← Product Code [IYN](/productcode/IYN) · K162902

# ARIETTA Prologue Diagnostic Ultrasound system and Transducers (K162902)

_Hitachi , Ltd. · IYN · Dec 15, 2016 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K162902

## Device Facts

- **Applicant:** Hitachi , Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Dec 15, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: FETAL; ABDOMINAL - (Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis)); INTRA-OPERATIVE (SPEC.)-(Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures)); INTRA-OPERATIVE (NEURO); LAPAROSCOPIC; PEDIATRIC; SMALL ORGAN (SPEC.)-(Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy): NEONATAL CEPHALIC: ADULT CEPHALIC; TRANS-RECTAL - (Includes imaging for guidance of trans-rectal biopsy); TRANS-VAGIAL -(Includes imaging for guidance of trans-vaginal biopsy); TRANS-ESOPHAGEAL (NON-CARDIAC & CARDIAC) -ADULT/PEDIATRIC; MUSCULOSKELETAL (CONVENT./SUPERFIC.); Other (spec.) -WOUND (Includes imaging for Cavernous/Non-Cavernous wounds); Other (spec) - GYNECOLOGICAL; CARDIAC: ADULT/ PEDIATRIC and PERIPHERAL VESSEL.

## Device Story

ARIETTA Prologue is a diagnostic ultrasound system using transducers to transmit/receive ultrasound energy. It processes reflected echoes via a computer system to generate images and Doppler data on a video monitor. Operated by trained personnel (doctors, sonographers) in clinical settings. Features include B-mode, PW/CW Doppler, Color Doppler, Amplitude Doppler, and Tissue Doppler Imaging. Output aids clinical decision-making through visualization of anatomy and fluid flow. Benefits include non-invasive diagnostic imaging and guidance for interventional procedures like biopsies.

## Clinical Evidence

Bench testing only. No clinical data required. Device evaluated for acoustic output, biocompatibility (ISO 10993-1), cleaning/disinfection effectiveness, electromagnetic compatibility, and electrical/mechanical safety.

## Technological Characteristics

System includes ultrasound transducers (convex, linear, sector), computer processing unit, and video display. Features 64 transmit channels, electric focusing (max 8 steps), and 2-beam receive processing. Connectivity includes DICOM (SR, QR). Power source includes built-in battery. Biocompatibility tested per ISO 10993-1. Software includes automated measurement functions (IMT, NT) and image enhancement (AIP/SIP, spatial compounding).

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- NOBLUS ([K142368](/device/K142368.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles, with flowing lines suggesting movement or energy. The figure is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 15, 2016

Hitachi, Ltd. % Ms. Angela Van Arsdale RA/QA Manager 10 Fairfield Boulevard WALLINGFORD CT 06492

Re: K162902

Trade/Device Name: ARIETTA Prologue Diagnostic Ultrasound System and Transducers Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: October 13, 2016 Received: October 18, 2016

Dear Ms. Van Arsdale:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael O'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

# 510(k) Number (if known)

K162902

Device Name

ARIETTA Prologue Diagnostic Ultrasound System and Transducers

#### Indications for Use (Describe)

ARIETTA Prologue is intended for use by trained personnel (doctor, sonographer, etc.) for the diagnostic ultrasound evaluation of following clinical applications: FETAL; ABDOMINAL - (Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis)); INTRA-OPERATIVE (SPEC.)-(Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures)); INTRA-OPERATIVE (NEURO); LAPAROSCOPIC; PEDIATRIC; SMALL ORGAN (SPEC.)-(Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy): NEONATAL CEPHALIC: ADULT CEPHALIC; TRANS-RECTAL - (Includes imaging for guidance of trans-rectal biopsy); TRANS-VAGIAL -(Includes imaging for guidance of trans-vaginal biopsy); TRANS-ESOPHAGEAL (NON-CARDIAC & CARDIAC) -ADULT/PEDIATRIC; MUSCULOSKELETAL (CONVENT./SUPERFIC.); Other (spec.) -WOUND (Includes imaging for Cavernous/Non-Cavernous wounds); Other (spec) - GYNECOLOGICAL; CARDIAC: ADULT/ PEDIATRIC and PERIPHERAL VESSEL.

The Modes of Operation are B mode, PW mode (Pulsed Wave Doppler), CW mode (Continuous Wave Dopler), Color Doppler, Amplitude Doppler (Color Flow Angiography) and TDI (Tissue Doppler Imaging), Omni Directional M mode.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### Device Name: ARIETTA Prologue

| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |  |  |  |  |  |
|--------------------------------------------------------------------------------------------------|--|--|--|--|--|
|                                                                                                  |  |  |  |  |  |

| Clinical Application      |                                    | Mode of Operation |    |     |     |                  |                   |                    |
|---------------------------|------------------------------------|-------------------|----|-----|-----|------------------|-------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       | B                 | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                         |                   |    |     |     |                  |                   |                    |
|                           | Fetal                              | P                 | P  | P   | P   | P                | P                 | P                  |
| Fetal Imaging<br>& Other  | Abdominal                          | Pa                | Pa | Pa  | Pa  | Pa               | Pa                | Pa                 |
|                           | Intra-operative (Spec.)            | Pb                | Pb | Pb  |     | Pb               | Pb                | Pb                 |
|                           | Intra-operative (Neuro.)           | P                 | P  | P   |     | P                | P                 | P                  |
|                           | Laparoscopic                       | P                 | P  | P   |     | P                | P                 | P                  |
|                           | Pediatric                          | P                 | P  | P   | P   | P                | P                 | P                  |
|                           | Small Organ (Spec.)                | Pd                | Pd | Pd  |     | Pd               | Pd                | Pd                 |
|                           | Neonatal Cephalic                  | P                 | P  | P   |     | P                | P                 | P                  |
|                           | Adult Cephalic                     | P                 | P  | P   | P   | P                | P                 | P                  |
|                           | Trans-rectal                       | Pe                | Pe | Pe  |     | Pe               | Pe                | Pe                 |
|                           | Trans-vaginal                      | Pf                | Pf | Pf  |     | Pf               | Pf                | Pf                 |
|                           | Trans-urethral                     |                   |    |     |     |                  |                   |                    |
|                           | Trans-esoph. (non-Card.)           | Pg                | Pg | Pg  | Pg  | Pg               | Pg                | Pg                 |
|                           | Musculo-skel. (Convent.)           | P                 | P  | P   |     | P                | P                 | P                  |
|                           | Musculo-skel. (Superfic.)          | P                 | P  | P   |     | P                | P                 | P                  |
|                           | Intra-luminal                      |                   |    |     |     |                  |                   |                    |
|                           | Other (Wound)                      | Ph                | Ph | Ph  |     | Ph               | Ph                | Ph                 |
|                           | Other (Gynecological)              | P                 | P  | P   |     | P                | P                 | P                  |
|                           | Cardiac Adult                      | P                 | P  | P   | P   | P                | P                 | P                  |
| Cardiac                   | Cardiac Pediatric                  | P                 | P  | P   | P   | P                | P                 | P                  |
|                           | Trans-esophageal (Adult/Pediatric) | Pg                | Pg | Pg  | Pg  | Pg               | Pg                | Pg                 |
|                           | Other (spec.)                      |                   |    |     |     |                  |                   |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  | P                 | P  | P   | P   | P                | P                 | P                  |
|                           | Other (spec.)                      |                   |    |     |     |                  |                   |                    |

N = new indication; P = previously cleared in K134016, K142368

* Combination of each operating mode, B, M, PWD, CWD and Color Doppler. B/B, B/M, B/CW, CFM-B/CFM-N, CFM-B/PW, CFM-B/CW

** Power Doppler (Color Flow Angiography), Tissue Doppler Imaging, Free Angular M-mode, Trapezoid mode

#### Additional Comments:

Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).

Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).

Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.

Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.

Subscript "e": Includes imaging for guidance of trans-rectal biopsy

Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.

Subscript "g": For Adult and pediatric patients

Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"

Prescription Use Only (Per 21 CFR 801.109)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

Page 2 of 18

510(K) Premarket Notification - ARIETTA Prologue

{4}------------------------------------------------

| Device Name: ARIETTA Prologue                                                                    |
|--------------------------------------------------------------------------------------------------|
| Transducer: C22P                                                                                 |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |

| Clinical Application      |                                    |    | Mode of Operation |     |     |                  |                      |                    |  |
|---------------------------|------------------------------------|----|-------------------|-----|-----|------------------|----------------------|--------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       | B  | M                 | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |  |
| Ophthalmic                | Ophthalmic                         |    |                   |     |     |                  |                      |                    |  |
|                           | Fetal                              | P  | P                 | P   |     | P                | P                    | P                  |  |
| Fetal Imaging<br>& Other  | Abdominal                          | Pa | Pa                | Pa  |     | Pa               | Pa                   | Pa                 |  |
|                           | Intra-operative (Spec.)            |    |                   |     |     |                  |                      |                    |  |
|                           | Intra-operative (Neuro.)           |    |                   |     |     |                  |                      |                    |  |
|                           | Laparoscopic                       |    |                   |     |     |                  |                      |                    |  |
|                           | Pediatric                          |    |                   |     |     |                  |                      |                    |  |
|                           | Small Organ (Spec.)                |    |                   |     |     |                  |                      |                    |  |
|                           | Neonatal Cephalic                  |    |                   |     |     |                  |                      |                    |  |
|                           | Adult Cephalic                     |    |                   |     |     |                  |                      |                    |  |
|                           | Trans-rectal                       |    |                   |     |     |                  |                      |                    |  |
|                           | Trans-vaginal                      |    |                   |     |     |                  |                      |                    |  |
|                           | Trans-urethral                     |    |                   |     |     |                  |                      |                    |  |
|                           | Trans-esoph. (non-Card.)           |    |                   |     |     |                  |                      |                    |  |
|                           | Musculo-skel. (Convent.)           |    |                   |     |     |                  |                      |                    |  |
|                           | Musculo-skel. (Superfic.)          |    |                   |     |     |                  |                      |                    |  |
|                           | Intra-luminal                      |    |                   |     |     |                  |                      |                    |  |
|                           | Other (Wound)                      |    |                   |     |     |                  |                      |                    |  |
|                           | Other (Gynecological)              |    |                   |     |     |                  |                      |                    |  |
|                           | Cardiac Adult                      |    |                   |     |     |                  |                      |                    |  |
| Cardiac                   | Cardiac Pediatric                  |    |                   |     |     |                  |                      |                    |  |
|                           | Trans-esophageal (Adult/Pediatric) |    |                   |     |     |                  |                      |                    |  |
|                           | Other (spec.)                      |    |                   |     |     |                  |                      |                    |  |
| Peripheral<br>Vessel      | Peripheral vessel                  |    |                   |     |     |                  |                      |                    |  |
|                           | Other (spec.)                      |    |                   |     |     |                  |                      |                    |  |

N = new indication; P = previously cleared in K142368

* Combination of each operating mode, B, M, PWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/PW **Power Doppler (Color Flow Angiography), Free Angular M-mode

Additional Comments:

Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).

Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).

Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.

Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.

Subscript "e": Includes imaging for guidance of trans-rectal biopsy

Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.

Subscript "g": For Adult and pediatric patients

Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"

Prescription Use Only (Per 21 CFR 801.109)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

Page 3 of 18

{5}------------------------------------------------

| Device Name: ARIETTA Prologue                                                                    |
|--------------------------------------------------------------------------------------------------|
| Transducer: C251                                                                                 |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |

| Clinical Application      |                                    |    | Mode of Operation |     |     |                  |                    |                    |
|---------------------------|------------------------------------|----|-------------------|-----|-----|------------------|--------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       | B  | M                 | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                         |    |                   |     |     |                  |                    |                    |
| Fetal Imaging<br>& Other  | Fetal                              | P  | P                 | P   |     | P                | P                  | P                  |
|                           | Abdominal                          | Pa | Pa                | Pa  |     | Pa               | Pa                 | Pa                 |
|                           | Intra-operative (Spec.)            |    |                   |     |     |                  |                    |                    |
|                           | Intra-operative (Neuro.)           |    |                   |     |     |                  |                    |                    |
|                           | Laparoscopic                       |    |                   |     |     |                  |                    |                    |
|                           | Pediatric                          | P  | P                 | P   |     | P                | P                  | P                  |
|                           | Small Organ (Spec.)                | Pd | Pd                | Pd  |     | Pd               | Pd                 | Pd                 |
|                           | Neonatal Cephalic                  |    |                   |     |     |                  |                    |                    |
|                           | Adult Cephalic                     |    |                   |     |     |                  |                    |                    |
|                           | Trans-rectal                       |    |                   |     |     |                  |                    |                    |
|                           | Trans-vaginal                      |    |                   |     |     |                  |                    |                    |
|                           | Trans-urethral                     |    |                   |     |     |                  |                    |                    |
|                           | Trans-esoph. (non-Card.)           |    |                   |     |     |                  |                    |                    |
|                           | Musculo-skel. (Convent.)           |    |                   |     |     |                  |                    |                    |
|                           | Musculo-skel. (Superfic.)          |    |                   |     |     |                  |                    |                    |
|                           | Intra-luminal                      |    |                   |     |     |                  |                    |                    |
|                           | Other (Wound)                      |    |                   |     |     |                  |                    |                    |
|                           | Other (Gynecological)              |    |                   |     |     |                  |                    |                    |
|                           | Cardiac Adult                      |    |                   |     |     |                  |                    |                    |
| Cardiac                   | Cardiac Pediatric                  |    |                   |     |     |                  |                    |                    |
|                           | Trans-esophageal (Adult/Pediatric) |    |                   |     |     |                  |                    |                    |
|                           | Other (spec.)                      |    |                   |     |     |                  |                    |                    |
| Peripheral                | Peripheral vessel                  |    |                   |     |     |                  |                    |                    |
| Vessel                    | Other (spec.)                      |    |                   |     |     |                  |                    |                    |

N = new indication; P = previously cleared in K142368

* Combination of each operating mode, B, M, PWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/CFM-M, CFM-B/PW

**Power Doppler (Color Flow Angiography), Tissue Doppler Imaging, Free Angular M-mode

Additional Comments:

Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).

Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).

Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.

Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.

Subscript "e": Includes imaging for guidance of trans-rectal biopsy

Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.

Subscript "g": For Adult and pediatric patients

Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"

Prescription Use Only (Per 21 CFR 801.109)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

## Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

Page 4 of 18

{6}------------------------------------------------

| Device Name: ARIETTA Prologue                                                                    |                   |
|--------------------------------------------------------------------------------------------------|-------------------|
| Transducer: C25P                                                                                 |                   |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |                   |
| Clinical Application                                                                             | Mode of Operation |

| Clinical Application      |                                    |    | Mode of Operation |     |     |                  |                    |                    |
|---------------------------|------------------------------------|----|-------------------|-----|-----|------------------|--------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       | B  | M                 | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                         |    |                   |     |     |                  |                    |                    |
|                           | Fetal                              | P  | P                 | P   |     | P                | P                  | P                  |
| Fetal Imaging<br>& Other  | Abdominal                          | Pa | Pa                | Pa  |     | Pa               | Pa                 | Pa                 |
|                           | Intra-operative (Spec.)            |    |                   |     |     |                  |                    |                    |
|                           | Intra-operative (Neuro.)           |    |                   |     |     |                  |                    |                    |
|                           | Laparoscopic                       |    |                   |     |     |                  |                    |                    |
|                           | Pediatric                          |    |                   |     |     |                  |                    |                    |
|                           | Small Organ (Spec.)                |    |                   |     |     |                  |                    |                    |
|                           | Neonatal Cephalic                  |    |                   |     |     |                  |                    |                    |
|                           | Adult Cephalic                     |    |                   |     |     |                  |                    |                    |
|                           | Trans-rectal                       |    |                   |     |     |                  |                    |                    |
|                           | Trans-vaginal                      |    |                   |     |     |                  |                    |                    |
|                           | Trans-urethral                     |    |                   |     |     |                  |                    |                    |
|                           | Trans-esoph. (non-Card.)           |    |                   |     |     |                  |                    |                    |
|                           | Musculo-skel. (Convent.)           |    |                   |     |     |                  |                    |                    |
|                           | Musculo-skel. (Superfic.)          |    |                   |     |     |                  |                    |                    |
|                           | Intra-luminal                      |    |                   |     |     |                  |                    |                    |
|                           | Other (Wound)                      |    |                   |     |     |                  |                    |                    |
|                           | Other (Gynecological)              |    |                   |     |     |                  |                    |                    |
| Cardiac                   | Cardiac Adult                      |    |                   |     |     |                  |                    |                    |
|                           | Cardiac Pediatric                  |    |                   |     |     |                  |                    |                    |
|                           | Trans-esophageal (Adult/Pediatric) |    |                   |     |     |                  |                    |                    |
|                           | Other (spec.)                      |    |                   |     |     |                  |                    |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  |    |                   |     |     |                  |                    |                    |
|                           | Other (spec.)                      |    |                   |     |     |                  |                    |                    |

N = new indication; P = previously cleared in K142368

* Combination of each operating mode, B, M, PWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/CFM-M, CFM-B/PW **Power Doppler (Color Flow Angiography), Free Angular M-mode

Additional Comments:

Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).

Subscript "b": Includes imaging of organs and structures exposed during neurosurgery and laparoscopic procedures).

Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.

Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.

Subscript "e": Includes imaging for guidance of trans-rectal biopsy

Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.

Subscript "g": For Adult and pediatric patients

Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"

Prescription Use Only (Per 21 CFR 801.109)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

Page 5 of 18

{7}------------------------------------------------

| Device Name: ARIETTA Prologue                                                                    |                   |
|--------------------------------------------------------------------------------------------------|-------------------|
| Transducer: C35                                                                                  |                   |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |                   |
| Clinical Application                                                                             | Mode of Operation |

| Clinical Application      |                                    | Mode of Operation |    |     |     |                  |                     |                    |
|---------------------------|------------------------------------|-------------------|----|-----|-----|------------------|---------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       | B                 | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe.) | Other**<br>(Spec.) |
|                           | Ophthalmic                         |                   |    |     |     |                  |                     |                    |
| Fetal Imaging<br>& Other  | Fetal                              | P                 | P  | P   |     | P                | P                   | P                  |
|                           | Abdominal                          | Pa                | Pa | Pa  |     | Pa               | Pa                  | Pa                 |
|                           | Intra-operative (Spec.)            |                   |    |     |     |                  |                     |                    |
|                           | Intra-operative (Neuro.)           |                   |    |     |     |                  |                     |                    |
|                           | Laparoscopic                       |                   |    |     |     |                  |                     |                    |
|                           | Pediatric                          | P                 | P  | P   |     | P                | P                   | P                  |
|                           | Small Organ (Spec.)                | Pd                | Pd | Pd  |     | Pd               | Pd                  | Pd                 |
|                           | Neonatal Cephalic                  |                   |    |     |     |                  |                     |                    |
|                           | Adult Cephalic                     |                   |    |     |     |                  |                     |                    |
|                           | Trans-rectal                       |                   |    |     |     |                  |                     |                    |
|                           | Trans-vaginal                      |                   |    |     |     |                  |                     |                    |
|                           | Trans-urethral                     |                   |    |     |     |                  |                     |                    |
|                           | Trans-esoph. (non-Card.)           |                   |    |     |     |                  |                     |                    |
|                           | Musculo-skel. (Convent.)           |                   |    |     |     |                  |                     |                    |
|                           | Musculo-skel. (Superfic.)          |                   |    |     |     |                  |                     |                    |
|                           | Intra-luminal                      |                   |    |     |     |                  |                     |                    |
|                           | Other (Wound)                      |                   |    |     |     |                  |                     |                    |
|                           | Other (Gynecological)              |                   |    |     |     |                  |                     |                    |
| Cardiac                   | Cardiac Adult                      |                   |    |     |     |                  |                     |                    |
|                           | Cardiac Pediatric                  |                   |    |     |     |                  |                     |                    |
|                           | Trans-esophageal (Adult/Pediatric) |                   |    |     |     |                  |                     |                    |
|                           | Other (spec.)                      |                   |    |     |     |                  |                     |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  |                   |    |     |     |                  |                     |                    |
|                           | Other (spec.)                      |                   |    |     |     |                  |                     |                    |

N = new indication; P = previously cleared in K142368

* Combination of each operating mode, B, M, PWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/CFM-M, CFM-B/PW

**Power Doppler (Color Flow Angiography), Tissue Doppler Imaging, Free Angular M-mode

Additional Comments:

Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).

Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).

Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.

Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.

Subscript "e": Includes imaging for guidance of trans-rectal biopsy

Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.

Subscript "g": For Adult and pediatric patients

Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"

Prescription Use Only (Per 21 CFR 801.109)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

## Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

Page 6 of 18

{8}------------------------------------------------

| Device Name: ARIETTA Prologue                                                                    |
|--------------------------------------------------------------------------------------------------|
| Transducer: C41B                                                                                 |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |

| Clinical Application      |                                    | Mode of Operation |    |     |     |                  |                      |                    |
|---------------------------|------------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       | B                 | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                         |                   |    |     |     |                  |                      |                    |
| Fetal Imaging<br>& Other  | Fetal                              | P                 | P  | P   |     |                  | P                    | P                  |
|                           | Abdominal                          |                   |    |     |     |                  |                      |                    |
|                           | Intra-operative (Spec.)            |                   |    |     |     |                  |                      |                    |
|                           | Intra-operative (Neuro.)           |                   |    |     |     |                  |                      |                    |
|                           | Laparoscopic                       |                   |    |     |     |                  |                      |                    |
|                           | Pediatric                          |                   |    |     |     |                  |                      |                    |
|                           | Small Organ (Spec.)                |                   |    |     |     |                  |                      |                    |
|                           | Neonatal Cephalic                  |                   |    |     |     |                  |                      |                    |
|                           | Adult Cephalic                     |                   |    |     |     |                  |                      |                    |
|                           | Trans-rectal                       | Pe                | Pe | Pe  |     | Pe               | Pe                   | Pe                 |
|                           | Trans-vaginal                      | Pf                | Pf | Pf  |     | Pf               | Pf                   | Pf                 |
|                           | Trans-urethral                     |                   |    |     |     |                  |                      |                    |
|                           | Trans-esoph. (non-Card.)           |                   |    |     |     |                  |                      |                    |
|                           | Musculo-skel. (Convent.)           |                   |    |     |     |                  |                      |                    |
|                           | Musculo-skel. (Superfic.)          |                   |    |     |     |                  |                      |                    |
|                           | Intra-luminal                      |                   |    |     |     |                  |                      |                    |
|                           | Other (Wound)                      |                   |    |     |     |                  |                      |                    |
|                           | Other (Gynecological)              | P                 | P  | P   |     |                  | P                    | P                  |
| Cardiac                   | Cardiac Adult                      |                   |    |     |     |                  |                      |                    |
|                           | Cardiac Pediatric                  |                   |    |     |     |                  |                      |                    |
|                           | Trans-esophageal (Adult/Pediatric) |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                      |                   |    |     |     |                  |                      |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  |                   |    |     |     |                  |                      |                    |
|                           | Other (spec.)                      |                   |    |     |     |                  |                      |                    |

N = new indication; P = previously cleared in K142368

* Combination of each operating mode, B, M, PWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/CFM-M, CFM-B/PW **Power Doppler (Color Flow Angiography), Free Angular M-mode

Additional Comments:

Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).

Subscript "b": Includes imaging of organs and structures exposed during neurosurgery and laparoscopic procedures).

Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.

Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.

Subscript "e": Includes imaging for guidance of trans-rectal biopsy

Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.

Subscript "g": For Adult and pediatric patients

Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"

Prescription Use Only (Per 21 CFR 801.109)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

Page 7 of 18

{9}------------------------------------------------

| Device Name: ARIETTA Prologue                                                                    |
|--------------------------------------------------------------------------------------------------|
| Transducer: C41V1                                                                                |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |

| Clinical Application      |                                    |    | Mode of Operation |     |     |                  |                    |                    |
|---------------------------|------------------------------------|----|-------------------|-----|-----|------------------|--------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       | B  | M                 | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                         |    |                   |     |     |                  |                    |                    |
|                           | Fetal                              | P  | P                 | P   |     | P                | P                  | P                  |
| Fetal Imaging<br>& Other  | Abdominal                          |    |                   |     |     |                  |                    |                    |
|                           | Intra-operative (Spec.)            |    |                   |     |     |                  |                    |                    |
|                           | Intra-operative (Neuro.)           |    |                   |     |     |                  |                    |                    |
|                           | Laparoscopic                       |    |                   |     |     |                  |                    |                    |
|                           | Pediatric                          |    |                   |     |     |                  |                    |                    |
|                           | Small Organ (Spec.)                |    |                   |     |     |                  |                    |                    |
|                           | Neonatal Cephalic                  |    |                   |     |     |                  |                    |                    |
|                           | Adult Cephalic                     |    |                   |     |     |                  |                    |                    |
|                           | Trans-rectal                       | Pe | Pe                | Pe  |     | Pe               | Pe                 | Pe                 |
|                           | Trans-vaginal                      | Pf | Pf                | Pf  |     | Pf               | Pf                 | Pf                 |
|                           | Trans-urethral                     |    |                   |     |     |                  |                    |                    |
|                           | Trans-esoph. (non-Card.)           |    |                   |     |     |                  |                    |                    |
|                           | Musculo-skel. (Convent.)           |    |                   |     |     |                  |                    |                    |
|                           | Musculo-skel. (Superfic.)          |    |                   |     |     |                  |                    |                    |
|                           | Intra-luminal                      |    |                   |     |     |                  |                    |                    |
|                           | Other (Wound)                      |    |                   |     |     |                  |                    |                    |
|                           | Other (Gynecological)              | P  | P                 | P   |     | P                | P                  | P                  |
|                           | Cardiac Adult                      |    |                   |     |     |                  |                    |                    |
| Cardiac                   | Cardiac Pediatric                  |    |                   |     |     |                  |                    |                    |
|                           | Trans-esophageal (Adult/Pediatric) |    |                   |     |     |                  |                    |                    |
|                           | Other (spec.)                      |    |                   |     |     |                  |                    |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  |    |                   |     |     |                  |                    |                    |
|                           | Other (spec.)                      |    |                   |     |     |                  |                    |                    |

N = new indication; P = previously cleared in K142368

* Combination of each operating mode, B, M, PWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/CFM-M, CFM-B/PW **Power Doppler (Color Flow Angiography), Free Angular M-mode

Additional Comments:

Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).

Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).

Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.

Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.

Subscript "e": Includes imaging for guidance of trans-rectal biopsy

Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.

Subscript "g": For Adult and pediatric patients

Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"

Prescription Use Only (Per 21 CFR 801.109)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

## Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

Page 8 of 18

{10}------------------------------------------------

| Device Name: ARIETTA Prologue                                                                    |                   |
|--------------------------------------------------------------------------------------------------|-------------------|
| Transducer: C42K                                                                                 |                   |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |                   |
| Clinical Application                                                                             | Mode of Operation |

| Clinical Application      |                                    | Mode of Operation |    |     |     |                  |                     |                    |
|---------------------------|------------------------------------|-------------------|----|-----|-----|------------------|---------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       | B                 | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe.) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                         |                   |    |     |     |                  |                     |                    |
|                           | Fetal                              |                   |    |     |     |                  |                     |                    |
| Fetal Imaging<br>& Other  | Abdominal                          |                   |    |     |     |                  |                     |                    |
|                           | Intra-operative (Spec.)            | Pb                | Pb | Pb  |     | Pb               | Pb                  | Pb                 |
|                           | Intra-operative (Neuro.)           |                   |    |     |     |                  |                     |                    |
|                           | Laparoscopic                       |                   |    |     |     |                  |                     |                    |
|                           | Pediatric                          |                   |    |     |     |                  |                     |                    |
|                           | Small Organ (Spec.)                | Pd                | Pd | Pd  |     | Pd               | Pd                  | Pd                 |
|                           | Neonatal Cephalic                  | P                 | P  | P   |     | P                | P                   | P                  |
|                           | Adult Cephalic                     |                   |    |     |     |                  |                     |                    |
|                           | Trans-rectal                       |                   |    |     |     |                  |                     |                    |
|                           | Trans-vaginal                      |                   |    |     |     |                  |                     |                    |
|                           | Trans-urethral                     |                   |    |     |     |                  |                     |                    |
|                           | Trans-esoph. (non-Card.)           |                   |    |     |     |                  |                     |                    |
|                           | Musculo-skel. (Convent.)           |                   |    |     |     |                  |                     |                    |
|                           | Musculo-skel. (Superfic.)          |                   |    |     |     |                  |                     |                    |
|                           | Intra-luminal                      |                   |    |     |     |                  |                     |                    |
|                           | Other (Wound)                      |                   |    |     |     |                  |                     |                    |
|                           | Other (Gynecological)              |                   |    |     |     |                  |                     |                    |
| Cardiac                   | Cardiac Adult                      |                   |    |     |     |                  |                     |                    |
| Cardiac                   | Cardiac Pediatric                  |                   |    |     |     |                  |                     |                    |
|                           | Trans-esophageal (Adult/Pediatric) |                   |    |     |     |                  |                     |                    |
|                           | Other (spec.)                      |                   |    |     |     |                  |                     |                    |
| Peripheral                | Peripheral vessel                  |                   |    |     |     |                  |                     |                    |
| Vessel                    | Other (spec.)                      |                   |    |     |     |                  |                     |                    |

N = new indication; P = previously cleared in K142368

* Combination of each operating mode, B, M, PWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/CFM-M, CFM-B/PW **Power Doppler (Color Flow Angiography), Free Angular M-mode

Additional Comments:

Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).

Subscript "b": Includes imaging of organs and structures exposed during neurosurgery and laparoscopic procedures).

Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.

Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.

Subscript "e": Includes imaging for guidance of trans-rectal biopsy

Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.

Subscript "g": For Adult and pediatric patients

Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"

Prescription Use Only (Per 21 CFR 801.109)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

Page 9 of 18

{11}------------------------------------------------

| Device Name: ARIETTA Prologue                                                                    |                   |
|--------------------------------------------------------------------------------------------------|-------------------|
| Transducer: L44                                                                                  |                   |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |                   |
| Clinical Application                                                                             | Mode of Operation |

| Clinical Application      |                                    | B  | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
|---------------------------|------------------------------------|----|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       |    |    |     |     |                  |                      |                    |
| Ophthalmic                | Ophthalmic                         |    |    |     |     |                  |                      |                    |
|                           | Fetal                              |    |    |     |     |                  |                      |                    |
| Fetal Imaging<br>& Other  | Abdominal                          | Pa | Pa | Pa  |     | Pa               | Pa                   | Pa                 |
|                           | Intra-operative (Spec.)            |    |    |     |     |                  |                      |                    |
|                           | Intra-operative (Neuro.)           |    |    |     |     |                  |                      |                    |
|                           | Laparoscopic                       |    |    |     |     |                  |                      |                    |
|                           | Pediatric                          | P  | P  | P   |     | P                | P                    | P                  |
|                           | Small Organ (Spec.)                | Pd | Pd | Pd  |     | Pd               | Pd                   | Pd                 |
|                           | Neonatal Cephalic                  |    |    |     |     |                  |                      |                    |
|                           | Adult Cephalic                     |    |    |     |     |                  |                      |                    |
|                           | Trans-rectal                       |    |    |     |     |                  |                      |                    |
|                           | Trans-vaginal                      |    |    |     |     |                  |                      |                    |
|                           | Trans-urethral                     |    |    |     |     |                  |                      |                    |
|                           | Trans-esoph. (non-Card.)           |    |    |     |     |                  |                      |                    |
|                           | Musculo-skel. (Convent.)           | P  | P  | P   |     | P                | P                    | P                  |
|                           | Musculo-skel. (Superfic.)          | P  | P  | P   |     | P                | P                    | P                  |
|                           | Intra-luminal                      |    |    |     |     |                  |                      |                    |
|                           | Other (Wound)                      |    |    |     |     |                  |                      |                    |
|                           | Other (Gynecological)              |    |    |     |     |                  |                      |                    |
|                           | Cardiac Adult                      |    |    |     |     |                  |                      |                    |
| Cardiac                   | Cardiac Pediatric                  |    |    |     |     |                  |                      |                    |
|                           | Trans-esophageal (Adult/Pediatric) |    |    |     |     |                  |                      |                    |
|                           | Other (spec.)                      |    |    |     |     |                  |                      |                    |
| Peripheral                | Peripheral vessel                  | P  | P  | P   |     | P                | P                    | P                  |
| Vessel                    | Other (spec.)                      |    |    |     |     |                  |                      |                    |

N = new indication; P = previously cleared in K142368

* Combination of each operating mode, B, M, PWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/CFM-M, CFM-B/PW

**Power Doppler (Color Flow Angiography), Free Angular M-mode, Trapezoid mode

Additional Comments:

Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).

Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).

Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.

Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.

Subscript "e": Includes imaging for guidance of trans-rectal biopsy

Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.

Subscript "g": For Adult and pediatric patients

Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"

Prescription Use Only (Per 21 CFR 801.109)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

## Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

Page 10 of 18

{12}------------------------------------------------

| Device Name: ARIETTA Prologue                                                                    |
|--------------------------------------------------------------------------------------------------|
| Transducer: L441                                                                                 |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |

| Clinical Application      |                                    |    | Mode of Operation |     |     |                  |                      |                    |
|---------------------------|------------------------------------|----|-------------------|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       | B  | M                 | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                         |    |                   |     |     |                  |                      |                    |
| Fetal Imaging<br>& Other  | Fetal                              |    |                   |     |     |                  |                      |                    |
|                           | Abdominal                          | Pa | Pa                | Pa  |     | Pa               | Pa                   | Pa                 |
|                           | Intra-operative (Spec.)            |    |                   |     |     |                  |                      |                    |
|                           | Intra-operative (Neuro.)           |    |                   |     |     |                  |                      |                    |
|                           | Laparoscopic                       |    |                   |     |     |                  |                      |                    |
|                           | Pediatric                          | P  | P                 | P   |     | P                | P                    | P                  |
|                           | Small Organ (Spec.)                | Pd | Pd                | Pd  |     | Pd               | Pd                   | Pd                 |
|                           | Neonatal Cephalic                  |    |                   |     |     |                  |                      |                    |
|                           | Adult Cephalic                     |    |                   |     |     |                  |                      |                    |
|                           | Trans-rectal                       |    |                   |     |     |                  |                      |                    |
|                           | Trans-vaginal                      |    |                   |     |     |                  |                      |                    |
|                           | Trans-urethral                     |    |                   |     |     |                  |                      |                    |
|                           | Trans-esoph. (non-Card.)           |    |                   |     |     |                  |                      |                    |
|                           | Musculo-skel. (Convent.)           | P  | P                 | P   |     | P                | P                    | P                  |
|                           | Musculo-skel. (Superfic.)          | P  | P                 | P   |     | P                | P                    | P                  |
|                           | Intra-luminal                      |    |                   |     |     |                  |                      |                    |
|                           | Other (Wound)                      |    |                   |     |     |                  |                      |                    |
|                           | Other (Gynecological)              |    |                   |     |     |                  |                      |                    |
| Cardiac                   | Cardiac Adult                      |    |                   |     |     |                  |                      |                    |
|                           | Cardiac Pediatric                  |    |                   |     |     |                  |                      |                    |
|                           | Trans-esophageal (Adult/Pediatric) |    |                   |     |     |                  |                      |                    |
|                           | Other (spec.)                      |    |                   |     |     |                  |                      |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  | P  | P                 | P   |     | P                | P                    | P                  |
|                           | Other (spec.)                      |    |                   |     |     |                  |                      |                    |

N = new indication; P = previously cleared in K142368

* Combination of each operating mode, B, M, PWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/CFM-M, CFM-B/PW **Power Doppler (Color Flow Angiography), Free Angular M-mode, Trapezoid mode

Additional Comments:

Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).

Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).

Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.

Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.

Subscript "e": Includes imaging for guidance of trans-rectal biopsy

Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.

Subscript "g": For Adult and pediatric patients

Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"

Prescription Use Only (Per 21 CFR 801.109)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

## Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

Page 11 of 18

{13}------------------------------------------------

| Device Name: ARIETTA Prologue                                                                    |                   |
|--------------------------------------------------------------------------------------------------|-------------------|
| Transducer: L44LA                                                                                |                   |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |                   |
| Clinical Application                                                                             | Mode of Operation |

| Clinical Application      |                                    | Mode of Operation |   |     |     |                  |                   |                    |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       | B                 | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                         |                   |   |     |     |                  |                   |                    |
|                           | Fetal                              |                   |   |     |     |                  |                   |                    |
| Fetal Imaging<br>& Other  | Abdominal                          |                   |   |     |     |                  |                   |                    |
|                           | Intra-operative (Spec.)            | P                 | P | P   |     | P                | P                 | P                  |
|                           | Intra-operative (Neuro.)           |                   |   |     |     |                  |                   |                    |
|                           | Laparoscopic                       | P                 | P | P   |     | P                | P                 | P                  |
|                           | Pediatric                          |                   |   |     |     |                  |                   |                    |
|                           | Small Organ (Spec.)                |                   |   |     |     |                  |                   |                    |
|                           | Neonatal Cephalic                  |                   |   |     |     |                  |                   |                    |
|                           | Adult Cephalic                     |                   |   |     |     |                  |                   |                    |
|                           | Trans-rectal                       |                   |   |     |     |                  |                   |                    |
|                           | Trans-vaginal                      |                   |   |     |     |                  |                   |                    |
|                           | Trans-urethral                     |                   |   |     |     |                  |                   |                    |
|                           | Trans-esoph. (non-Card.)           |                   |   |     |     |                  |                   |                    |
|                           | Musculo-skel. (Convent.)           |                   |   |     |     |                  |                   |                    |
|                           | Musculo-skel. (Superfic.)          |                   |   |     |     |                  |                   |                    |
|                           | Intra-luminal                      |                   |   |     |     |                  |                   |                    |
|                           | Other (Wound)                      |                   |   |     |     |                  |                   |                    |
|                           | Other (Gynecological)              |                   |   |     |     |                  |                   |                    |
| Cardiac                   | Cardiac Adult                      |                   |   |     |     |                  |                   |                    |
|                           | Cardiac Pediatric                  |                   |   |     |     |                  |                   |                    |
|                           | Trans-esophageal (Adult/Pediatric) |                   |   |     |     |                  |                   |                    |
|                           | Other (spec.)                      |                   |   |     |     |                  |                   |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  |                   |   |     |     |                  |                   |                    |
|                           | Other (spec.)                      |                   |   |     |     |                  |                   |                    |

N = new indication; P = previously cleared in K142368

* Combination of each operating mode, B, M, PWD and Color Doppler. B/B, B/M, B/PW, CFM-B/CFM-M, CFM-B/CFM-M, CFM-B/PW **Power Doppler (Color Flow Angiography), Free Angular M-mode, Trapezoid mode

Additional Comments:

Subscript "a": Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).

Subscript "b": Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures).

Subscript "c": Includes thyroid, parathyroid, breast, scrotum, and penis.

Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis, and imaging for guidance of biopsy.

Subscript "e": Includes imaging for guidance of trans-rectal biopsy

Subscript "f": Includes imaging for guidance of trans-vaginal biopsy.

Subscript "g": For Adult and pediatric patients

Includes imaging for Cavernous/Non-Cavernous wounds Subscript "h"

Prescription Use Only (Per 21 CFR 801.109)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

## Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

Page 12 of 18

{14}------------------------------------------------

| Device Name: ARIETTA Prologue                                                                     |                    |
|---------------------------------------------------------------------------------------------------|--------------------|
| Transducer: L53K                                                                                  |                    |
| Intended use:  Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |                    |
| (1iminal American                                                                                 | Mode of Overention |

| Clinical Application      |                                    | Mode of Operation |    |     |     |                  |                    |                    |
|---------------------------|------------------------------------|-------------------|----|-----|-----|------------------|--------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)       | B                 | M  | PWD | CWD | Color<br>Doppler | Combined*<br>(Spe) | Other**<br>(Spec.) |
| Ophthalmic                | Ophthalmic                         |                   |    |     |     |                  |                    |                    |
| Fetal Imaging<br>& Other  | Fetal                              |                   |    |     |     |                  |                    |                    |
|                           | Abdominal                          |                   |    |     |     |                  |                    |                    |
|…

---

**Source:** [https://fda-staging.innolitics.com/device/K162902](https://fda-staging.innolitics.com/device/K162902)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).