K162968 · Vortran Medical Technology 1, Inc. · BTL · Apr 10, 2017 · Anesthesiology
Device Facts
Record ID
K162968
Device Name
VORTRAN GO2VENT
Applicant
Vortran Medical Technology 1, Inc.
Product Code
BTL · Anesthesiology
Decision Date
Apr 10, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5925
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
This Device is to be used by properly trained personnel to deliver emergency, short term, constant flow, pressure cycled ventilator support on patients weighing 10kg and above.
Device Story
VORTRAN GO2VENT is a gas-powered, pressure-cycled emergency ventilator providing short-term ventilatory support. Input is a 50 psig gas source; device transforms flow via a pulmonary modulator (exhalation valve) and internal spring-loaded diaphragm mechanism. Output is constant flow ventilation in pressure control or pressure support modes. Operated by trained personnel in emergency settings; device is secured to patient. Clinicians adjust rate and pressure dials to achieve desired tidal volume, I-time, and I:E ratio. Device includes a redundant pop-off valve and FiO2 entrainment control (50% or 100%). Benefits include portable, simple-to-use ventilatory support for patients ≥10kg. MR conditional for 3-Tesla or less.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including performance verification against ASTM F920-93 and ISO 10651-5 standards, MR compatibility testing (ASTM F2052), and fatigue testing of Beryllium-Copper springs.
Technological Characteristics
Gas-powered, pressure-cycled ventilator. Materials: K-Resin, Polycarbonate, Beryllium-Copper springs, HDPE, Silicone, Nylon. MR Conditional (3-Tesla, 10,000-gauss/cm). Dimensions/form factor: handheld/portable. Connectivity: None. Sterilization: Not specified. No software/algorithm.
Indications for Use
Indicated for emergency, short-term, constant flow, pressure-cycled ventilatory support in patients weighing 10kg and above. To be used by trained personnel.
Regulatory Classification
Identification
A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are connected and appear to be part of a single, flowing design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 10, 2017
Vortran Medical Technology 1, Inc. James Lee Executive Vice President & COO 21 Goldenland Court, #100 Sacramento, California 95834
Re: K162968
Trade/Device Name: VORTRAN® GO2VENT™ Regulation Number: 21 CFR 868.5925 Regulation Name: Powered Emergency Ventilator Regulatory Class: Class II Product Code: BTL Dated: March 2, 2017 Received: March 10, 2017
Dear James Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tina
Kiang-S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162968
Device Name
VORTRAN® GO2VENT™
## Indications for Use (Describe)
This Device is to be used by properly trained personnel to deliver emergency, short term, constant flow, pressure cycled ventilator support on patients weighing 10kg and above.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary / K162968 - VORTRAN® GO2VENT™
### I. SUBMITTER
VORTRAN® Medical Technology 1, Inc. 21 Goldenland Court, Suite 100 Sacramento, CA 95834 USA
Phone: (800) 434-4034 Fax: (916) 648-9751
Contact Person: James Lee Date Prepared: March 31, 2017
### II. DEVICE
Name of Device: VORTRAN® GO2VENT™ Common or Usual Name: Ventilator, Emergency, Powered (Resuscitator) Classification Name: Powered Emergency Ventilator Regulation Number: 21 CFR 868.5925 Regulatory Class: II Product Code: BTL Classification Advisory Committee: Anesthesiology Review Advisory Committee: Anesthesiology
### III. PREDICATE DEVICE
| K041473 | VORTRAN® Automatic Resuscitator (VAR-Plus)<br>(Primary Predicate Device) |
|---------|----------------------------------------------------------------------------|
| K153733 | VORTRAN® Manometer<br>(Reference Device – For Material Compatibility Only) |
## IV. DEVICE DESCRIPTION
The VORTRAN® GO2VENT™ provides short term, constant flow, pressure cycled ventilatory support in either pressure control or pressure support modes on patients weighing 10kg and above. In the pressure support mode, the rate dial of the VORTRAN® GO2VENT™ is set so that the baseline pressure is above the set PEEP allowing the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP. The device includes the pulmonary modulator (an exhalation valve that opens at PIP and closes at PEEP) and a patient connector tee to supply gas flow, entrain additional air, and provides a redundant pop-off valve for patient care. The working mechanism of the VORTRAN® GO2VENTTM consists of a moving diaphragm which adds or subtracts spring force when it is moved from a horizontal to a vertical position, the addition or subtraction of spring force will affect the PIP setting by 1~3 cm-H2O. The VORTRAN® GO2VENT™ will function in any position as long as the final adjustments are made in a secured position (strapped or taped to the patient).
{4}------------------------------------------------
# 510(k) Summary / K162968 - VORTRAN® GO2VENTTM
The VORTRAN® GO2VENT™ is not an ICU stand alone ventilator with multiple monitoring features. Set up and use of the VORTRAN® GO2VENT™ is simple. Set desired flow and adjust pressure dial to obtain desired I-time and/or tidal volume (see tidal volume chart in instructions), and adjust rate dial to obtain desired rate and I to E ratio.
| Device Model Number: | 6123-10 |
|---------------------------|-------------------------------------------------------|
| Device Accessories: | Elbow Flex Hose, Oxygen Tubing, Pressure Manometer |
| Interaction with Patient: | The Elbow Flex Hose has indirect contact with patient |
### INDICATIONS FOR USE V.
This device is intended to be use by properly trained personnel to deliver emergency, short term, constant flow, pressure cycled, ventilatory support on patients weighing 10kg and above.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
# A. Technical Modifications:
MR conditional - The new model of the VORTRAN® Automatic Resuscitator has been modified with new springs made of Beryllium-Copper instead of the Stainless-Steel. This change was verified (in the Shellock MR Testing on Vortran Products Report) to make the device MR Conditional. The report concluded that the VORTRAN® GO2VENT™ can be used in an MRI environment with a static magnetic field of 3-Tesla or less, and a spatial gradient magnetic field of 10,000-gauss/cm or less. Shellock performs their tests based on ASTM F2052 - Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment. We have verified that the performance of the beryllium-copper springs is substantially equivalent to the stainless-steel springs.
New Entrainment Connectors for the 50% and 100% FiO2 Delivery - The new device connectors include a combination knob that can be rotated from the 100% FiO2 (a fully closed position of entrainment ports) to the 50% FiO2 (a position of having entrainment ports open) versus the two connectors of the predicate device that the 100% FiO2 connector should be completely removed to access the 50% FiO2 connector. The modification facilitates the FiO2 change from 50% to 100% FiO2 and vice versa. All testing data have shown and verified that the change in the knobs has not degraded the performance and the device delivers the required specifications.
{5}------------------------------------------------
# 510(k) Summary / K162968 – VORTRAN® GO2VENT™
The tables below show the technological differences and similarities of our VORTRAN® GO2VENT™ compared to the predicate device (VAR-Plus).
# Operational Characteristics:
| | NEW DEVICE | PREDICATE DEVICE |
|--------------------------------------------------|-----------------------------------------------------------------|--------------------------------------------|
| Device Name | VORTRAN® GO2VENT™ | VORTRAN® Automatic Resuscitator (VAR-Plus) |
| 510(k) Number | N/A | K041473 |
| Maximum Inspiratory Flow | 40 L/min<br>Same as predicate device | 40 L/min |
| Ventilatory Frequency | Auto-adjusting to lung capacity [c]<br>Same as predicate device | Auto-adjusting to lung capacity [c] |
| Peak Pressure Range | 10 - 45 cm-H2O [d]<br>Same as predicate device | 10-45 cm-H2O [d] |
| PEEP | 2 - 9 cm-H2O [d]<br>Same as predicate device | 2 - 9 cm-H2O [d] |
| Required Source Pressure | 50 psig<br>Same as predicate device | 50 psig |
| Dead Space | 4 ± 3 mL<br>Same as predicate device | 4 ± 3 mL |
| Inspiratory Resistance | 3 ± 1 cm-H2O / L / sec<br>Same as predicate device | 3 ± 1 cm-H2O / L / sec |
| Expiratory Resistance | 3 ± 1 cm-H2O / L / sec<br>Same as predicate device | 3 ± 1 cm-H2O / L / sec |
| High Pressure Pop-off | Yes, 60 cm H2O<br>Same as predicate device | Yes, 60 cm H2O |
| Visual or Audible Indication<br>of High Pressure | Yes<br>Same as predicate device | Yes<br>Same as predicate device |
{6}------------------------------------------------
# 510(k) Summary / K162968 – VORTRAN® GO2VENT™
# Functional Characteristics:
| | NEW DEVICE | | PREDICATE DEVICE | |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Device Name | VORTRAN® GO2VENT™ | | VORTRAN® Automatic Resuscitator (VAR-Plus) | |
| 510(k) Number | N/A | | K041473 | |
| FiO2 Devliery | 50% FiO2 by entraining room air | FiO2 of >85% when supplied with 100% O2 | 50% FiO2 by entraining room air | FiO2 of >85% when supplied with 100% O2 |
| | Supply 100% O2<br>[A] | Image: B<br>Supply 100% O2<br>[B] | Supply 100% O2<br>[C] | Supply 100% O2<br>[D] |
| Method of<br>Changing<br>Delivery of<br>FiO2 when<br>supplied with<br>100% O2 | [A] Delivers FiO2 at<br>50% by entraining<br>room air when<br>setting the selector<br>dial to the 50%<br>mark on the new<br>patient tee | [B] Delivers FiO2 of<br>>85% by NOT<br>entraining room air<br>when setting the<br>selector dial to the<br>100% mark on the<br>new patient tee | [C] Delivers FiO2 at<br>50% by entraining<br>room air when the<br>100% adaptor is<br>removed, exposing<br>the 50%<br>entrainment nozzle<br>on the patient tee | [D] Delivers FiO2 of<br>>85% by NOT<br>entraining room air<br>when the 100%<br>adaptor is<br>connected to the<br>patient tee |
| Oxygen<br>Concentration | Delivers FiO2 >85% or 50% by entraining room air<br>Same as predicate device | | Delivers FiO2 >85% or 50% by entraining room air<br>[e] | |
Image /page/6/Figure/5 description: This image shows a comparison between a new device and a predicate device. The new device is labeled as "VORTRAN® GO2VENT™" with a 510(k) number of N/A. The predicate device is labeled as "VORTRAN® Automatic Resuscitator (VAR-Plus)" with a 510(k) number of K041473. The image shows the devices in two different scenarios: 50% FiO2 by entraining room air and FiO2 of >85% when supplied with 100% O2.
{7}------------------------------------------------
# 510(k) Summary / K162968 – VORTRAN® GO₂VENT™
# Material of Construction:
| | NEW DEVICE | PREDICATE DEVICE | |
|----------------------------|--------------------------------------------------|-----------------------------------------------|----------------------|
| Device Name | VORTRAN® GO2VENT™ | VORTRAN® Automatic Resuscitator<br>(VAR-Plus) | VORTRAN® Manometer |
| 510(k) Number | N/A | K041473 | K153733 |
| Date SE Decision | NA | July 15, 2004 | September 16, 2016 |
| Housing | K-Resin®, Polycarb<br>Same as predicate device | K-Resin® | Polycarb |
| Internal Springs | Beryllium-Copper [f]<br>Same as predicate device | 302 Stainless Steel | Beryllium-Copper [f] |
| Pressure and Rate<br>Dials | HDPE<br>Same as predicate device | DOW® HDPE 12450 color blue | Acetal |
| One-Way Valve<br>Body | HDPE<br>Same as predicate device | DOW® HDPE 12450 | N/A |
| Flapper Valve | Silicone<br>Same as predicate device | Silicon Polymer | N/A |
| Diaphragm | Silicone<br>Same as predicate device | Silicone | SILICONE/<br>Natural |
| Disk Center | Nylon<br>Same as predicate device | Nylon | N/A |
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility Testing ●
| Device Nature of Body Contact Category: | | | Duration of Contact for the Device: | |
|-----------------------------------------|-------------|----------------------------------|----------------------------------------------------------------------|------------------------------|
| External Communicating | | | Up to 30 days | |
| | Part<br>No. | Part Description | Patient Contact Status<br>Substantial Equivalence<br>Device (510K #) | Material/Color |
| 1. | 6007 | 100% - 50% VAR Nozzle | Indirect patient contact<br>K153733 | Polycarbonate/White |
| 2. | 6008 | FiO2 Controller Knob | Indirect patient contact<br>K041473 | HDPE/Green |
| 3. | 6009 | Entrainment Barrel | Indirect patient contact<br>K153733 | Polycarbonate/White |
| 4. | 2015 | One-Way Valve Body | Indirect patient contact<br>K041473 | HDPE/Natural |
| 5. | 2016 | One-Way Valve Flapper | Indirect patient contact<br>K153733 | Silicone/Red |
| 6. | 6005 | Pop-Off Valve Cap | No patient contact | Polycarbonate/Blue |
| 7. | 6010 | Pop-Off Valve Spring | No patient contact | Beryllium Copper |
| 8. | 2012 | Pop-Off Valve Piston | Indirect patient contact<br>K041473 | HDPE/Red |
| 9. | 6006 | Patient Tee | Indirect patient contact<br>K041473 | K-Resin®/Clear/Blue |
| 10. | 2291 | Manometer Assembly | Indirect patient contact<br>K153733 | N/A |
| 11. | 6003 | Modulator Bottom, Single<br>Port | Indirect patient contact<br>K153733 | Polycarbonate/Clear/<br>Blue |
| 14. | 2182B | Diaphragm | No patient contact | Silicone/Natural |
| | 2181B | Hard Center | Indirect patient contact<br>K041473 | Nylon/Natural |
{8}------------------------------------------------
# 510(k) Summary / K162968 - VORTRAN® GO2VENT™
#### Biocompatibility Testing (Continued) ●
| Device Nature of Body Contact Category:<br>External Communicating | | | Duration of Contact for the Device:<br>Up to 30 days | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|----------------------|-------------------------------------------------------------------|------------------------------|
| | Part No. | Part Description | Patient Contact Status<br>Substantial Equivalence Device (510K #) | Material/Color |
| 15. | 6004 | Pressure Dial Spring | No patient contact | Beryllium Copper |
| 17. | 6002 | Modulator Top | No patient contact | Polycarbonate/Clear/<br>Blue |
| 18. | 6017 | Rate Dial | No patient contact | HDPE/Blue |
| 19. | 6016 | Pressure Dial | No patient contact | HDPE/Blue |
| Discussion:<br>1) Rationale for Not Needing Biocompatibility Testing: Considering the material<br>used in the components of the new device (VORTRAN® GO2VENT™) and the fact<br>that the same material used as those in the predicated devices manufactured by<br>VORTRAN Medical and produced with the same manufacturing process, the<br>material of the new device is compatible and requires no additional biocompatibility | | | | |
#### Animal Study ●
None
- Non-Clinical Performance Testing/Performance Data/Compliance with Performance Standards
The VORTRAN® GO2VENT™, just as the VAR-Plus, meets the "Standard Specification for Minimum Performance and Safety Requirements for Resuscitators Intended for Use with Humans" ASTM Designation: F 920 -93. A separate non-clinical test showed that all requirements were also met for ISO 10651-5, "Lung ventilators for medical use - Particular requirements for basic safety and essential performance - Part 5: Gas-powered emergency resuscitators". The determination for substantial equivalence was based on a comparison of performance data between the new device and predicate device (FDA 510(k) No. K041473). The predicate device performance tests were documented. To demonstrate substantial equivalence, the new device was also tested using "ASTM F920-93 (Reapproved 1999)". In addition, the features that are considered technological differences were tested. The function of the new FiO2 entrainment feature (at 50% and 100% FiO2 settings) and device shelf-life were tested with the results documented. The new Beryllium-Copper springs were tested for performance and fatigue.
{9}------------------------------------------------
# 510(k) Summary / K162968 - VORTRAN® GO2VENT™
- . Clinical Studies None
#### VII. CONCLUSIONS
- The VORTRAN® GO2VENT™ is substantially equivalent to a predicate ● device: The VAR-Plus. The results from the nonclinical tests show that the device achieves predefined acceptance criteria for all assessments that were previously performed on the predicate device. VORTRAN® GO2VENT™, just as the VAR-Plus, meets the "Standard Specification for Minimum Performance and Safety Requirements for Resuscitators Intended for Use with Humans" ASTM Designation: F 920 - 93. All requirements were also met for ISO 10651-5, "Lung ventilators for medical use - Particular requirements for basic safety and essential performance - Part 5: Gaspowered emergency resuscitators". The VORTRAN® GO2VENT™ has been shown to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.