K162956 · Contex International Technologies (Canada), Inc. · QYX · Sep 22, 2017 · Cardiovascular
Device Facts
Record ID
K162956
Device Name
NOWCARDIO SYSTEM
Applicant
Contex International Technologies (Canada), Inc.
Product Code
QYX · Cardiovascular
Decision Date
Sep 22, 2017
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The NowCardio™ System is an ambulatory cardiac monitoring system prescribed by a physician or other qualified healthcare professional for out-patient individuals with suspected non-lethal arrhythmias such as: Atrial Fibrillation, Bradycardia, Tachycardia, Pause. It continuously monitors, records, and stores electrocardiographic (ECG) data. If equipped with a transmitter, the NowCardio will also periodically transmit ECG Cardiac Event to a remote computer server. Both recorded and transmitted ECG data are available to HealthCare Providers for analysis and reporting. NowCardio™ is intended to be used with adult patients 22 years and older, but not designed for pediatric patients.
Device Story
NowCardio™ System is a wearable, wireless ambulatory cardiac monitor for continuous ECG recording (up to 32 days). System components: wearable sensor; cradle for configuration/charging/data download; Remote Data Management System (RDMS) server. Optional transmitter sends ECG data to RDMS via Bluetooth/cellular. Healthcare providers access RDMS to analyze, interpret, and report on ECG results. Device aids in detection of non-lethal arrhythmias; supports patient-triggered events. Used in outpatient settings; operated by patients (wearable) and healthcare professionals (analysis).
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including system validation, mechanical verification, software/firmware verification, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-1), and defibrillation-proof testing.
Technological Characteristics
Wearable sensor, cradle, and remote server/software. Connectivity via Bluetooth and cellular. Continuous ECG recording with periodic transmission. Complies with IEC 60601-1 (safety), IEC 60601-2-47 (ambulatory ECG), IEC 62304 (software lifecycle), and ISO 14971 (risk management).
Indications for Use
Indicated for adult patients (22+ years) requiring ambulatory cardiac monitoring for suspected non-lethal arrhythmias, including atrial fibrillation, bradycardia, tachycardia, and pauses. Not for pediatric use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with a stylized image of an eagle, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" being the largest and boldest, followed by "U.S. FOOD & DRUG" and then "ADMINISTRATION" in a smaller font size. The logo is simple and professional, conveying the organization's authority and mission.
November 15, 2023
Contex International Technologies (canada), Inc. Ricardo Rodriguez Coo 2798 Thamesgate Drive, Units 5&6 Mississauga, ONTARIO L4T 4E8 Canada
Re: K162956
Trade/Device Name: Nowcardio System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including St-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: QYX, MHX, DSI
Dear Ricardo Rodriguez:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated September 22, 2017. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Kozen, OHT2: Office of Cardiovascular Devices, 301-796-5813, jennifer.kozen@fda.hhs.gov.
Sincerely,
Jennifer W. Shih -S
Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of a human figure.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 22, 2017
Contex International Technologies (Canada), Inc. Ricardo Rodriguez Chief Operating Officer 2798 Thamesgate Drive, Units 5&6 Mississauga, CA L4T 4E8 Ontario
Re: K162956
Trade/Device Name: NowCardio™ System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, DSI Dated: August 14, 2017 Received: August 16, 2017
Dear Ricardo Rodriguez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Hillemann
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162956
Device Name NowCardio™ System
Indications for Use (Describe)
The Now Cardio™ System is an ambulatory cardiac monitoring system prescribed by a physician or other qualified healthcare professional for out-patient individuals with known non-lethal arrhythmias such as:
- · Atrial Fibrillation
- · Bradycardia
- Tachycardia
- · Pause
It continuously monitors, records, and stores electrocardiographic (ECG) data. If equipped with a transmitter, the Now Cardio™ will also periodically transmit ECG Cardiac Event to a remote computer server. Both recorded and transmitted ECG data are available to HealthCare Providers for analysis and reporting.
Now Cardio™ is intended to be used with adult patients 22 years and older, but not designed for pediatric patients.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### I. Submitter:
| Company: | Contex International Technologies Inc. |
|----------------|----------------------------------------|
| Address | 2798 Thamesgate Drive Unit 5&6 |
| | Mississauga, ON L4T 4E8 Canada |
| Contact: | Ricardo Rodriguez |
| Phone: | 905-677-8324 ext. 202 |
| Fax: | 905-676-8324 |
| Email: | ricardo@contexcanada.com |
| Date Prepared: | August 03, 2017 |
#### II. Device:
| Trade Name: | NowCardio™ System |
|----------------------|---------------------------------------------------------------------------------------------------------------|
| Common Name: | NowCardio™ |
| Classification Name: | Arrhythmia Detector and Alarm |
| Product Code: | DSI, MHX |
| Regulation: | 21 CFR 870.1025<br>Patient Physiological Monitor (with arrhythmia detection)<br>Arrhythmia detector and alarm |
| Class: | Class II, Special Controls |
#### III. PREDICATE DEVICE(S):
- 1. NUVANT MCT System, cleared February 6, 2014 (NUVANT™ MCT) K133701 rebranded by Medtronic as SEEQ™ MCT. This is the primary predicate device.
- 2. BodyGuardian System, manufactured by Preventice, Inc. and cleared under 510(k) number K121197.
#### IV. DEVICE DESCRIPTION:
The NowCardio™ System is a wearable, wireless arrhythmia detection system that provides continuous cardiac monitoring functionality for an extended period of up to 32 days. The NowCardio System is prescribed by healthcare providers to monitor, record, and report on a patient's electrocardiographic (ECG) data. The NowCardio™ system consists of: (1) a wearable electronic component that collects and monitors a patient's ECG data referred to as a NowCardio™ Sensor; (2) a custom cradle, referred to as the NowCardio™ Cradle, that is used to configure/charge the NowCardio™ Sensor in addition to providing ECG data download capability; and (3) a Remote Data Management System (RDMS) which is a remote server with software that will allow
{5}------------------------------------------------
qualified healthcare providers to receive, analyze, interpret, and report on the results of the ECG test. The NowCardio™ System can be equipped with an optional Transmitter. The NowCardio™ Transmitter is used to transmit data acquired from the NowCardio™ Sensor to the RDMS as cardiac events are detected or patient triggered.
### INTENTED USE AND INDICATIONS FOR USE V.
The NowCardio™ System is an ambulatory cardiac monitoring system prescribed by a physician or other qualified healthcare professional for out-patient individuals with suspected non-lethal arrhythmias such as:
- . Atrial Fibrillation
- . Bradycardia
- . Tachycardia
- . Pause
It continuously monitors, records, and stores electrocardiographic (ECG) data. If equipped with a transmitter, the NowCardio will also periodically transmit ECG Cardiac Event to a remote computer server. Both recorded and transmitted ECG data are available to HealthCare Providers for analysis and reporting.
NowCardio™ is intended to be used with adult patients 22 years and older, but not designed for pediatric patients.
### VI. SUBSTANTIAL EQUIVALENCE
Both predicates and the NowCardio™ System are ambulatory cardiac monitors that feature body worn Sensors that monitor and continuously record and periodically transmit ECG patient data. Both predicates and the NowCardio™ System will transmit the ECG data from the Sensor to Transmitter/Phone using Bluetooth, this data is then sent to a remote server with software that will allow HealthCare providers to access, review, and produce reports. There are no substantial differences between technological characteristics of the devices.
The indications for use for both the NowCardio™ System and its predicates state that the devices will provide monitoring of non-lethal cardiac arrhythmias. There are no substantial differences between the Indications for use of the devices.
Similar performance testing has been conducted on the NowCardio™ System and its predicates. There are no substantial differences in performance testing of the devices. Therefore, the NowCardio™ System is determined to be substantially equivalent to the predicate devices.
{6}------------------------------------------------
The following table compares the NowCardio™ System to the predicate devices with respect to intended use, technological characteristics and principles of operation, etc.
| ITEM | Subject Device:<br>NowCardio™<br>System<br>K162956 | Primary Predicate 1:<br>Medtronic SEEQ™<br>K133701 | Predicate 2:<br>BodyGuardian<br>System<br>K121197 |
|--------------------------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Regulation Number | 21 CFR 870.1025 | Same | Same |
| FDA classification | Class II | Same | Same |
| Product Code(s) | DSI | DSI | MHX, DSI |
| Classification<br>Name: | Arrhythmia<br>Detector and Alarm | Same | Same |
| Patient<br>Environment: | Ambulatory | Same | Same |
| Patient Population: | Non-pediatric, non-<br>critical care patients | Same | Same |
| Transmission<br>characteristics: | Bluetooth and<br>cellular connections | Same | Same |
| Key System<br>Components: | (1) Wearable<br>Sensor<br>(2) Transmitter<br>(3) Remote Server<br>and Software | (1) Wearable Sensor<br>(2) Transmitter<br>(3) Remote Server<br>and Software | (1) Wearable<br>Sensor<br>(2) Smart phone/<br>Transmitter<br>(3) Remote Server<br>and Software |
| Technical<br>Operating<br>Characteristics: | Continuous<br>Recording with<br>periodic<br>transmission of<br>ECG data. | Same | Same |
| Non-Lethal<br>Arrhythmia<br>Detection: | Yes | Yes | Yes |
| Event Trigger | Yes | Yes | Yes |
Table 1: Substantial Equivalence Summary Table
### VII. PERFORMANCE DATA
Nonclinical testing was conducted in full compliance to design control requirements and included the following:
{7}------------------------------------------------
- System Validation testing .
- . Mechanical verification testing
- Software verification testing .
- Firmware verification testing .
- Electrical safety and EMC testing .
- Biocompatibility testing .
- Defibrillation-proof applied parts testing .
The scope of the nonclinical testing summarized in Table 2 demonstrates that the NowCardio™ System is in compliant with FDA-recognized standards and guidance documents.
| Standard | Description | Release Date |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| IEC 60601-1 | Medical Electrical Equipment - General<br>requirements for basic safety and<br>essential performance | Ed. 3 |
| IEC 60601-2-47 | Particular requirements for the basic<br>safety and essential performance of<br>ambulatory electrocardiographic systems | 2012 |
| IEC 60601-1-2 | General requirements for basic safety<br>and essential performance – Collateral<br>standard: Electromagnetic compatibility<br>– Requirements and tests | 2007 |
| ISO 14971 | Application of risk management to<br>medical devices | 2007 |
| IEC 60601-1-6 | Medical electrical equipment – Part 1-6:<br>General requirements for safety –<br>Collateral standard: Usability | 2010 |
| IEC 60601-1-11 | The Collateral Standard for the Home<br>Healthcare Environment | 2010 |
| IEC 62304 | Medical device software - Software life<br>cycle processes | 2006 |
| ISO 15223-1 | Medical devices - Symbols to be used<br>with medical device labels, labeling and<br>information to be supplied | 2007 |
| ISO 10993-1 | Biological evaluation of medical devices<br>– Part 1: Evaluation and testing | 2003 |
| IEC 62366 | Medical devices – Application of<br>usability engineering to medical devices | 2007 |
| EN 1041 | Information supplied by the<br>manufacturer of medical devices | 2008 |
Table 2: Nonclinical Testing Summary
{8}------------------------------------------------
| Guidance for<br>Industry<br>and FDA Staff | Class II Special Controls Guidance<br>Document: Arrhythmia Detector<br>and Alarm | October 28, 2003 |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------|------------------|
| Guidance for<br>Industry<br>and FDA Staff | Providing Information about<br>Pediatric Uses of Medical Devices | May 1, 2014 |
| Guidance for<br>Industry<br>and FDA Staff | Guidance for Industry and FDA on<br>Alternative to Certain Prescription<br>Device Labeling Requirements | 1/21/2000 |
| Guidance for<br>Industry<br>and FDA Staff | Content of Premarket Submissions for<br>Management of Cybersecurity in<br>Medical Devices | October 2, 2014 |
| Guidance for<br>Industry<br>and FDA Staff | Design Considerations for Devices<br>Intended for Home Use | August 8, 2016 |
| Guidance for<br>Industry<br>and FDA Staff | Class II Special Controls Guidance<br>Document: Electrocardiograph<br>Electrodes | July 21, 2011 |
| IEC 60601-1 | clauses 8.5.5.1a, 8.5.5.1b, 8.5.5.2 for<br>DEFIBRILLATION-PROOF APPLIED<br>PARTS testing | 2005/A1:2012 |
# VIII. CONCLUSIONS
The NowCardio™ System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.