite 210 Iselin. NJ 08830
Re: K163107 Trade/Device Name: Lifestyles Zero Lubricated Latex Condom Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: HIS Dated: November 4, 2016 Received: November 7, 2016
Dear Robert Mahler,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163107
Device Name
Lifestyles Zero Lubricated Latex Male Condom
Indications for Use (Describe)
The Lifestyles Zero Lubricated Latex Male Condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510k Summary - K163107
#### Submitter:
Ansell Healthcare Products LLC. 111 Wood Avenue South, Suite 210 Iselin, NJ 08830, USA
## Contact Person:
Robert S. Mahler Director Regulatory Affairs for the Americas Phone: 732-345-2174 Rob.mahler@ansell.com
#### Date Prepared
January 23, 2017
#### Device Name:
| Proprietary Name: | Lifestyles Zero Lubricated Latex Male Condom |
|----------------------------|----------------------------------------------|
| Common Name: | Condom, Natural Rubber Latex |
| Classification Regulation: | 21 CFR 884.5300 |
| Device Class: | Class II |
| Product Code: | HIS; condom |
| Classification Panel: | Obstetrics/Gynecology |
| Classification Name: | Condom |
#### Reason for 510(k) Submission:
New product.
#### Predicate Device:
K941258 – Latex Condom with Silicone Oil Lubricant. The predicate device has not been subject to a design related recall.
{4}------------------------------------------------
## Device Description:
This condom is a single use and non-sterile device intended for in home use by a consumer. The condom is made of a natural rubber latex sheath, which completely covers the penis with closely fitted membrane. The condom provides a physical barrier to fluids which helps reduce the risk of pregnancy and the transmission of sexually-transmitted infections. The device is a straight walled, cylindrical, lubricated condom with a reservoir tip. Condom dimensions are length 180+/- 10 mm, width 52±2mm, and thickness 0.045+/-0.005 mm. The condom is designed to conform to the requirements of ASTM D3492:2015 - Standard Specification for Rubber Contraceptives (Male Condoms).
## Indications for Use:
The Lifestyles Zero Lubricated Latex Male Condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).
## Technological Characteristics:
The Lifestyles Zero Lubricated Latex Male Condom has the following technological characteristics as compared to ASTM or equivalent standards
| Technological<br>Characteristic | Standard/Test/FDA<br>Guidance | Specification |
|---------------------------------|-------------------------------|---------------------------------------------------------|
| Dimensions | ASTM D3492-15 | |
| ■ Length | 170mm to 210 mm | $180 +/- 10$ mm |
| ■ Width | Maximum 57 mm | $52\pm2$ mm |
| ■ Thickness | Minimum 0.03 mm | $0.045+/-0.005$ mm |
| Burst Properties | ASTM D3492-15 | |
| ■ Air Burst Test<br>Pressure | Minimum 18L | Meets ASTM D3492-15 for<br>Air Burst test pressure |
| ■ Air Burst Test<br>Volume | Minimum 1 kPa | Meets ASTM D3492-15 for<br>Air Burst test volume |
| Leakage AQL | ASTM D3492-15 | Meets ASTM D3492-15 for<br>Leakage AQL of 0.25 |
| Package integrity AQL | ASTM D3492-15 | Meets ASTM D3492-15 for<br>Package Integrity AQL of 2.5 |
| Lubricant Quantity | Not defined | 400 – 600 mg |
{5}------------------------------------------------
# Substantial Equivalence:
| | Predicate Device | Subject Device | Substantial<br>Equivalence |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| Trade Name | Latex Condom with silicone oil lubricant | Lifestyles ZERO | Not applicable |
| 510(k) Number | K941258 | K163107 | Not applicable |
| Submitter | Suretex Ltd. | Ansell Healthcare Products LLC | Yes |
| Product Code | HIS | HIS | Yes |
| Regulation<br>Number | 21 CFR 884.5300 | 21 CFR 884.5300 | Yes |
| Regulation<br>Name | Condom | Condom | Yes |
| Indications for<br>Use | This device is a latex condom indicated for contraception and as an aid to prevention of sexually-transmitted diseases (STDs) | The Lifestyles Zero Lubricated Condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs) | Yes; minor difference in language of indications statements does not alter intended use |
| Prescription<br>or Over-The<br>Counter-Use | Over-The-Counter-Use | Over-The-Counter-Use | Yes |
| Materials | Natural rubber latex | Natural rubber latex | Yes |
| Form | Straight wall, nipple end | Straight wall, nipple end | Yes |
| Lubricant<br>Coating | Silicone gel | Silicone gel | Yes |
| Design | Single use | Single use | Yes |
| Dimensions | ASTM D3492-15 | ASTM D3492-15 | Yes |
| Length | $180 +/- 10 mm$ | $180 +/- 10 mm$ | Yes |
| Width | $52\pm2 mm$ | $52\pm2 mm$ | Yes |
| Thickness | $0.060-080$ | $0.045+/-0.005 mm$ | Different |
| Burst Properties | ASTM D3492-15 | ASTM D3492-15 | Yes |
| | Predicate Device | Subject Device | Substantial<br>Equivalence |
| Air Burst Test<br>Pressure | Minimum 18L | Minimum 18L | Yes |
| Air Burst Test<br>Volume | Minimum 1 kPa | Minimum 1 kPa | Yes |
| Leakage AQL | ASTM D3492-15 | ASTM D3492-15 | Yes |
| Package<br>integrity AQL | ASTM D3492-15 | ASTM D3492-15 | Yes |
| Lubricant<br>Quantity | 400-600 mg | 400-600 mg | Yes |
| Sterilization | Non-Sterile | Non-Sterile | Yes |
| Shelf Life | 5 years | 5 years | Yes |
| Biocompatibility | "Under the conditions of the<br>study, not an irritant" and<br>"Under the conditions of the<br>study, not a sensitizer" | "Under the conditions of the<br>study, not an irritant" and<br>"Under the conditions of the<br>study, not a sensitizer" | Yes |
{6}------------------------------------------------
The basic technological characteristics of the Lifestyles Zero Lubricated Latex Male Condom are similar to the predicate device. The Lifestyles Zero Lubricated Latex Male Condom and the predicate are made of natural rubber latex, both are straight walled, nipple ended, lubricated condom with an integral ring at the open end. The condom designs conform to the requirements of ASTM D3492:2015 - Standard Specification for Rubber Contraceptives (Male Condoms) and ISO 4074:2015 Natural rubber latex male condoms- Requirements and test methods. The only difference between the subject device and the predicate is the thickness of the condom: 0.040-0.050 mm for this device.
This difference in the technological characteristics does not raise different questions of safety or efficacy in the subject device.
## Performance Data:
The subject condom has been tested in conformance with the FDA guidance document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process" Guidance for Industry and Food and Drug Administration Staff," issued June 16, 2016. The subject device is indicated for contact with a skin and/or mucosal membrane surfaces for 24 hours or less. Therefore, the device was evaluated for cytotoxicity, irritation, sensitization and acute
{7}------------------------------------------------
systemic toxicity as described below. The condom was demonstrated to be not toxic (local or systemic), sensitizing, or locally irritating.
## Cytotoxicity
Lifestyles® Zero® Lubricated Latex Condom was tested per ISO 10993-5 Biological Evaluation of Medical Devices - Part 5. Tests for In Vitro Cytotoxicity
Cytotoxicity test were conducted using the elution method Utilizing serial dilutions of device extract ranged from undiluted to a ratio of 1:32. Severe cytotoxic reactivity was observed for extracts at undiluted and dilutions of 1:2, whereas no cytotoxic reactivity was found for any extracts at lower dilutions.
Such findings are not unexpected for synthetic and natural rubber latexes, and do not necessarily translate to local or systemic toxicity in clinical use. This was the case for the subject condom, where cytotoxic reactions were observed in cell culture assays but there was no evidence of systemic toxicity from the animal studies conducted (as described below).
## Irritation and sensitization
Lifestyles® Zero® Lubricated Latex Condom was testing per ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10, Tests for Irritation and Sensitization
Based on the results obtained in the study, it is concluded that the Lifestyles® Zero® Lubricated Male Latex Condom is not considered an irritant or a sensitizer.
## Vaginal irritation:
The test article was evaluated for vaginal irritation per ISO 10993-10:2010. The extracts of the test article did not show any signs of vaginal irritation and the average mucosal vaginal irritation scores were within acceptable limits. Test data demonstrates that the device is nonirritating.
## Penile irritation:
The test article was evaluated for penile irritation test per ISO 10993-10:2010. No signs of penile irritation were evident macroscopically or microscopically. The penile irritation indices obtained were within the acceptable limit. Test data demonstrates that the device is non-irritating.
## Sensitization:
The test article was assessed for sensitization using guinea pig maximization test per ISO 10993-10:2010. Under the conditions of the study, the device was demonstrated to not be a sensitizer.
{8}------------------------------------------------
## Acute systemic toxicity
Lifestyles® Zero® Lubricated Latex Condom conforms to the requirements of ISO 10993-11 Biological Evaluation of Medical Devices - Part 11, Tests for Systemic Toxicity.
The test article was evaluated for Acute Systemic Toxicity test. There was no mortality or evidence of systemic toxicity resulted from the test article extract. Test data showed that the test article extract did not show any systemic toxicity.
# Physical Testing:
The Lifestyles® Zero® Lubricated Latex Condom was tested and met all the requirements of ASTM D3492-15 Standard Specification for Rubber Contraceptives (Male Condoms).
## Shelf Life Testing:
The Lifestyles® Zero® Lubricated Latex Condom has a five-year shelf life based on the results of accelerated stability evaluations conducted as required in 21 CFR 801.435. Five-year expiration dating will be verified with ongoing real time stability evaluations as required in 21 CFR 801.435.
# Conclusion:
The subject and predicate devices have the same intended use and similar technological characteristics. The subject device is different from the predicate device in that it is thinner. However, this difference does not raise different questions of safety or effectiveness. The difference was assessed through biocompatibility testing and mechanical performance testing. Performance data demonstrated that the subject device is as safe and effective as the predicate. Therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.