K162887 · Baxter Healthcare Corporation · KPO · Jan 6, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K162887
Device Name
PrismaSATE
Applicant
Baxter Healthcare Corporation
Product Code
KPO · Gastroenterology, Urology
Decision Date
Jan 6, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PrismaSATE solutions are indicated for use as a dialysate in Continuous Renal Replacement Therapy.
Device Story
PrismaSATE dialysis solutions are ready-to-use, sterile, non-pyrogenic dialysate solutions for CRRT; used for fluid/solute removal in acute renal failure or drug poisoning. Solutions consist of sodium chloride, potassium chloride, magnesium chloride, calcium chloride, dextrose, and a bicarbonate buffer (with hydrochloric acid as pH-adjusting excipient). Packaged in two-compartment polyolefin bags; final solution obtained by breaking a peel seal to mix compartments prior to use. Prescribed by physicians based on patient blood electrolyte levels; used with commercially available CRRT systems. Normalizes blood plasma water. This submission expands existing bicarbonate-buffered/HCl-excipient formulations into polyolefin packaging.
Clinical Evidence
No clinical data. Bench testing only (design verification).
Indicated for use as a dialysate in Continuous Renal Replacement Therapy (CRRT) for the treatment of acute renal failure and other cases necessitating fluid or solute removal, such as drug poisoning.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
PrismaSATE dialysis solution formulations buffered with bicarbonate using hydrochloric acid (K120333)
PrismaSATE dialysis solutions buffered with bicarbonate with lactate in polyolefin primary packaging (K072908)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 6, 2017
Baxter Healthcare Corporation Kristen Bozzelli Manager, Regulatory Affairs 32650 North Wilson Road Round Lake, IL 60073
Re: K162887
> Trade/Device Name: PrismaSATE dialysis solutions Regulation Number: 21 CFR§ 876.5820 Regulation Name: Hemodialysis System and Accessories Regulatory Class: II Product Code: KPO Dated: December 8, 2016 Received: December 12, 2016
Dear Kristen Bozzelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) TBDK162887
Device Name PrismaSATE dialysis solutions
Indications for Use (Describe)
PrismaSATE solutions are indicated for use as a dialysate in Continuous Renal Replacement Therapy.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be italicized, with the letters slanting slightly to the right. The word is the primary focus of the image, and the background is plain.
### Section 5. 510(k) Summary
January 5, 2017
#### OWNER:
Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015
#### CONTACT PERSON:
Kristen Bozzelli Manager, Global Regulatory Affairs Baxter Healthcare Corporation 32650 N Wilson Road Round Lake, IL 60073 Telephone: (224) 240-3075 Fax: (224) 240-4119
#### IDENTIFICATION OF THE DEVICE:
Common Name: Dialysate Solutions for Continuous Renal Replacement Therapy (CRRT) Ready to Use Sterile Dialysate. Trade Name or Proprietary Name: PrismaSATE Classification Panel: Gastroenterology/Urology Classification: 21 CFR 876.5820 Dialysate concentrate for hemodialysis (liquid or powder) Class: Class II Product Code: KPO
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is slightly italicized, giving the word a sense of movement. The background is plain white, which makes the blue color stand out.
#### PREDICATE DEVICE:
The predicate devices are the PrismaSATE dialysis solution formulations buffered by bicarbonate using hydrochloric acid as an excipient (pH adjustor) in PVC packaging cleared under K120333, and PrismaSATE dialysis solutions buffered by bicarbonate with lactate in polyolefin primary packaging cleared under K072908.
| Device | Company | Predicate 510(k) | Clearance Date |
|------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|------------------|------------------|
| PrismaSATE dialysate<br>solutions, bicarbonate<br>buffered with HCl<br>excipient in PVC<br>packaging | Baxter Healthcare<br>(formerly Gambro Renal<br>Products, Inc.) | K120333 | May 31, 2012 |
| PrismaSate dialysate<br>solutions bicarbonate<br>buffered with lactate in<br>polyolefin packaging | Baxter Healthcare<br>(formerly Gambro Renal<br>Products, Inc.) | K072908 | February 2, 2008 |
| | Table 1. Predicate Device(s) | |
|--|------------------------------|--|
| | | |
### DESCRIPTION OF THE DEVICE:
PrismaSATE dialysis solutions are ready to use sterile dialysate solutions for use in Continuous Renal Replacement Therapy (CRRT) for the treatment of acute renal failure and in other cases necessitating fluid or solute removal, such as in the case of drug poisoning with dialyzable or filterable substances. The solutions are intended to be used with commercially available CRRT systems as dialysate. A physician prescribes the chemical composition of the solution to be used. The solutions are sterile, and packaged in flexible bags. The composition of the fluid used in renal therapies mirrors normal plasma water, since normalization of the blood is the objective.
PrismaSATE solutions, like other currently available dialysate solutions, consist of various quantities of the following chemicals: sodium chloride, potassium chloride, magnesium chloride, calcium chloride, dextrose, and a buffer of lactose and/or bicarbonate. The solution is prescribed by a physician; the physician selects the appropriate formulation based on the patient's blood electrolyte level and desired traits. PrismaSATE dialysate solutions are offered in a range of electrolyte concentrations which mirror the range of electrolytes in plasma. The solution may be buffered using lactate, bicarbonate using lactic acid as a pH adjustor, or bicarbonate using hydrochloric
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly blurred, but still legible.
acid as a pH-adjusting excipient.. No changes to the available solution formulations are described in this Special 510(k).
The bicarbonate buffered PrismaSATE solutions currently on the market may be packaged in Polyvinyl Chloride (PVC) or in Polyolefin (non-PVC) two compartment bags. The final solution is obtained by breaking a frangible (PVC) or peel seal (non-PVC) between the compartments and mixing both solutions just prior to use. Neither the bag design nor the bag materials are being changed; they are not the subject of this Special 510(k).
The purpose of this Special 510(k) is to expand the PrismaSATE solution formulations buffered with bicarbonate using hydrochloric acid as a pH-adjusting excipient into the polyolefin primary packaging materials. The formulations that are the subject of this submission are described in the Table 2 below.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is in blue color. The letters are slightly italicized, giving the word a sense of movement.
| Composition for small compartment (contains 250 ml) | | | | | | |
|------------------------------------------------------|------------------|--------------------|---------------------|---------------------|-------|-------|
| Formulation | Calcium Chloride | Magnesium Chloride | Water for Injection | | | |
| BzK 2/0 | - | 2.44 | q.s. | | | |
| BzK 4/2.5 | 3.68 | 3.05 | q.s. | | | |
| BzK 0/3.5 | 5.15 | 2.03 | q.s. | | | |
| Composition for large compartment (contains 4750 ml) | | | | | | |
| Formulation | Sodium Chloride | Potassium Chloride | Sodium Bicarbonate | Water for Injection | | |
| BzK 2/0 | 6.46 | 0.157 | 3.09 | q.s. | | |
| BzK 4/2.5 | 6.46 | 0.314 | 3.09 | q.s. | | |
| BzK 0/3.5 | 6.46 | - | 3.09 | q.s. | | |
| Composition after reconstitution in mmol/l | | | | | | |
| Formulation | Na+ | K+ | Ca2+ | Mg2+ | HCO3- | Cl- |
| BzK 2/0 | 140 | 2 | 0 | 0.6 | 32 | 111.2 |
| BzK 4/2.5 | 140 | 4 | 1.25 | 0.75 | 32 | 116.0 |
| BzK 0/3.5 | 140 | 0 | 1.75 | 0.5 | 32 | 112.5 |
| Composition after reconstitution in mEq/l | | | | | | |
| Formulation | Na+ | K+ | Ca2+ | Mg2+ | HCO3- | Cl- |
| BzK 2/0 | 140 | 2 | 0 | 1.2 | 32 | 111.2 |
| BzK 4/2.5 | 140 | 4 | 2.5 | 1.5 | 32 | 116.0 |
| BzK 0/3.5 | 140 | 0 | 3.5 | 1.0 | 32 | 112.5 |
## Table 2. PrismaSATE dialysate solution bicarbonate buffered with HCl excipient, packaged in polyolefin
## INDICATIONS FOR USE:
PrismaSATE solutions are indicated for use as a dialysate in Continuous Renal Replacement Therapy.
## TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The proposed device has the same technological characteristics as the predicate device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font has a slight shadow effect, giving it a 3D appearance. The word is the logo for the healthcare company Baxter International.
| Features | Predicate Device<br>PrismaSATE<br>K120333 | Predicate Device<br>PrismaSATE<br>K072908 | Proposed Device |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Indications for use | Gambro PrismaSATE<br>Solutions are indicated<br>for use as a dialysate in<br>Continuous Renal<br>Replacement Therapy | Gambro PrismaSATE<br>Solutions are indicated<br>for use as a dialysate in<br>Continuous Renal<br>Replacement Therapy | Same:<br>PrismaSATE Solutions<br>are indicated for use as a<br>dialysate in Continuous<br>Renal Replacement<br>Therapy. |
| Sterile | Steam sterilized, to a<br>Sterilization Assurance<br>Level of at least 10-6. | Steam sterilized, to a<br>Sterilization Assurance<br>Level of at least 10-6. | Same:<br>Steam sterilized, to a<br>Sterilization Assurance<br>Level of at least 10-6. |
| Non-Pyrogenic | Non-pyrogenic solution | Non-pyrogenic solution | Same:<br>Non-pyrogenic solution |
| Single Use | Yes | Yes | Same: Yes |
| Solution Formulation | PrismaSATE solution<br>formulations buffered with<br>bicarbonate using<br>hydrochloric acid as<br>excipient | PrismaSATE solutions<br>buffered with bicarbonate<br>with lactate | Same as predicate K120333 |
| Primary Packaging<br>material | PVC | Polyolefin | Same as predicate<br>K072908 : Polyolefin |
| Shelf life | 12 months | 18 months | Same as predicate<br>K072908: 18 months |
#### Table 3. Device Comparison Table
# DISCUSSION OF NONCLINICAL TESTS:
A risk analysis was performed on the proposed combination of PrismaSATE solution formulations buffered with bicarbonate using hydrochloric acid as a pH-adjusting excipient (cleared under K120333) in the polyolefin primary packaging materials (cleared under K072908). No new risks were identified.
Baxter Healthcare Corporation performed design verification for the modification. The result met its acceptance criteria, and supports that the proposed device is appropriately designed for its intended use.
#### Performance Data:
No performance data is included in this Special 510(k).
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly italicized, giving it a sense of movement. The background is plain white, which makes the blue text stand out.
# Biocompatibility:
No new materials are introduced in this Special 510(k). No biocompatibility data is necessary.
# CONCLUSION:
The design control activities demonstrate that the subject device is substantially equivalent and performs comparably to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.