K163106 · Medos International SARL · GEI · Dec 1, 2016 · General, Plastic Surgery
Device Facts
Record ID
K163106
Device Name
CODMAN Integrated Bipolar Cord and Tubing Set
Applicant
Medos International SARL
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 1, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CODMAN Integrated Bipolar Cord and Tubing Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation.
Device Story
Device consists of integrated bipolar cord and irrigation tubing set; includes inlet spike, on/off clamp, silicone tubing, male luer connector, plugs, and female socket connector. Used in clinical settings by surgeons/clinicians to connect bipolar forceps to electrosurgical generators and irrigation modules. Device delivers electrical energy for coagulation and fluid for irrigation simultaneously during surgery. Single-use, sterile, nonpyrogenic fluid pathway. Benefits include streamlined surgical setup and consistent delivery of irrigation and power to specialized forceps.
Clinical Evidence
Bench testing only. No clinical studies performed. Testing included electrical functionality (continuity, resistance), irrigation leak testing, mating part compatibility, pull testing, simulated use, electrical safety/EMC (IEC 60601-1, 60601-2-2, 60601-1-2), sterilization validation (ISO 11135), shelf-life (accelerated aging), and biocompatibility (ISO 10993). All tests passed.
Technological Characteristics
Integrated bipolar cord and irrigation tubing. Materials: ABS, PVC, silicone, polyethylene, nickel/tin-plated brass, polyacetal. Energy: High-frequency electrosurgical. Connectivity: Wired connection to generator/irrigation module. Sterilization: Ethylene oxide (SAL 10^-6). Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, ISO 10993, ISO 11135.
Indications for Use
Indicated for providing simultaneous irrigation and energy to bipolar forceps designed for irrigation. Intended for use with Codman/Malis CMC-II and CMC III I.E.C. Irrigation Modules and Codman/Malis Bipolar Coagulators.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Codman Integrated Irrigation Tubing and Bipolar Cord Set (K052449)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 1, 2016
Medos International SARL Ms. Megan Palumbo Senior Regulatory Affairs Specialist Chemin-blanc 38 2400 Le Locle, Switzerland
Re: K163106
Trade/Device Name: Codman Integrated Bipolar Cord and Tubing Set Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: November 4, 2016 Received: November 7, 2016
Dear Ms. Palumbo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
Sincerely,
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Jennifer R. Stevenson -A For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K163106
#### Device Name
CODMAN Integrated Bipolar Cord and Tubing Set
Indications for Use (Describe)
The CODMAN Integrated Bipolar Cord and Tubing Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 – 510(k) Summary
I. Submitter Codman & Shurtleff, Inc. 325 Paramount Drive Raynham, MA 02767
Establishment Registration Number: 1226348
On behalf of: Medos International SARL Chemin-Blanc 38 2400 LeLocle, Switzerland
Establishment Registration Number: 3008114965
Contact: Megan Palumbo Phone: (508) 828-3571 Fax: (508) 977-6979
Date of Submission: November 4, 2016
# II. Device
| II. Device | Device Proprietary Name | Codman Integrated Bipolar Cord and Tubing Set |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| | Common Name | Irrigation Tubing and Bipolar Cord |
| | Classification Name | Electrosurgical, Cutting & Coagulation & Accessories<br>(21 CFR 878.4400) |
| | Regulatory Classification | II |
| | Product Code | GEI |
| III. Predicate<br>Device | The predicate device for this submission is the Codman Integrated Irrigation<br>Tubing and Bipolar Cord Set (K052449), which was cleared on October 17,<br>2005. | |
| IV. Device<br>Description | The CODMAN Integrated Bipolar Cord and Tubing Set consists of an inlet<br>spike with protective cap, an on/off clamp, silicone tubing and fittings, a male<br>luer connector, two plugs, and a female socket connector. The set is packaged | |
V. Indications The CODMAN Integrated Bipolar Cord and Tubing Sets are intended to for Use provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation.
sterile for single use with a nonpyrogenic fluid pathway.
{4}------------------------------------------------
- the same intended use, ●
- use the same fundamental scientific technology, .
- incorporates the same basic cord and tube set design, ●
- . incorporates similar materials,
- packaged and sterilized using the same packaging design, materials, . and processes.
The minor differences between the predicate and subject device are:
- a slight difference in device design, .
- minor formulation changes to the existing materials (material types are . remaining the same), and
- . change in shelf life.
The table below details the comparison of the predicate and subject devices.
| Comparison of the Predicate and Subject Device | | |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| | Predicate Device:<br>Codman Integrated Irrigation Tubing<br>and Bipolar Cord Set (K052449) | Subject Device:<br>Codman Integrated Bipolar Cord and Tubing<br>Set (This Submission) |
| FDA Regulatory Information | | |
| Manufacturer | Codman & Shurtleff, Inc. | Same as predicate |
| FDA Product<br>Code | GEI | Same as predicate |
| Classification | Class II - 21 CFR 878.4400 | Same as predicate |
| Classification<br>Name | Electrosurgical, Cutting & Coagulation &<br>Accessories | Same as predicate |
| Indications for Use | | |
| Indications for<br>Use | The CODMAN Integrated Bipolar Cord<br>and Tubing Sets are intended to provide<br>irrigation and energy simultaneously to<br>bipolar forceps specifically designed for<br>irrigation.<br>[They are intended for use with the<br>Codman/Malis CMC-II and the<br>Codman/Malis CMC III I.E.C. Irrigation<br>Modules and the Codman/Malis Bipolar<br>Coagulators.] | Same as predicate |
| Device Design | | |
| Single Use | Yes | Same as predicate |
| Device Design | Integrated disposable bipolar cord with an<br>irrigation tube. | Same as predicate |
| | Comprised of irrigation inlet spike with<br>protective cap, male luer connector, female | Same as predicate |
| Comparison of the Predicate and Subject Device | | |
| | Predicate Device: | Subject Device: |
| | Codman Integrated Irrigation Tubing<br>and Bipolar Cord Set (K052449) | Codman Integrated Bipolar Cord and Tubing<br>Set (This Submission) |
| Devices for which<br>this Accessory is<br>to be used with | socket connector, on/off clamp, silicone<br>tubing and fittings, and two bipolar<br>generator plug components.<br>They are intended for use with the<br>Codman/Malis CMC-II and the<br>Codman/Malis CMC III I.E.C. Irrigation<br>Modules and the Codman/Malis Bipolar<br>Coagulators. | Same as predicate |
| Device Color | <b>Bipolar Cord</b><br>Transparent tubing surrounding copper<br>cord with white plugs and socket<br><br><b>Irrigation Tubing</b><br>Transparent tubing and connectors with<br>white inlet spike | Same as predicate |
| Device Materials | | |
| Spike | ABS | ABS |
| PVC Tubing | PVC | PVC |
| Tubing Adaptor | PVC | Gravity Feed:<br>PVC<br><br>Rotary Pump:<br>ABS |
| Silicone Tubing | Silicone | Silicone |
| Extruded PVC/<br>Copper Cable | PVC and Copper | PVC and Copper |
| Male Slip Luer | PVC | PVC |
| Bushing | PVC | PVC |
| Spike Cap | Polyethylene | Polyethylene |
| Tubing Clamp | Polyethylene, Polysterene, Polypropylene | Celanese and Polyacetal |
| Male Slip Luer<br>Cap | Polyethylene | Polyethylene |
| Banana Plug Pin | nickel plate brass | nickel plate brass |
| Banana Plug<br>Overmold | PVC | PVC |
| Socket | nickel plated brass | tin plated brass |
| Socket Overmold | PVC | PVC |
| Other | | |
| Electrical Safety<br>Testing | Testing in accordance with:<br>• ANSI/AAMI HF18:2001 | Testing in accordance with<br>• IEC 60601-1<br>• IEC 60601-1-2<br>• IEC 60601-2-2 |
| Packaging<br>Configuration<br>(Sterile Barrier) | Peel Pouch | Same as predicate |
| Comparison of the Predicate and Subject Device | | |
| | Predicate Device:<br>Codman Integrated Irrigation Tubing<br>and Bipolar Cord Set (K052449) | Subject Device:<br>Codman Integrated Bipolar Cord and Tubing<br>Set (This Submission) |
| Packaging<br>Materials<br>(Sterile Barrier) | Transparent bottom stock formable film-<br>thermoform and 1073B Tyvek top stock<br>thermoform | Same as predicate |
| Packaging Unit | 10 units per box | Same as predicate |
| Sterility<br>Assurance Level<br>(SAL) | 10-6 | Same as predicate |
| Sterilization<br>Method | Ethylene oxide | Same as predicate |
| Shelf Life | 5 years | 1 year |
{5}------------------------------------------------
{6}------------------------------------------------
# VII. Performance Data
The following performance testing has been conducted in support of the substantial equivalence determination. All testing was performed on final sterile devices unless otherwise specified.
Please see the Summary of Testing table below for a summary of testing and testing results. Additional information regarding the Bench, Electrical Safety, Sterilization, Shelf Life and Biocompatibility Testing follows.
| Summary of Testing | |
|------------------------------------------------------------|--------|
| Test | Result |
| Bench Testing | |
| Electrical Functionality Test - Continuity and Resistance | Pass |
| Irrigation Functionality Test - Leak (Dry) | Pass |
| Mating Part - Part 1: Pins | Pass |
| Mating Part - Part 2: Socket | Pass |
| Mating Part - Part 3: Luer | Pass |
| Mating Part - Part 4: Spike | Pass |
| Pull Testing | Pass |
| Irrigation Simulated Use Test | Pass |
| Simulated Use Pull Test | Pass |
| Electrical Safety and EMC per IEC 60601 | |
| High Frequency Leakage Current - 60601-1 and 60601-2-2 | Pass |
| High Frequency Dielectric Strength - 60601-1 and 60601-2-2 | Pass |
| Main Frequency Dielectric Strength - 60601-1 and 60601-2-2 | Pass |
| Cable Flexing - 60601-1 and 60601-2-2 | Pass |
| Electromagnetic Compatibility - 60601-1-2 | Pass |
| Sterilization | |
| Bioburden Validation | Pass |
| Bioburden Test | Pass |
| EO/ECH Residuals | Pass |
| LAL Validation | Pass |
{7}------------------------------------------------
| Shelf Life - Functionality Testing Complete after One Year Accelerated Aging | |
|------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Electrical Functionality Test - Continuity and Resistance | Pass |
| Irrigation Functionality Test - Leak (Dry) | Pass |
| Mating Part - Part 1: Pins | Pass |
| Mating Part - Part 2: Socket | Pass |
| Mating Part - Part 3: Luer | Pass |
| Mating Part - Part 4: Spike | Pass |
| Pull Testing | Pass |
| Biocompatibility | |
| Cytotoxicity - ISO MEM Elution | Pass |
| Delayed Type Sensitization - ISO Maximization in Guinea<br>Pigs | Pass |
| Intracutaneous Reactivity - ISO Intracutaneous Study in<br>Rabbits | Pass |
| Acute Systemic Toxicity - ISO Systemic Toxicity in Mice | Pass |
| Pyrogenicity Testing - USP Rabbit Pyrogen Study, Material<br>Mediated | Pass |
| Hemolysis - ASTM F756 (Extract only) | Pass |
| Chemical Characterization of Extractables (ISO 10993-18) | Low level of extractables |
| Toxicology Risk Assessment | Extractables do not represent<br>a significant toxicological risk<br>to patients |
# Bench Testing
Bench testing data demonstrated that the Codman Integrated Bipolar Cord and Tubing Set performed according to its description and intended use, and established the performance characteristics of this device. The Codman Integrated Bipolar Cord and Tubing Set passed each bench test conducted, as compared to the device characteristics and performance specifications of the predicate device. Clinical testing was not required to establish substantial equivalence.
Results of verification and validation testing conducted on Codman Integrated Bipolar Cord and Tubing Set demonstrated that the proposed device performed as designed, is suitable for the intended use, and is substantially equivalent to the predicate device.
Performance bench testing was conducted in alignment with FDA's guidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016 and internal requirements.
# Electrical Safety and Electromagnetic Compatibility Testing
Electrical safety testing was conducted in accordance with IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance and IEC 60601-2-2:2009/C1:2014 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories. Electromagnetic Compatibility Testing
{8}------------------------------------------------
(EMC) was conducted in accordance with IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests. The Codman Integrated Bipolar Cord and Tubing Set passed all electrical safety and EMC testing.
### Sterilization
The Codman Integrated Bipolar Cord and Tubing Sets are sterilized via ethylene oxide. The Cord and Tubing Sets have been validated to ensure a sterility assurance level (SAL) of 10t in accordance with ISO 11135:2014 Sterilization of health care products - Ethylene Oxide - Requirements for the Development, Validation and Routine Control of a Sterilization Process for Medical Devices and ISO 11737-1:2006, "Sterilization of medical devices -Microbiological methods - Part 1: Determination of a population of microorganisms on products." Sterilization testing also demonstrated that ethylene oxide residuals can be reduced to an acceptable level in accordance with ISO 10993-7:2008 Biological Evaluation of Medical Devices - Ethylene Oxide Sterilization Residuals and the proposed device can be successfully adopted into Codman's existing sterilization cycle.
#### Shelf-Life Testing
Shelf-life testing was conducted in accordance with FDA's guidance document Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016 and internal requirements. The Codman Integrated Bipolar Cord and Tubing Set was subjected to accelerated aging. The aging studies established that the device and packaging remain functional and maintain sterility for 1 year.
# Biocompatibility Testing
Biocompatibility with ISO 10993-1:2009/AC:2010, Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process, and FDA's guidance documents, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued June 16, 2016 and Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016.
#### Animal Studies
No animal studies were performed as appropriate verification and validation of the modified device was achieved from the results of the bench testing, biocompatibility evaluation, and electrical/safety testing.
#### Clinical Studies
No clinical studies were performed as appropriate verification and validation of the modified device was achieved from the results of the bench testing.
{9}------------------------------------------------
biocompatibility evaluation, and electrical/safety testing.
VIII. Based upon the same indications for use, intended use, fundamental scientific Conclusion technology, material types, comparison to the predicate device, and testing conducted, it is concluded that the subject device, the Codman Bipolar Cord and Tubing Set, is substantially equivalent to the predicate device, the Codman Integrated Irrigation Tubing and Bipolar Cord Set (K052449), and therefore does not raise any new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.