VITEK® 2 AST-Gram Negative Tobramycin is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Negative Tobramycin is a qualitative test. Tobramycin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Citrobacter spp. Enterobacter spp. Escherichia coli Klebsiella spp. Morganella morganii Pseudomonas aeruginosa Proteus mirabilis Proteus vulgaris Providencia spp. Serratia spp. The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.
Device Story
VITEK® 2 AST-Gram Negative Tobramycin card; automated quantitative antimicrobial susceptibility test. Input: isolated bacterial colonies in standardized saline suspension. Operation: card contains 64 microwells with dehydrated media and specific antibiotic concentrations (8, 16, 64 µg/mL). Inoculum rehydrates media via vacuum filling; card incubated at 35.5°C. Growth monitored optically 4x/hour via light transmittance. Instrument calculates MIC based on growth threshold. Output: MIC value and interpretive category (S, I, R). Used in clinical labs; operated by technicians. Results aid clinicians in selecting appropriate antimicrobial therapy for Gram-negative infections.
Clinical Evidence
Performance evaluated using 1000 clinical and challenge isolates (924 Enterobacteriaceae, 76 P. aeruginosa). Compared to CLSI broth microdilution reference method. Overall essential agreement (EA) 98.4%, category agreement (CA) 96.8%. No very major errors, 1 major error (0.1%), 31 minor errors (3.1%). Trending analysis showed a tendency for lower MIC values compared to reference; addressed via labeling. Growth failure rate 0.3%.
Technological Characteristics
Miniaturized broth microdilution card. Automated incubation and optical growth monitoring. System platforms: VITEK 2 (automated fill/seal) and VITEK 2 Compact (manual fill/seal).
Indications for Use
Indicated for antimicrobial susceptibility testing of isolated colonies of clinically significant aerobic Gram-negative bacilli (Citrobacter spp., Enterobacter spp., E. coli, Klebsiella spp., M. morganii, P. aeruginosa, P. mirabilis, P. vulgaris, Providencia spp., Serratia spp.) to Tobramycin.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
{0}
Page 1 of 13
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K163083
B. Purpose for Submission:
To obtain a substantial equivalence determination for Tobramycin for testing of gram negative bacilli on the VITEK® 2 and VITEK® 2 Compact Antimicrobial Susceptibility Test (AST) Systems
C. Measurand:
The VITEK® 2 AST-Gram Negative card contains the following concentrations of Tobramycin: 8, 16, and 64 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The Tobramycin MIC reporting range for the card is ≤ 1 – ≥ 16 µg/mL.
D. Type of Test:
Automated quantitative or qualitative antimicrobial susceptibility test for Tobramycin
E. Applicant:
bioMérieux, Inc.
F. Proprietary and Established Names:
VITEK® 2 AST-Gram Negative Tobramycin (≤ 1 – ≥ 16 µg/mL)
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1645: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System
2. Classification:
Class II
3. Product code(s):
{1}
LON – Fully automated short-term incubation cycle antimicrobial susceptibility system
LTW – Susceptibility Test Cards, Antimicrobial
LTT – Panels, Test, Susceptibility, Antimicrobial
4. Panel:
Microbiology (83)
H. Intended Use/Indications for Use:
1. Intended Use (s):
The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.
2. Indications for Use:
VITEK® 2 AST-Gram Negative Tobramycin is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Negative Tobramycin is a quantitative test. Tobramycin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
Active in vitro and in clinical infections
Citrobacter species
Enterobacter species
Escherichia coli
Klebsiella species
Morganella morganii
Pseudomonas aeruginosa
Proteus mirabilis
Proteus vulgaris
Providencia species
Serratia species
The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant
Page 2 of 13
{2}
yeast.
3. Special conditions for use statement(s):
For Prescription Use Only
Limitations:
"Perform an alternative method of testing prior to reporting of results for the following antibiotic/organisms combination(s):
Tobramycin: Providencia stuartii"
"The ability of the VITEK® 2 AST card to detect resistance with the following combination(s) is unknown because resistant strains were not available at the time of comparative testing:
Tobramycin: Citrobacter koseri, Morganella morganii, and Proteus vulgaris"
"The ability of the VITEK® 2 AST card to detect resistance with the following combination(s) is unknown because an insufficient number of resistant strains were available at the time of comparative testing:
Tobramycin: Enterobacter aerogenes, Enterobacter cloacae, and Proteus mirabilis"
4. Special instrument requirements:
VITEK® 2 and VITEK® 2 Compact Systems using VITEK® 2 Systems 8.01 software
I. Device Description:
The VITEK® 2 AST card utilizes a miniaturized, abbreviated, and automated version of the doubling dilution technique for determining the minimal inhibitory concentration (MIC) on the VITEK® 2 and VITEK® 2 Compact Systems.
Each VITEK® 2 AST card has 64 microwells containing dehydrated media with or without an antimicrobial. A control well that contains only dehydrated culture medium is resident on all cards, with the remaining wells containing premeasured dried amounts of a specific antibiotic combined with culture medium. VITEK® 2 AST-GN Tobramycin has 8, 16, and 64 μg/mL concentrations included in the test. The dried media in the card is rehydrated with the organism suspension during the vacuum filling process and is then sealed. The VITEK® 2 System has two options for preparing the AST card inoculum including an auto-dilution or manual dilution method. The inoculated, sealed card is automatically loaded into the carousel of the instrument and incubated (35.5°C) for the duration of the test. In contrast, the VITEK® 2 Compact instrument has a manual inoculation and sealing operation.
Growth of the test organism is optically monitored and measured four times per hour
Page 3 of 13
{3}
throughout the incubation cycle (up to 18 hours for GN cards). Interpretive calls are calculated by the instrument once a predefined growth threshold is obtained. When the test is complete, the VITEK® 2 card is automatically ejected from the carousel into a waste container. A report is generated that contains the minimal inhibitory concentration (MIC) value with an interpretation category of susceptible, intermediate, or resistant for each antimicrobial on the card. The MIC result reporting range for Tobramycin on the VITEK® 2 AST-GN card is ≤ 1 - ≥ 16 μg/mL.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
VITEK® 2 AST-GN Piperacillin-Tazobactam
2. Predicate 510(k) number(s):
K113200
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device: VITEK® 2 AST-GN Tobramycin (K163083) | Predicate Device: VITEK® 2 AST-GN Piperacillin-Tazobactam (K113200) |
| Intended Use | The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. | Same |
| Test Method | Automated quantitative antimicrobial susceptibility test for use with the VITEK® 2 and VITEK® 2 Compact Systems to determine the in vitro susceptibility of Gram negative bacilli | Same |
| Inoculum | Standardized saline suspension of test organism | Same |
| Test Card | VITEK® 2 Gram Negative Susceptibility Test Card | Same |
| Instrument | VITEK® 2 and VITEK® 2 Compact Systems | Same |
| Analysis Algorithm | Growth pattern analysis | Same |
| Detection Method | Light optics using optical scanner | Same |
| Report | Automatically generated and contains | Same |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device: VITEK® 2 AST-GN Tobramycin (K163083) | Predicate Device: VITEK® 2 AST-GN Piperacillin-Tazobactam (K113200) |
| | the MIC value with an interpretive category (S, I, R) for each antimicrobial on the test card | |
| Differences | | |
| Item | Device: VITEK® 2 AST-GN Tobramycin (K163083) | Predicate Device: VITEK® 2 AST-GN Piperacillin-Tazobactam (K113200) |
| Antimicrobial Agent | Tobramycin | Piperacillin-tazobactam |
| Antimicrobial Concentrations | 8, 16, 64 μg/mL | 2/4, 8/4, 24/4, 32/4, 32/8, 48/8 μg/mL |
| Reporting Range | ≤ 1 - ≥ 16 μg/mL | ≤ 4 - ≥ 128 μg/mL |
| Analysis Algorithm | Unique to Tobramycin | Unique to Piperacillin-tazobactam |
# K. Standard/Guidance Document Referenced (if applicable)
FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)
- CLSI M07-A9, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard-Ninth Edition" Vol. 32 No. 2 (January 2012)
- CLSI M100-S24, "Performance Standards for Antimicrobial Susceptibility Testing"; Twenty-fourth Informational Supplement, Vol. 33 No. 1 (January 2014)
# L. Test Principle:
The VITEK® 2 and VITEK® 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics within the systems use visible light to directly measure organism growth within each of the 64 microwells.
Transmittance optics are based on an initial light reading of a well before significant growth has begun. Every 15 minutes throughout the incubation cycle (defined period of time based on the VITEK® 2 card), light transmittance readings of each well measures organism growth by the amount of light that is prevented from passing through the well. At the completion of the incubation period, the MIC values and their associated interpretive category results for each antimicrobial on the test card are displayed in an automatically generated report.
# M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
{5}
A reproducibility study for the VITEK® 2 AST-GN card with Tobramycin was conducted at three clinical sites using ten isolates of Gram-negative bacilli consistent with the Intended Use. Testing was performed on three separate days and in triplicate for a total of 270 data points. The isolates tested in the reproducibility study included Serratia marcescens (three isolates), Escherichia coli (two isolates), Klebsiella pneumoniae pneumoniae (three isolates), and Citrobacter braakii (two isolates). Inocula were prepared using both the auto-dilution and manual dilution methods for testing in the VITEK® 2 System. Inocula were prepared by the manual dilution method only for use with the VITEK® 2 Compact. The mode MIC value was determined and the reproducibility was calculated based on MIC values that fell within +/- one doubling dilution from the mode MIC value. Both intra-site and inter-site reproducibility for Tobramycin was calculated. There were some off-scale MIC results; accordingly, reproducibility was calculated for best- and worst-case scenarios per the AST Special Controls Guidance.
For the VITEK® 2 auto-dilution method, overall reproducibility was 100% and 97.78% for best- and worst-case scenarios, respectively. There were a total of 6 off-scale MIC results.
For the VITEK® 2 manual dilution method, overall reproducibility was 100% and 95.2% for best- and worst-case scenarios, respectively.
For the VITEK® 2 Compact manual dilution method, overall reproducibility was 100% and 95.9% for best- and worst-case scenarios, respectively. There were a total of 11 off-scale MIC results.
The combined reproducibility results for all three sites were acceptable.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
**Inoculum Density Control**
The DensiCHEK™ Plus was used to standardize the inoculum to a 0.5 MacFarland standard. The instrument was standardized daily with all results recorded at each site. Calibration values were within the expected range.
**Purity Check**
A purity check of all organisms was performed on the dilution tube used to prepare the VITEK® 2 card inoculum. Only those cultures that were pure were evaluated in the study.
**Growth Failure Rate**
Of the 961 clinical isolates that were evaluated (810 Enterobacteriaceae and 147 P.
Page 6 of 13
{6}
aeruginosa), three isolates did not grow in the VITEK® 2 Tobramycin test using the VITEK® 2 System. Complete test results were available for 957 isolates out of 961 clinical isolates. The growth failure rate was 0.3% and was acceptable.
All 76 challenge organisms grew in the VITEK® 2 GN card with Tobramycin using both the auto-dilution and manual dilution methods of the VITEK® 2 and Compact Systems. Therefore, 1033 isolates had VITEK® 2 AST results available.
## Quality Control (QC) Testing
Two FDA/CLSI recommended QC organisms were tested throughout the comparative testing at three study sites. The organisms tested were Escherichia coli ATCC 25922 and Pseudomonas aeruginosa ATCC 27853. These recommended QC organisms were tested a minimum of 20 times/site by both the VITEK® 2 AST-GN card with Tobramycin and with the CLSI broth microdilution reference method.
Both the auto-dilution and the manual dilution methods were within the expected range >95% of the time. In instances where any organism was out of range for the reference method, all testing data from that particular day was invalid and repeated. The following table (Table 1) provides a summary of QC results.
Table 1. Quality Control Results Summary for VITEK® 2 (Auto-Dilution and Manual Dilution Methods) and VITEK® 2 Compact (Manual Dilution Method)
| | | VITEK® 2 Auto-Dilution | | VITEK® 2 Manual Dilution | | VITEK® 2 Compact Manual Dilution | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Organism | Conc. (μg/mL) | Test | Reference | Test | Reference | Test | Reference |
| E. coli ATCC 25922 Expected Range 0.25 - 1μg/mL | ≤0.0625 | | | | | | |
| | 0.125 | | | | | | |
| | 0.25 | | | | | | |
| | 0.5 | | 166 | | 83 | | 83 |
| | ≤1* | 225 | 59 | 112 | 29 | 112 | 29 |
| | 2 | | | | | | |
| | 4 | | | | | | |
| | 8 | | | | | | |
| | 16 | | | | | | |
| | 32 | | | | | | |
| | ≥64 | | | | | | |
| | | | | | | | |
| P. aeruginosa ATCC 27853 Expected Range 0.25 - 1μg/mL | ≤0.0625 | | | | | | |
| | 0.125 | | | | | | |
| | 0.25 | | 141 | | 67 | | 67 |
| | 0.5 | | 71 | | 33 | | 33 |
| | ≤1* | 216 | 8 | 103 | 7 | 103 | 7 |
| | 2 | | | | | | |
| | 4 | | | | | | |
| | 8 | 4 | | 4 | | 4 | |
| | 16 | | | | | | |
| | 32 | | | | | | |
{7}
| | ≥64 | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
*The lowest dilution of the VITEK® 2 Tobramycin MIC range is ≤1 μg/mL. Obtaining this value was considered as an indicator that the quality control test results were acceptable.
The expected range for *E. coli* ATCC 25922 with Tobramycin is 0.25 – 1 μg/mL. Even though the Tobramycin concentrations included in the VITEK® 2 AST-GN card are 8, 16, and 64 μg/mL, the reporting range is ≤1 – ≥16 μg/mL. Therefore, all results for *E. coli* ATCC 25922 were off scale for the VITEK® 2 and VITEK® 2 Compact Systems as both VITEK® systems report the lowest end of the scale as ≤1 μg/mL. *P. aeruginosa* ATCC 27853 was also tested to verify the performance of the device. Many results for this QC strain were also off scale (≤1 μg/mL). Overall, performance of the QC organisms was found to be acceptable.
There was no specific recommendation for any additional QC strains to be tested, but the sponsor’s internal QC process assures that the devices are produced and manufactured appropriately. To inform the end-user of this point, the sponsor included the following footnote to the QC table in the device labeling:
"The VITEK® 2 Gram Negative Tobramycin does not include the full CLSI/FDA-recommended dilution ranges for QC testing with *E. coli* ATCC 25922 and *P. aeruginosa* ATCC 27853".
In addition, two Gram-positive organisms and one Gram-negative organism were tested as supplemental quality control organisms throughout comparative testing at each study site by the reference method only. This was done to provide further assurance of the performance of the broth microdilution panels. The organisms tested were:
- *Enterococcus faecalis* ATCC 29212
- *Staphylococcus aureus* ATCC 29213
- *Klebsiella pneumoniae* subspecies *pneumoniae* ATCC 700603
Performance of these additional QC organisms was evaluated and deemed acceptable.
d. Detection limit:
Not applicable
e. Analytical Specificity:
Not applicable
f. Assay cut-off:
Not applicable
Page 8 of 13
{8}
# 2. Comparison studies:
a. Method comparison with predicate device:
Results obtained with the bioMérieux VITEK® 2 AST-Gram Negative card with Tobramycin were compared to results obtained with the CLSI broth microdilution reference panel. The following concentrations of Tobramycin are contained in the VITEK® 2 AST-GN test card: 8, 16, and 64 µg/mL (equivalent standard method concentration by efficacy in µg/mL) and the reporting range is ≤ 1 – ≥ 16 µg/mL (i.e., ≤ 1, 2, 4, 8, ≥ 16). The reference panel contained two-fold serial dilutions with a range of ≤ 0.0625 to ≥ 128 µg/mL. The testing conditions for the reference method consisted of the following:
- Medium – Mueller Hinton broth with appropriate dilutions of antimicrobial solution added
- Inoculum – Direct colony suspension
- Incubation – 35°C ambient air incubator; 16-20 hours for all organisms
All test inocula used for the evaluation of VITEK® 2 AST-GN Tobramycin and the reference method were standardized using the DensiCHEK™ Plus instrument. The cards were inoculated with each test organism by auto-dilution for reading by the VITEK® 2 System and by manual dilution for reading on the VITEK® 2 System and Compact instrument. Reference broth microdilution panels were inoculated in adherence with CLSI document, M07-A9.
The VITEK® 2 Tobramycin susceptibility test was evaluated by three independent clinical sites located in different geographic regions of the United States. Every clinical isolate was tested one time by auto-dilution on the VITEK® 2 System and the reference method using the same initial standardized suspension. The study evaluated 961 clinical isolates and 76 challenge isolates for a combined data set of 1037 isolates. Of the 1037 clinical and challenge isolates that were tested by the VITEK® 2 card with Tobramycin, 1033 isolates yielded AST results. However, 33 of these isolates were removed from analysis due to inherent forcing rules applied by the VITEK® 2 software for non-indicated Salmonella and Shigella species. Thus, the combined performance of the VITEK® 2 Tobramycin test as compared to the reference method is shown in Table 2 for 1000 isolates (924 Enterobacteriaceae and 76 P. aeruginosa).
The majority of clinical isolates were isolated from fresh clinical specimens (825 isolates, 85.8%) and 136 were stock isolates (14.2%).
A total of 76 challenge organisms (51 Enterobacteriaceae and 26 P. aeruginosa) were evaluated at one external site. The challenge set was tested with both the auto-dilution and manual dilution options of the VITEK® 2 System and with the manual dilution method on the VITEK® 2 Compact System. Overall performance of the challenge organisms is shown in Table 2.
Page 9 of 13
{9}
Table 2. Performance ${}^{ \ddagger }$ of Clinical and Challenge Isolates,VITEK ${}^{ \ddagger }$ 2 Auto-Dilution Method
| Organism Group | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #R | Min | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Enterobacteriaceae | | | | | | | | | | | | |
| Clinical | 777 | 762 | 98.1 | 71 | 56 | 78.9 | 748 | 96.3 | 35 | 28 | 1 | 0 |
| Challenge | 51 | 51 | 100.0 | 4 | 4 | 100.0 | 50 | 98.0 | 20 | 1 | 0 | 0 |
| Combined | 828 | 813 | 98.2 | 75 | 60 | 80.0 | 798 | 96.4 | 55 | 29 | 1 | 0 |
| Pseudomonas aeruginosa | | | | | | | | | | | | |
| Clinical | 147 | 146 | 99.3 | 1 | 0 | 0.0 | 146 | 99.3 | 6 | 1 | 0 | 0 |
| Challenge | 25 | 25 | 100.0 | 1 | 1 | 100.0 | 24 | 96.0 | 5 | 1 | 0 | 0 |
| Combined | 172 | 171 | 99.4 | 2 | 1 | 50.0 | 170 | 98.8 | 11 | 2 | 0 | 0 |
| Enterobacteriaceae, Pseudomonas aeruginosa | | | | | | | | | | | | |
| Clinical | 924 | 908 | 98.3 | 72 | 56 | 77.8 | 894 | 96.9 | 41 | 29 | 1 | 0 |
| Challenge | 76 | 76 | 100.0 | 5 | 5 | 100.0 | 74 | 97.4 | 25 | 2 | 0 | 0 |
| Combined | 1000 | 984 | 98.4 | 77 | 61 | 79.2 | 968 | 96.8 | 66 | 31 | 1 | 0 |
$^{1}$ EA - Essential Agreement (+/- 1 doubling dilution)
CA - Category Agreement
EVAL - Evaluable isolates
R - Resistant isolates
Min - Minor discrepancies
Maj - Major discrepancies
Vmj - Very major discrepancies
Essential agreement was calculated when the VITEK® 2 System results were within +/- one doubling dilution of the reference method results. Category agreement was calculated when the VITEK® 2 System result interpretations agreed exactly with the reference method result interpretations. Evaluable results were defined as when both the reference method results and the VITEK® 2 System results were on-scale.
Evaluable results were also defined as when the reference method results were on-scale and off-scale VITEK® 2 System results clearly did not agree within the accepted +/- one doubling dilution.
The overall performance using the auto-dilution method of the VITEK® 2 System demonstrated an essential agreement of $98.4\%$ and an overall category agreement of $96.8\%$ . There were no very major errors, 1 major error (0.1% error rate, 1/864 susceptible organisms), and 31 minor errors (3.1% error rate).
The performance based on combined clinical and challenge data was acceptable.
# Evaluable P. aeruginosa Isolates:
P. aeruginosa provided evaluable results for a total of two isolates. Of these two isolates, only one was within essential agreement (EA) for a percent EA of evaluable isolates of $50\%$ . Because of the limited data submitted with evaluable MIC data, the performance of this device with isolates that have MICs near the breakpoints could not be determined and the following footnote was added in the labeling (Product Information Manual):
"Due to an insufficient number of on-scale isolates available for comparative testing, the performance of VITEK® 2 Gram Negative Tobramycin is unknown for P. aeruginosa isolates with MICs in the range of $4 - 16\mu \mathrm{g} / \mathrm{mL}$ . If critical to patient care, isolates with MICs of $4 - 16\mu \mathrm{g} / \mathrm{mL}$ should be retested using another method".
{10}
Page 11 of 13
# Resistant Organisms:
A total of 66 resistant isolates were identified in the combined clinical and challenge study of Tobramycin for the VITEK® 2 auto-dilution method out of 1000 organisms (6.6%). However, the following organisms had no resistant isolates available during comparative testing: *Citrobacter koseri*, *Morganella morganii*, and *Proteus vulgaris*. This was addressed by adding the following limitation in the package insert:
> “The ability of the VITEK® 2 AST card to detect resistance with the following combination(s) is unknown because resistant strains were not available at the time of comparative testing:
> Tobramycin: *Citrobacter koseri*, *Morganella morganii*, and *Proteus vulgaris*”
In addition, there was an insufficient number of resistant isolates tested during the comparative study for the following organisms: *Enterobacter aerogenes*, *Enterobacter cloacae*, and *Proteus mirabilis*. This was addressed by adding the following limitation in the package insert:
> “The ability of the VITEK® 2 AST card to detect resistance with the following combination(s) is unknown because an insufficient number of resistant strains were available at the time of comparative testing:
> Tobramycin: *Enterobacter aerogenes*, *Enterobacter cloacae*, and *Proteus mirabilis*”
# Challenge Data – Auto and Manual Dilution:
The challenge set of 76 isolates (51 *Enterobacteriaceae* and 25 *Pseudomonas aeruginosa*) was evaluated on the VITEK® 2 System and the VITEK® 2 Compact System with the manual dilution method only.
Overall performance is shown in Table 3.
Table 3. Performance of Challenge Isolates, VITEK® 2 and VITEK® 2 Compact-Manual Dilution
| Organism Group | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #R | Min | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| VITEK® 2 System | | | | | | | | | | | | |
| Auto-dilution | 76 | 76 | 100.0 | 5 | 5 | 100.0 | 76 | 97.4 | 25 | 2 | 0 | 0 |
| Manual dilution | 76 | 75 | 98.7 | 7 | 6 | 85.7 | 76 | 97.4 | 25 | 2 | 0 | 0 |
| VITEK® 2 Compact | | | | | | | | | | | | |
| Manual dilution | 76 | 76 | 100.0 | 6 | 6 | 100.0 | 76 | 97.4 | 25 | 2 | 0 | 0 |
The performance of the challenge isolates using the VITEK® 2 Tobramycin test on both the VITEK® 2 and VITEK® 2 Compact Systems was considered acceptable.
{11}
Page 12 of 13
# MIC Trends:
Using the combined clinical and challenge data for Enterobacteriaceae, an analysis of trending was conducted. This trending calculation takes into account MIC values that are determined to be $\leq 1$ or $\geq 1$ doubling dilution compared to the reference method irrespective whether the device MIC values are on-scale or not. The data for 122 results constitute the evaluable data for trend analysis which is presented in Table 4.
Table 4. Trending Analysis of Evaluable Enterobacteriaceae Clinical and Challenge Isolate Results, VITEK® 2 Auto-Dilution Method
| Enterobacteriaceae | # Eval Isolates | ≥ 1 dilution lower | Exact | ≥ 1 dilution higher |
| --- | --- | --- | --- | --- |
| Combined Clinical and Challenge | 122 | 71 | 28 | 23 |
| | | 58.2%* | 23.0% | 18.9%* |
| | | 95% CI: 49.3% to 66.6% | 95% CI: 16.4% to 31.2% | 95% CI: 12.9% to 26.7% |
*Difference: 39.3%
95% CI: 27.5% to 49.6%
Using the combined clinical and challenge data for *P. aeruginosa*, an analysis of trending was conducted. This trending calculation takes into account MIC values that are determined to be $\leq 1$ or $\geq 1$ doubling dilution compared to the reference method irrespective whether the device MIC values are on-scale or not. The data for 13 results constitute the evaluable data for trend analysis which is presented in Table 5.
Table 5. Trending Analysis of Evaluable *P. aeruginosa* Clinical and Challenge Isolate Results, VITEK® 2 Auto-Dilution Method
| P. aeruginosa | # Eval Isolates | ≥ 1 dilution lower | Exact | ≥ 1 dilution higher |
| --- | --- | --- | --- | --- |
| Combined Clinical and Challenge | 13 | 11 | 0 | 2 |
| | | 84.6%* | 0% | 15.4%* |
| | | 95% CI: 57.8% to 95.7% | | 95% CI: 4.3% to 42.2% |
*Difference: 69.2%
95% CI: 31.3% to 84.9%
A lower MIC reading trend was observed in the overall performance of *Enterobacteriaceae* and *Pseudomonas aeruginosa* compared to CLSI broth microdilution, which raises concerns for potential very major errors.
To address the low trending and the potential occurrence of very major error(s) for Tobramycin when testing clinical and challenge *Enterobacteriaceae* and *P. aeruginosa* isolates with the VITEK® 2 System, the following statement was added as a footnote in the labeling (Product Information Manual):
“VITEK® 2 Tobramycin MIC values for evaluable Enterobacteriaceae and *P. aeruginosa* isolates tended to be one or more doubling dilutions lower compared to reference broth micro-dilution”.
b. Matrix comparison:
{12}
Not applicable
3. Clinical studies:
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Table 6. Interpretive Criteria for Tobramycin (FDA Drug Label)
| Organism | FDA Interpretive Criteria for Tobramycin MIC (μg/mL) | | |
| --- | --- | --- | --- |
| | S | I | R |
| Enterobacteriaceae | ≤4 | 8 | ≥16 |
| Pseudomonas aeruginosa | ≤4 | 8 | ≥16 |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.