Stryker Navigation System with CranialMap software application, Stryker CranialMap Planning software application, Stryker CranialMask Tracker
Applicant
Stryker Corporation
Product Code
HAW · Neurology
Decision Date
Feb 22, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Stryker Navigation System, with the CranialMap software application, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer-assisted surgery. The system is indicated for any medical condition in which the use of computer-assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. The system should be operated only by trained personnel such as surgeons and clinic staff. The system assists in the positioning of instruments for cranial procedures, including: Cranial bionsies Puncture of abscesses Craniotomies Craniectomies Resection of tumors and other lesions Removal of foreign objects Skull base procedures Transnasal neurosurgical procedures Transphenoidal pituitary surgery Shunt placement, including pediatric shunt placement Placement of electrodes for recording, stimulation and lesion generation Endoscopic Sinus Surgery (ESS) Intranasal procedures Ear implant procedures Craniofacial procedures Skull reconstruction procedures Orbital cavity reconstruction procedures The user should consult Chapter "System Accuracy" of the safety information to assess if the accuracy of the system is suitable for their needs.
Device Story
Stryker Navigation System provides image-guided planning and intraoperative guidance for cranial surgery. Inputs include preoperative/intraoperative medical images (CT, CTA, MR, MRA, fMRI, PET, SPECT). System uses wireless optical tracking (IR LED) to determine position of navigated instruments relative to patient anatomy. Output is 3D/orthogonal images, target guidance, and microscope injection displayed on a monitor. Used in OR by surgeons/staff. CranialMap Planning software allows preoperative planning on compliant computers. CranialMask Tracker (non-invasive, skin-applied) enables automatic patient registration. System assists in precise instrument positioning, potentially improving surgical accuracy and patient outcomes.
Clinical Evidence
Bench testing only. No clinical data required. Verification and validation testing included system-level software verification, accuracy testing, planning software verification, and component-level code verification. CranialMask Tracker evaluated for communication, visibility, flexibility, adhesion, battery life, electrical safety, EMC, biocompatibility, shelf-life, and sterilization. Mean navigation accuracy is 1.11 mm (positional) and 1.02° (trajectory).
Technological Characteristics
Stereotaxic image-guided system. Components: platform (computer, camera, IO-Tablet), software, navigated instruments, accessories. Tracking: active wireless optical (IR LED). Materials: flexible PCB mask, adhesive backing. Power: AC 100/240V, battery for instruments. Connectivity: wireless tracking, USB. Sterilization: EO (mask). Software: Windows 8.1, rule-based navigation/planning algorithms.
Indications for Use
Indicated for patients undergoing open or percutaneous cranial procedures, including biopsies, abscess puncture, craniotomies, craniectomies, tumor/lesion resection, foreign object removal, skull base/transnasal/transphenoidal/craniofacial/skull reconstruction/orbital cavity procedures, shunt placement, electrode placement, ESS, intranasal, and ear implant procedures. Operated by trained surgeons and clinic staff.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 22, 2017
Stryker Corporation Brittney Larsen Staff Regulatory Affairs Specialist Bötzinger Straße 41 Freiburg Baden-Wurttemberg, D-79111 DE
Re: K162929
Trade/Device Name: Stryker Navigation System With Cranialmap Software Application, Stryker Cranialmap Planning Software Application, Stryker Cranialmask Tracker Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: January 20, 2017 Received: January 23, 2017
Dear Ms. Larsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Hoffmann -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
### K162929
Device Name
Stryker Navigation System with CranialMap software application
Indications for Use (Describe)
The Stryker Navigation System, with the CranialMap software application, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer-assisted surgery.
The system is indicated for any medical condition in which the use of computer-assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.
The system should be operated only by trained personnel such as surgeons and clinic staff.
The system assists in the positioning of instruments for cranial procedures, including:
- Cranial bionsies
- Puncture of abscesses
- Craniotomies
- Craniectomies
- Resection of tumors and other lesions
- Removal of foreign objects
- Skull base procedures
- Transnasal neurosurgical procedures
- Transphenoidal pituitary surgery
- Shunt placement, including pediatric shunt placement
- Placement of electrodes for recording, stimulation and lesion generation
- Endoscopic Sinus Surgery (ESS)
- Intranasal procedures
- Ear implant procedures
- Craniofacial procedures
- Skull reconstruction procedures
- Orbital cavity reconstruction procedures
The user should consult Chapter "System Accuracy" of the safety information to assess if the accuracy of the system is suitable for their needs.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Form Approved: OMB No. 0910-0120 Food and Drug Administration
## Indications for Use
Expiration Date: January 31, 2017 See PRA Statement below.
# 510(k) Number (if known)
K162929
Device Name
Stryker CranialMap Planning software application
Indications for Use (Describe)
The Stryker CranialMap Planning software application, when used on a compliant computer, is intended as a planning software to enable open or percutaneous computer-assisted surgery.
The software application is indicated for any medical condition in which the use of computer-assisted planning may be appropriate.
The software application should be operated only by trained personnel such as surgeons and clinic staff.
This can include the following cranial procedures:
- Cranial biopsies
- Puncture of abscesses
- Craniotomies
- Craniectomies
- Resection of tumors and other lesions
- Removal of foreign objects
- Skull base procedures
- Transnasal neurosurgical procedures
- Transphenoidal pituitary surgery
- Shunt placement, including pediatric shunt placement
- Placement of electrodes for recording, stimulation and lesion generation
- Endoscopic Sinus Surgery (ESS)
- Intranasal procedures
- Ear implant procedures
- Craniofacial procedures
- Skull reconstruction procedures
- Orbital cavity reconstruction procedures
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}------------------------------------------------
### This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|-----------------------------------------|-----------------------------------|
| Food and Drug Administration | Expiration Date: January 31, 2017 |
| Indications for Use | See PRA Statement below. |
| Number (if known) | |
|-------------------|--|
|-------------------|--|
# 510(k) N
## K162929
Device Name Stryker CranialMask Tracker
#### Indications for Use (Describe)
The Stryker CranialMask Tracker is intended to be used as an accessory to the Stryker Navigation System, with the CranialMap software application. It is placed onto the patient's face on top of the skin.
- In combination with preoperative and intraoperative imaging devices, it enables automatic patient registration for open or percutaneous computer assisted surgery.
- The Stryker CranialMask Tracker can be used as a noninvasive patient tracker to support open or percutaneous cranial procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 5 of 5
{7}------------------------------------------------
# 510(K) SUMMARY
| | | | Contact Information |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|------------------------------------|
| 510(k) Submitter | Stryker Leibinger GmbH & Co. KG<br>Bötzinger Straße 41<br>79111 Freiburg, Germany<br>Ph: +1 (269) 389-5293<br>Fax: +1 (269) 389-5412 | | |
| FDA Establishment<br>Registration No. | 8010177 | | |
| Contact Person | Brittney Larsen MS, RAC<br>Ph: +1 (269) 389-5293<br>Fax: +1 (269) 389-5412<br>brittney.larsen@stryker.com | | |
| Date Prepared | October 18, 2016 | | |
| | | | Subject Information |
| Trade Name | Stryker Navigation System with CranialMap software application;<br>Stryker CranialMap Planning software application; and<br>Stryker CranialMask Tracker. | | |
| Common Name | Stereotaxic instrument | | |
| Classification | II, non-exempt | | |
| Classification<br>Name and Product<br>Code | Neurological Stereotaxic Instrument<br>(21 CFR 882.4560, Product code HAW) | | |
| | Predicate Information | | |
| 510(k) Number | Product<br>Code | Trade Name | Manufacturer |
| K131214 | HAW | CranialMap Neuro | Stryker Leibinger GmbH<br>& Co. KG |
| | | | Reference Information |
| K141941 | OLO | Stryker SpineMask<br>Tracker | Stryker Leibinger GmbH<br>& Co. KG |
{8}------------------------------------------------
### Device Description
The devices subject to this Traditional 510(k) are within the Stryker Navigation System with CranialMap software application. Subject devices include the CranialMap software application, CranialMap Planning software application, and CranialMask Tracker.
## System Overview
The Stryker Navigation System with CranialMap software is a computer-assisted, stereotaxic, image-guided, planning and intraoperative guidance system intended to enable open or percutaneous computer-assisted surgery. It assists the surgeon in precise positioning of instrumentation during Cranial surgeries. The system provides intraoperative guidance to the surgeon by displaying the position of navigated surgical instruments relative to the medical images.
The Stryker Navigation System with CranialMap software is comprised of a platform, CranialMap software application, navigated instruments, and accessories. The predicate CranialMap Neuro navigation system includes the same parts. The platform consists of a computer, camera, and IO-Tablet (input/output). The CranialMap software is dedicated for the Cranial procedures as defined in the indications for use and identical to the predicate. Required navigated instruments include: a patient tracker; a pointer, suction tube, or seeker; shunt placement tool, and a frameless guide. A battery is the only required accessory of the system, and is used in the navigated instruments only.
## CranialMap software application
The CranialMap software application is a required part of the Stryker Navigation System. It is installed by a Stryker representative on the platform. The CranialMap software is used with a Stryker platform and interfaces with Stryker navigated instruments and accessories. The software displays the intraoperative location of navigated surgical instruments relative to imported medical images via wireless optical tracking technology. The subject CranialMap software does not provide direct or indirect patient-contact.
CranialMap is an interactive software application, which provides the functions needed to conduct the indicated cranial procedures. The software application implements the methods for planning, patient registration, and instrument navigation. Furthermore, the software guides the user through preoperative and intraoperative workflow steps. The significant changes made to the subject CranialMap software when compared with the predicate include adding planes and mirroring functionalities and adding compatibility with the CranialMask Tracker.
## CranialMap Planning software application
The CranialMap Planning software is an optional accessory to the Stryker Navigation System with CranialMap software application. The CranialMap Planning software application is a standalone software device intended for use on compliant computers as defined in the Safety Information document. It is a new device introduced with this Traditional 510(k).
The subject CranialMap Planning software application is an identical subset of functionality copied from the subject CranialMap software. The planning version serves to preoperatively plan a navigated surgery, outside of the operating room, on a compliant computer. It does not provide registration functionality or surgical navigation guidance,
{9}------------------------------------------------
and it does not interface with instruments or accessories. The preplanned patient record can only be exported to the Stryker Navigation System with CranialMap software. The subject CranialMap Planning software does not provide direct or indirect patient-contact.
## CranialMask Tracker
The CranialMask Tracker is a non-invasive patient tracker. It is indicated as an optional accessory to the Stryker Navigation System with CranialMap software application. The subject CranialMask Tracker directly contacts the patient skin, which is identical to the predicate Patient Registration Mask.
The CranialMask Tracker includes a flexible printed circuit board (PCB) mask and closed housing unit. It consists of tracking diodes, green status light emitting diode (LED), communication receiver, battery, an on/off switch, and adhesive backing. The electronic components in the communication unit of the CranialMask Tracker are identical to the components used in the reference device, SpineMask Tracker, cleared by K141941.
This Traditional 510(k) requests clearance for changes made to the CranialMap software used with the Stryker Navigation System. It also seeks clearance for the new CranialMap Planning software and CranialMask Tracker devices as optional accessories to the system. Indication for Use
## Stryker Navigation System with CranialMap software application
The Stryker Navigation System, with the CranialMap software application, is intended as a planning and intraoperative guidance system to enable open or percutaneous computerassisted surgery.
The system is indicated for any medical condition in which the use of computer-assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.
The system should be operated only by trained personnel such as surgeons and clinic staff.
The system assists in the positioning of instruments for cranial procedures, including:
- Cranial biopsies
- Puncture of abscesses
- Craniotomies
- Craniectomies
- Resection of tumors and other lesions
- Removal of foreign objects
- Skull base procedures
- Transnasal neurosurgical procedures
- Transphenoidal pituitary surgery
- Shunt placement, including pediatric shunt placement
- Placement of electrodes for recording, stimulation and lesion generation
- Endoscopic Sinus Surgery (ESS)
- Intranasal procedures
- Ear implant procedures
- Craniofacial procedures
- Skull reconstruction procedures
{10}------------------------------------------------
- Orbital cavity reconstruction procedures
The user should consult the Accuracy section of the Safety Instructions to assess if the accuracy of the system is suitable for their needs.
## CranialMap Planning software application
The Stryker CranialMap Planning software application, when used on a compliant computer, is intended as a planning software to enable open or percutaneous computerassisted surgery.
The software application is indicated for any medical condition in which the use of computer-assisted planning may be appropriate.
The software application should be operated only by trained personnel such as surgeons and clinic staff.
This can include the following cranial procedures:
- Cranial biopsies
- Puncture of abscesses
- Craniotomies
- Craniectomies
- Resection of tumors and other lesions
- Removal of foreign objects
- Skull base procedures
- Transnasal neurosurgical procedures
- Transphenoidal pituitary surgery
- Shunt placement, including pediatric shunt placement
- Placement of electrodes for recording, stimulation and lesion generation
- Endoscopic Sinus Surgery (ESS)
- Intranasal procedures
- Ear implant procedures
- Craniofacial procedures
- Skull reconstruction procedures
- Orbital cavity reconstruction procedures
## CranialMask Tracker
The Stryker CranialMask Tracker is intended to be used as an accessory to the Stryker Navigation System, with the CranialMap software application. It is placed onto the patient's face on top of the skin.
- In combination with preoperative and intraoperative imaging devices, it enables automatic patient registration for open or percutaneous computer assisted surgery. - The Stryker CranialMask Tracker can be used as a noninvasive patient tracker to support open or percutaneous cranial procedures.
{11}------------------------------------------------
| Technological Characteristic Comparison Summary | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Characteristic | Predicate Device (K131214) | Subject Device | |
| Type of Use | Prescription use only | No change | |
| Condition for Use | Single-use (battery only)<br>Reusable | No change | |
| Sterility | Sterile (battery only)<br>Non-sterile | No change | |
| Mode/Mechanism of Action | Image-guided planning and surgery Active wireless optical tracking via infrared (IR) LED signal detection by the camera Instrument tracking relative to patient anatomical images Spatial computation method | No change | |
| Imaging Modalities | Computed tomography (CT) Computed tomography angiography (CTA) Magnetic resonance (MR) Magnetic resonance angiography (MRA) Magnetic resonance imaging (MRI) Functional magnetic resonance imaging (fMRI), including Diffusion Tensor Imaging (DTI) Position emission tomography (PET) Single-photon emission computed tomography (SPECT) | No change | |
| External Power Source | Alternating current (AC) power supply, 100/240V (Volts) and 50/60Hz (Hertz) Off-the-shelf (OTS) uninterruptable power supply (UPS) for power interruptions less than or equal to six (6) minutes | No change | |
| Intended user group | Trained clinical staff | No change | |
| Intended use environment | Operating room (OR) | No change | |
| System Accuracy | Mean navigation accuracy of ±2mm and angular axis displacement of ±2° | No change | |
| Input | Analog video input (AVI) | No change | |
| Output | 3D image, anatomic orthogonal images, endoscopy analog video image, target guidance image, microscope injection | No change | |
| User Interface | Monitor with high resolution display screen Virtual keyboard | No change | |
| Technological Characteristic Comparison Summary | | | |
| Characteristic | Predicate Device (K131214) | Subject Device | |
| Camera Working Space | IO-Tablet with touch screen, USB ports, and CD/DVD drive Buttons on navigated instruments 1.25m (meters) | No change | |
| Computer Operating System | Windows XP Embedded (SPC 3.0) (Stryker Personal Computer) Windows 8.1 (SPC3.1) Off-The-Shelf (OTS) Service Pack 3 | <b>Software</b><br>No subject change. Windows 8.1 was implemented to maintain the latest technology. This change did not exceed the threshold requiring FDA premarket notification as per 21 CFR 807.81(a)(3). Windows XP and OTS service pack 3 remain identical to the predicate. | |
| Software Version | CranialMap software application Version 2.0 | The subject device software version 3.0 was modified, for marketing purposes, in the following manner: 1) addition planes and mirroring functionality, and 2) compatibility with the new CranialMask Tracker. Planes and mirroring offer additional software functionality to users during navigated surgery. The CranialMask Tracker is a new device introduced in this submission and intended as an optional accessory to the Stryker Navigation System with CranialMap software application. | |
| Standalone Software | None | The new CranialMap Planning software is indicated for use with a compliant computer as defined in user documentation. It is identical to the full version CranialMap software, except the registration and navigation features are disabled. It only allows for limited functionality that is within the predicate scope. | |
| Technological Characteristic Comparison Summary | | | |
| Characteristic | Predicate Device (K131214) | Subject Device | |
| Collision Zones | Segments and annotations can be defined Collision check option for segments, annotations, and functional overlays during navigation | No change | |
| Hardware | | | |
| Microscope Navigation Interfaces | Zeiss Pentero Zeiss Pentero 800 Zeiss Pentero 900 Leica M525, Leica M530 Leica M720 | No subject change. External, compatible microscope list updated to add more clarity and due to Zeiss portfolio obsolescence. The microscope and navigation system interface remains identical to the predicate. No software changes were made. This change was made since the predicate clearance, however did not require FDA notification as per 21 CFR 807.81(a)(3). | |
| Required System Components | Platform; Software application; Tracker; Pointer, suction tube, or seeker; Shunt placement tool; Frameless guide; and Battery | No change | |
| Mask Tracker | | | |
| Conditions of use | Flexible mask: single-use<br>Communication unit: Reusable | | |
| Anatomical Location | Patient's face | | |
| Battery | Lithium battery | | |
| Patient-contacting | Yes | No change | |
| Sterility | Sterile | No change | |
| Sterilization method | Ethylene oxide (EO) | No change | |
| Sterility assurance level | 10-6 | No change | |
| Sterilization cycle | Number 45 | No change | |
| | | Single-use.<br>Identical to reference device. | |
| Technological Characteristic Comparison Summary | | | |
| Characteristic | Predicate Device (K131214) | Subject Device<br>Performance Data | |
| | | Nonclinical | |
| The changes subject to this 510(k) include: software modifications to the CranialMap software, and the introduction of the CranialMap Planning software and CranialMask Tracker. Non-clinical verification and validation testing performed with the subject devices focused on the software changes implemented, and evaluated the visibility and communication, flexibility and adhesion, and working battery time performance of the CranialMask Tracker. | | | |
| The software changes have been tested in accordance with the Stryker software development life cycle procedure. Verification and validation testing activities were performed at the software system, integration, and component levels to show sufficient implementation of the changes as per the specifications.<br>Software testing conducted includes:<br>System level software verification; Accuracy testing; Planning software verification with compliant computers; and Component level code verification. | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.