ReliaCatch Specimen Retrieval Bag 10mm; ReliaCatch Specimen Retrieval Bag 12mm

K163102 · Covidien · GCJ · Nov 30, 2016 · Gastroenterology, Urology

Device Facts

Record IDK163102
Device NameReliaCatch Specimen Retrieval Bag 10mm; ReliaCatch Specimen Retrieval Bag 12mm
ApplicantCovidien
Product CodeGCJ · Gastroenterology, Urology
Decision DateNov 30, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The specimen retrieval bag has application for safe, convenient removal of tissue specimens such as the appendix, an ectopic pregnancy, gallbladder, gallstones, lymph nodes, ovaries, small sections of bowl and other tissue structures.

Device Story

ReliaCatch is a manual surgical instrument for tissue specimen retrieval during endoscopic procedures. Device consists of a long cylindrical tube, a handle, and a rip-stop nylon bag. Unlike the predicate, the bag is packaged outside the tube and must be retracted into the tube before trocar insertion. The device uses a suture loop mechanism to cinch the bag, whereas the predicate used a ring and metal spring. The bag is mounted on 'forks' that are removed prior to cinching and specimen extraction. Operated by surgeons in an OR setting, the device facilitates safe removal of tissue through trocar sleeves. The design change to a suture loop and fork-based detachment allows for complete removal of the retrieval mechanism from the bag before extraction, potentially simplifying the procedure and improving specimen containment.

Clinical Evidence

Bench testing only. Performance testing included leak rate evaluation, trocar insertion/removal forces, pouch deployment/retraction/cinch forces, pouch burst pressure, and human factors/usability testing. All tests met objective acceptance criteria.

Technological Characteristics

Components comply with ISO 10993-1. Features a long cylindrical tube, handle, and rip-stop nylon bag. Employs a suture loop for cinching and a fork-based mechanism for bag detachment. Manual operation; no energy source or software.

Indications for Use

Indicated for the removal of tissue specimens including appendix, ectopic pregnancy, gallbladder, gallstones, lymph nodes, ovaries, and small sections of bowel during surgical procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is composed of three human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 30, 2016 Covidien Ms. Kate Brown Associate, Regulatory Affairs 60 Middleton Ave. North Haven, Connecticut 06473 Re: K163102 Trade/Device Name: ReliaCatch Specimen Retrieval Bag 10mm; ReliaCatch Specimen Retrieval Bag 12mm Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: November 2, 2016 Received: November 7, 2016 Dear Ms. Brown: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Jennifer R. Stevenson -A For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K163102 Device Name ReliaCatch™M Specimen Retrieval Bag #### Indications for Use (Describe) The specimen retrieval bag has application for safe, convenient removal of tissue specimens such as the appendix, an ectopic pregnancy, gallbladder, gallstones, lymph nodes, ovaries, small sections of bowel and other tissue structures. | Type of Use (Select one or both, as applicable) | <span> <b> <span style="text-decoration: underline;">Prescription Use</span> </b> (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "Medtronic" in a bold, blue font. The letters are closely spaced and appear to be sans-serif. The logo is simple and modern, with a dot above the "i". # 510(k) Summary This 510(k) summary of data used to demonstrate substantial equivalence is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.98 | NAME: | Covidien | |----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ADDRESS: | 60 Middletown Avenue<br>North Haven, Connecticut 06473 USA | | CONTACT PERSON: | Kate Brown<br>Associate, Regulatory Affairs | | PHONE NUMBER: | (203) 492-8350 | | FAX NUMBER: | (203) 492-5029 | | DATE PREPARED: | 11/02/2016 | | TRADE/PROPRIETARY NAME: | ReliaCatch™ 10mm Specimen Retrieval Bag /<br>ReliaCatch™ 12mm Specimen Retrieval Bag | | COMMON/USUAL NAME: | Specimen Retrieval Device | | CLASSIFICATION NAME: | Endoscope and Accessories per 21 CFR §<br>876.1500 | | PRODUCT CODE: | GCJ | | CLASSIFICATION PANEL NAME: | General and Plastic Surgery | | FDA PANEL NUMBER: | 79 | | DEVICE CLASS: | Pursuant to 21 CFR § 876.1500 an endoscope and<br>accessories is a Class II device | | PREDICATE DEVICE(S): | EndoCatch™ Gold previously cleared as Auto<br>Suture™ disposable endoscopic specimen pouch<br>[K922123] | | DEVICE DESCRIPTION: | The ReliaCatch™ 10mm Specimen Retrieval Bag<br>and ReliaCatch™ 12mm Specimen Retrieval Bag<br>consist of a long cylindrical tube and a rip-stop<br>nylon bag that encapsulates and contains<br>specimens for removal from the body cavity. The<br>device contains a handle used to retract and deploy<br>the bag. The specimen bag is designed for<br>introduction and use through all appropriately sized<br>trocar sleeves. | {4}------------------------------------------------ # Medtronic ## INTENDED USE: The specimen retrieval bag has application for safe, convenient removal of tissue specimens such as the appendix, an ectopic pregnancy, gallbladder, gallstones, lymph nodes, ovaries, small sections of bowl and other tissue structures. SUMMARY COMPARING THE TECHNOLOGICAL CHARACTERISTICS OF THE PROPOSED AND PREDICATE DEVICE(S) The predicate device EndoCatch™ is packaged with the bag inside of the tube. Upon opening the package, the device is inserted into the trocar where the baq is deployed and the specimen retrieved. In the new design, the ReliaCatch™ device is packaged with the bag outside of the tube. Upon opening the package, the bag must initially be retracted into the tube prior to inserting into the trocar, and then deployed, and specimen retrieved. Additionally, the subject device will utilize a suture loop to cinch the bag once the specimen is retrieved, where the predicate utilized a ring. The predicate design consisted of a bag attached to a stainless steel spring. Once specimen is retrieved, the bag is cinched by pulling the gold ring and detaching the bag from the metal spring. The metal spring would then be retracted into the tube and the string cut, leaving the bag in the body cavity while the EndoCatch™ device is removed from the trocar. The subject device consists of a bag on "forks" which allows for the complete removal of the forks from the bag prior to cinching the bag and removing the specimen from the body cavity. Lastly, the suture material was changed to better perform in conjunction with the new rip-stop nylon bag material. MATERIALS: All components of the ReliaCatch™ specimen retrieval bag are comprised of materials which are in accordance with ISO 10993-1. {5}------------------------------------------------ # Medtronic # PERFORMANCE DATA: Summary of testing includes description of test methods, sample size, acceptance criteria, and pass/fail results. Objective acceptance criteria were defined for each test. Performance testing included: - Leak Rate Evaluation ● - Trocar Insertion Force ● - Trocar Removal Force ● - Pouch Deployment Force ● - Pouch Retraction Force - . Pouch Cinch Force - Pouch Burst Pressure - Human Factors/Usability # CONCLUSION: The results of the performance and comparative testing described above demonstrate that the ReliaCatch™ specimen retrieval bag is substantially equivalent to the predicate device, EndoCatch™ Gold specimen pouch cleared under [K922123].
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...