K162830 · Surgical Information Sciences, Inc. · LLZ · Feb 14, 2017 · Radiology
Device Facts
Record ID
K162830
Device Name
SIS Software
Applicant
Surgical Information Sciences, Inc.
Product Code
LLZ · Radiology
Decision Date
Feb 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K162830 · Feb 14, 2017
SIS Software
Surgical Information Sciences, Inc.
Retrospective clinical MRI imaging database (1.5T, 3T, and 7T MRI)
The sponsor used a retrospective database of clinical MRI images to develop and validate a machine-learning algorithm for the visualization of the subthalamic nucleus (STN).
Machine learning; Retrospective database; MRI imaging; Algorithm development
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Pivotal validation testing; Retrospective validation study using a separate dataset
34 subjects with both clinical MRI (1.5T and 3T) and High Field (7T) MRI; Sample Size: 34 subjects (68 STNs)
Not applicable for this study
Center of mass distance, surface distance, and Dice coefficient values
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Subthalamic nuclei visualization
—
90% of center of mass distances and surface distances not greater than 2.0mm
95% of center of mass distances and 100% of surface distances not greater than 2.0mm
10 subject datasets retrospectively selected from the locked SIS reference database.
—
34 subjects (68 STNs) scanned with both clinical MRI (1.5T and 3T) and High Field (7T) MRI.
—
Indications for Use
SIS Software is an application intended for use in the viewing, presentation of medical imaging, including different modules for image fusion, and intraoperative functional planning where the 3D output can be used with stereotactic image quided surgery or other devices for further processing and visualization. The device can be used in conjunction with other clinical methods as an aid in visualization of the subthalamic nuclei (STN).
Device Story
SIS Software is a medical imaging application for viewing, presentation, and intraoperative functional planning. It takes standard clinical MRI (1.5T or 3T) as input and uses a machine learning algorithm trained on a reference database of high-resolution 7T MRI images to identify and segment the subthalamic nucleus (STN). The device generates a patient-specific 3D anatomical model of the STN, which is exported to stereotactic image-guided surgery systems. Used by surgeons, neurologists, and radiologists in clinical settings, the software supplements standard clinical methods to improve visualization of brain structures not easily seen on standard MRI. By providing more accurate anatomical localization, the device assists in neurosurgical planning, potentially improving surgical precision and patient outcomes.
Clinical Evidence
Validation study used 34 subjects (68 STNs) scanned with both clinical (1.5T/3T) and high-field (7T) MRI. Data was independent of the training/development set. Primary endpoints: center of mass distance, surface distance, and Dice coefficient. Results: 95% of center of mass distances ≤ 2.0mm (p<0.0001 vs. standard of care); 100% of surface distances ≤ 2.0mm. Mean Dice coefficient was 0.64. Bench testing using leave-one-out cross-validation on 10 subjects showed 1mm average distance error.
Technological Characteristics
Software-based medical imaging application. Uses machine learning and image processing algorithms to generate 3D segmented anatomical models from 1.5T or 3T MRI scans by referencing a database of 7T MRI images. Performs image fusion, slice-by-slice segmentation, and 3D object conversion. Supports export to stereotactic planning systems. Operates on standard clinical imaging hardware.
Indications for Use
Indicated for medical professionals, including surgeons, neurologists, and radiologists, as an aid in the visualization of the subthalamic nuclei (STN) and for intraoperative functional planning using medical imaging.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three human profiles facing to the right, stacked on top of each other, with flowing lines extending from the bottom profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 14, 2017
Surgical Information Sciences, Inc. % Ms. Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th Floor PHILADELPHIA PA 19103
Re: K162830
Trade/Device Name: SIS Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 12, 2017 Received: January 12, 2017
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last paqe
510(k) Number (if known)
K162830
Device Name
SIS Software
Indications for Use (Describe)
SIS Software is an application intended for use in the viewing, presentation of medical imaging, including different modules for image fusion, and intraoperative functional planning where the 3D output can be used with stereotactic image quided surgery or other devices for further processing and visualization. The device can be used in conjunction with other clinical methods as an aid in visualization of the subthalamic nuclei (STN).
Typical users of the SIS Software are medical professionals, including but not limited to surgeons, neurologists and radiologists.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| <div> <span style="text-decoration: overline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="text-decoration: none;"></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
### Surgical Information Sciences, Inc.'s SIS Software Device
#### Sponsor's Name, Address, Telephone Number, Contact Person and Date Prepared
Surgical Information Sciences. Inc. 60 South 6th Street, Suite 2410 Minneapolis, MN 55402 Contact Person: Mark Headrick Phone: (612) 335-8683 E-mail: mark.headrick@surqicalis.com
Date Prepared: January 12, 2017
### Trade Name of Device
SIS Software
### Common or Usual Name / Classification Name
Picture Archiving and Communication System (Product Code: LLZ; 21 C.F.R. 892.2050)
### Predicate and Reference Devices
Medtronic's StealthStation (K050438) Medtronic's Stealth Viz Advanced Planning Application with StealthDTl Package (K081512) Brainreader ApS' Neuroreader Image Processing Software (K140828) (Reference device)
# Intended Use / Indications for Use
SIS Software is an application intended for use in the viewing, presentation and documentation of medical imaging, including different modules for image processing, image fusion, and intraoperative functional planning where the 3D output can be used with stereotactic image guided surgery or other devices for further processing and visualization. The device can be used in coniunction with other clinical methods as an aid in visualization of the subthalamic nuclei (STN).
Typical users of the SIS Software are medical professionals, including but not limited to surgeons, neurologists and radiologists.
# Technological Characteristics
SIS Software uses machine learning and image processing to enhance standard clinical images for the visualization of the subthalamic nucleus ("STN"). The SIS Software supplements the information available through standard clinical methods, providing additional, adjunctive information to surgeons, neurologists and radiologists for use in visualization and planning stereotactic surgical procedures. SIS Software provides a patient
{4}------------------------------------------------
specific, 3D anatomical model of the patient's own brain structures that supplements other clinical information to facilitate visualization in neurosurqical procedures. The software makes use of the fact that some structures in the brain are not easily visualized in 1.5T or 3T clinical MRJ, but are better visualized using high-resolution and high-contrast 7T MRI.
The company's software methodology relies on a reference database of high-resolution brain images (7T MRI) and standard clinical brain images (1.5T or 3T MRI). The 7T images allow visualization of anatomical structures that are then used to find regions of interest within the brain (i.e., the STN) on a patient's clinical image.
SIS visualization is incorporated in the standard clinical MR data, thereby not changing the current standard-of-care workflow protocol and does not require any additional visualization software or hardware platforms.
### Performance Data
Developmental testing of the reference database was performed using 10 subject datasets that were retrospectively selected from the locked SIS reference database. For each of the selected datasets, the patient's 7T MRI and labeled structures were excluded from the dataset, such that only the corresponding clinical MRI image remained for the test. The subject machine-learning method then predicted the subthalamic nucleus (STN) on these removed datasets and this prediction was compared to the STN as segmented on the 7T image of that subject to validate the prediction when the tested subject was removed from the dataset (standard leave-one-out statistical procedure). The average distance between the predicted and the original (on the 7T) was 1mm, the actual size of the pixel (data resolution). The overlap between the 3D predicted and the original STN was significantly better (p<0.05) in comparison to the overall of a standard atlas and the original STN.
The pivotal validation testing of the subject device, including the reference database, included images from 34 subjects to validate the performance of the SIS Software. A set of 68 STNs (from 34 subjects) were scanned with both clinical MRI (1.5T and 3T) and High Field (7T) MRI. None of the 68 STNs were part of the company's database for algorithm development and none were used to optimize or design the company's software. Thus, this validation data set was completely separate from the data set that was used for development. The software development was frozen and labeled before tested on this validation set.
Three measurements were used to compare the SIS visualization via the subject software and ground truth STNs (manually segmented clinical images and 7T images superimposed): (1) Center of mass distance; (2) Surface distance; and (3) Dice coefficient values.
In sum, 90% of the center of mass distances and surface distances were below 1.9mm and 0.8mm, respectively. Specifically, 95% of the center of mass distances and 100% of the surface distances were not greater than 2.0mm. Thus, the study met the pre-specified criteria of 90% of center of mass distances and surface distances not greater than 2.0mm. Furthermore, the proportion of visualizations not greater than 2.0mm was conservatively estimated from the literature to be 20%. Therefore, the rate of successful visualizations from SIS Software (95% of the center of mass distances not greater than 2.0mm) is significantly greater than the standard of care (p<0.0001). The corresponding confidence intervals are as follows:
{5}------------------------------------------------
- (a) 90% of the center of mass distances and surface distances were below 1.9mm and 0.8mm, respectively (95% CI: 80.74 - 95.56%);
- (b) 95% of the center of mass distances were not greater than 2.0mm (95% C): 86.91 98.37%):
- (c) 100% of the surface distances were not greater than 2.0mm (95% Cl: 94.25 100%).
In addition, the Dice coefficient in this dataset was 0.64, which was expected given the small size of the STN.
In sum, the SIS Software performed as intended and clinical validation data results observed were as expected.
# Substantial Equivalence
With regard to technological characteristics, both the SIS Software and the predicates (K050438; K081512) are all software applications that can be used for visualization, presentation and documentation of medical imaging, including different modules for image processing, image fusion, intraoperative functional planning where the 2D or 3D output can be used with stereotactic image guided surgery or other devices for further processing and visualization. In addition, the SIS Software, like the identified predicate and reference devices, use proprietary algorithms to generate 3D segmented anatomical models from patient's MRI scans.
The SIS Software and predicate devices also all perform image fusion of datasets using automated or manual image matching techniques. Below provides a summary comparison between the SIS Software and the predicate and reference devices.
| | SIS Software | Medtronic<br>Navigation,<br>Inc.'s<br>Stealth Viz<br>Advanced<br>Planning<br>Application with<br>StealthDTI<br>Package<br>(K081512) | Medtronic<br>Navigation,<br>Inc.'s<br>StealthStation<br>(K050438) | Braindreader<br>ApS'<br>NeuroReader<br>Medical<br>Image<br>Processing<br>Software<br>(K140828) |
|--------------------------------------------------------------------------------------------------------------------|--------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Allows for importing of<br>digital imaging sets | Yes | Yes | Yes | Yes |
| Uses proprietary<br>software algorithm to<br>generate 3D segmented<br>anatomical models from<br>patient's MR scans | Yes | Yes | Yes | Yes |
| Allows for review and<br>analysis of data in<br>various 2D and 3D<br>presentation formats | Yes | Yes | Yes | Yes |
| | | | | SIS Software Technological Characteristics Comparison Table | |
|--|--|--|--|-------------------------------------------------------------|--|
|--|--|--|--|-------------------------------------------------------------|--|
{6}------------------------------------------------
| | SIS Software | Medtronic<br>Navigation,<br>Inc.'s<br>Stealth Viz<br>Advanced<br>Planning<br>Application with<br>StealthDTI<br>Package<br>(K081512) | Medtronic<br>Navigation,<br>Inc.'s<br>StealthStation<br>(K050438) | Braindreader<br>ApS'<br>NeuroReader<br>Medical<br>Image<br>Processing<br>Software<br>(K140828) |
|-------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Performs image fusion of<br>datasets using<br>automated or manual<br>image matching<br>technique | Yes; atlas-based<br>mapping (patient<br>specific);<br>reference<br>database | Yes; atlas-based<br>mapping | Yes, atlas-based<br>mapping | Yes; reference<br>database |
| Segments structures in<br>images with manual and<br>automated tools and<br>converts them into 3D<br>objects for display | Yes | Yes | Yes | Yes |
| Creates hybrid datasets<br>by filing in segmented<br>regions slice-by-slice on<br>anatomical datasets | Yes | Yes | Yes | Yes |
| Exports results to<br>planning system | Yes | No | Yes | Yes |
In sum, the SIS Software has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate and reference devices. The minor technological differences between the subject device and its predicate and reference devices raise no different questions of safety or effectiveness. Performance data demonstrate that the SIS Software performs as intended and is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.