Smart Stitching software is indicated for use to allow post-capture positioning and joining of individual images to produce an additional composite image for two dimensional image review.
Device Story
Smart Stitching software processes multiple medical images (up to 5) to create a single composite image. Inputs include DICOM 3.0, TIFF, and Raw image formats. The device uses an auto-stitching function based on overlapping areas to join images; manual stitch editing is available for fine adjustments. It provides additional functions for image retrieval, storage, transmission, and enhancement (brightness, contrast, sharpness, zoom). The software does not control image acquisition hardware. It is intended for use by healthcare professionals to facilitate diagnostic review. The output is a composite image used to assist in clinical decision-making by providing a long-length view of anatomy.
Clinical Evidence
No clinical testing performed. Bench testing only. Software validation (TS-0001 through TS-0017) verified input/output functions, auto-stitching performance, and software compatibility against predetermined acceptance criteria.
Technological Characteristics
Software-based image processing system. Supports DICOM 3.0, TIFF, and Raw formats. Features include auto-stitching based on image overlap, manual stitch editing, and image enhancement tools (brightness, contrast, sharpness). Connectivity via LAN with DICOM interface. Developed in accordance with IEC 62304 (software life-cycle) and ISO 14971 (risk management).
Indications for Use
Indicated for post-capture positioning and joining of individual medical images to produce a composite image for 2D review.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Carestream Health Inc DR Long Length Imaging Software (K130567)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three parallel lines forming a wing-like shape above a series of human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 28, 2017
OSKO, Inc. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025
Re: K163048
Trade/Device Name: Smart Stitching Software System, Image Processing, Radiological Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 25, 2017 Received: July 26, 2017
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K163048
Device Name Smart Stitching Software System, Image Processing, Radiological
### Indications for Use (Describe)
Smart Stitching software is indicated for use to allow postioning and joining of individual images to produce an additional composite image for two dimensional image review.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"><b> </b></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for OSKO. The logo consists of the word "OSKO" in a bold, sans-serif font, with the letters in a dark blue color. To the right of the word "OSKO" is a stylized graphic that resembles a wave or a sine wave. The graphic is composed of two curved lines that intersect in the middle, creating a symmetrical shape. The colors of the graphic are light blue and dark blue.
# 510(k) Summary
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### Date 510k summary prepared: June 22, 2017
#### I. SUBMITTER
| Submitter's Name | OSKO, Inc. |
|------------------------|----------------------------------------------------------------------------------|
| Submitter's Address | 8085 NW 90TH ST, Medley, Florida 33166, USA |
| Submitter's Telephone | + 1 305-599-7161 |
| Contact person | Name: Thomas Kim<br>Position: Vice President<br>E-mail: thomas@medisoneconet.com |
| Official Correspondent | Dave Kim (davekim@mtech-inc.net) |
| Address | 8310 Buffalo Speedway, Houston, TX 77025 |
| Telephone | +713-467-2607 |
| Fax: | +713-583-8988 |
#### II. DEVICE
| Trade/proprietary Name | Smart Stitching software |
|------------------------|---------------------------------------------|
| Common or Usual Name | System, Image Processing, Radiological |
| Regulation Name | Picture archiving and communications system |
| Regulation Number | 21 CFR 892.2050 (Product Code:LLZ) |
| Regulatory Class | Class II |
#### III. PREDICATE DEVICE
| Primary Manufacturer | Cedara Software Corporation |
|----------------------|---------------------------------------------|
| Device Name | Cedara I-Soft View |
| 510(k) Number | K022881 |
| Regulation Name | Picture archiving and communications system |
| Regulation Number | 21 CFR 892.2050 (Product Code:LLZ) |
| Regulatory Class | Class II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for OSKO. The logo consists of the word "OSKO" in a bold, sans-serif font, with the letters in a dark blue color. To the right of the word "OSKO" is a stylized graphic that resembles a butterfly or a stylized plus sign, with the top and bottom portions in a darker blue and the left and right portions in a lighter blue.
#### IV. REFERENCE DEVICE
Primary Manufacturer Device Name
Carestream Health Inc DR Long Length Imaging Software
510(k) Number K130567 Regulation Name Regulation Number Regulatory Class Class II
Solid Stat X-ray Imager 21 CFR 892.1650 (Product Code:MQB)
#### V. DEVICE DESCRIPTION
Smart Stitching provides a function to read multiple medical images at one time by stitching them to one image based on overlapping areas. It supports DICOM 3.0 which is standard of medical image format as well as Tiff and Raw images. It provides additional functions such as image retrieval, storage, and transmission. Smart Stitching is designed to facilitate diagnosis by easily stitching multiple medical images, up to 5 images, from multiple captures to one image.
Smart Stitching software does not serve as acquisition software nor has control function for acquisition setting.
#### VI. INDICATIONS FOR USE:
Smart Stitching software is indicated for use to allow post-capture positioning and joining of individual images to produce an additional composite image for two dimensional image review.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for OSKO. The logo consists of the word "OSKO" in a bold, sans-serif font, with the letters in a dark blue color. To the right of the word "OSKO" is a stylized graphic that resembles a butterfly or a stylized "X". The graphic is composed of two curved shapes, one in a darker shade of blue and the other in a lighter shade of blue, creating a sense of depth and movement.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VII. PREDICATE DEVICE
| Feature | New Device<br>Smart Stitching<br>K163048 | Reference Device<br>DR Long Length<br>Imaging Software<br>K130567 | Predicate Device<br>Ceara I-SoftView<br>K022881 |
|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | OSKO, Inc. | Carestream Health<br>Inc. | Cedara Software |
| Intended Use | Smart Stitching<br>software is indicated<br>for use to allow post-<br>capture positioning<br>and joining of<br>individual images to<br>produce an additional<br>composite image for<br>two dimensional<br>image review. | The software intended<br>use is to allow post-<br>capture positioning<br>and joining of<br>individual images to<br>produce an additional<br>composite image. The<br>software is sold only<br>for use on Carestream<br>Health System<br>solutions and is<br>available as an<br>integrated feature<br>within the DirectView<br>software | "Two and three dimensional<br>image review, manipulation,<br>analysis and therapy planning<br>capabilities that support<br>image management display<br>needs in the medical<br>environment from multiple<br>locations within and outside<br>the hospital"<br>"Productivity-Enhancing<br>Second Console Workstations<br>- Workstations designed to<br>perform automated, routine<br>tasks such as image review,<br>printing and archiving<br>as well as post processing<br>capabilities that enable<br>special services for referring<br>physicians."<br>"Diagnostic Review<br>Workstations - Workstations<br>designed to assist radiologists<br>and surgeons in conducting<br>primary diagnostic review<br>and planning through<br>flexible and interactive<br>manipulation of multi-<br>modality softcopy images.<br>This includes the use of<br>prosthetic template overlays"<br>"Physician's Review<br>Workstations - Workstations<br>designed to give easy and<br>economic access to multi-<br>modality softcopy images in<br>multiple locations within |
| | | | and outside the hospital. (e.g.<br>teleconferencing,<br>teleradiology etc.)" |
| How to store<br>patient data an<br>diagnostic<br>documents | Be comprised of a<br>central database to store<br>patient data and<br>diagnostic documents | Be comprised of a<br>central database to<br>store patient data and<br>diagnostic documents | Be comprised of a central<br>database to store patient data<br>and diagnostic documents |
| Type of image<br>format | Tiff, Raw, and DICOM<br>file | DICOM file | DICOM file |
| Ability to<br>provide the<br>management,<br>storage and<br>processing and<br>display of<br>patient,<br>diagnostic<br>image data | Provide the<br>management, storage<br>and processing and<br>display of patient,<br>diagnostic image data | Provide the<br>management, storage<br>and processing and<br>display of patient,<br>diagnostic image data | Provide the management,<br>storage and processing and<br>display of patient, diagnostic<br>image data |
| Function to<br>enhance image<br>brightness,<br>contrast and<br>sharpness | Provide functions to<br>enhance image, zoom<br>in/out, brightness,<br>contrast and sharpness | Provide functions to<br>enhance image<br>brightness, contrast<br>and sharpness | Provide functions to<br>enhance image brightness,<br>contrast and sharpness |
| Function to<br>search to<br>retrieve<br>medical<br>records lined<br>to a specific<br>patient | a search function to<br>retrieve medical records<br>linked to a specific<br>patient | a search function to<br>retrieve medical<br>records linked to a<br>specific patient | a search function to retrieve<br>medical records linked to a<br>specific patient |
| Function to<br>export and<br>import data | Include a function to<br>export and import data | Include a function to<br>export and import data | Include a function to export<br>and import data |
| Function to<br>print out the<br>stored data | Can produce printouts<br>of the stored data | Can produce printouts<br>of the stored data | Can produce printouts of the<br>stored data |
| Function to<br>connect to<br>LAN and have<br>DICOM<br>interface | Can connect to LAN<br>and have DICOM<br>interface (by default or<br>as a separate option) | Can connect to LAN<br>and have DICOM<br>interface (by default or<br>as a separate option) | Can connect to LAN and<br>have DICOM interface (by<br>default or as a separate<br>option) |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for OSKO. The word "OSKO" is written in a bold, dark blue font on the left side of the logo. To the right of the word, there is a stylized graphic that resembles a butterfly or a stylized "X". The graphic is composed of two intersecting shapes, one in a darker shade of blue and the other in a lighter shade of blue.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for OSKO. The logo consists of the word "OSKO" in a bold, dark blue font on the left side. To the right of the word, there is a stylized graphic that resembles two intersecting sine waves or curves, one in a lighter blue and the other in a darker blue. The overall design is clean and modern.
#### VIII. COMPARISON DISCUSSION
The indications for use statement for the subject device is similar to indications for use of the predicate device based on the description and the proposed labeling contained in this submission. Both the subject and the predicate software aligns captured images to stitch and generate a longlength composite image, while fine adjustments can be made using the manual stitch editor. The differences in the indications for use for the subject and predicate device, such as the image file type, do not affect the intended diagnostic use of the software. Smart Stitching software has the same technological characteristics as the reference device and the predicate device, "DR Long Length Imaging Stitching" and Cedara I-Softview "Accustitch" respectively. These devices allow a composite image to be formed from multiple general radiography images acquired separately. The new software device raises no new issues of safety or effectiveness.
## No clinical testing is performed.
The complete system configuration has been assessed and tested by the manufacturer and passed all in house testing criteria. The software validation test was designed to evaluate all input functions, output functions, and actions performed by Smart Stitching Software. Each operational mode and the imaging stitching process are documented in the software validation report.
The auto stitching function and SW compatibility test have been conducted and the result demonstrated that the software functions as intended. In addition, the software validation testing ,from TS-0001 through TS-0017, verified and validated the risk analysis and individual performance results were within the predetermined acceptance criteria.
## Safety and Performance Data:
- IEC 62304 Medical device software Software life-cycle processes : 2006
- ISO 14971 Medical Devices Application of risk management to medical device : 2007
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for OSKO. The logo consists of the word "OSKO" in bold, blue letters. To the right of the word is a blue graphic that resembles a sine wave or a stylized butterfly. The graphic is made up of two different shades of blue.
- NEMAPS 3.1-3.20 - Digital Imaging and Communications in Medicine (DICOM) Set (Radiology)
#### CONCLUSIONS IX.
Based on the information above, it is the sponsor's opinion that Smart Stitching software is substantially equivalent to the predicate software in design and performance. The differences between Smart Stitching software and its predicate do not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.