99.6% of the WMH volume identical to manual labeling
12 datasets of different subjects, each consisting of a baseline exam and 1 to 3 follow-up exams.
>1 (manual labelers)
—
—
Indications for Use
Quantib™ Brain is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ Brain output consists of segmentations, visualizations and volumetric measurements of grey matter (GM), white matter (WM), and cerebrospinal fluid (CSF). The output also visualizes and quantifies white matter hyperintensity (WMH) candidates. Users need to review and if necessary, edit WMH candidates using the provided tools, before validation of the WMHs. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ Brain is a post-processing plugin for the GE Advantage Workstation (AW 4.7) or AW Server (AWS 3.2) platforms.
Device Story
Quantib™ Brain 1.2 is a post-processing software plugin for GE Advantage Workstation/Server platforms. It processes 3D T1-weighted and T2-weighted FLAIR MR images to perform automatic segmentation, visualization, and volumetric quantification of brain tissues (GM, WM, CSF) and WMH candidates. The device uses an atlas-based segmentation approach. Users review and edit WMH candidates before final validation. The software supports longitudinal analysis by comparing multiple exams of the same patient, calculating differences in volumes and identifying new or disappearing WMH via voxel-overlap analysis. The output provides quantitative reports to assist radiologists and medical professionals in clinical assessment. The device is intended for use in clinical settings by trained professionals to provide complementary diagnostic information.
Clinical Evidence
Bench testing only. Performance of the new longitudinal WMH labeling algorithm was validated using 12 datasets (baseline plus 1-3 follow-up exams). The automatic labeling of WMHs showed 99.6% agreement with manual labeling of WMH volumes. Existing core segmentation algorithms were unchanged.
Technological Characteristics
Post-processing software plugin for GE AW 4.7/AWS 3.2. Uses atlas-based segmentation (3D T1-weighted and T2-weighted FLAIR MR inputs). Features include cross-sectional and longitudinal volumetric analysis. Developed in compliance with ISO 14971, IEC 62304, and IEC 62366. Software-based, non-invasive.
Indications for Use
Indicated for trained medical professionals to perform automatic labeling, visualization, and volumetric quantification of brain structures (GM, WM, CSF) and white matter hyperintensity (WMH) candidates from MR images to aid in quantitative reporting.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping and connected by flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 6, 2017
Quantib B.V. % Mr. Rudolf Scholte CEO Westblaak 106 Rotterdam, Zuid-Holland 3012KM NETHERLANDS
Re: K163013 Trade/Device Name: Ouantib" Brain 1.2 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 13, 2016 Received: December 14, 2016
Dear Mr. Scholte:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163013
Device Name QuantibTM Brain 1.2
#### Indications for Use (Describe)
Quantib™ Brain is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ Brain output consists of segmentations, visualizations and volumetric measurements of grev matter (GM), white matter (WM), and cerebrospinal fluid (CSF). The output also visualizes white matter hyperintensity (WMH) candidates. Users need to review and if necessary, edit WMH candidates using the provided tools, before validation of the WMHs. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ Brain is a post-processing plugin for the GE Advantage Workstation (AW 4.7) or AW Server (AWS 3.2) platforms.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Image /page/3/Picture/1 description: The image shows a logo for Quantib Image Informal Medicine. The logo has a graphic on the left side that is made up of several small squares in different shades of blue, orange, and white. The text "Quantib" is on the right side of the graphic in a teal color. Below the word "Quantib" is the text "Image Informal Medicine" in a smaller font size.
## I. SUBMITTER
Quantib B.V. Westblaak 106 3012 KM Rotterdam Phone: (+31) 108 41 17 49 Contact Person: Rudolf Scholte Date Prepared: October 27th, 2016
## II. DEVICE
Name of Device: Quantib™ Brain 1.2 Common or Usual Name: Quantib™ Brain Classification Name: System, image processing, radiology (892.2050) Regulatory Class: II Product Code: Picture archiving and communication system (LLZ)
# III. Predicate device
Device: Quantib™ Brain 1 Manufacturer: Quantib B.V. 510(k) Reg. No: K153351 This predicate has not been subject to a design-related recall. Regulatory Class: II Product Code: Picture archiving and communication system (LLZ)
## IV. DEVICE DESCRIPTION
Quantib™ Brain is post-processing analysis software for the GE Advantage Workstation (AW 4.7) and AW Server (AWS 3.2) platforms using Volume Viewer Apps. 12.3 Ext 8 (or higher). It is intended for automatic labeling, visualization, and volumetric quantification of identifiable brain structures from magnetic resonance images (a 3D T1-weighted MR image, with an additional T2-weighted FLAIR MR image for white matter hyperintensities (WMH) segmentation system relies on a number of atlases each consisting of a 3D T1-weighted MR image and a label map dividing the MR image into different tissue segments. Quantib™ Brain provides quantitative information on both the absolute and relative volume of the segmented regions. The automatic WMH segmentation is to be reviewed and if necessary, edited by the user before validation of the segmentation, after which volumetric information is accessible. Longitudinal analysis can be performed for the brain tissue segmentation and WMH segmentation in order to compare multiple exams of an individual patient. Quantib Brain is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the radiology specialist in quantitative reporting.
{4}------------------------------------------------
## V. INTENDED USE
#### Intended use Quantib™ Brain 1.2
Quantib™ Brain is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ Brain output consists of segmentations, visualizations and volumetric measurements of grey matter (GM), white matter (WM), and cerebrospinal fluid (CSF). The output also visualizes and quantifies white matter hyperintensity (WMH) candidates. Users need to review and if necessary, edit WMH candidates using the provided tools, before validation of the WMHs. lt is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ Brain is a post-processing plugin for the GE Advantage Workstation (AW 4.7) or AW Server (AWS 3.2) platforms.
#### Intended use predicate device
The intended use of the modified device is equal to the intended use of the previously cleared predicate device [K153351].
#### VI. DEVICE MODIFICATIONS
Quantib™ Brain 1.2 is an update of Quantib™ Brain 1 (the predicate device). The differences are the following:
- 1. User interface modifications of the White Matter Hyperintensities cross sectional analysis.
In the predicate device, each segmented white matter hyperintensity was listed as a separate item in the on-screen table containing all measurements. In case of many hyperintensities (e.g. more than 20), this table becomes over-populated. The software was changed to only display one measurement which contains the sum of all hyperintensities. Individual volumes can still be assessed in the viewers. There is no impact of this change on the final report. The report shows in both the predicate and modified device the total white matter hyperintensity volume and the burden (number of hyperintensities).
- 2. Two additional review steps to facilitate longitudinal comparison of Brain Volumetry and White Matter Hyperintensities.
This modification facilitates the comparison of Quantib Brain analyses for MRI studies of multiple time points of the same patient. Two or more time points can be loaded and all cross sectional measurements will be linked in the on-screen measurement table and presented as differences using mathematical subtraction. The viewers allow for side-by-side visualization of the MRI scans and the segmentations. The final report is extended by adding the measurement differences between time points and the number of appearing and disappearing hyperintensities. This modification did not require changes in the underlying automated segmentation algorithms and all extra information is derived from the cross sectional measurements. To help the user in distinguishing between consistent, new and
{5}------------------------------------------------
disappearing WMH, a straight forward algorithm based on overlapping voxels was added to the software for which the performance was assessed.
## VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Based on these two modifications, the following technological characteristics are the same:
- Intended use and Indications for use
- . Target users, anatomical site and usage location
- Design
- Compatibility with the environment and other devices
- . Performance of core segmentation (brain volumetry and WMH) algorithms
- . Reported measures of cross sectional brain volumetry and WMH analysis
- Required input for cross sectional analysis
The following technological characteristics are different:
- . Performance: Assessment of the performance of the newly added WMH longitudinal labeling algorithm was added.
- . Reported measures: New measures related to longitudinal comparison were added. These measures are directly derived from already existing cross sectional metrics using mathematical subtraction.
- Required input: More than one exam of the same patient (longitudinal data) can now be used.
#### VIII. PERFORMANCE DATA
#### 1. Quality and safety
Quantib™ Brain was designed in compliance with the following process standards:
- ISO 14971 – Medical devices - Application of risk management to medical devices
- IEC 62304 – Medical device software – Software life cycle processes
- IEC 62366 – Medical devices - Application of usability engineering to medical devices
The following verification and validation reviews and tests are performed:
- Requirements review
- . Architectural review
- Unit tests
- Functional tests
- . Story verification
- . Code analysis
- Code review
- System tests
- . Integration tests
- . Regression tests
{6}------------------------------------------------
- Algorithm tests
- Tool validation
- . SOUP analysis
The changes made in Quantib™ Brain 1.2 do not affect the safety of the device. This conclusion is based on:
- . Failure mode and effects analysis on the added functionality.
- Risk category classification of new software components.
## 2. Algorithm performance
To help the user in distinguishing between consistent, new and disappearing WMH, a straight forward algorithm based on overlapping voxels was added to the software for which the performance was assessed. This algorithm was validated on 12 datasets of different subjects, each consisting of a baseline exam and 1 to 3 follow-up exams. WMHs were labeled as consistent, new and disappearing in a comparison of follow-up exams to their preceding exam. The automatic labeling of WMHs was for 99.6% of the WMH volume identical to manual labeling of these WMHs.
The performance of the already existing algorithms did not change.
## VIIII. CONCLUSIONS
By virtue of its intended use and physical and technological characteristics, Quantib™ Brain 1.2 is substantially equivalent to a device that has been approved for marketing in the United States. The performance data shows that Quantib™ Brain 1.2 is as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.