For storage of soft (hydrophilic), hard and rigid gas permeable contact lenses during chemical disinfection. Use for storage during chemical disinfection only. DO NOT USE WITH HEAT.
Device Story
Contact Lens Case is a simple, non-sterile storage container for soft, hard, and rigid gas permeable contact lenses. Device consists of a polypropylene base with two wells (4.8 ml capacity) and two polyethylene screw-on lids. Used by patients in home or clinical settings to hold lenses while submerged in chemical cleaning/disinfection solutions. Device provides physical containment for chemical disinfection process; does not perform disinfection itself. Patient retrieves lenses from case after disinfection cycle. Benefits include safe storage and maintenance of lens hydration during chemical cleaning.
Clinical Evidence
No clinical tests were performed. Evidence consists of bench testing, including biocompatibility (ISO 10993-5 cytotoxicity, ISO 10993-10 irritation/sensitization, ISO 10993-11 systemic toxicity) and leakage testing (240 samples, 100% pass rate).
Technological Characteristics
Materials: Polypropylene (base), Polyethylene (lid). Dimensions: 63x32x16mm. Capacity: 4.8ml per well. Design: Two-well base with screw-on caps. R/L markings included. Not sold sterile. Biocompatibility per ISO 10993-5, ISO 10993-10, and ISO 10993-11.
Indications for Use
Indicated for storage of soft (hydrophilic), hard, and rigid gas permeable contact lenses during chemical disinfection. Contraindicated for heat disinfection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a symbol that appears to be three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 9, 2016
Ningbo Yinzhou Zonghai Artware Co., Ltd. c/o Mr. Charles Mack IRC 7808 Rush Creek Drive Pasco. WA 99301
Re: K162869
Trade/Device Name: Contact Lens Case Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LRX Dated: September 28, 2016 Received: October 13, 2016
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162869
Device Name Contact Lens Case
Indications for Use (Describe)
For storage of soft (hydrophilic), hard and rigid gas permeable contact lenses during chemical disinfection. Use for storage during chemical disinfection only. DO NOT USE WITH HEAT.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
### Date: September 30, 2016
1. Company and Correspondent making the submission:
Name - Ningbo Yinzhou Zonghai Artware Co., Ltd. Address - Honglianchi, Guangsheng Village, Jishigang Town, Yinzhou District, Ningbo City Zhejiang Province, China 315172 Tel: +86-574-88003277 Fax: +86-574-88003255 Contact - Mr. Weng Changhai General Manager Email – : sales1@zonghai.net
US Correspondent Name- IRC, Charles Mack Address- 2091 Oak Drive, Lake Havasu City, Arizona 86406 Tel: 931-625-4938 Principal Engineer Email- charliemack@irc-us.com
- 2. Device :
| Trade/proprietary name | : Contact Lens Case |
|------------------------|----------------------|
| Common Name | : Contact Lens Case |
| Classification Name | : case, contact lens |
- 3. Predicate Devices :
| Manufacturer | Predicate Device | 510(k)<br>Number | Submitted Device |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------|------------------|-------------------|
| Ni Hau<br>Industrial Co.,<br>Ltd. | Cat Contact Lens Case<br>Sunglasses Contact Lens<br>Case<br>1st Contact Lens Case<br>Geeky Eye Contact Lens<br>Case | K142717 | Contact Lens Case |
- 4. Classifications Names & Citations : 21CFR 886.5928, LRX, Contact Lens Case
{4}------------------------------------------------
5. Description :
The Contact Lens Case is a simple device made with a polypropylene base and a polyethylene lid to store soft (hydrophilic), hard, and rigid gas permeable contact lenses during chemical disinfections. This device is manufactured only for chemical disinfections with approved chemical cleaning agents to clean contact lenses.
The physical style is simple with a base volume of 4.8 milliliter and dimensions of 6.25 mm wide and 9.25 mm high. This provides sufficient space to insert and retrieve the contact lenses. It also provides sufficient space to introduce the chemical cleaning solution and adequately cover the lenses for cleaning.
### 6. Indication for use:
For storage of soft (hydrophilic), hard, and rigid gas permeable contact lenses during chemical disinfections. Use for storage during chemical disinfections only. DO NOT USE WITH HEAT
### 7. Testing summary:
Biocompatibility testing was conducted as noted below:
ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity. The purpose of the test is to assess the potential cytotoxicity of test article Contact Lens Case in the In Vitro Cytotoxicity Test using L929 mouse fibroblast cells (From ATCC), which was sensitive to extractable cytotoxic articles. Under the conditions of this study, the test articles Contact Lens Case extract did not show potential toxicity to L-929 cells with polar and non-polar extracts.
ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. The test was designed to evaluate the potential of a test article extract to cause ocular irritation. According to what observed, the response of ocular on testing side does not exceed that on the control side. Thus, it is identified as grade 0. All animals were not found abnormal clinical symptoms. except ocular reactions. All animals were normal weight change. Under the conditions of this study, the test article extracts showed no significant evidence of causing ocular irritation.
ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. The purpose of this test is to evaluate the potential of a test article extract to cause skin sensitization in the guinea pig. All animals were not found abnormal clinical symptoms except skin reactions. All animals were normal weight change. The positive rate of all test group animals was 0%. The positive rate of all negative control animals was 0%. Under the conditions of this study, the test article extract showed no significant evidence of causing skin sensitization in the guinea pig.
{5}------------------------------------------------
ISO 10993-11 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity. The purpose of this testing was to evaluate the potential of a test article extract to cause acute systemic toxicity in mouse. No animals were found with any abnormal clinical symptoms.
All animals had normal weight change. Under the conditions of this study, the test article extracts showed no significant evidence of causing acute systemic toxicity in the mice.
Leakage testing was conducted to the contact lens case samples. 240 combinations of different tops and bottoms were filled to 2/3 with liquid. Each set was turned upside down for 15 minutes and the tests were repeated 3 times.
None of the tested contact lens cases showed any leakage and all passed the leakage tests successfully.
The test results show the Contact Lens Case complies with the requirement as defined.
8. Comparison with predicate device:
Ningbo Yinzhou Zonghai Artware Co., Ltd. believes that the Contact Lens Case is substantially equivalent to the (K142717) Ni Hau Industrial Co., Ltd. Cat Contact Lens Case; Sunglasses Contact Lens Case; 1st Contact Lens Case; Geeky Eye Contact Lens Case.
{6}------------------------------------------------
| Element of<br>comparison | Subject Device | Predict Device |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | Ningbo Yinzhou Zonghai Artware<br>Co., Ltd. | Ni Hau Industrial Co., Ltd. |
| FDA510(K) Number | N/A | K142717 |
| Device Name | Contact Lens Case | Cat Contact Lens Case<br>Sunglasses Contact Lens Case<br>1st Contact Lens Case<br>Geeky Eye Contact Lens Case |
| Classification | CFR 886.5928, Class II | CFR 886.5928, Class II |
| Product Code | LRX | LRX |
| Indications for Use | For storage of soft (hydrophilic),<br>hard, and rigid gas permeable<br>contact lenses during chemical<br>disinfections. Use for storage<br>during chemical disinfections only.<br>DO NOT USE WITH HEAT | For storage of soft (hydrophilic),<br>hard, and rigid gas permeable<br>contact lenses during chemical<br>disinfection. Use for storage during<br>chemical disinfections only. DO NOT<br>USE WITH HEAT. |
| Case Dimension | 63x32x16mm | Cat: 65 x 37 x 14mm<br>Sunglasses: 71 x 31 x 14mm<br>1st Contact: 60 x 39.6 mm<br>Geeky eyes: 80 x 32 x 21mm |
| Single Well<br>Capacity Volume | 4.8ml | Cat: 3ml<br>Sunglasses: 2.5ml<br>1st Contact: 3ml<br>Geeky eyes: 2.8ml |
| Device Material | BASE: Polypropylene(PP)<br>LID: Polyethylene(PE) | Acrylonitrile Butadene Styrene(ABS) |
| Device<br>Construction | Shaped base with two wells to hold<br>the fluid and two screw lids for the<br>Left and Right eyes of contact<br>lenses. | Shaped base with two wells to hold<br>the fluid and two screw lids for the<br>Left and Right eyes of contact<br>lenses. |
| Screw On Caps | Yes | Yes |
| R/L indications<br>on chamber bottoms<br>and/or top | Yes | Yes |
| Biocompatibility | Complying with ISO10993-1:<br>- Cytotoxicity (ISO 10993-5)<br>- Skin sensitization (ISO10993-10)<br>- Ocular Irritation (ISO 10993-10)<br>-SystematicToxicity (ISO10993-11) | PASS the following test:<br>- Cytotoxicity (ISO 10993-5)<br>- Eye Irritation (ISO 10993-10)<br>- Systematic Toxicity (ISO10993-11) |
| Disinfection Type | Chemical disinfection, not heat<br>disinfection. | Chemical disinfection, not heat<br>disinfection. |
| Sterilization | Not sold sterile | Not sold sterile |
{7}------------------------------------------------
## 9. Safety and Performance Data:
| Serial<br>Number/Version | Standard and Description | Scope |
|--------------------------|----------------------------------------------------------------------------------------------------|-----------------------------|
| Performance Standards | | |
| ISO10993-5: 2009 | Biological evaluation of medical devices -- Part 5: Tests<br>for In Vitro cytotoxicity | Biocompatibility<br>Testing |
| ISO10993-10: | Biological evaluation of medical devices - Part 10: Tests<br>for irritation and skin sensitization | Biocompatibility<br>Testing |
| ISO10993-10: | Biological evaluation of medical devices - Part 10: Tests<br>for ocular irritation | Biocompatibility<br>Testing |
| ISO10993-11: | Biological evaluation of medical devices - Part 11: Tests<br>for systemic toxicity | Biocompatibility<br>Testing |
- 10. Clinical: No clinical tests were performed.
The basic design of the submitted device specifications is substantially the same 11. as the predicate, with differences in the aesthetic design and plastic material. Though the plastic material is different, biocompatibility testing has verified that the submitted device is safe for use.
All the labeling and characteristics of the submitted OTC Contact Lens Case is the same as the predicate devices and most typical OTC contact lens cases currently on the market. The submitted device and predicate both are used for storage and disinfection by chemical solution of soft (hydrophilic), hard, and rigid gas permeable contact lenses.
- 11. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Ningbo Yinzhou Zonghai Artware Co., Ltd. concludes that the Contact Lens Case is substantially equivalent to predicate devices as described herein.
END
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.