OARtrac System with Patient Specific Reusable Universal PSD Sensors
Applicant
Radiadyne, LLC
Product Code
NZT · Radiology
Decision Date
Jun 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The OARtrac® System with patient specific, reusable, pre-calibrated PSD sensors is intended for use in photon radiation therapy to monitor and validate radiation dose during External Beam Therapy and HDR Brachytherapy to the surface of the skin or the rectal prostatic interface.
Device Story
OARtrac System uses plastic scintillation detectors (PSD) to monitor/validate radiation dose during external beam therapy and HDR brachytherapy. Sensors placed on skin (with bolus) or in rectum (via endorectal balloon) measure photon radiation; data transmitted to system for comparison against planned dose. Operated by radiation oncologists/technicians in clinical settings. Output provides dose verification data; allows clinicians to assess treatment delivery and adjust plans if necessary. Benefits include improved accuracy of radiation dose delivery to targeted areas, potentially sparing surrounding organs at risk. Subject device introduces reusability (up to five times) for PSD sensors, requiring validated cleaning/disinfection protocols.
Clinical Evidence
No clinical data. Bench testing only. Validation included biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), EMC (IEC 60601-12), software V&V (IEC 62304), shelf-life (ASTM F1980-07), risk analysis (ISO 14971), dose range verification, and cleaning/disinfection validation for reusability.
Technological Characteristics
Materials: Polyurethane plastic scintillation detectors (PSD). Energy: Photon energy range 0.37-18 MeV. Dose rate range 1.3-17.3 cGy/s. Dose range 27-1200 cGy. Accuracy: +/-6%, 2 sigma. Connectivity: Not specified. Sterilization: Not applicable (reusable via cleaning/disinfection). Software: IEC 62304 compliant.
Indications for Use
Indicated for cancer patients undergoing photon radiation therapy. Used as an adjunct to treatment planning to measure and validate radiation dose to targeted body areas. Indicated for use when adhered to skin with a bolus or inserted into the rectum via a specifically designed endorectal balloon device.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, overlaid with three human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 1, 2017
RadiaDyne, LLC % Mr. Stuart Goldman Senior Consultant Emergo Group 2500 Bee Cave Road, Bldg. 1, Suite 300 AUSTIN TX 78746
Re: K162954
Trade/Device Name: OARtrac® System with Patient Specific Reusable PSD Sensors Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: NZT Dated: May 3, 2017 Received: May 8, 2017
Dear Mr. Goldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael D. O'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162954
Device Name
OARtrac® System with Patient Specific Reusable PSD Sensors
Indications for Use (Describe)
The OARtrac® System with patient specific, reusable, pre-calibrated PSD sensors are intended for use during cancer treatments to measure photon radiation therapy as an adjunct to treatment planning measurement and validation of radiation dose received by the patient to the targeted area of their body, and indicated to the skin with a bolus, or inserted into the rectum to measure the rectal prostatic interface via a specifically designed endorectal balloon device.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 – 510(k) Summary
## OARtrac® System with Patient Specific Reusable PSD Sensors
# K162954
#### 1. Submission Sponsor
RadiaDyne, LLC
10801 Hammerly Blvd.
Suite 220
Houston, TX 77043
USA
Phone number: (281) 759-9600
Contact: John Isham
Title: President & CEO
#### 2. Submission Correspondent
Emergo Global Consulting, LLC
2500 Bee Cave Road, Bldg. 1, Suite 300
Austin, TX 78746
Office Phone: (512) 327-9997
Contact: Stuart R. Goldman, Senior Consultant, RA
Email: project.management@emergogroup.com
#### 3. Date Prepared
May 29, 2017
#### 4. Device Identification
| Trade/Proprietary Name: | OARtrac® System with Patient Specific Reusable PSD Sensors |
|-------------------------|------------------------------------------------------------|
| Common/Usual Name: | Radiation Dose Verification System |
| Classification Name: | Medical charged-particle radiation therapy system |
{4}------------------------------------------------
| Regulation Number: | 892.5050 |
|-----------------------|------------------------------------------------|
| Product Code: | NZT (dosimeter, ionizing radiation, implanted) |
| Device Class: | Class II |
| Classification Panel: | Radiology |
### 5. Legally Marketed Predicate Devices
OARtrac® System with Skin Sensors (K150719)
OARtrac® System (K141154)
#### 6. Device Description
The OARtrac® System with patient specific, reusable, pre-calibrated PSD sensors is intended for use in photon radiation therapy to monitor and validate radiation dose during External Beam Therapy and HDR Brachytherapy to the surface of the skin or the rectal prostatic interface. This dose verification information obtained during the treatment is then used to compare with the planned dose that the Radiation Oncologists expect to provide to their patient. The OARtrac® System itself does not stop the radiation treatment to the patient, or change the radiation delivery, but only provides dose data which a trained Radiation Oncologist can decipher and use to adjust a patient's treatment plan accordingly.
#### 7. Indication for Use Statement
The OARtrac® System with patient specific, reusable, pre-calibrated PSD sensors are intended for use during cancer treatments to measure photon radiation therapy as an adjunct to treatment planning permitting measurement and validation of radiation dose received by the patient to the targeted area of their body, and indicated for use when adhered to the skin with a bolus, or inserted into the rectum to measure the rectal prostatic interface via a specifically designed endorectal balloon device.
#### 8. Substantial Equivalence Discussion
RadiaDyne has chosen its OARtrac® System (K141154) that was cleared for use at the prostatic rectal interface, and their OARtrac® System with Skin Sensors (K150719) that was cleared for use on the surface of the patient's skin as the predicate devices for its OARtrac® System with Patient Specific Reusable PSD Sensors that is the subject of this 510(k) submission. The following table compares the subject device against the two referenced predicate devices with respect to intended use, indications for use, technology and performance testing between them, thus demonstrating the basis for determination of substantial equivalence between these three devices.
{5}------------------------------------------------
| Device | RadiaDyne | | Similarities/Differences | | |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Trade Name: | OARtrac® System<br>with Patient Specific Reusable PSD Sensors | OARtrac® System<br>with Skin<br>Sensors | OARtrac® System | - | |
| 510(k): | Pending | K150719 | K141154 | - | |
| Product Code: | NZT | NZT | NZT<br>PCT | Same | |
| Regulation: | §892.5050 | §892.5050 | §892.5050<br>§892.5720 | Same | |
| Class: | II | II | II | Same | |
| Intended Use: | The OARtrac® System with patient<br>specific, reusable, pre-calibrated PSD<br>sensors is intended for use in photon<br>radiation therapy to monitor and validate<br>radiation dose during External Beam<br>Therapy and HDR Brachytherapy to the<br>surface of the skin or the rectal prostatic<br>interface. | The OARtrac®<br>System with<br>Skin Sensors is<br>intended for<br>use in photon<br>beam<br>radiation<br>therapy to<br>monitor and<br>verify<br>radiation<br>treatment<br>dose at the<br>entrance to<br>the skin. | The<br>OARtrac®<br>System is<br>intended<br>for use in<br>photon<br>beam<br>radiation<br>therapy to<br>monitor<br>and verify<br>radiation<br>treatment<br>dose to the<br>surrounding organs at<br>risk,<br>specifically<br>the<br>prostatic<br>rectal<br>interface,<br>during<br>prostate<br>external<br>beam<br>radiation<br>treatment. | The subject and<br>predicate devices act as<br>a radiation dose<br>verification system, to<br>the prostatic rectal<br>interface, and the skin<br>of the patient, that<br>have been exposed to<br>radiation treatment. | |
| Indications for<br>Use: | The OARtrac® System with patient<br>specific, reusable, pre-calibrated PSD<br>sensors are intended for use during cancer<br>treatments to measure photon radiation<br>therapy as an adjunct to treatment<br>planning permitting measurement and<br>validation of radiation dose received by<br>the patient to the targeted area of their<br>body, and indicated for use when adhered | The OARtrac®<br>System pre-<br>calibrated skin<br>sensors are<br>specifically<br>indicated for<br>use during<br>cancer<br>treatments to | The<br>OARtrac®<br>System is<br>specifically<br>indicated<br>for male<br>prostate<br>cancer<br>treatment | The subject device has<br>expanded indications<br>where the OARtrac®<br>System PSD sensors are<br>now reusable up to five<br>times in total on the<br>same patient. | |
| | the rectum to measure the rectal prostatic | photon beam | photon | | |
| | interface via a specifically designed | therapy as an | beam | | |
| endorectal balloon device. | | adjunct to | therapy as | | |
| | | treatment | an adjunct | | |
| | | planning | to | | |
| | | permitting | treatment | | |
| | | measurement | planning | | |
| | | of radiation | permitting | | |
| | | dose received | measurem | | |
| | | on the surface | ent of in- | | |
| | | of the skin. | vivo | | |
| | | OARtrac® | radiation | | |
| | | System pre- | dose | | |
| | | calibrated skin | received on | | |
| | | sensors are | the | | |
| | | indicated for | anterior | | |
| | | use when | surface of a | | |
| | | adhered to the | | | |
| | | surface of the | modified | | |
| | | | prostate | | |
| | | skin using | Endorectal | | |
| | | medical grade | Balloon | | |
| | | adhesive and | (ERB) | | |
| | | with a medical | device to | | |
| | | grade bolus | monitor | | |
| | | buildup placed | and verify | | |
| | | directly on top | the | | |
| | | of the sensor. | surroundin | | |
| | | | g organs at | | |
| | | | risk, | | |
| | | | specifically | | |
| | | | the | | |
| | | | prostatic | | |
| | | | rectal | | |
| | | | interface. | | |
| Material: | Plastic scintillation detectors (PSD) are<br>made from polyurethane. | Plastic<br>scintillation | Plastic<br>scintillation | Same | |
| | | detectors | detectors | | |
| | | (PSD) are | (PSD) are | | |
| | | made from | made from | | |
| | | polyurethane. | polyuretha | | |
| | | | ne. | | |
| Implantable: | | No | No | Same | |
| Body Location: | No<br>Prostatic rectal interface and the skin of | Skin | Placed in | Same | |
| | the patient. | contacting. | rectum; | | |
| | | | | | |
| | | | inserted | | |
| | | | inside<br>prostate | | |
| | | | | | |
| | | | Endorectal<br>balloon. | | |
| Sterile: | No | No | No | Same | |
| Single Use: | No | Yes | Yes | The PSD sensors used | |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.