K163006 · bioMerieux, Inc. · LON · Jan 24, 2017 · Microbiology
Device Facts
Record ID
K163006
Device Name
VITEK 2 AST-GN Tigecycline (<0.5->8 ug/mL)
Applicant
bioMerieux, Inc.
Product Code
LON · Microbiology
Decision Date
Jan 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Indications for Use
VITEK® 2 AST-Gram Negative Tigeovcline is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Negative Tigecycline is a quantitative test. Tigecycline has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Citrobacter freundii Enterobacter cloacae Escherichia coli Klebsiella oxytoca Klebsiella pneumonia In vitro data available but clinical significance is unknown: Citrobacter koseri Enterobacter aerogenes Serratia marcescens The VITEK@2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococus spp., Enterococcus spp., and clinically significant yeast.
Device Story
VITEK 2 AST-GN Tigecycline is an automated, miniaturized, abbreviated doubling dilution test for determining minimum inhibitory concentrations (MIC) of tigecycline against Gram-negative bacilli. Input: isolated bacterial colonies diluted in saline. Process: VITEK 2 system automatically fills, seals, and incubates cards; optical scanner measures light attenuation to monitor growth over time; analysis program calculates MIC based on growth kinetics. Output: MIC values and interpretive category results. Used in clinical laboratories; operated by technicians. Benefits: provides standardized, automated susceptibility testing to guide antimicrobial therapy.
Clinical Evidence
Bench testing only. Performance evaluated using fresh and stock clinical isolates and challenge strains compared to CLSI broth microdilution reference method (16-20 hrs incubation). Results: 98.9% overall Essential Agreement and 98.6% overall Category Agreement. Reproducibility and Quality Control testing met acceptance criteria.
Technological Characteristics
Miniaturized broth microdilution card containing antimicrobial agents. Sensing principle: automated optical growth monitoring of wells over time. Connectivity: integrated with VITEK 2/VITEK 2 Compact systems. Software: automated analysis of growth kinetics to determine MIC. Sterilization: not specified.
Indications for Use
Indicated for antimicrobial susceptibility testing of isolated colonies of clinically significant aerobic Gram-negative bacilli, Staphylococcus spp., Enterococcus spp., and clinically significant yeast. Specifically validated for Tigecycline susceptibility testing of Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Citrobacter koseri, Enterobacter aerogenes, and Serratia marcescens.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K163006
B. Purpose for Submission:
To obtain a substantial equivalence determination for Tigecycline for testing of Gram negative bacilli on the VITEK®2 and VITEK®2 Compact Antimicrobial Susceptibility Test (AST) Systems.
C. Measurand:
The VITEK 2 AST-Gram Negative card contains the following concentrations of Tigecycline: 1.5, 4 and 8 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The MIC result reporting range for the card is ≤0.5 - ≥8 µg/mL.
D. Type of Test:
Automated quantitative antimicrobial susceptibility for Tigecycline.
E. Applicant:
bioMérieux, Inc.
F. Proprietary and Established Names:
VITEK® 2 AST- GN Tigecycline (≤0.5 - ≥8 µg/mL)
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System
2. Classification:
Class II
3. Product code:
{1}
LON - Fully automated short-term incubation cycle antimicrobial susceptibility system.
LTW - Susceptibility Test Cards, Antimicrobial
LTT - Panels, Test, Susceptibility, Antimicrobial
4. Panel:
83 Microbiology
H. Intended Use:
1. Intended use(s):
The VITEK® 2 Antimicrobial Susceptibility Tests (AST) are intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic Gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., and clinically significant yeast.
2. Indication(s) for use:
VITEK® 2 AST-Gram Negative Tigecycline is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Negative Tigecycline is a quantitative test. Tigecycline has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
Active in vitro and in clinical infections
- Citrobacter freundii
- Enterobacter cloacae
- Escherichia coli
- Klebsiella oxytoca
- Klebsiella pneumonia
In vitro data available but clinical significance is unknown
- Citrobacter koseri
- Enterobacter aerogenes
- Serratia marcescens
The VITEK® 2 Antimicrobial Susceptibility Tests (AST) are intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic Gram-negative bacilli,
2
{2}
Staphylococcus spp., Enterococcus spp., Streptococcus spp., and clinically significant yeast.
3. Special conditions for use statement(s):
Prescription use only
Limitation:
"The ability of the AST card to detect resistance with the following combination(s) is unknown because resistant strains were not available at the time of comparative testing:
Tigecycline: Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Citrobacter koseri, Enterobacter aerogenes, Serratia marcescens."
4. Special instrument requirements:
VITEK® 2 and VITEK® 2 Compact Systems
I. Device Description:
The VITEK® 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK® 2 AST card contains 64 wells. A control well(s) which contain only nutrient medium is resident on all cards. The remaining wells contain premeasured portions of antimicrobials combined with the nutrient media. The isolate to be tested is diluted to a standardized concentration with 0.45% to 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System will automatically dilute the bacterial suspension to prepare an inoculum for susceptibility cards. Then the VITEK® 2 will fill, seal and place the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time (up to 24 hours for Streptococcus Species). The analysis program determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. This data is used to determine the minimum inhibitory concentration or "MIC" values for the anti-microbial agent. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.
VITEK® 2 AST-GN Tigecycline has the following concentrations in the card: 1.5, 4 and 8 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The MIC result range for the VITEK® 2 is ≤ 0.5 - ≥ 8 µg/mL.
J. Substantial Equivalence Information:
{3}
1. Predicate device name(s):
VITEK® 2 AST-GN Cefepime
2. Predicate $510(\mathrm{k})$ number(s):
K161227
3. Comparison with predicate:
Table 1. Comparison with the Predicate Device
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | VITEK® 2 AST-Gram Negative Tigecycline is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Negative Tigecycline is a quantitative test. Tigecycline has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.Active in vitro and in clinical infections:Citrobacter freundiiEnterobacter cloacaeEscherichia coliKlebsiella oxytocaKlebsiella pneumoniaIn vitro data available but clinical significance is unknown:Citrobacter koseri | VITEK® 2 Gram Negative Cefepime is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Gram Negative Cefepime is a quantitative test. Cefepime has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.Active in vitro and in clinical infections:Enterobacter spp.Escherichia coliKlebsiella pneumoniaeProteus mirabilisPseudomonas aeruginosaIn vitro data available but clinical significance is unknown:Citrobacter koseriCitrobacter freundiiPantoea agglomeransKlebsiella oxytoca |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Antimicrobial | Concentration of antimicrobial in the test wells of the VITEK® 2 AST card and the analysis algorithms are unique for each antimicrobial. Tigecycline | Concentration of antimicrobial in the test wells of the VITEK® 2 AST card and the analysis algorithms are unique for each antimicrobial. Cefepime |
| Antimicrobial concentrations | 1.5, 4, 8 | 0.25, 1, 4, 16, 32 |
| Reporting Range | ≤0.5 - ≥8 | ≤0.12 - ≥32 |
| Analysis algorithm | Unique to Tigecycline | Unique to Cefepime |
{5}
6
K. Standard/Guidance Document Referenced (if applicable):
CLSI M100-S024: Performance Standards for Antimicrobial Susceptibility Testing
CLSI M07-A9: Methods for Dilution Antimicrobial Susceptibility Test (AST) Systems Aerobically
Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems
L. Test Principle:
The VITEK® 2 and VITEK® 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics used in the systems use visible light to directly measure organism growth. Transmittance optics are based on an initial light reading of a well before significant growth has begun. Periodic light transmittance samplings of the same well measure organism growth by how much light is prevented from going through the well. The VITEK® 2 System monitors the growth of each well in the card over a defined period of time. An interpretive call is made between 4 and 16 hours for a “rapid” read but may be extended to 18 hours in some instances. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic on the card.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
A reproducibility study was conducted at three sites using ten isolates of gram-negative bacilli that were consistent with the intended use. Isolates were tested in triplicate over three days for a total of 270 data points. The isolates tested in the reproducibility study included *Enterobacter cloacae* (two isolates), *Enterobacter aerogenes* (two isolates), *Serratia marcescens* (three isolates), and *Klebsiella pneumoniae* (three isolates). Inocula were prepared both manually and using automatic dilution for testing in the VITEK2. Inocula were prepared manually for testing in the VITEK2 Compact. The mode MIC value was determined and the reproducibility was calculated based on MIC values falling within +/- 1 dilution of the mode MIC value.
Using VITEK2 and automatic dilution, best case reproducibility was 99.6%; worst case reproducibility was 99.3%.
Using VITEK2 and manual dilution, all results were on scale and the reproducibility was 100%.
Using VITEK® 2 Compact and manual dilution, best case reproducibility was 99.6%; worst case reproducibility was 97.8%.
{6}
The reproducibility results were acceptable.
b. Linearity/assay reportable range:
N/A
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Inoculum Density Check. The inoculum density was monitored using the DensiCHEK Plus™ instrument. The DensiCHEK Plus™ was standardized weekly with all results recorded and in expected range.
Purity Check: A purity check of all organisms was performed at the time of VITEK2 card inoculation. Only results obtained with pure cultures were evaluated.
Growth Failure Rate: During the course of the study there were no growth failures in the VITEK 2 AST-Gram Negative cards.
Quality Control Testing. The CLSI-recommended QC organism (E. coli ATCC 25922) was tested using both the VITEK 2 card and the reference method at each site using both the automatic dilution and manual dilution methods for the VITEK 2 and using the manual dilution method for the VITEK 2 Compact.
The expected range for E. coli ATCC 25922 with tigecycline is 0.03 – 0.25 µg/mL. The tigecycline concentrations included in the VITEK 2 AST-Gram Negative card are 1.5, 4, and 8 µg/mL and the reporting range is ≤0.5 – ≥8 µg/mL. Therefore, all results for the QC strain were off-scale for the VITEK 2 and VITEK 2 Compact Systems and were reported as ≤0.5 µg/mL (Table 2). In response to an FDA request for additional information regarding production and process controls to verify the performance of the device, the sponsor indicated that no other QC strains suitable for the Enterobacteriaceae claim could be identified that would cover the expected tigecycline MIC ranges based on CLSI or FDA drug label. Therefore, there was no specific recommendation for any additional QC strains to be tested, but the sponsor’s internal QC process assures that the devices are produced and manufactured appropriateley. Furthermore, and to inform the end-user of this point, the sponsor included the following footnote to the QC table in the device labeling:
"The VITEK 2 Gram Negative tigecycline does not include the full CLSI/FDA-recommended dilution ranges for QC testing with this organism."
7
{7}
Table 2. Quality Control Results for VITEK 2 with Automatic and Manual Dilution Inoculation Methods and for VITEK 2 Compact with the Manual Dilution Inoculation Method.
| | | VITEK 2 Automatic Dilution | | VITEK 2 Manual Dilution | | VITEK 2 Compact Manual Dilution | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Organism | Conc. (μg/mL) | Test | Ref. | Test | Ref. | Test | Ref. |
| E. coliATCC 25922Expected Range:0.03 – 0.25 μg/mL | ≤0.0156 | | | | | | |
| | 0.03125 | | | | | | |
| | 0.0625 | | 110 | | 56 | | 56 |
| | 0.125 | | 112 | | 55 | | 55 |
| | 0.25 | | 2 | | 1 | | 1 |
| | (≤)0.5* | 225 | | 112 | | 112 | |
| | 1 | | | | | | |
| | 2 | | 1 | | | | |
| | 4 | | | | | | |
| | 8 | | | | | | |
| | ≥16 | | | | | | |
*The lowest dilution of the VITEK 2 MIC range is 0.5 μg/mL. Obtaining this value was considered as an indicator that the quality control test results were acceptable.
The quality control results were acceptable.
d. Detection limit:
N/A
e. Analytical specificity:
N/A
f. Assay cut-off:
N/A
2. Comparison studies:
a. Method comparison with predicate device:
Results obtained with the bioMerieux VITEK AST - Gram Negative card with tigecycline were compared to result obtained with the CLSI broth microdilution reference panel. The VITEK 2 AST-Gram Negative card with tigecycline contains the following concentrations of tigecycline: 1.5, 4 and 8 μg/mL (equivalent standard
{8}
method concentration by efficacy in $\mu \mathrm{g} / \mathrm{mL}$ and the reporting range is $\leq 0.5 - \geq 8$ $\mu \mathrm{g} / \mathrm{mL}$ . The reference panel contained two-fold serial dilutions with a range of 0.0156 to $32~\mu \mathrm{g} / \mathrm{mL}$ .
Test inocula were standardized using the DensiCHEK Plus instrument. VITEK 2 AST - Gram Negative cards were inoculated using automatic dilution (for reading on the VITEK 2 instrument) or using a manual dilution method (for reading on the VITEK 2 instrument or on the VITEK 2 COMPACT instrument). Reference panels were inoculated as outlined in the CLSI document M07-A9.
A total of 637 Enterobacteriaceae clinical isolates were evaluated at three sites with VITEK 2 AST - Gram Negative cards inoculated by automatic dilution and interpreted using the VITEK 2 instrument. The majority of isolates were fresh (569 isolates, $89\%$ ); 68 isolates $(11\%)$ were stock isolates.
A total of 75 challenge isolates were tested at one site. The challenge set was tested with both card inoculation options (automatic dilution and manual dilution) on the VITEK 2 system and with the manual dilution on the VITEK 2 Compact system.
For MICs interpreted using the VITEK 2 system an inoculated using the automatic dilution method, the combined results from the clinical and challenge testing demonstrated a combined EA of $98.9\%$ and CA of $98.6\%$ (Table 3). A total of 96 isolates were determined to have evaluable results; the EA of the evaluable results was $91.7\%$ . There were 2 major errors $(0.3\%)$ and no very major errors identified. The performance based on combined clinical and challenge data was acceptable.
Table 3: Performance of Clinical and Challenge Isolates, VITEK 2 Automatic Dilution Method
| | Tot | No. EA | EA% | Eval Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | No. S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Clinical | 637 | 629 | 98.7 | 78 | 70 | 89.7 | 630 | 98.9 | 1 | 636 | 5 | 2 | 0 |
| Challenge | 75 | 75 | 100 | 18 | 18 | 100 | 72 | 96.0 | 1 | 71 | 3 | 0 | 0 |
| Combined | 712 | 704 | 98.9 | 96 | 88 | 91.7 | 702 | 98.6 | 2 | 707 | 8 | 2 | 0 |
EA - Essential Agreement (+/- 2 dilutions) min - minor discrepancies
CA - Category Agreement maj - major discrepancies
EVAL - Evaluable isolates vmj - very major discrepancies
R or NS - Resistant or non-susceptible isolates
Essential agreement (EA) occurs when there is agreement between the results of the reference method and that of VITEK2 test card within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the VITEK2 test card and the reference method. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the VITEK2 test card.
Challenge isolates interpreted using the VITEK2 and inoculated using the manual dilution method demonstrated an EA of $98.7\%$ and a CA of $94.7\%$ (Table 4). A total of 17 isolates were determined to have evaluable results; the EA of the evaluable results was $94.1\%$ . There were no major or very major errors identified.
{9}
10
Table 4: Performance of Challenge Isolates, VITEK2 Manual Dilution Method
| | Tot | No. EA | EA% | Eval Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | No. S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Challenge | 75 | 74 | 98.7 | 17 | 16 | 94.1 | 71 | 94.7 | 1 | 71 | 4 | 0 | 0 |
Challenge isolates interpreted using the VITEK 2 Compact and inoculated using the manual dilution method demonstrated an EA of 100.0% and a CA of 96.0% (Table 5). A total of 18 isolates were determined to have evaluable results; the EA of the evaluable results was 100%. There were no major or very major errors identified.
Table 5: Performance of Challenge Isolates, VITEK2 Compact, Manual Dilution Method
| | Tot | No. EA | EA% | Eval Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | No. S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Challenge | 75 | 75 | 100 | 18 | 18 | 100 | 72 | 96.0 | 1 | 71 | 3 | 0 | 0 |
Only two resistant isolates were tested with this device, one clinical isolate and one challenge isolate. Due to the lack of available resistant isolates at the time of comparative testing the sponsor was asked to include the following limitation in their device labeling:
The ability of the AST card to detect resistance with the following combination(s) is unknown because resistant strains were not available at the time of comparative testing:
Tigecycline: Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumonia, Citrobacter koseri, Enterobacter aerogenes, Serratia marcescens
Trending. Analysis of trending of MIC values indicated that overall for all Enterobacteriaceae no trending was observed. However, when evaluated by species MICs for K. pneumoniae tended to be one dilution higher than the reference method and MICs for S. marcescens tended to be one dilution lower than the reference method. The following footnote to the performance table was added to the device labeling.
Overall VITEK 2 Tigecycline MICs showed no trending. However, compared to reference broth microdilution, K. pneumoniae results tended to be one dilution higher; results for S. marcescens tended to be one dilution lower.
b. Matrix comparison:
N/A
{10}
3. Clinical studies:
a. Clinical Sensitivity:
N/A
b. Clinical specificity:
N/A
c. Other clinical supportive data (when a. and b. are not applicable):
N/A
4. Clinical cut-off:
N/A
5. Expected values/Reference range:
Table 6. Interpretive Criteria for Tigecycline (FDA Drug Label)
| Organism | FDA Interpretive Criteria for Tigecycline MIC (μg/mL) | | |
| --- | --- | --- | --- |
| | S | I | R |
| Enterobacteriaceae | ≤ 2 | 4 | ≥ 8 |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.