The Aero Chrome* Select Breathable Performance Surgical Gowns are sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Aero Chrome* Select Breathable Performance Surgical Gowns meet the Level 4 requirements of the AAMI PB70:2012 Liquid Barrier classifications.
Device Story
Surgical gown; single-use; sterile or non-sterile. Constructed from SFSMS (Spunbond/Film/Spunbond/Meltblown/Spunbond) laminate fabric. Worn by healthcare professionals in surgical settings to provide barrier protection against microorganisms, body fluids, and particulates. Provides AAMI Level 4 liquid barrier protection in critical zones and ASTM F1671-compliant protection in non-critical back zones. Protects both patient and clinician; reduces cross-contamination risk.
Clinical Evidence
Bench testing only. No clinical data. Testing included flammability (16 CFR 1610), biocompatibility (ISO 10993-5, -10, -7), laser ignition resistance (ISO 11810-1), AAMI PB70:2012 Level 4 liquid barrier performance, ASTM F1671 viral penetration resistance, tensile strength (ASTM D5034), abrasion resistance (WSP 20.5), and water vapor transmission (WSP 70.4). All tests passed.
Technological Characteristics
SFSMS (Spunbond/Film/Spunbond/Meltblown/Spunbond) laminate fabric; moisture-vapor breathable; polypropylene/polyethylene blend. AAMI PB70:2012 Level 4 barrier. Sterilization via Ethylene Oxide (EtO). Single-use. No software or electronic components.
Indications for Use
Indicated for healthcare professionals as sterile, single-use surgical apparel to protect patients and workers from microorganisms, body fluids, and particulate matter. Meets AAMI PB70:2012 Level 4 liquid barrier requirements.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, stacked on top of each other. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 21, 2017
Halyard Health, Inc. David M. Lee, JD Associate Director, Regulatory Affairs 5405 Windward Parkway Alpharetta, Georgia 30004
Re: K162930
Trade/Device Name: Aero Chrome* Select Breathable Performance Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FYA Dated: March 15, 2017 Received: March 17, 2017
Dear David M. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|------------------------------------------------------------------------------------------------|
| Food and Drug Administration | | Form Approved: OMB No. 0910-0120 | |
| | Indications for Use | | Expiration Date: January 31, 2017 |
| | | | See PRA Statement below. |
| 510(k) Number (if known) | K162930 | | |
| Device Name | Aero Chrome* Select Breathable Performance Surgical Gown | | |
| Indications for Use (Describe) | | | |
| The Aero Chrome* Select Breathable Performance Surgical Gowns are sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Aero Chrome* Select Breathable Performance Surgical Gowns meet the Level 4 requirements of the AAMI PB70:2012 Liquid Barrier classifications. | | | |
| Type of Use (Select one or both, as applicable) | | | |
| | <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
| | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | |
| | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | |
| | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | | |
| | Department of Health and<br>Human Services Food and<br>Drug Administration<br>Office of Chief Information<br>Officer Paperwork<br>Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov | | |
| | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | |
| FORM FDA 3881 (8/14) | Page 1 of 1 | | EF<br>PSC Publishing Services (301) 443-6740 |
{3}------------------------------------------------
## 510(k) Summary
| Date Summary<br>was Prepared: | April 14, 2017 |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Submitter: | David M. Lee, JD<br>Associate Director of Regulatory Affairs<br>HALYARD HEALTH<br>5405 Windward Parkway<br>Alpharetta, GA 30004<br>470 448 5182<br>David.lee@hyh.com |
| Primary Contact<br>for this 510(k)<br>Submission: | David M. Lee, JD<br>Associate Director of Regulatory Affairs,<br>HALYARD HEALTH<br>5405 Windward Parkway<br>Alpharetta, GA 30004<br>470 448 5182<br>David.lee@hyh.com |
| Device Trade<br>Name: | Aero Chrome* Select Breathable Performance Surgical Gown<br>(hereinafter described as "Aero Chrome* Select") |
| Device Common<br>Names | Surgical Gown |
| Regulation: | 878.4040 |
| Device Product<br>Codes and<br>Classification<br>Names: | FYA<br>Class II<br>Surgical Apparel |
| Predicate Device | The Aero Chrome* Select Breathable Performance Surgical<br>Gown under submission is substantially equivalent to the<br>predicate device, the Aero Chrome* Breathable Performance<br>Surgical Gown (K153255). |
| Device<br>Description: | The Aero Chrome* Select Breathable Performance Surgical<br>Gown has a Spunbond/Film/Spunbond/Meltblown/Spunbond<br>design (SFSMS) that provides AAMI Level 4 liquid barrier<br>protection in the critical zones of the gown. The back of the<br>Aero Chrome* Select Breathable Performance Surgical<br>Gown in the non-critical zone also has the same<br>Spunbond/Film/Spunbond/Meltblown/Spunbond fabric (SFSMS)<br>that provides ASTM 1671 liquid barrier protection. The Aero<br>Chrome* Select Breathable Performance Surgical Gown is<br>single use, disposable medical device that will be provided in a |
{4}------------------------------------------------
variety of sizes and sterile and non-sterile packaging configurations.
- The Aero Chrome* Select Breathable Performance Surgical Gowns Intended Use: are sterile, single-use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Aero Chrome* Select Breathable Performance Surgical Gowns meet the Level 4 requirements of the AAMI PB70:2012 Liquid Barrier classifications.
### Product Codes and Descriptions
#### STERILE CODES
| PRODUCT STOCK CODE | DESCRIPTION | SIZE |
|--------------------|----------------------------------------------------------------|----------|
| 44699 | Aero Chrome* Select<br>Breathable Performance<br>Surgical Gown | Large |
| 44706 | Aero Chrome* Select<br>Breathable Performance<br>Surgical Gown | X Large |
| 44707 | Aero Chrome* Select<br>Breathable Performance<br>Surgical Gown | XX Large |
#### NON-STERILE CODES
| PRODUCT STOCK CODE | DESCRIPTION | SIZE |
|--------------------|----------------------------------------------------------------|----------|
| 44696NS | Aero Chrome* Select<br>Breathable Performance<br>Surgical Gown | Large |
| 44697NS | Aero Chrome* Select<br>Breathable Performance<br>Surgical Gown | X Large |
| 44698NS | Aero Chrome* Select<br>Breathable Performance<br>Surgical Gown | XX Large |
#### Summary of Non-Clinical Testing
| Standard or Reference | Test Method | Data<br>Generated | Meets<br>Requirement |
|--------------------------------------------------------------------------------------------|---------------------------|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Standard for the Flammability for Clothing<br>Textiles | 16 CFR 1610 | Flammability | Pass |
| Standard or Reference | Test Method | Data<br>Generated | Meets<br>Requirement |
| ISO 10993-5:2009 Biological Evaluation of<br>Medical Devices-Part 5: | Cytotoxicity | Cytotoxicity | Pass, under the<br>conditions of the<br>study non-<br>cytotoxic |
| ISO 10993-10:2010 Biological Evaluation of<br>Medical Devices-Part 10 | Skin Irritation<br>Study | Irritation | Pass, under the<br>conditions of the<br>study a<br>negligible irritant |
| ISO 10993-10:2010 Biological Evaluation of<br>Medical Devices-Part 10 | Sensitization Test | Sensitization | Pass, under the<br>conditions of the<br>study non-<br>sensitizing |
| ISO 10993-7:2008, Ethylene Oxide<br>sterilization residuals | EO residuals | EO residuals | Pass |
| Laser Ignition Resistance | ISO 11810-1 (2005) | Laser<br>resistance | Pass |
| AAMI Liquid Barrier Performance in Critical<br>Zone<br>(body, sleeves, and ties) | AAMI PB70:2012<br>Level 4 | Resistance to<br>Penetration by<br>Blood-Borne<br>Pathogens<br>Using Phi-X174<br>Bacteriophage | Pass |
| AAMI Liquid Barrier Performance Level 1 in<br>non-critical raglan sleeve to shoulder seams | AAMI PB70:2012<br>Level 1 | Water<br>resistance | Pass |
| Liquid Barrier Performance in Non-Critical<br>Zone fabric on the back of gown | ASTM F1671 | Resistance to<br>Penetration by<br>Blood-Borne<br>Pathogens<br>Using Phi-X174<br>Bacteriophage | Pass |
| Grab Tensile, Peak Stretch, and Peak<br>Energy - Nonwovens | ASTM D5034<br>(2009) | Tensile<br>Strength | Pass |
| Abrasion Resistance and Surface Bonding of<br>SMS Laminates | WSP 20.5 (2008) | Abrasion<br>resistance | Pass |
| Synthetic Blood Penetration<br>(fabric only in critical and non-critical zone) | ASTM-1670-08<br>(2008) | Resistance to<br>penetration | Pass |
| Mass Per Area ( <i>Basis Weight</i> ) of Materials | D3776 (2009) | Fabric basis<br>weight | Pass |
| Water Vapor Transmission Rate Through<br>Nonwovens and Plastic Films | WSP<br>70.4 (2008) | Water Vapor<br>Transmission | Pass |
| Degree Peel Strength of Laminated<br>Nonwovens - Raw Materials | STM-00197(2010) | Peel Strength | Pass |
| Resistance to Linting Dry Particle Generation | INDA WSP 160.1<br>(2009) | Particulate | Pass |
{5}------------------------------------------------
Test results established that the product met the predetermined specifications.
{6}------------------------------------------------
## Substantial Equivalence Table
| Attribute | Predicate Device: K153255<br>Aero Chrome* Breathable<br>Performance Surgical Gown (AAMI<br>Liquid Barrier Level 4) | Device under submission: Aero<br>Chrome* Select Breathable<br>Performance Surgical Gown<br>(AAMI Liquid Barrier Level 4) |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Aero Chrome* Breathable<br>Performance Surgical Gowns are<br>sterile, single use surgical apparel<br>intended to be worn by healthcare<br>professionals to help protect both the<br>patient and the healthcare worker from<br>the transfer of microorganisms, body<br>fluids, and particulate matter. The Aero<br>Chrome* Breathable Performance<br>Surgical Gowns meet the Level 4<br>requirements of the AAMI PB70:2012<br>Liquid Barrier classifications.<br>The Aero Chrome* Breathable<br>Performance Surgical Gowns are also<br>sold as bulk non-sterile, single use<br>items, to repackager/relabeler<br>establishments for further packaging<br>and Ethylene Oxide (EtO) sterilization | The Aero Chrome* Select Breathable<br>Performance Surgical Gowns are<br>sterile, single use surgical apparel<br>intended to be worn by healthcare<br>professionals to help protect both the<br>patient and the healthcare worker from the<br>transfer of microorganisms, body fluids,<br>and particulate matter. The Aero<br>Chrome* Select Breathable<br>Performance Surgical Gowns meet the<br>Level 4 requirements of the AAMI<br>PB70:2012 Liquid Barrier classifications. |
| How<br>supplied | Sterile (10-6) or bulk non-sterile | Same |
| Sterilization<br>Method | Ethylene Oxide | Same |
| SAL | 10-6 | Same |
| Gown color | Gray | Same |
| Gown sizes | Small, Large, X-Large, XX-Large,<br>XXX-Large, L X-Long, XXL X- Long | Large, X-Large, and XX-Large |
| Construction<br>Overview | The Aero Chrome* surgical gowns<br>under submission are manufactured<br>from a moisture-vapor breathable,<br>repellent, non-woven fabric using a<br>polymer blend of polypropylene and<br>polyethylene. The front body and<br>sleeve fabric is a three-layer film<br>laminate. This fabric is an SFSMS<br>design Spunbond/Film/Spunbond-<br>Meltblown-Spunbond that is adhesively<br>bonded together. The Aero Chrome*<br>surgical gowns in the critical zone meet<br>AAMI-4 liquid barrier requirements,<br>while back of gown in the non-critical<br>zone is composed of a breathable SMS<br>fabric with an AAMI level 1 liquid<br>barrier protection. | Same, except the fabric on the back of<br>the gown in the non-critical zone is made<br>with the same Spunbond/Film/Spunbond-<br>Meltblown-Spunbond fabric (SFSMS) as<br>used on the front of the gown in the<br>critical zone. This fabric on the back of<br>the gown in the non-critical zone meets<br>ASTM F1671.<br>Overall finished gown dimensions are not<br>changing from the predicate device. |
| Not made<br>with Natural<br>Rubber<br>Latex | Yes | Same |
| Attribute | Predicate Device: Aero Chrome* Performance Surgical Gown | Device under submission in this 510(k): Aero Chrome Select Breathable Performance Surgical Gowns (AAMI Liquid Barrier Level 4) |
| Performance Testing in Critical Zone | ANSI/AAMI PB70: 2012 Level 4 Liquid Barrier Requirements - Pass | Same |
| | WSP 70.4 (STM-00164)-Water Vapor Transmission Rate - Pass | Same |
| | WSP 160.1 (STM-00353) Test Method for Resistance to Linting: Dry Particle Generation with a Modified Gelbo Flex Unit -Pass | Same |
| | ASTM D 5034 - 2009 Standard Test Method for Breaking Strength and Elongation of Textile Fabrics (Grab Test) - Pass | Same |
| | ASTM F1670:2008 for fabric in critical zone-Pass | Same |
| | STM- 00149 (WSP 20.5 2008), Abrasion Resistance Test - Pass | Same |
| | Biocompatibility per ISO 10993 – Pass, the device under the conditions of the study is non-cytotoxic, non-irritant, and non-sensitizing. | Same, device under the conditions of the study was non-cytotoxic, non-irritant, and non-sensitizing. |
| | 16 CFR, Chapter II--Consumer Product Safety Commission Part 1610 -Standard for The Flammability of Clothing Textiles Class I - Pass | Same |
| | Laser Ignition Resistance/ ISO 11810- Pass | Same |
| | Performance Testing on fabric in Non-Critical Zone on back of gown | AAMI level 1-Pass |
{7}------------------------------------------------
#### Performance Testing
Conclusion The performance testing submitted for the Aero Chrome* Select Breathable Performance Surgical Gown demonstrates substantial equivalence to the predicate Aero Chrome* Breathable Performance Surgical Gown (K153255) in intended use, design, materials, performance, and biocompatibility attributes. Based on the comparisons to the predicate device, the subject device described in this 510(k) Summary, Aero Chrome* Select Breathable Performance Surgical Gown, does not raise any different questions or new issues of safety and effectiveness.
*Registered Trademark or Trademark of Halyard Health, Inc., or its affiliates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.