Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs)
K162982 · Medline Industries, Inc. · LZA · Feb 27, 2017 · General Hospital
Device Facts
Record ID
K162982
Device Name
Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs)
Applicant
Medline Industries, Inc.
Product Code
LZA · General Hospital
Decision Date
Feb 27, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable, non-sterile, ambidextrous nitrile examination glove; features inner coating of colloidal oatmeal USP; beaded cuff design. Used in clinical settings by healthcare examiners to prevent cross-contamination between patient and examiner. Input: physical barrier requirement during patient examination or chemotherapy drug handling. Output: physical protection for the examiner's hand. Device provides barrier protection against biological contaminants and specific chemotherapy agents. Healthcare providers use the glove as a standard personal protective equipment (PPE) item; output affects clinical decision-making by enabling safe handling of hazardous chemotherapy drugs and reducing infection transmission risk. Benefits include hand protection and skin comfort via colloidal oatmeal.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility evaluated per ISO 10993-10 (skin irritation, sensitization). Physical performance tested per ASTM D6319-10. Chemotherapy drug permeation tested per ASTM D6978-05, with breakthrough times reported for 38 drugs.
Technological Characteristics
Nitrile material with colloidal oatmeal USP inner coating. Ambidextrous, beaded cuff, white color. Complies with ASTM D6319-10 (physical properties, dimensions, residual powder) and ASTM D5151-06 (freedom from holes). Chemotherapy resistance tested per ASTM D6978-05. Non-sterile, single-use.
Indications for Use
Indicated for use as a disposable patient examination glove worn on the examiner's hand to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Tested for Use with Chemotherapy Drugs (K160562)
Reference Devices
Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile. Tested for Use with Chemotherapy Drugs (K151750)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 27, 2017
Medline Industries, Inc. Jennifer Mason Senior Regulatory Affairs Specialist One Medline Place Mundelein, Illinois 60060
Re: K162982
Trade/Device Name: Medline Nitrile Powder-free Examination Glove With Colloidal Oatmeal USP (tested For Use With Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: January 31, 2017 Received: January 31, 2017
Dear Jennifer Mason:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Susan Runne DDS, MA
For Tina Kiang, Ph.D. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162982
#### Device Name
Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs)
Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Blenoxane (Bleomycin) 15 mg/ml >240 minutes Bortezomib 1 mg/ml >240 minutes Busulfan 6 mg/ml >240 minutes Carboplatin 10.0 mg/ml >240 minutes Carmustine (BCNU) 3.3 mg/ml 11.9 minutes Cetuximab (Erbitux) 2 mg/ml >240 minutes Cisplatin 1.0 mg.ml >240 minutes Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 minutes Cytarabine 100 mg/ml >240 minutes Dacarbazine (DTIC)10.0 mg/ml >240 minutes Daunorubicin 5 mg/ml >240 minutes Docetaxel 10.0 mg/ml >240 minutes Doxorubicin Hydrochloride 2.0 mg/ml >240 minutes Ellence 2 mg/ml >240 minutes Etoposide (Toposar) 20.0 mg/ml >240 minutes Fludarabine 25 mg/ml >240 minutes Fluorouracil 50.0 mg/ml >240 minutes Gemcitabine (Gemzar) 38 mg/ml >240 minutes Idarubicin 1 mg/ml >240 minutes Ifosfamide 50.0 mg/ml >240 minutes Irinotecan 20.0 mg/ml >240 minutes Mechlorethamine HCI 1.0 mg/ml >240 minutes Melphalan 5 mg/ml >240 minutes Methotrexate 25 mg/ml >240 minutes Methotrexate 50 mg/ml >240 minutes Mitomycin C 0.5 mg/ml >240 minutes Mitoxantrone 2.0 mg/ml >240 minutes Oxaliplatin 5 mg/ml >240 minutes Paclitaxel (Taxol) 6.0 mg/ml >240 minutes Paraplatin 10 mg/ml >240 minutes Pemetrexed Disodium 25 mg/ml >240 minutes Raltitrexed 0.5 mg/ml >240 minutes Rituximab 10 mg/ml >240 minutes Thiotepa 10.0 mg/ml 18.6 minutes Trisonex 0.1 mg/ml >240 minutes Vidaza (5-Azacytidine) 25 mg/ml >240 minutes Vinblastine 1 mg/ml >240 minutes
{3}------------------------------------------------
Vinorelbine 10 mg/ml >240 minutes Vincristine Sulfate 1.0 mg/ml >240 minutes
Please note that the following drugs have extremely low permeation times: Carmustine (BCNU) (3.3 mg/ml) 11.9 minutes Thiotepa (10.0 mg/ml) 18.6 minutes
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image features the Medline logo, set against a blue background. The logo consists of the word "MEDLINE" in a bold, sans-serif font, positioned to the left of a white, stylized starburst symbol. The starburst has four points, with the top and bottom points being longer than the side points. A gray bar is visible on the right side of the image.
# K162982 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]
### Submitter / 510(k) Sponsor
Medline Industries, Inc. 1 Medline Place Mundelein, IL 60060
Registration Number: 1417592
### Contact Person
Jennifer Mason Senior Regulatory Affairs Specialist Phone: 847-643-3652 Email: jamason@medline.com
Summary Preparation Date February 27, 2017
Type of 510(k) Submission Traditional
### Device Name / Classification
Name of Device: Patient Examination Glove (Tested for Use with Chemotherapy Drugs) Proprietary Name: Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs) Common Name: Patient Examination Glove Classification Name: Patient Examination Glove Product Code: LZA, LZC Classification Panel: General Hospital Regulation #: 21 CFR 880.6250
### Predicate Device
Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Tested for Use with Chemotherapy Drugs K160562
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image features the logo for Medline, a medical supply company. The logo consists of the word "MEDLINE" in bold, white sans-serif font, positioned to the left of a stylized white cross. The cross is oriented diagonally, with its top point extending upwards and to the right. The background is a solid dark blue color, providing a strong contrast to the white text and cross. A small gray rectangle is visible in the upper right corner of the image.
e Medline Place
#### Reference Device
Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile. Tested for Use with Chemotherapy Drugs K151750
#### Device Description
The Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs) are non-sterile single use only, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between the patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, white colored nitrile gloves featuring an inner coating of colloidal oatmeal USP. The gloves are offered in five sizes, extra-small, small, medium, large and extra-large.
The gloves are designed and manufactured per ASTM D6319-10 standard and tested for use with chemotherapy drugs per ASTM D6978-05 (reapproved 2013).
### Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Chemotherapy Drug Tested | Breakthrough Time<br>(in minutes) |
|---------------------------------------|-----------------------------------|
| Blenoxane 15 mg/ml | >240 |
| Bortezomib 1 mg/ml | >240 |
| Busulfan 6 mg/ml | >240 |
| Carmustine (BCNU) 3.3 mg/ml | 11.9 |
| Cetuximab (Eributux) 2 mg/ml | >240 |
| Cisplatin 1.0 mg.ml | >240 |
| Cyclophosphamide (Cytoxan) 20.0 mg/ml | >240 |
| Cytarabine 100 mg/ml | >240 |
| Dacarbazine (DTIC)10.0 mg/ml | >240 |
| Daunorubicin 5 mg/ml | >240 |
| Docetaxel 10.0 mg/ml | >240 |
| Doxorubicin Hydrochloride 2.0 mg/ml | >240 |
| Ellence 2 mg/ml | >240 |
| Etoposide (Toposar) 20.0 mg/ml | >240 |
| Fludarabine 25 mg/ml | >240 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, set against a blue background. To the right of the word is a white star-like symbol with four points. The background is a solid blue color, providing a stark contrast to the white text and symbol.
| Fluorouracil 50.0 mg/ml | >240 |
|---------------------------------|------|
| Gemcitabine (Gemzar) 38 mg/ml | >240 |
| Idarubicin 1 mg/ml | >240 |
| Ifosfamide 50.0 mg/ml | >240 |
| Irinotecan 20.0 mg/ml | >240 |
| Mechlorethamine HCI 1.0 mg/ml | >240 |
| Melphalan 5 mg/ml | >240 |
| Methotrexate 25 mg/ml | >240 |
| Methotrexate 50 mg/ml | >240 |
| Mitomycin C 0.5 mg/ml | >240 |
| Mitoxantrone 2.0 mg/ml | >240 |
| Oxaliplatin 5 mg/ml | >240 |
| Paclitaxel (Taxol) 6.0 mg/ml | >240 |
| Paraplatin 10 mg/ml | >240 |
| Pemetrexed Disodium 25 mg/ml | >240 |
| Raltitrexed 0.5 mg/ml | >240 |
| Rituximab 10 mg/ml | >240 |
| Thiotepa 10.0 mg/ml | 18.6 |
| Trisonex 0.1 mg/ml | >240 |
| Vidaza (5-Azacytidine) 25 mg/ml | >240 |
| Vinblastine 1 mg/ml | >240 |
| Vinorelbine 10 mg/ml | >240 |
| Vincristine Sulfate 1.0 mg/ml | >240 |
| | |
Please note that the following drugs have extremely low permeation times:
Carmustine (BCNU) (3.3 mg/ml) 11.9 minutes
Thiotepa (10.0 mg/ml) 18.6 minutes
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Medline. The logo consists of the word "MEDLINE" in a sans-serif font, with the letters stacked on top of each other. To the right of the word is a white starburst symbol. The background of the logo is a solid blue color.
## Summary of Technological Characteristics
### TABLE 1: Comparison of Proposed and Predicate Devices
| Device<br>Characteristic | Proposed<br>Device | Predicate<br>Device | Reference<br>Device | Comparison<br>Analysis |
|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Product Name | Medline Nitrile<br>Powder-Free<br>Examination Glove<br>with Colloidal<br>Oatmeal USP (Tested<br>for Use with<br>Chemotherapy Drugs) | Nitrile Powder Free<br>Examination Glove<br>with Colloidal<br>Oatmeal USP Tested<br>for Use with<br>Chemotherapy Drugs | Powder Free Nitrile<br>Patient Examination<br>Glove, White<br>Colored, Non Sterile.<br>Tested for Use with<br>Chemotherapy Drugs | N/A |
| 510(k) Reference | K162982 | K160562 | K151750 | N/A |
| Product Owner | Medline Industries,<br>Inc. | Hartalega | Kossan | Different |
| Product Code | LZA, LZC | LZA., LZC | LZA, LZC | Same |
| Intended Use | A patient examination<br>glove is a disposable<br>device intended for<br>medical purposes that<br>is worn on the<br>examiner's hand to<br>prevent contamination<br>between patient and<br>examiner.<br>These gloves were<br>tested for use with<br>chemotherapy drugs<br>as per ASTM D6978-<br>05 (Reapproved 2013)<br>Standard Practice for<br>Assessment of<br>Medical Gloves to<br>Permeation by<br>Chemotherapy Drugs | A non-sterile<br>disposable device<br>intended for medical<br>purpose that is worn<br>on the examiner's<br>hand to prevent<br>contamination<br>between patient and<br>examiner. These<br>gloves were tested<br>for use with<br>Chemotherapy Drugs<br>in accordance with<br>ASTM D6978-05,<br>Standard Practice for<br>Assessment of<br>Resistance of<br>Medical Gloves to<br>Permeation by<br>Chemotherapy Drugs | A patient<br>examination glove is<br>a disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's<br>hand to prevent<br>contamination<br>between patient and<br>examiner.<br>These gloves were<br>tested for use with<br>chemotherapy drugs<br>as per ASTM<br>D6978-05<br>(Reapproved 2013)<br>Standard Practice for<br>Assessment of<br>Medical Gloves to<br>Permeation by<br>Chemotherapy Drugs | Same |
| Regulation<br>Number | 21 CFR 880.6250 | 21 CFR 880.6250 | 21 CFR 880.6250 | Same |
| Materials | Nitrile<br>Colloidal oatmeal | Nitrile<br>Colloidal oatmeal | Nitrile | Same as K160562 |
| Color | White | White | White | Same |
| Sizes | x-small, small, medium, large, x-large | x-small, small, medium, large, x-large | x- small, small, medium, large, x-large, xx-large | Same |
| Dimensions – Length | Complies with ASTM D6319-10<br>230mm min. | Complies with ASTM D6319-10<br>240mm min. | Complies with ASTM D6319-10<br>230mm min. | Same as K151750 and similar to K160562 which is 10mm longer than the subject device. |
| Dimensions – Width | Complies with ASTM D6319-10<br>X-small – 70±10mm<br>Small – 80±10mm<br>Medium – 95±10mm<br>Large – 110±10mm<br>X-large – 120±10mm | Complies with ASTM D6319-10<br>Medium – 95mm | Complies with ASTM D6319-10<br>X-small – 70-80mm<br>Small – 80-90mm<br>Medium – 90-100mm<br>Large – 101-11mm<br>X-large – 111-121mm<br>XX-large – 121-131mm | Same as K160562 and similar to K151750. |
| Dimensions – Thickness | Complies with ASTM D6319-10<br>Palm – 0.05mm min.<br>Finger – 0.05mm min. | Complies with ASTM D6319-10<br>Palm – 0.07mm min.<br>Finger – 0.07mm min. | Complies with ASTM D6319-10<br>Palm – 0.05mm min.<br>Finger – 0.05mm min. | Same as K151750 and similar to K160562 |
| Physical Properties | Complies with ASTM D6319-10 minimum: | Complies with ASTM D6319-10 minimum: | Complies with ASTM D6319-10 minimum: | Same |
| | Tensile Strength:<br>Before Aging ≥14 MPa, min.<br>After Aging ≥14 MPa, min.<br><br>Elongation:<br>Before Aging 500%, min.<br>After Aging 400%, min. | Tensile Strength:<br>Before Aging ≥14 MPa, min.<br>After Aging ≥14 MPa, min.<br><br>Elongation:<br>Before Aging 500%, min.<br>After Aging 400%, min. | Tensile Strength:<br>Before Aging ≥14 MPa, min.<br>After Aging ≥14 MPa, min.<br><br>Elongation:<br>Before Aging 500%, min.<br>After Aging 400%, min. | |
| Freedom from Holes | Complies with ASTM D6319-10 and ASTM D5151-06<br>(reapproved 2011) G-1, AQL 1.5 | Complies with ASTM D6319-10 and ASTM D5151-06 (reapproved 2011) AQL 1.5 | Complies with ASTM D6319-10 and ASTM D5151-06 (reapproved 2011) G-1, AQL 1.5 | Same |
| Powder or Powder-Free | Powder-Free | Powder-Free | Powder-Free | Same |
| Residual Powder | Complies with ASTM D6319-10 <2mg per glove | Complies with ASTM D6319-10 <2mg per glove | Complies with ASTM D6319-10 <2mg per glove | Same |
| Contact<br>Durations | Limited ≥24 hours | Limited ≥24 hours | Limited ≥24 hours | Same |
| Biocompatibility | AAMI/ANSI/ISO<br>10993-10:<br>Not a skin irritant<br>Not a skin sensitizer | AAMI/ANSI/ISO<br>10993-10:<br>Not a skin irritant<br>Not a skin sensitizer | AAMI/ANSI/ISO<br>10993-10:<br>Not a skin irritant<br>Not a skin sensitizer | Same |
| Sterile vs. Non-<br>Sterile | Non-sterile | Non-sterile | Non-sterile | Same |
| Chemotherapy<br>Drugs Tested with<br>Minimum<br>Breakthrough<br>Detection Time as<br>tested per ASTM<br>D6978 | Blenoxane 15 mg/ml<br>>240 min. | | | Different |
| | Bortezomib 1 mg/ml<br>>240 min. | | | Different |
| | Busulfan 6 mg/ml<br>>240 min. | | | Different |
| | Carmustine (BCNU)<br>3.3 mg/ml<br>11.9 min. | Carmustine<br>3.3 mg/ml<br>10.1 min. | Carmustine (BCNU)<br>3.3 mg/ml<br>10.1 min. | Same |
| | Cetuximab (Erbitux)<br>2 mg/ml<br>>240 min. | | | Different |
| | Cisplatin 1.0 mg.ml<br>>240 min. | Cisplatin 1.0 mg.ml<br>>240 min. | Cisplatin 1.0 mg.ml<br>>240 min. | Same |
| | Cyclophosphamide<br>(Cytoxan) 20.0 mg/ml<br>>240 min. | Cyclophosphamide<br>(Cytoxan) 20 mg/ml<br>>240 min. | Cyclophosphamide<br>(Cytoxan) 20 mg/ml<br>>240 min. | Same |
| | Cytarabine 100 mg/ml<br>>240 min. | | Cytarabine 100<br>mg/ml<br>>240 min. | Different |
| | Dacarbazine<br>(DTIC)10.0 mg/ml<br>>240 min. | Dacarbazine<br>(DTIC)10.0 mg/ml<br>>240 min. | Dacarbazine<br>(DTIC)10.0 mg/ml<br>>240 min. | Same |
| | Daunorubicin 5<br>mg/ml<br>>240 min. | | | Different |
| | Docetaxel 10.0 mg/ml<br>>240 min. | | | Different |
| | Doxorubicin<br>Hydrochloride 2.0<br>mg/ml<br>>240 min. | Doxorubicin<br>Hydrochloride 2.0<br>mg/ml<br>>240 min. | Doxorubicin<br>Hydrochloride 2.0<br>mg/ml<br>>240 min. | Same |
| | Ellence 2 mg/ml<br>>240 min. | | | Different |
| | Etoposide (Toposar)<br>20.0 mg/ml<br>>240 min. | Etoposide (Toposar)<br>20.0 mg/ml<br>>240 min. | Etoposide (Toposar)<br>20.0 mg/ml<br>>240 min. | Same |
| | Fludarabine 25 mg/ml<br>>240 min. | | | Different |
| Fluorouracil 50.0<br>mg/ml<br>>240 min. | Fluorouracil 50.0<br>mg/ml<br>>240 min. | Fluorouracil 50.0<br>mg/ml<br>>240 min. | Same | |
| Gemcitabine<br>(Gemzar) 38 mg/ml<br>>240 min. | | | Different | |
| Idarubicin 1 mg/ml<br>>240 min. | | | Different | |
| Ifosfamide 50.0<br>mg/ml<br>>240 min. | | Ifosfamide (IFEX)<br>50 mg/ml<br>>240 min. | Different | |
| Irinotecan 20.0 mg/ml<br>>240 min. | | | Different | |
| Mechlorethamine<br>HCI 1.0 mg/ml<br>>240 min. | | | Different | |
| Melphalan 5 mg/ml<br>>240 min. | | | Different | |
| Methotrexate 25<br>mg/ml<br>>240 min. | Methotrexate 25<br>mg/ml<br>>240 min | Methotrexate 25<br>mg/ml<br>>240 min. | Same | |
| Methotrexate 50<br>mg/ml<br>>240 min. |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.