K162992 · Karl Storz Endoscopy America, Inc. · LJH · Jul 7, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K162992
Device Name
Karl Storz UROMAT E.A.S.I.
Applicant
Karl Storz Endoscopy America, Inc.
Product Code
LJH · Gastroenterology, Urology
Decision Date
Jul 7, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Indications for Use
The UROMAT E.A.S.I. is used for the introduction of irrigation fluids into organs and operating fields as well as the suctioning off of irrigation and bodily fluids, secretions and tissue during endoscopic interventions.
Device Story
UROMAT E.A.S.I. is a microprocessor-controlled, peristaltic double-roller pump for irrigation and suction during endoscopic procedures. It provides continuous fluid flow for urological interventions or independent flow regulation for laparoscopic interventions. The device features a touch screen interface and digital display. It monitors pressure via a sensor at the pump output; software enforces a hardcoded 200 mm Hg pressure limit, with a hardware alarm triggered if pressure exceeds this by 40 mm Hg. Used in clinical settings by medical professionals, the device assists in maintaining visibility and managing fluids in the operating field. It benefits patients by enabling controlled irrigation and efficient suction during minimally invasive surgery.
Clinical Evidence
Bench testing only. Provided data includes electrical safety, electromagnetic compatibility (EMC), software verification and validation, usability testing, biocompatibility (cytotoxicity and leachables/extractability), and sterilization/shelf-life/packaging validation.
Technological Characteristics
Peristaltic double-roller pump; microprocessor-controlled; touch screen interface; pressure sensor at pump output; irrigation pressure 20-200 mm Hg; suction flow 100-1800 ml/min. Complies with federal and international safety/performance standards.
Indications for Use
Indicated for patients undergoing endoscopic interventions requiring irrigation and suction of fluids, secretions, and tissue.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 7, 2017
KARL STORZ Endoscopy America, Inc. % Jack Rogers Consultant Safis Solutions LLC 303 N. Alabama, Suite 210 Indianapolis, IN 46204-2132
Re: K162992
> Trade/Device Name: Karl Storz UROMAT E.A.S.I. Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: LJH Dated: May 22, 2017 Received: May 24, 2017
Dear Jack Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162992
Device Name Karl Storz UROMAT E.A.S.I
#### Indications for Use (Describe)
The UROMAT E.A.S.I. is used for the introduction of irrigation fluids into organs and operating fields as well as the suctioning off of irrigation and bodily fluids, secretions and tissue during endoscopic interventions.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
#### 1. Applicant Information
Karl Storz Endoscopy America, Inc. 2151 E. Grand Avenue El Segundo, CA 90245
Contact Person: Serkan Sezer, VP Global Quality and Regulatory Phone: 424-218-8201
Date Summary Prepared: September 22, 2016
#### 2. Device/Classification
| Device Trade Name: | Karl Storz Uromat E.A.S.I. |
|--------------------|----------------------------------------------------------------|
| Common Name: | System, Irrigation, Urological |
| Classification: | 21CFR 876.5130, Class 2<br>Urological catheter and accessories |
| Panel: | Gastroenterology/Urology |
| Product Code(s): | LJH |
#### 3. Predicate Devices
| 510(k) number: | K981615 | K010569 |
|--------------------|----------------------------------------------------------------|------------------------------------------------------|
| Device Trade Name: | Karl Storz Uropump | Karl Storz Universal Laparomat |
| Common Name: | System, Irrigation, Urological | Laparoscope, General & Plastic Surgery |
| Classification: | 21CFR 876.5130, Class 2<br>Urological catheter and accessories | 21CFR 876.1500, Class 2<br>Endoscope and accessories |
| Panel: | Gastroenterology/Urology | General & Plastic Surgery |
| Product Code(s): | LJH | GCJ |
#### 4. Device Description
The UROMAT E.A.S.I. is a peristaltic (double roller), microprocessor controlled, irrigation and suction pump with continuous fluid flow for urological interventions. Alternatively, the two pumps can be flowoperated independently of each other for flow regulation, e.g., for laparoscopic interventions.
#### 5. Intended Use / Indications for Use
The UROMAT E.A.S.I. is used for the introduction of irrigation fluids into organs and operating fields as well as the suctioning off of irrigation and bodily fluids, secretions and tissue during endoscopic interventions.
{4}------------------------------------------------
| Device | Karl Storz<br>UROMAT E.A.S.I. | Karl Storz<br>UROPUMP | Karl Storz<br>Universal Laparomat |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Basic Design | Peristaltic (double roller)<br>pump;<br>Microprocessor controlled | Peristaltic (single roller)<br>pump;<br>Microprocessor controlled | Gear pump;<br>microprocessor controlled |
| Selection of<br>Application(s) | Dual purpose pump for<br>irrigation and suction:<br>1) continuous flow for<br>urologic interventions;<br>including use in lithotripsy<br>(KS CALCUSON)<br>2) independent pump flow<br>and regulation for<br>laparoscopic interventions | Dual purpose pump for<br>irrigation and suction,<br>including when used in<br>lithotripsy (KS CALCUSON) | Single mode |
| Overpressure<br>Safety | Software does not allow<br>pressure over 200 mm Hg<br>(hardcode).<br>Hardware alarm is triggered<br>when the maximum pressure<br>of 200 mm Hg is exceeded by<br>40 mm Hg. | Software alarm triggered<br>when the present pressure<br>setpoint value is exceeded by<br>20 mmHg.<br>Hardware alarm is triggered<br>when the maximum pressure<br>of 200 mm Hg is exceeded by<br>40 mm Hg. | Electronic monitoring of<br>the maximum output<br>pressure via the torque of<br>the gear pump.<br>Double monitoring of the<br>suction pressure. |
| Primary<br>Pressure<br>Measurement | Pressure sensor at the<br>pressure dome measures the<br>pressure at the pump output. | Pressure sensor at the<br>pressure dome measures the<br>pressure at the pump output. | Electronic monitoring of<br>the maximum output<br>pressure via the torque of<br>the gear pump. |
| Irrigation<br>Pressure | 20...200mm Hg for<br>Continuous Flow (CF), Single<br>Flow (SF) | 0 – 200 mm Hg | 0 – 1800 mm Hg |
| Irrigation Flow | 10...250 ml/min for<br>Continuous Flow (CF) | 0 – 400 ml/min | 0 – 3000 ml/min (LAP) |
| Suction<br>Pressure | N/A | Max -0.75 bar | 0 – 800 mbar (LAP) |
| Suction Flow | 100...1800 ml/min for<br>Suction/Irrigation and<br>Lithotripsy | 0-1000 ml/min<br>Suction flow rate is adjustable<br>by the user and is preset<br>before Uropump is used.<br>Flow rate is key parameter for<br>suction. | 0 - (-)8 bar<br>Suction under pressure |
| User Interface | Touch screen, digital display | Keys and LED bar graphs | Bar graph display |
## 6. Comparison of Technological Characteristics with the Predicate Devices
{5}------------------------------------------------
#### 7. Performance Data - Nonclinical testing
The following test data were provided in support of the substantial equivalence determination:
- o Uromat E.A.S.I.
- Electrical safety and electromagnetic compatibility (EMC ) o
- Software Verification and Validation Testing O
- O Usability
- Tubing sets:
- Biocompatibility including Cytotoxicity and Leachables/Extractability (Chemical Analysis) O
- Sterilization, shelf life and packaging verification and validation o
#### 8. Conclusion
The Karl Storz (KS) Uromat E.A.S.I. is substantially equivalent to the Karl Storz predicate devices.
The KS Uromat E.A.S.I. is substantially equivalent to the KS UROPUMP with regard to use and design characteristics. Both are used for irrigation and suction of fluids in ureteroscopic procedures.
The KS UROMAT E.A.S.I. is substantially equivalent to the KS KSEA Universal Laparomat with regard to use and design characteristics. Both are used for irrigation of fluids in laparoscopic procedures.
Minor technological differences between the new and predicate devices do not raise any new concerns of safety and effectiveness.
The new device has the same intended use and operating principles, with similar design features, and functional and performance characteristics as the predicate devices. Based upon the performance data, it has been demonstrated that the Uromat E.A.S.I. performs as intended in the specified use conditions. The device is designed to comply with relevant federal and international safety and performance standards.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.