← Product Code [PEK](/productcode/PEK) · K162849

# Huvex Interspinous Fixation System (K162849)

_Dio Medical Co., Ltd. · PEK · Feb 16, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K162849

## Device Facts

- **Applicant:** Dio Medical Co., Ltd.
- **Product Code:** [PEK](/productcode/PEK.md)
- **Decision Date:** Feb 16, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Huvex Interspinous Fixation System is a single-level, posterior, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/ attachment to the spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), and/or tumor. The Huvex Interspinous Fixation System is intended for use at one level, in conjunction with autogenous bone graft, and not intended for stand-alone use.

## Device Story

Huvex Interspinous Fixation System is a posterior, non-pedicle, single-level supplemental fixation device. Components include left/right plates, pin, bolt, inner cap, center bar, and set screw. Left plate features poly-axial bar with bone graft window for fusion; right plate locks to bar via set screw. Used in non-cervical spine (T1-S1) as adjunct to fusion with autogenous bone graft; not for stand-alone use. Implanted by surgeons in clinical settings to stabilize spinous processes. Provides mechanical support to facilitate spinal fusion; benefits patients by addressing instability from degenerative disc disease, spondylolisthesis, trauma, or tumor.

## Clinical Evidence

Bench testing only. Performance demonstrated via static and dynamic compression, static torsion, and static tension testing per modified ASTM F1717, and static grip testing per modified ASTM F1798.

## Technological Characteristics

Titanium alloy (Ti-6Al-4V ELI) per ASTM F136. Components: left/right plates, pin, bolt, inner cap, center bar, set screw. Poly-axial bar design. Non-sterile, single-use implants. Reusable surgical instruments provided. Mechanical fixation via spinous process attachment.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- Globus Medical - Sp-Fix Spinous Process Fixation Plate ([K102195](/device/K102195.md))
- Medtronic - CD Horizon SPIRE™ Spinous Process Plate ([K032037](/device/K032037.md))
- NuVasive - Affix I & II Spinous Process Plate system ([K131238](/device/K131238.md) & [K143388](/device/K143388.md))
- Alphatec Spine - Bridgepoint Spinous Process System ([K103205](/device/K103205.md))
- Life Spine - Aileron posterior fusion system ([K140236](/device/K140236.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 16, 2017

DIO Medical Co., Ltd. Mr. Jaewon Jang Manager 8770 West Bryn Mawr Avenue, Suite 1250 Chicago, Illinois 60631

Re: K162849

Trade/Device Name: Huvex Interspinous Fixation System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: PEK Dated: January 27, 2017 Received: January 30, 2017

Dear Mr. Jang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K162849

Device Name Huvex Interspinous Fixation System

#### Indications for Use (Describe)

The Huvex Interspinous Fixation System is a single-level, posterior, non-pedicle supplemental fixation device intended for use in the non-cervical spine (TI-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/ attachment to the spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), and/or tumor. The Huvex Interspinous Fixation System is intended for use at one level, in conjunction with autogenous bone graft, and not intended for stand-alone use.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

 Registration for Part 40 CFR 266, Subpart S Burner
  On-Site Combustion Unit (No CFR)
|× | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

## DIO Medical Co., Ltd.

## Huvex Interspinous Fixation System

Sponsor: DIO Medical Co., Ltd. 8770 West Bryn Mawr Avenue, Suite 1250 Chicago, Illinois 60631 Official Contact: Jaewon Jang Phone: 847-795-1078 ext. 103 Date Prepared: February 15, 2017 Device Name: Huvex Interspinous Fixation System Common Name: Spinous Process Plate Classification Spinal interlaminal fixation orthosis Name: Classification 21 CFR 888.3050 Number: Product PEK, class II Code/Classification: Description: The Huvex Interspinous Fixation System consists of a left plate, a right plate, pin, bolt, inner cap, center bar, and set screw. Each of these components is provided in several sizes to allow for the construction of five different Huvex Interspinous Fixation implant sizes. The left plate is provided assembled with the poly axial bar. The bar has a bone graft window to allow fusion between spinous process. Poly axial bar is also designed to fit the anatomical characteristics of the spinous process. The right plate is designed to be combined with left plate fixed to spinous process. Right plate contains a set screw to lock the right plate to the poly axial bar. The Huvex Interspinous Fixation System components are supplied non-sterile, are single use and are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136. In addition to the implants a set of reusable surgical instruments are provided.

Both implant and instruments have trays that are used for handling and storage.

{4}------------------------------------------------

| Intended Use:                        | The Huvex Interspinous Fixation System is a single-level, posterior, non-pedicle<br>supplemental fixation device intended for use in the non-cervical spine (T1-S1) as ar<br>adjunct to fusion in skeletally mature patients. It is intended for plate<br>fixation/attachment to the spinous processes for the purpose of<br>achieving supplemental fusion in the following conditions: degenerative disc<br>disease (defined as back pain of discogenic origin with degeneration of the disc<br>confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e.,<br>fracture or dislocation), and/or tumor. The Huvex Interspinous Fixation System is<br>intended for use at one level, in conjunction with autogenous bone graft, and not<br>intended for stand-alone use. |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Data:                    | Static & Dynamic compression, Static torsion, and Static tension testing in<br>accordance with a modified ASTM F1717 setup, in addition Static grip testing<br>with a modified ASTM F1798 setup were conducted to demonstrate substantial<br>equivalence to the predicate system(s).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Predicate Device:                    | Primary predicate: Globus Medical- Sp-Fix Spinous Process Fixation Plate<br>(K102195)<br>Additional predicates: Medtronic - CD Horizon SPIRE™ Spinous Process Plate<br>(K032037), NuVasive - Affix I & II Spinous Process Plate system (K131238 &<br>K143388), Alphatec Spine - Bridgepoint Spinous Process System (K103205) and Life<br>Spine - Aileron posterior fusion system (K140236)                                                                                                                                                                                                                                                                                                                                                                                                   |
| Performance and<br>SE Determination: | The Huvex Interspinous Fixation System has been demonstrated to be<br>substantially equivalent to the predicate system(s) with respect to technical<br>characteristics, performance, and intended use. The information provided within<br>this premarket notification supports substantial equivalence of the subject device<br>to the predicate device(s).                                                                                                                                                                                                                                                                                                                                                                                                                                  |

---

**Source:** [https://fda-staging.innolitics.com/device/K162849](https://fda-staging.innolitics.com/device/K162849)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com