K163049 · Antarma Dba Golnit Sutures · NBY · Jun 21, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163049
Device Name
GOLNIT Non-Absorbable PTFE Surgical Suture
Applicant
Antarma Dba Golnit Sutures
Product Code
NBY · General, Plastic Surgery
Decision Date
Jun 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5035
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GOLNIT Non-Absorbable PTFE Surgical Suture is a removable, non-absorbable surgical suture intended for use in the approximation and ligation of soft tissue in the oral cavity including fixation of barrier membranes.
Device Story
GOLNIT Non-Absorbable PTFE Surgical Suture is a sterile, single-use, non-absorbable monofilament suture manufactured from 100% virgin expanded polytetrafluoroethylene (ePTFE). The device is used by clinicians for soft tissue approximation and ligation in the oral cavity, including barrier membrane fixation. The suture is supplied in various lengths and diameters (U.S.P. 2-0 to 6-0), either loose or swaged to medical-grade stainless steel needles of various curvatures and tips. The device is intended for short-term implantation (up to 30 days). It is provided in a double-pouch sterile packaging system and sterilized via ethylene oxide. The clinician uses the suture to secure tissue or membranes; the device remains in place until removed. The material is biocompatible, non-cytotoxic, and non-pyrogenic, providing a stable, inert mechanical closure for oral surgical sites.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including USP performance requirements (tensile strength, needle attachment, diameter) and biocompatibility testing (cytotoxicity, irritation, sensitization, and pyrogenicity per ISO 10993).
Technological Characteristics
Material: 100% virgin expanded polytetrafluoroethylene (ePTFE) monofilament. Suture sizes: U.S.P. 2-0 to 6-0 (oversized diameter). Needle: Medical-grade stainless steel. Sterilization: Ethylene oxide (ISO 11135). Packaging: Double-pouch (ISO 11607-1/2, EN 868-5). Biocompatibility: ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization/irritation).
Indications for Use
Indicated for approximation and ligation of soft tissue in the oral cavity, including fixation of barrier membranes. Not for use in central nervous system or central circulatory system procedures.
Regulatory Classification
Identification
Nonabsorbable expanded polytetrafluoroethylene (ePTFE) surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from ePTFE and is intended for use in soft tissue approximation and ligation, including cardiovascular surgery. It may be undyed or dyed with an approved color additive and may be provided with or without an attached needle(s).
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
Osteogenics Biomedical, Inc. Cytoplast™ Sutures (K003028)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 21, 2017
Antarma dba GOLNIT Sutures Ms. Armine Badalyan Chief Executive Officer 1204 Avenue U, #2030 Brooklyn, New York 11229
Re: K163049
Trade/Device Name: GOLNIT Non-absorbable PTFE Surgical Suture Regulation Number: 21 CFR 878.5035 Regulation Name: Nonabsorbable expanded polytetrafluoroethylene surgical suture Regulatory Class: Class II Product Code: NBY Dated: May 24, 2017 Received: May 31, 2017
Dear Ms. Badalyan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163049
Device Name
GOLNIT Non-Absorbable PTFE Surgical Suture
Indications for Use (Describe)
GOLNIT Non-Absorbable PTFE Surgical Suture is a removable, non-absorbable surgical suture intended for use in the approximation and ligation of soft tissue in the oral cavity including fixation of barrier membranes.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Golnit Sutures. The logo is pink and features a stylized letter G on the left. To the right of the G is the word "GOLNIT" in large, bold letters, with the word "SUTURES" in smaller letters underneath.
## 5.1. 510(k) Summary
In accordance with 21 CFR 807.92, the following summary of information is provided.
| Date: | June 16, 2017 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Device Owner: | Antarma dba GOLNIT Sutures |
| Company Contact: | Armine Badalyan<br>Chief Executive Officer, Antarma LLC<br>1204 Avenue U, #2030<br>Brooklyn, NY 11229<br>718-219-0731<br>armine@golnit.com |
| Proprietary Name: | GOLNIT Non-Absorbable Surgical Suture |
| Common Names: | Non-absorbable Surgical Suture, Synthetic:<br>• Non-absorbable, Expanded, Polytetrafluoroethylene (PTFE) Suture |
| Classification Names: | Suture, Surgical, Non-absorbable, Polytetrafluoroethylene |
| Regulation Numbers: | 21 CFR 878.5035 |
| Regulatory Class: | Class II |
| Product Codes: | NBY |
| Review Panel: | General & Plastic Surgery |
| 510 (k) Predicate Devices: | New Device |
|------------------------------|--------------------------------|
| Osteogenics Biomedical, Inc. | GOLNIT Non-Absorbable PTFE |
| Cytoplast™ Sutures | Surgical Suture: |
| K003028 | • Non-absorbable, Expanded, |
| | Polytetrafluoroethylene (PTFE) |
| | Surgical Suture |
### 5.2 Device Description
GOLNIT Non-absorbable PTFE Surgical Suture is a sterile, disposable, non-absorbable suture that meets the requirements of the United States Pharmacopeia (U.S.P.) except for
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for GOLNIT SUTURES. The logo consists of a stylized letter "G" on the left, followed by the word "GOLNIT" in large, bold, pink letters. Below "GOLNIT" is the word "SUTURES" in smaller, pink letters.
diameter. The suture is manufactured of a 100% polytetrafluoroethylene (PTFE) polymer as monofilament strands that are uncoated, undyed, and of various diameters and lengths that may be crimped or swaged to a standard, medical grade suture needle. The suture sizes range from U.S.P. 2-0 to 6-0.
The surgical suture incorporates a monofilament manufactured from 100% virgin polytetrafluoroethylene (PTFE) resin that is extruded as an expanded porous (ePTFE) monofilament that is soft and supple.
GOLNIT Non-Absorbable PTFE Surgical Suture is sterilized by ethylene oxide. The GOLNIT surgical suture is intended for single use only. The surgically invasive device is indicated to remain in the body for up to 30 days and is biocompatible.
GOLNIT Non-Absorbable PTFE Surgical Suture is a removable, non-absorbable surgical suture intended for use in the approximation and ligation of soft tissue in the oral cavity including fixation of barrier membranes. The device is not intended for invasive procedures related to central nervous system and central circulatory system.
### 5.3 Device Format
The GOLNIT Non-absorbable PTFE Surgical Suture is packaged into labelled two-pouch packaging system consisting of medical grade heat-sealable pouches for single use only. The pouches conform to the international product standards and norms: ISO 11607-1:2006. ISO 11607-2:2006, EN 868-5:2009.
The suture final configuration consists of specified length and diameters of PTFE monofilaments with attached standard, medical grade suture needles coiled in double pouches, sterilized by ethylene oxide, cartoned in labeled white board boxes and shipping containers.
### 5.4 Special Controls
FDA Guidance "'Class II. Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA was followed during the preparation of this submission.
### 5.5 Intended Use
GOLNIT Non-Absorbable PTFE Surgical Suture is a removable, non-absorbable surgical suture intended for use in the approximation and ligation of soft tissue in the oral cavity including fixation of barrier membranes.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for GOLNIT SUTURES. The logo consists of a stylized letter "G" in a square on the left, followed by the word "GOLNIT" in large, bold letters. Below "GOLNIT" is the word "SUTURES" in smaller letters. The entire logo is in a pink color.
### 5.6 Technological Characteristics
GOLNIT Non-absorbable Surgical PTFE Sutures have been designed to meet the requirements for tensile strength, and needle attachment strength as specified in current USP for non-absorbable sutures. Testing is performed on each lot of product to verify that these USP requirements have been met.
Sizing designations for GOLNIT Non-absorbable Surgical PTFE Suture does not follow the standard United States Pharmacopeia (USP) nomenclature used for many commercially available nonabsorbable sutures, differing in the diameter-strength relationship. The sutures in the current submission are oversized in diameter.
### 5.7. Subject Device vs. Predicate Device
The determination of substantial equivalence for the suture device compared to the predicate devices was based on a detailed device description, including intended use, color and materials composition, and sterilization. In addition, performance testing and conformance with voluntary performance standards, e.g. ANSVAAMI/ISO 10993-1 Biological Evaluation of Medical Devices, USP 29 - Non-Absorbable Surgical Sutures, and the FDA "Class II Special Controls Guidance Document: Surgical Sutures: Guidance for Industry and FDA, December 19,2002 demonstrates compliance to U.S.P. Requirements.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Golnit Sutures. The logo is pink and features a stylized "G" on the left side, enclosed in a square. To the right of the "G" is the word "GOLNIT" in large, bold letters, with the word "SUTURES" underneath in smaller letters.
GOLNIT Non-Absorbable PTFE Surgical Suture 510 (k) Premarket Notification
| Predicate Devices vs. GOLNIT Non-absorbable PTFE Surgical Sutures - Substantial Equivalence<br>Comparison | | | |
|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | New Device | Predicate Device | Discussion |
| Device Name | GOLNIT Non-absorbable PTFE Surgical Suture | Osteogenics Biomedical, Inc. CytoplastTM Sutures | New device vs Predicate device |
| FDA Device Premarket Notification | New device | K003028 | New device vs Predicate device |
| FDA Device Classification | Class II per 21 CFR 878.5035 | Same:<br>Class II per 21 CFR 878.5035 | Same |
| Product Code | NBY | NBY | Same |
| Intended Use | GOLNIT Non-Absorbable PTFE<br>Surgical Suture is a<br>removable, non-<br>absorbable surgical<br>suture intended for use in<br>the approximation and<br>ligation of soft tissue in<br>the oral cavity including<br>fixation of barrier<br>membranes. | The CytoplastTM Suture is<br>intended for use in the<br>approximation and ligation<br>of soft tissue in the oral<br>cavity, including fixation<br>of barrier membrane. | The new device has the<br>same Intended Use as<br>Cytoplast Suture<br>(K003028) for oral cavity<br>suturing. |
| How Supplied | PTFE: Sterile<br>monofilament, thread,<br>undyed (natural) in<br>various sizes with<br>attached needles<br>Approximate U.S.P.<br>Sizes:2-0 to 6-0 | PTFE: Sterile<br>monofilament, thread,<br>undyed (natural) in various<br>sizes with attached needles<br>U.S.P. Sizes: 2-0 to 4-0 | The How Supplied for the<br>new device is comparable<br>to the predicate device,<br>CytoplastTM Sutures, in<br>terms of USP sizes (gauge)<br>and attachment of needles. |
| Use | Single Use Only | Single Use Only | Same |
| Predicate Devices vs. GOLNIT Non-absorbable PTFE Surgical Sutures - Substantial Equivalence<br>Comparison | | | |
| Feature | New Device | Predicate Device | Discussion |
| Color | Natural (undyed) | Natural (undyed) | Same |
| Material<br>Composition | Polytetrafluoroethylene<br>(ePTFE) polymer (100%),<br>expanded and uncoated<br>monofilaments | Polytetrafluoroethylene<br>(PTFE) polymer,<br>monofilament, uncoated,<br>high density monofilament | The polymer format of the<br>new device is expanded<br>whereas the Cytoplast™<br>Suture polymer is extruded<br>as a high density PTFE<br>monofilament. |
| Needle Choices | Stainless steel: Needle tips:<br>Taper, Reverse Cutting,<br>Taper-Cutting,<br>Needle Curvature: $\frac{1}{2}$ , 3/8,<br>5/8, $\frac{1}{4}$ , and straight<br>Length: various | Similar: Stainless steel<br>needles with various<br>needles points and needle<br>curvature | The types of GOLNIT<br>needle tips and shapes are<br>similar to the predicate<br>devices and provides a<br>range of eyeless suture<br>needle designs for user<br>convenience or preference. |
| Suture<br>Characteristics | Non-absorbable and does<br>not undergo any changes in<br>vivo. | Not absorbed and no<br>significant changes known<br>to occur in vivo. | The new device test results<br>indicate requirements met<br>for Biocompatibility per<br>ISO 10993 per FDA<br>Special Controls Guidance<br>Class II Special Controls<br>Guidance Document:<br>Surgical Sutures; Section<br>7. The sutures are<br>biocompatible for<br>implantation of short<br>duration less than 30 days. |
| Suture<br>Diameter,<br>Suture Length | Meets U.S.P. requirements<br>except for variance in | Meets U.S.P. requirements | New device with the<br>expanded PTFE |
| Predicate Devices vs. GOLNIT Non-absorbable PTFE Surgical Sutures - Substantial Equivalence Comparison | | | |
| Feature | New Device | Predicate Device | Discussion |
| Needle<br>Attachment,<br>Strength, Knot<br>Pull Tensile | diameters. | | monofilament (designated<br>as ePTFE) differs from<br>USP requirements for<br>diameter. |
| Packaging | Dry packaged in tear-open<br>double, paper/PET/CPP<br>pouches | Dry packaged in<br>paper/polyester-propylene<br>tear-open pouch | New device has same or<br>equivalent design to the<br>predicate devices. |
| Safety | Biocompatible per<br>ISO10993 for<br>cytotoxicity, maximum<br>sensitization and<br>irritation. | Biologically inert, non-<br>cytotoxic | New device has same or<br>equivalent biocompatibility<br>to the predicate devices<br>tested per ISO 10993-5 for<br>cytotoxicity and 10993-10<br>for maximum sensitization<br>and irritation. |
| | Non-pyrogenic | Not determined | The new device has been<br>tested to verify non-<br>pyrogenicity characteristic<br>per USP. There is no impact<br>on the product safety. |
| | Sterilization per ISO<br>11135 | Sterilized by Ethylene Oxide | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Golnit Sutures. The logo is pink and features a stylized "G" on the left side of a square. To the right of the square, the word "GOLNIT" is written in large, bold letters. Below "GOLNIT", the word "SUTURES" is written in smaller letters.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for Golnit Sutures. The logo is pink and features a stylized letter G on the left side of the logo. To the right of the G is the word "GOLNIT" in large, bold letters, and below that is the word "SUTURES" in smaller letters.
GOLNIT Non-Absorbable PTFE Surgical Suture 510 (k) Premarket Notification
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for GOLNIT SUTURES. The logo features a stylized letter "G" on the left, enclosed in a square. To the right of the "G" is the word "GOLNIT" in a bold, sans-serif font, with the word "SUTURES" underneath in a smaller font. The entire logo is in a pink color.
GOLNIT Non-Absorbable PTFE Surgical Suture 510 (k) Premarket Notification
#### 5.8. Performance Data:
Per FDA's Special Control Guidance Document Surgical Sutures, performance testing including testing in accordance to USP for nonabsorbable suture and biocompatibility testing of the suture material in accordance to ISO 10993-1 has been performed to further ensure substantial equivalence to the predicate devices. Biocompatibility testing performed on each suture material type for implantable material included Cytotoxicity, Irritation and Sensitiation, and Pyrogenicity.
All testing performed has demonstrated that the GOLNIT PTFE suture is as safe and effective as the predicate, meets current performance requirements for nonabsorbable surgical sutures unless stated otherwise in labeling, and the sutures are substantially equivalent to the predicate devices.
Design verification testing (U.S.P. <871> - Sutures - Needle Attachment; U.S.P. <881> -Tensile Strength; U.S.P. < 861> -Sutures -Diameter) was conducted to demonstrate the performance of the subject devices to prove conformance to the requirements of U.S.P. standards except for suture diameter. Stability testing was performed to support the proposed shelf life.
The materials selected were based on known biocompatibility (per ISO 10993) and are substantially equivalent to the materials used in the predicate devices listed. Biocompatibility testing performed on each suture material type for implantable material. Testing included: Cytotoxicity, Irritation, Sensitization and Pyrogenicity. The PTFE sutures were non-cytotoxic, non-irritant and non-sensitizer material.
#### 6. Conclusion
Physical properties and functionality testing assured the safety and effectiveness of the subject device within its intended uses. Results of the non-clinical testing demonstrate conformance with the U.S.P. standards and requirements for Non-Absorbable Surgical Sutures except for the variance in suture diameter. There are no new questions concerning the safety and effectiveness of the device. These results support the safety and effectiveness of the devices for their intended use.
Based on the similarities to the predicate devices identified in this submission as well as the outcome of design verification, and that the new device has the same intended use, the same fundamental technology, and the same principle of operation, we conclude that the new device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.