PHA S1 Spinal System

K163011 · Phamedica, Inc. · NKB · Jun 29, 2017 · Orthopedic

Device Facts

Record IDK163011
Device NamePHA S1 Spinal System
ApplicantPhamedica, Inc.
Product CodeNKB · Orthopedic
Decision DateJun 29, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Phamedica PHA S1 Spinal System is intended for posterior non-cervical pedicle fixation of Thoracic, Lumbar, Sacral/ Iliac Spine in skeletally mature patients, as an adjunct to fusion using autogeneous bone graft or all of the following indications: Degenerative Disc Disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) Spondylolisthesis Trauma (i.e., fracture or dislocation) Spinal Stenosis Curvatures (i.e., scoliosis, kyphosis, and/or lordosis) Tumor Pseudoarthrosis Failed previous fusion in skeletally mature patients.

Device Story

The PHA S1 Spinal Fixation system is a surgical implant set used for spinal stabilization and fusion. Components include poly-axial and mono-axial screws, reduction screws, rods, set screws, and transverse/offset connectors. The system is implanted by a surgeon during spinal procedures to provide mechanical stabilization while bone graft promotes fusion. It is used in clinical settings (OR) for patients with spinal disorders. The device provides structural support to the spine, facilitating the healing process and correcting spinal deformities.

Clinical Evidence

Bench testing only. The device underwent static compression bending, static torsion, and dynamic compression bending testing per ASTM F1717-15. Results confirmed the system's mechanical strength is sufficient for its intended use.

Technological Characteristics

Constructed from Titanium Alloy (Ti6AL4V). Components include poly-axial/mono-axial screws, reduction screws, rods, set screws, and connectors. Mechanical testing performed per ASTM F1717-15. No software or electronic components.

Indications for Use

Indicated for skeletally mature patients requiring posterior non-cervical pedicle fixation of the Thoracic, Lumbar, or Sacral/Iliac spine as an adjunct to fusion. Conditions include degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a cascading manner. The profiles are rendered in a dark color, creating a sense of depth and unity. Encircling the profiles is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 29, 2017 Phamedica, Inc. % Mr. Kenneth C. Maxwell II Regulatory and Quality Specialist Empirical Consulting, LLC 4628 Northpark Drive Colorado Springs, Colorado 80918 Re: K163011 Trade/Device Name: PHA S1 Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: June 13, 2017 Received: June 14, 2017 Dear Mr. Maxwell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K163011 Device Name PHA S1 Spinal Fixation Indications for Use (Describe) Phamedica PHA S1 Spinal System is intended for posterior non-cervical pedicle fixation of Thoracic, Lumbar, Sacral/ Iliac Spine in skeletally mature patients, as an adjunct to fusion using autogeneous bone graft or all of the following indications: · Degenerative Disc Disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) - · Spondylolisthesis - · Trauma (i.e., fracture or dislocation) - · Spinal Stenosis - · Curvatures (i.e., scoliosis, kyphosis, and/or lordosis) - Tumor - Pseudoarthrosis - · Failed previous fusion in skeletally mature patients. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 5.510(K) SUMMARY | Submitter's Name: | Phamedica, Inc. | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter's Address: | 15 Newtown Rd<br>Plainview NY 11803 | | Submitter's Telephone: | 719.291.6874 | | Contact Person: | Kenneth C. Maxwell II<br>Empirical Consulting, LLC<br>719.337.7579 | | Date Summary was Prepared: | 29 June 2017 | | Trade or Proprietary Name: | PHA S1 Spinal Fixation | | Common or Usual Name: | Orthosis, Spinal Pedicle Fixation<br>Orthosis, Spondylolisthesis Spinal Fixation<br>Orthosis, Spinal Pedicle Fixation, For Degenerative Disc<br>Disease | | Classification: | Class II per 21 CFR §888.3070 Device Classification | | Product Code: | NKB | | Classification Panel: | Division of Orthopedic Devices | ## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION: Phamedica PHA S1 Spinal Fixation system consists of Poly Axial screws, Mono Axial Screws, Poly Axial and Mono Axial Reduction Screws, Rods, Set Screws, Transverse and Offset connectors made from Titanium Alloy Ti6AL4V which can be variously assembled for stabilization and to promote bone fusion during the normal healing process following surgical correction of disorders of the spine. ## INDICATIONS FOR USE Phamedica PHA S1 Spinal System is intended for posterior non-cervical pedicle fixation of Thoracic, Lumbar, Sacral/Iliac Spine in skeletally mature patients, as an adjunct to fusion using autogeneous bone graft or allograft for all of the following indications: - Degenerative Disc Disease (defined as back pain of discogenic origin with ● degeneration of the disc confirmed by history and radiographic studies) - Spondylolisthesis - Trauma (i.e., fracture or dislocation) ● - Spinal Stenosis ● - Curvatures (i.e., scoliosis, kyphosis, and/or lordosis) ● - Tumor ● - Pseudoarthrosis ● - Failed previous fusion in skeletally mature patients. {4}------------------------------------------------ The indications for use for the PHA S1 Spinal System is similar to that of the predicate devices listed in Table 5-1. #### Table 5-1 Predicate Devices | 510k<br>Number | Trade or Proprietary or<br>Model Name | Manufacturer | Predicate<br>Type | |----------------|---------------------------------------|--------------|-------------------| | K000236 | Synergy VLS Open | Interpore | Primary | | K111940 | S 100 Pedicle Screw System | Renovis | Additional | ### PERFORMANCE DATA The PHA S1 Spinal System has been tested in the following test modes: - Static compression bending per ASTM F1717-15 ● - Static torsion per ASTM F1717-15 . - Dynamic compression bending per ASTM F1717-15 ● The results of this non-clinical testing show that the strength of the PHA S1 Spinal System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices. ## CONCLUSION The overall technology characteristics and mechanical performance data lead to the conclusion that the PHA S1 Spinal System is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...