The Plasma Edge System single use bipolar resection electrodes are used for the ablation and hemostasis of tissues under endoscopic control, in association with endoscopic accessories. They are intended for endoscopic surgeries with saline irrigation, in the field of urology. The use of the Plasma Edge System is restricted to surgeons, specialized in urological surgery, for specific use in: · Transurethral resection of prostate (TURP) for benign prostatic hyperplasia - · Transurethral incision of the prostate (TUIP) or bladder neck - · Transurethral resection of bladder tumors (TURBT) - · Cystodiathermy · Transurethral Vaporization of the prostate (TUVP/TVP) for benign prostatic hyperplasia, and for Transurethral Vaporization of bladder tumors. (MVVS and MVV models only)
Device Story
Plasma EDGE System is a manual, single-use bipolar electrosurgical device for urological endoscopic procedures. Input: high-frequency (HF) electrical energy delivered via a compatible generator. Operation: HF energy ionizes saline irrigation fluid at the active electrode tip, creating plasma to cut, coagulate, or vaporize tissue. Components: single-use electrodes (side or front load), reusable working elements (compatible with Storz, ACMI, Olympus optics), and an adapter for HF generator connection. Used in clinical settings by urological surgeons. Output: physical tissue ablation/hemostasis. Benefits: enables precise endoscopic resection and vaporization in saline media, reducing risks associated with non-conductive irrigation fluids.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and sterilization validation (ISO 17665). Performance testing included end-of-life simulation, breakdown simulation, working element compatibility, and ex vivo tissue testing to demonstrate safety and efficiency.
Technological Characteristics
Bipolar electrosurgical electrode; active tip connected via wires through ceramic tubes. Materials: biocompatible per ISO 10993. Energy: High-frequency (HF) electrical. Connectivity: manual connection to bipolar HF generator via adapter. Sterilization: Ethylene oxide (electrodes), Autoclave (reusable working elements). Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-2-18.
Indications for Use
Indicated for urological endoscopic surgeries requiring tissue ablation and hemostasis, including TURP, TUIP, bladder neck incision, TURBT, cystodiathermy, and TUVP/TVP for benign prostatic hyperplasia or bladder tumors. Restricted to specialized urological surgeons.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 23, 2017
Chirurgie Innovation % Mark Job Official Correspondent Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K163090 Trade/Device Name: Plasma EDGE System Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: FAS, GEI, FDC Dated: May 15, 2017 Received: May 16, 2017
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Charles Viviano -S
For
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K163090
Device Name
Plasma EDGE System
Indications for Use (Describe)
The Plasma Edge System single use bipolar resection electrodes are used for the ablation and hemostasis of tissues under endoscopic control, in association with endoscopic accessories. They are intended for endoscopic surgeries with saline irrigation, in the field of urology. The use of the Plasma Edge System is restricted to surgeons, specialized in urological surgery, for specific use in:
· Transurethral resection of prostate (TURP) for benign prostatic hyperplasia
- · Transurethral incision of the prostate (TUIP) or bladder neck
- · Transurethral resection of bladder tumors (TURBT)
- · Cystodiathermy
· Transurethral Vaporization of the prostate (TUVP/TVP) for benign prostatic hyperplasia, and for Transurethral Vaporization of bladder tumors. (MVVS and MVV models only)
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
K163090 Page 1 of 7
| Section 5: 510(k) Summary | | | |
|---------------------------|----------------------|--------------------------------|------------|
| Applicant | Chirurgie Innovation | | |
| Device : | Plasma Edge System | Traditional 510(k) application | |
| | | Rev. 4 | Page 1 / 7 |
# SECTION 5:
# 510(K) SUMMARY
| Submitter | Chirurgie Innovation<br>27 Place Victor Schoelcher<br>F-91300 Massy FRANCE |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contacts | Guillaume Noury<br>CEO<br>regulatory@orange.fr<br>Phone : +33 6 16 16 63 30 |
| Date | 07/29/2016 |
| Trade Name | Plasma EDGE System |
| Common name | Bipolar electrodes |
| Classification Name | Electrosurgical cutting and coagulation device and accessories<br>Resectoscope, working element |
| Class | II |
| Product Code | Classification product code : FAS<br>Subsequent product code : GEI / FDC |
| CFR section | 876.4300 / 878.4400 / 876.1500 |
| Device panel | Regulation Medical Specialty : Gastroenterology / Urology<br>510k Review Panel : General & Plastic Surgery |
| Legally marketed<br>predicate devices | K994166: Axipolar Resectoscope Electrode - manufactured by<br>Gyrus Medical Ltd.<br>K120567: Gyrus ACMI PK® Button Electrode - manufactured by<br>Gyrus Medical Ltd. |
## Product Description
The Plasma Edge system is a manual surgical device, consisting of a single-use electrode with cable range, an active and passive working element reusable and an adaptator to an HF generator compatible.
The electrodes consist of an active tip, two wires threaded through ceramic tubes to connect the active tip to the body of the loop, allowing the HF energy to reach the active tip.
{4}------------------------------------------------
| Section 5: 510(k) Summary | |
|--------------------------------|----------------------|
| Applicant | Chirurgie Innovation |
| Device : | Plasma Edge System |
| Traditional 510(k) application | |
| Rev. 4 | Page 2 / 7 |
It has to be used with continuous flow irrigation of saline solution (NaCl 0,9%) that reaches the operative site through a resectoscope. The HF energy delivered from the generator to the electrode ionizes the gas of the saline solution, creating a plasma for the cutting, coagulation and vaporization of tissues.
### Electrodes range:
The list of electrodes is divided in 2 ranges: SIDE LOAD and FRONT LOAD.
Each range got is specific way to be assembly into the working element. The only physical difference between both ranges is the electrode cable connexion.
End user can be used to both versions so we decided to offer the choice between them.
#### Working element range:
The working elements are compatible with the 3 main existing optics brands: STORZ, ACMI and OLYMPUS.
The working elements will be divided in 2 families: side load connection and front load connection.
#### Adaptor:
This will allow to connect the Plasma EDGE electrode range to a bipolar HF generator system
## Indications for use
The Plasma Edge System single use bipolar resection electrodes are used for the ablation and hemostasis of tissues under endoscopic control, in association with endoscopic accessories. They are intended for endoscopic surgeries with saline irrigation, in the field of urology. The use of the Plasma Edge System is restricted to surgeons, specialized in urological surgery, for specific use in:
- Transurethral resection of prostate (TURP) for benign prostatic hyperplasia
- Transurethral incision of the prostate (TUIP) or bladder neck
- Transurethral resection of bladder tumors (TURBT)
{5}------------------------------------------------
| Section 5: 510(k) Summary | | | | |
|---------------------------|--------------------|----------------------|--------------------------------|------------|
| Applicant | | Chirurgie Innovation | | |
| Device : | Plasma Edge System | | Traditional 510(k) application | |
| | | | Rev. 4 | Page 3 / 7 |
- Cystodiathermy
• Transurethral Vaporization of the prostate (TUVP/TVP) for benign prostatic hyperplasia, and for Transurethral Vaporization of bladder tumors. (MVVS and MVV models only)
## Performance testing
#### Risk analysis:
Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO-14971:2007
Performance tests were carried out to ensure that the system functions as intended and meets design specifications. The following performance tests and usability studies were conducted:
#### Biocompatibility testing
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, `Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995. The testing included the following tests: • Biological Safety – toxicology: AAMI ANSI ISO 10993-1:2009; Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process · Cytotoxicity: AAMI ANSI ISO 10993-5:2009; Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity • Chemical Analysis: AAMI ANSI ISO 10993-12:2012; Biological evaluation of medical devices - Part 12: Sample Preparation and Reference Materials
### Electrical safety
Electrical Safety was tested by an independent laboratory according to AAMI/ANSI ES 60601-1:2005 and C1:2009 and A2:2010; Medical Electrical Equipment - Part 1.1 General requirements for safety and essential performance.
### Electromagnetic compatibility (EMC)
{6}------------------------------------------------
| Section 5: 510(k) Summary | | |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Applicant | Chirurgie Innovation | |
| Device : | Plasma Edge System | |
| | Traditional 510(k) application | |
| | Rev. 4 Rev. 4 | Page 4 / 7 |
Electromagnetic compatibility with the existing HF generator was tested by an independent laboratory according to IEC 60601-1 -2 : 2007
### Cleaning and sterilization validation for the reusable working element range:
Steam sterilization validation was tested by an independent laboratory following AAMI TIR N°12, ISO 17664 and ISO 17665
Cleaning and disinfection validation was tested by an independent laboratory following AAMI TIR Nº30
### Bench top validation testing
End of life simulation report, breakdown simulation report, working element compatibility report and test on ex vivo tissues have been tested on the Plasma Edge System to demonstrate the product safety and the efficiency.
#### Summary of Sterilization and Shelf Life Discussion
The electrodes are delivered in a sterile state and are intended for single patient use only. The validated sterilization method used is ethylene oxide. The product has a shelf life of three (3) years.
## List of standards
| Standards | Description |
|--------------|------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-1 | Biological evaluation of medical devices - Part 1: Evaluation and testing<br>within a risk management process |
| ISO 10993-5 | Biological evaluation of medical devices -- Part 5: Tests for In Vitro<br>cytotoxicity |
| ISO 10993-7 | Biological evaluation of medical devices - part 7: ethylene oxide<br>sterilization residuals [including: technical corrigendum 1 (2009)] |
| ISO 10993-10 | Biological evaluation of medical devices - Part 10: Tests for irritation<br>and skin sensitization |
{7}------------------------------------------------
K163090
Page 5 of 7
| Section 5: 510(k) Summary | |
|--------------------------------|----------------------|
| Applicant | Chirurgie Innovation |
| Device : | Plasma Edge System |
| Traditional 510(k) application | |
| Rev. 4 | Page 5 / 7 |
| Standards | Description |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-11 | Biological evaluation of medical devices - Part 11: Tests for systemic<br>toxicity |
| ISO 11135 | Sterilization of health-care products - ethylene oxide - requirements for<br>the development, validation and routine control of a sterilization process<br>for medical devices |
| ISO 11607-1 | Packaging for terminally sterilized medical devices - Part 1:<br>Requirements for materials, sterile barrier systems and packaging<br>systems |
| ISO 11607-2 | Packaging for terminally sterilized medical devices - Part 2: Validation<br>requirements for forming, sealing and assembly processes |
| ISO 17664<br>ISO 17665-1<br>AAMI TIR<br>N°12 | Sterilization of medical devices - Information to be provided by the<br>manufacturer for the processing of resterilizable medical devices<br>Sterilization of health care products -- Moist heat -- Part 1:<br>Requirements for the development, validation and routine control of a<br>sterilization process for medical devices |
| AAMI TIR<br>N°30 | Cleaning and disinfection validation: following FDA guidance<br>"Reprocessing medical device in health care settings: validation<br>methods and labeling guidance for industry and food administration<br>staff issued March 17, 2015" |
| IEC 60601-1 | Medical electrical equipment - Part 1: General requirements for basic<br>safety and essential performance |
| IEC 60601-1-2 | General requirements for basic safety and essential performance -<br>Collateral Standard: Electromagnetic disturbances<br>- Requirements and tests |
| IEC 60601-1-6 | Medical electrical equipment - Part 1-6: General requirements for basic<br>safety and essential performance - Collateral standard: Usability |
| IEC 60601-2-2 | Medical electrical equipment - Part 2-2: Particular requirements for the<br>basic safety and essential performance of high frequency surgical<br>equipment and high frequency surgical accessories |
{8}------------------------------------------------
| Section 5: 510(k) Summary | | |
|---------------------------|----------------------|--------------------------------|
| Applicant | Chirurgie Innovation | |
| Device : | Plasma Edge System | Traditional 510(k) application |
| | Rev. 4 | Page 6 / 7 |
| Standards | Description |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-2-18 | Medical electrical equipment - Part 2-18: Particular requirements for the<br>basic safety and essential performance of endoscopic equipment |
| IEC 62366 | Medical devices – Application of usability engineering to medical devices |
| ISO 14971 | Medical devices - application of risk management to medical devices. |
| ISO 15223-1 | Medical devices - symbols to be used with medical device labels,<br>labeling, and information to be supplied - part 1 : general requirements |
## Comparison of Technological characteristics:
| ELECTRODE | Proposed device | Predicate device | Predicate device |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| | Plasma Edge (cutting loop electrode / vaporization electrode) | Axipolar<br>Resectoscope<br>Electrode<br>(K994166) | Gyrus ACMI PK®<br>Button Electrode<br>(K120567) |
| Indications for use | - Transurethral resection of prostate (TURP)<br>- Transurethral incision of the prostate (TUIP) or bladder neck<br>- Transurethral resection of bladder tumours (TURBT)<br>- Cystodiathermy<br>-Transurethral electrovaporization of the prostate | - Transurethral resection of prostate (TURP)<br>- Transurethral incision of the prostate (TUIP) or bladder neck<br>- Transurethral resection of bladder tumours (TURBT)<br>- Cystodiathermy<br>- | - Transurethral electrovaporization of the prostate |
| Sterilization | Eto | Eto | Eto |
{9}------------------------------------------------
K163090 Page 7 of 7
| Section 5: 510(k) Summary | | | | | |
|---------------------------|--------------------|----------------------|--------------------------------|------------|--|
| Applicant | | Chirurgie Innovation | | | |
| Device : | Plasma Edge System | | Traditional 510(k) application | | |
| | | | Rev. 4 | Page 7 / 7 | |
| ELECTRODE | Proposed device | Predicate device | Predicate device |
|----------------------------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| | Plasma Edge (cutting loop electrode / vaporization electrode) | Axipolar<br>Resectoscope<br>Electrode<br>(K994166) | Gyrus ACMI PK®<br>Button Electrode<br>(K120567) |
| Disposable | Yes | | Yes |
| Energy type | High frequency | | High frequency |
| Mode | Bipolar | | Bipolar |
| User interface | Footswitch | | Footswitch |
| Use only in<br>Conductive<br>Media | The electrode is to be activated only when immersed in a conductive media such as standard saline solution | | The electrode is to be activated only when immersed in a conductive media such as standard saline solution |
| Electrode<br>manipulation | Working element | | Working element |
| Working<br>element<br>compatibility | Storz<br>Olympus<br>ACMI | | Storz<br>Olympus<br>ACMI |
| Sterilization of<br>working<br>element | Autoclave | | Autoclave |
## Conclusion:
There is no difference between the Plasma EDGE resection system and the predicate devices in terms of intended use, principle of operation, and the technology used for device performance. In other words, there is no difference technically, clinically and biologically.
The Plasma EDGE resection system was subjected to verification testing to confirm device performance. There is no new technology and no difference that would raise new or different questions of safety or efficiency. Comparative performance testing demonstrate that the device performed as well as, or better than, the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.