K162822 · Arkray Factory, Inc. · PDJ · Jun 29, 2017 · Clinical Chemistry
Device Facts
Record ID
K162822
Device Name
ADAMS A1c HA-8180V, CALIBRATOR 80
Applicant
Arkray Factory, Inc.
Product Code
PDJ · Clinical Chemistry
Decision Date
Jun 29, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1373
Device Class
Class 2
Indications for Use
The ADAMS A1c HA-8180V system is intended for the quantitative determination of hemoglobin A1c (IFCC mmol/mol and NGSP %) in human whole blood and hemolysate samples using Ion-Exchange high-performance liquid chromatography (HPLC). Hemoglobin A1c measurements obtained from the ADAMS A1c HA-8180V are used as an aid in the diagnosis of diabetes mellitus, an aid in the identification of individuals who may be at risk of developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus. The ADAMS H1c HA-8180V is intended for professional use only. The CALIBRATOR 80 is for the calibration of the ADAMS A1c HA-8180V system used for the quantitative determination of hemoglobin A1c in human whole blood and hemolysate samples. For In Vitro Diagnostic Use Only.
Device Story
The ADAMS A1c HA-8180V is a professional-use, automated HPLC system for HbA1c quantification. It accepts anticoagulated whole blood (K2/K3-EDTA) or manual hemolysates. The system performs automated 1:101 dilution using hemolysis washing solution. Samples are injected into a column where hemoglobin fractions are separated via reversed-phase cation exchange chromatography using a three-eluent mobile phase of increasing ionic strength. Fractions are detected via dual-wavelength (420 nm and 500 nm) LED colorimetric sensors. A microcomputer processes the resulting chromatograms to identify and quantify hemoglobin peaks. Output includes HbA1c in NGSP (%) and IFCC (mmol/mol) units. Results assist clinicians in diagnosing diabetes and monitoring long-term glycemic control. The system includes bi-level lyophilized calibrators and requires professional operation in clinical settings.
Clinical Evidence
No clinical trials were performed. Analytical performance was established via bench testing, including precision (reproducibility studies across 20 days), linearity (3.0-19.0% HbA1c), and interference testing (evaluating hemoglobin variants A2, C, D, E, F, S and common substances). Method comparison against the NGSP reference method (Tosoh G8) using 143 samples showed high correlation (R²=0.998) and acceptable bias.
Technological Characteristics
Reversed-phase cation exchange HPLC; dual-wavelength (420/500 nm) LED colorimetric detection. Reagents include phosphate-buffered eluents and hemolysis washing solution. Dimensions/form factor: floor-standing/benchtop analyzer. Connectivity: internal barcode reader for sample ID and calibration. Software: microcomputer-based peak identification and quantification. Sterilization: N/A (reagents are chemical solutions).
Indications for Use
Indicated for quantitative determination of HbA1c in human whole blood and hemolysate samples to aid in diagnosis of diabetes mellitus, identification of individuals at risk for diabetes, and monitoring long-term blood glucose control in patients with diabetes. Intended for professional clinical laboratory use.
Regulatory Classification
Identification
A hemoglobin A1c test system is a device used to measure the percentage concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c is used as an aid in the diagnosis of diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.
Special Controls
(b) Classification. Class II (special controls). Hemoglobin A1c test systems must comply with the following special controls: 1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. 2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity and interference, including the following: i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0%, 6.5%, 8.0% and 12% hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least 3 lots of the device and 3 instruments, as applicable. ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the new device and compare results of the new device to results of the standardized test method. Results must demonstrate little or no bias versus the standardized method. iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6%. iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2 and Hemoglobin S. 3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA.
(2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity, and interference, including the following:
(i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0 percent, and 12 percent hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least three lots of the device and three instruments, as applicable.
(ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the device and compare results of the new device to results of a standardized test method. Results must demonstrate little or no bias versus the standardized method.
(iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6 percent.
(iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S.
(3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.
Predicate Devices
Tosoh Bioscience, Inc. Automated Glycohemoglobin Analyzer HLC-723G8 (k131580)
Tosoh Bioscience, Inc. Hemoglobin A1c Calibrator Set (k071132)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k162822
B. Purpose for Submission:
New device
C. Measurand:
Whole Blood Glycosylated Hemoglobin (HbA1c)
D. Type of Test:
Reversed-Phase Cation Exchange High Performance Liquid Chromatography (HPLC)
E. Applicant:
ARKRAY Factory, Inc.
F. Proprietary and Established Names:
ADAMS A1c HA-8180V
CALIBRATOR 80
G. Regulatory Information:
| Regulation Description | Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- | --- |
| Hemoglobin A1c Test System | PDJ | Class II | 21 CFR 862.1373 | Chemistry 75 |
| Glycosylated Hemoglobin Assay | LCP | Class II | 21 CFR 864.7470 | Hematology 81 |
| Calibrator | JIT | Class II | 21 CFR 862.1150 | Chemistry 75 |
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below
2. Indication(s) for use:
The ADAMS A1c HA-8180V system is intended for the quantitative determination of hemoglobin A1c (IFCC mmol/mol and NGSP %) in human whole blood and hemolysate samples using Ion-Exchange high-performance liquid chromatography (HPLC).
Hemoglobin A1c measurements obtained from the ADAMS A1c HA-8180V are used as an aid in the diagnosis of diabetes mellitus, an aid in the identification of individuals who may be at risk of developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus. The ADAMS H1c HA-8180V is intended for professional use only.
The CALIBRATOR 80 is for the calibration of the ADAMS A1c HA-8180V system used for the quantitative determination of hemoglobin A1c in human whole blood and hemolysate samples.
For In Vitro Diagnostic Use Only.
Special conditions for use statement(s):
Hemoglobin A1c tests should not be used as a diagnostic aid for diabetes in patients with a hemoglobinopathy but normal red cell turnover (e.g. sickle cell trait).
Hemoglobin A1c tests should not be used as a diagnostic aid in patients with abnormal red cell turnover (e.g. anemias from hemolysis and iron deficiency), or patients with iron deficiency and hemolytic anemia, various hemoglobinopathies, thalassemias, hereditary spherocytosis, malignancies, and severe chronic hepatic and renal disease.
Hemoglobin A1c tests should not be used as a diagnostic aid in pregnant women as it reflects the average blood glucose level over the preceding 3 months (average life of a red blood cell), and therefore may be falsely low during pregnancy or any other condition associated with recent onset of hyperglycemia and/or decreased red cell survival.
Hemoglobin A1c testing should not be used to replace glucose testing for type 1 diabetes in pediatric patients and in pregnant women.
Hemoglobin A1c testing should not be used in the diagnosis of gestational diabetes.
Hemoglobin A1c testing should not be used in patients with homozygous sickle cell trait, hemolytic anemia or other hemolytic diseases, or in patients with a recent significant
2
{2}
blood loss.
In cases of rapidly evolving type 1 diabetes the increase of HbA1c values might be delayed compared to the acute increase in glucose concentrations. In these conditions diabetes mellitus must be diagnosed based on plasma glucose concentration and/or the typical clinical symptoms.
The results of interference studies performed to evaluate the effect of hemoglobin variants on the HbA1c measurement show there is no significant interference for HbA2 concentrations of less than or equal to 16%, HbC concentrations of less than or equal to 39%, HbD concentrations of less than or equal to 36%, HbE concentrations of less than or equal to 30%, HbF concentrations of less than or equal to 30%, and HbS concentrations of less than or equal to 40%.
For prescription use only.
3. Special instrument requirements:
ADAMS A1c HA-8180V analyzer.
I. Device Description:
The ADAMS HA-8180V HbA1c system includes two levels of calibrators, diluents used to reconstitute the calibrators, three eluent solutions, a washing solution, and a control dilution set (used for sample dilutions and control material preparation). The components are described below:
CALIBRATOR 80: A boxed set of two levels of calibrator material, CALIBRATOR 80 Low and CALIBRATOR 80 High, which are single use, lyophilized calibrators from human-sourced hemoglobin used to calibrate the HA-8180V analyzer. CALIBRATOR 80 Low has a concentration of approximately 5.5% HbA1c (NGSP) and CALIBRATOR 80 High has a concentration of approximately 11.5% HbA1c (NGSP).
CALIBRATOR 80 Diluent: Phosphate-buffered containing surfactant used to reconstitute CALIBRATOR 80 Low and High.
Eluent 80A, Eluent 80B, and Eluent 80CV: Phosphate-buffered solution containing <0.06% sodium azide as preservative used in the chromatographic separation of hemoglobin fractions by elution through column.
Hemolysis Washing Solution 80H: Phosphate-buffered solution containing a surfactant and <0.02% sodium azide as a preservative, used for on-board hemolysis of whole blood samples.
Diluent 80: Phosphate-buffered solution containing a surfactant used for the dilution of liquid controls and whole blood samples.
{3}
Reconstitute 80: Phosphate-buffered solution containing a surfactant, used for the dilution of lyophilized controls.
The calibrator set reagents are provided in glass bottles capped with a rubber stopper and plastic cover. The eluents are provided in pouches consisting of a polyethylene (PE), aluminum foil, and polyethylene terephthalate (PET) laminate. The hemolysis washing solution and diluent 80 are provided in PE bottles. The hemolysis washing solution utilizes a polypropylene (PP) cap, while the diluent 80 uses a PE inner cap and PP outer cap. The reconstituent 80 reagent is provided in a PP bottle with a PP cap.
All human source materials were tested by FDA approved methods and found to be negative for the presence of HBs Ag and antibody to HIV1/HIV2, and HCV
J. Substantial Equivalence Information:
1. Predicate device name(s):
Tosoh Bioscience, Inc. Automated Glycohemoglobin Analyzer HLC-723G8
Tosoh Bioscience, Inc. Hemoglobin A1c Calibrator Set
2. Predicate 510(k) number(s):
k131580
k071132
3. Comparison with predicate:
4
{4}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Candidate Device
ARKRAY ADAMS
A1c HA-8180V | Predicate Device
Tosoh Glycohemoglobin Analyzer HLC-723G8
(k131580) |
| Intended Use | For the quantitative measurement of HbA1c to be used as an aid in diagnosis of diabetes, an aid in the identification of patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus. | Same |
| Assay Principle | Reversed-phase cation exchange high performance liquid chromatography | Cation-exchange high performance liquid chromatography |
| Detection Method | Dual wavelength (420 nm and 500 nm) LED colorimetric detector) | Dual wavelength (415 nm and unknown) LED colorimetric detector |
| HbA1c Reporting Units | %HbA1c, mmol/mol, chromatogram | Same |
| Specimen Type | Human whole blood, hemolysate sample | Human whole blood |
| Matrix | K3-EDTA and K2-EDTA whole blood | K3-EDTA whole blood |
| Measuring range | 4-15% HbA1c (20-140 mmol/mol) HbA1c | 4-16.9% (20-161 mmol/mol) HbA1c |
| Standardization/Traceability | Traceable to the Diabetes Control and Complications Trial (DCCT) reference method and IFCC. Certified via the National Glycohemoglobin Standardization Program (NGSP). | Same |
{5}
| Calibrator Similarities and Differences | | |
| --- | --- | --- |
| Characteristics | Candidate Device CALIBRATOR 80 | Predicate Device Tosoh Bioscience, Inc. Hemoglobin A1c Calibrator Set (k071132) |
| Intended Use | For the calibration of the HbA1c measuring system. | Same |
| Levels | Bi-level calibrators (low and high) with human-source material. | Same |
| HbA1c Concentrations | Level 1: 5.0-6.0%
Level 2: 10.0-11.0% | Level 1: 5.5-6.5%
Level 2: 10.5-11.5% |
| Matrix | Lyophilized calibrator from human-sourced hemoglobin. | Buffed human red blood cells, 2 mg/mL human hemoglobin |
| Stability | Shelf life: 18 moths at 2-8°C, Open bottle: 8 hrs. at 2-8 °C | Shelf life: 2 years at 2-8°C, Open bottle: one week at 2-8 °C |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP07-A2, Interference Testing in Clinical Chemistry –second edition; Approved Guideline
CLSI EP05-A2, Evaluation of Precision Performance of Quantitative Measurement Methods-second edition; Approved Guideline
CLSI-EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
{6}
7
L. Test Principle:
Hemoglobin test specimens are prepared by automated dilution (1:101) of anticoagulated whole blood in hemolysis washing solution 80H. Alternatively, hemolysates may be prepared “off-line” by manual dilution of anticoagulated whole blood with the diluent 80 provided in control dilution set 80. The sample is then injected into the column, which contains a filter to remove sample impurities. The differing polarity and ionic charges of the released hemoglobin fractions bind with variable strength to the negatively charged stationary phase of the column. The mobile phase, consisting of three eluents of increasing ionic strength, is injected into the column. Hemoglobin fractions eluted from the column are identified and quantified using light absorption, and are measured at 420 nm and 500 nm. The dual wavelength colorimetric analysis of the separated peaks is processed by a microcomputer to obtain peak identification and hemoglobin fraction quantitation.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
A precision study was performed in accordance to CLSI EP05 by testing 4 concentrations of unaltered K2-EDTA whole blood samples at the following targeted HbA1c values: ~5%, ~6.5%, ~8% and ~12%. Three levels of controls were also tested. Each sample was tested as whole blood and as hemolysates. All samples were run on three HA-8180V analyzers, each with three lots of each reagent. The aliquots of each sample type were tested in duplicate in two runs on each analyzer per day over a course of 20 days for a total of 720 results per sample. The results are summarized in the tables below for both whole blood and hemolysate applications in NGSP units (%HbA1c) and IFCC units (mmol/mol).
{7}
Whole Blood Application: ADAMS HA-8180V analyzer # 1 (NSGP)
| Mean %HbA1c | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient1, 5.2% | 0.01 | 0.3% | 0.00 | 0.0% | 0.00 | 0.0% | 0.00 | 0.0% | 0.01 | 0.3% |
| Patient 2, 6.4% | 0.02 | 0.3% | 0.01 | 0.2% | 0.00 | 0.0% | 0.05 | 0.8% | 0.05 | 0.8% |
| Patient 3, 7.9% | 0.03 | 0.4% | 0.03 | 0.4% | 0.01 | 0.2% | 0.06 | 0.7% | 0.07 | 0.9% |
| Patient 4, 11.8% | 0.03 | 0.3% | 0.02 | 0.1% | 0.04 | 0.3% | 0.07 | 0.6% | 0.08 | 0.7% |
| Control 1, 5.2% | 0.03 | 0.7% | 0.03 | 0.7% | 0.02 | 0.3% | 0.02 | 0.4% | 0.05 | 1.1% |
| Control 2, 9.0% | 0.03 | 0.3% | 0.03 | 0.3% | 0.01 | 0.2% | 0.01 | 0.1% | 0.04 | 0.5% |
| Control 3, 13.4% | 0.02 | 0.2% | 0.04 | 0.3% | 0.03 | 0.2% | 0.02 | 0.1% | 0.06 | 0.4% |
{8}
Whole Blood Application: ADAMS HA-8180V analyzer # 2 (NSGP)
| Mean % HbA1c | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient 1, 5.2% | 0.02 | 0.4% | 0.01 | 0.2% | 0.00 | 0.0% | 0.01 | 0.2% | 0.03 | 0.5% |
| Patient 2, 6.4% | 0.01 | 0.1% | 0.00 | 0.0% | 0.00 | 0.0% | 0.06 | 0.9% | 0.06 | 0.9% |
| Patient 3, 7.9% | 0.03 | 0.4% | 0.02 | 0.3% | 0.01 | 0.1% | 0.08 | 1.0% | 0.09 | 1.1% |
| Patient 4, 11.8% | 0.05 | 0.4% | 0.01 | 0.1% | 0.03 | 0.2% | 0.10 | 0.9% | 0.12 | 1.0% |
| Control 1, 5.2% | 0.02 | 0.4% | 0.03 | 0.5% | 0.01 | 0.3% | 0.04 | 0.9% | 0.06 | 1.1% |
| Control 2, 9.0% | 0.02 | 0.2% | 0.03 | 0.4% | 0.01 | 0.2% | 0.04 | 0.5% | 0.06 | 0.6% |
| Control 3, 13.4% | 0.03 | 0.2% | 0.03 | 0.3% | 0.04 | 0.3% | 0.05 | 0.4% | 0.08 | 0.6% |
Whole Blood Application: ADAMS HA-8180V analyzer # 3 (NSGP)
| Mean % HbA1c | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient 1, 5.2% | 0.02 | 0.4% | 0.02 | 0.4% | 0.00 | 0.0% | 0.01 | 0.2% | 0.03 | 0.6% |
| Patient 2, 6.4% | 0.02 | 0.4% | 0.01 | 0.2% | 0.00 | 0.0% | 0.06 | 0.9% | 0.06 | 1.0% |
| Patient 3, 7.9% | 0.03 | 0.4% | 0.02 | 0.2% | 0.02 | 0.3% | 0.05 | 0.7% | 0.07 | 0.9% |
| Patient 4, 11.8% | 0.02 | 0.2% | 0.03 | 0.2% | 0.02 | 0.2% | 0.11 | 1.0% | 0.12 | 1.0% |
| Control 1, 5.2% | 0.02 | 0.5% | 0.02 | 0.3% | 0.01 | 0.2% | 0.02 | 0.4% | 0.04 | 0.8% |
| Control 2, 9.1% | 0.03 | 0.4% | 0.03 | 0.3% | 0.03 | 0.3% | 0.05 | 0.5% | 0.07 | 0.8% |
| Control 3, 13.4% | 0.03 | 0.2% | 0.04 | 0.3% | 0.04 | 0.3% | 0.08 | 0.6% | 0.10 | 0.7% |
{9}
Whole Blood Application: ADAMS HA-8180V analyzers combined (NSGP)
| Mean % HbA1c | Repeatability | | Between Run | | Between Day | | Between Lot | | Between Instrument | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Pt 1, 5.2% | 0.02 | 0.3% | 0.02 | 0.4% | 0.01 | 0.3% | 0.01 | 0.1% | 0.00 | 0.0% | 0.03 | 0.6% |
| Pt 2, 6.5% | 0.02 | 0.4% | 0.02 | 0.3% | 0.02 | 0.3% | 0.04 | 0.7% | 0.02 | 0.2% | 0.06 | 0.9% |
| Pt 3, 7.9% | 0.02 | 0.2% | 0.02 | 0.2% | 0.02 | 0.2% | 0.05 | 0.7% | 0.01 | 0.1% | 0.06 | 0.8% |
| Pt 4, 11.9% | 0.03 | 0.3% | 0.03 | 0.2% | 0.03 | 0.3% | 0.09 | 0.7% | 0.03 | 0.3% | 0.11 | 0.9% |
| Cntrl 1, 5.2% | 0.02 | 0.5% | 0.05 | 1.1% | 0.05 | 1.0% | 0.04 | 0.7% | 0.01 | 0.3% | 0.09 | 1.7% |
| Cntrl 2, 9.1% | 0.03 | 0.3% | 0.04 | 0.4% | 0.04 | 0.5% | 0.03 | 0.3% | 0.02 | 0.2% | 0.07 | 0.8% |
| Cntrl 3, 13.6% | 0.03 | 0.2% | 0.05 | 0.4% | 0.05 | 0.4% | 0.05 | 0.4% | 0.03 | 0.3% | 0.10 | 0.8% |
Whole Blood Application: ADAMS HA-8180V analyzer # 1 (IFCC)
| Mean mmol/mol | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient 1, 33 | 0.2 | 0.7% | 0.2 | 0.7% | 0.0 | 0.0% | 0.3 | 0.9% | 0.4 | 1.4% |
| Patient 2, 47 | 0.3 | 0.6% | 0.2 | 0.5% | 0.1 | 0.2% | 0.4 | 0.9% | 0.5 | 1.2% |
| Patient 3, 62 | 0.3 | 0.5% | 0.1 | 0.2% | 0.1 | 0.1% | 0.6 | 1.0% | 0.7 | 1.1% |
| Patient 4, 105 | 0.3 | 0.3% | 0.2 | 0.2% | 0.2 | 0.2% | 0.7 | 0.7% | 0.8 | 0.8% |
| Control 1, 33 | 0.2 | 0.7% | 0.2 | 0.7% | 0.1 | 0.2% | 0.2 | 0.5% | 0.4 | 1.1% |
| Control 2, 75 | 0.3 | 0.4% | 0.3 | 0.3% | 0.2 | 0.2% | 0.0 | 0.0% | 0.4 | 0.6% |
| Control 3, 123 | 0.2 | 0.2% | 0.4 | 0.3% | 0.2 | 0.2% | 0.2 | 0.2% | 0.6 | 0.5% |
{10}
Whole Blood Application: ADAMS HA-8180V analyzer # 2 (IFCC)
| Mean mmol/mol | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient 1, 33 | 0.2 | 0.7% | 0.1 | 0.3% | 0.0 | 0.0% | 0.5 | 1.4% | 0.5 | 1.6% |
| Patient 2, 47 | 0.3 | 0.7% | 0.0 | 0.0% | 0.2 | 0.5% | 0.5 | 1.1% | 0.7 | 1.4% |
| Patient 3, 63 | 0.4 | 0.6% | 0.2 | 0.4% | 0.2 | 0.3% | 0.8 | 1.2% | 0.9 | 1.4% |
| Patient 4, 106 | 0.5 | 0.5% | 0.3 | 0.3% | 0.2 | 0.2% | 1.1 | 1.0% | 1.2 | 1.2% |
| Control 1, 33 | 0.2 | 0.7% | 0.3 | 0.8% | 0.1 | 0.4% | 0.1 | 0.5% | 0.4 | 1.2% |
| Control 2, 75 | 0.3 | 0.3% | 0.4 | 0.5% | 0.0 | 0.0% | 0.5 | 0.7% | 0.7 | 0.9% |
| Control 3, 123 | 0.3 | 0.3% | 0.3 | 0.3% | 0.3 | 0.3% | 0.6 | 0.5% | 0.8 | 0.7% |
Whole Blood Application: ADAMS HA-8180V analyzer # 3 (IFCC)
| Mean mmol/mol | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient 1, 33 | 0.2 | 0.6% | 0.2 | 0.5% | 0.0 | 0.0% | 0.5 | 1.6% | 0.6 | 1.8% |
| Patient 2, 47 | 0.3 | 0.7% | 0.2 | 0.3% | 0.2 | 0.4% | 0.3 | 0.7% | 0.5 | 1.1% |
| Patient 3, 63 | 0.3 | 0.5% | 0.2 | 0.4% | 0.3 | 0.4% | 0.6 | 1.0% | 0.8 | 1.3% |
| Patient 4, 106 | 0.3 | 0.3% | 0.3 | 0.3% | 0.3 | 0.2% | 1.3 | 1.2% | 1.4 | 1.3% |
| Control 1, 33 | 0.2 | 0.6% | 0.2 | 0.5% | 0.2 | 0.6% | 0.1 | 0.4% | 0.4 | 1.1% |
| Control 2, 75 | 0.3 | 0.4% | 0.3 | 0.4% | 0.3 | 0.4% | 0.5 | 0.6% | 0.7 | 0.9% |
| Control 3, 123 | 0.3 | 0.2% | 0.4 | 0.3% | 0.4 | 0.3% | 0.9 | 0.7% | 1.1 | 0.9% |
{11}
Whole Blood Application: ADAMS HA-8180V analyzers combined (IFCC)
| Mean mmol/mol | Repeatability | | Between Run | | Between Day | | Between Lot | | Between Instrument | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Pt 1, 33 | 0.22 | 0.7% | 0.17 | 0.5% | 0.06 | 0.2% | 0.44 | 1.3% | 0.07 | 0.2% | 0.53 | 1.6% |
| Pt 2, 47 | 0.31 | 0.7% | 0.14 | 0.3% | 0.27 | 0.6% | 0.35 | 0.7% | 0.27 | 0.6% | 0.61 | 1.3% |
| Pt 3, 63 | 0.32 | 0.5% | 0.20 | 0.3% | 0.24 | 0.4% | 0.64 | 1.0% | 0.21 | 0.3% | 0.81 | 1.3% |
| Pt 4, 106 | 0.39 | 0.4% | 0.29 | 0.3% | 0.33 | 0.3% | 1.00 | 0.9% | 0.31 | 0.3% | 1.20 | 1.1% |
| Cntrl 1, 33 | 0.22 | 0.7% | 0.22 | 0.7% | 0.14 | 0.4% | 0.14 | 0.4% | 0.13 | 0.4% | 0.39 | 1.2% |
| Cntrl 2, 75 | 0.28 | 0.4% | 0.31 | 0.4% | 0.27 | 0.4% | 0.32 | 0.4% | 0.23 | 0.3% | 0.63 | 0.8% |
| Cntrl 3, 123 | 0.28 | 0.2% | 0.40 | 0.3% | 0.43 | 0.4% | 0.53 | 0.4% | 0.34 | 0.3% | 0.90 | 0.7% |
Hemolysate Application: ADAMS HA-8180V analyzer # 1 (NGSP)
| Mean % HbA1c | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient 1, 5.2% | 0.02 | 0.4% | 0.03 | 0.5% | 0.01 | 0.2% | 0.00 | 0.1% | 0.04 | 0.7% |
| Patient 2, 6.4% | 0.01 | 0.2% | 0.02 | 0.3% | 0.02 | 0.3% | 0.05 | 0.8% | 0.06 | 0.9% |
| Patient 3, 7.9% | 0.03 | 0.4% | 0.02 | 0.2% | 0.02 | 0.3% | 0.04 | 0.6% | 0.06 | 0.8% |
| Patient 4, 11.9% | 0.03 | 0.3% | 0.03 | 0.3% | 0.02 | 0.2% | 0.07 | 0.6% | 0.08 | 0.7% |
| Control 1, 5.2% | 0.03 | 0.5% | 0.07 | 1.3% | 0.06 | 1.2% | 0.05 | 0.9% | 0.11 | 2.1% |
| Control 2, 9.1% | 0.03 | 0.3% | 0.02 | 0.2% | 0.04 | 0.4% | 0.01 | 0.2% | 0.05 | 0.6% |
| Control 3, 13.5% | 0.03 | 0.2% | 0.05 | 0.4% | 0.04 | 0.3% | 0.03 | 0.3% | 0.08 | 0.6% |
{12}
Hemolysate Application: ADAMS HA-8180V analyzer # 2 (NGSP)
| Mean % HbA1c | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient 1, 5.2% | 0.01 | 0.2% | 0.02 | 0.4% | 0.02 | 0.3% | 0.01 | 0.2% | 0.03 | 0.6% |
| Patient 2, 6.4% | 0.02 | 0.3% | 0.01 | 0.1% | 0.02 | 0.3% | 0.05 | 0.8% | 0.06 | 0.9% |
| Patient 3, 7.9% | 0.02 | 0.2% | 0.02 | 0.3% | 0.01 | 0.1% | 0.06 | 0.8% | 0.07 | 0.9% |
| Patient 4, 11.9% | 0.03 | 0.3% | 0.02 | 0.2% | 0.03 | 0.3% | 0.10 | 0.8% | 0.11 | 0.9% |
| Control 1, 5.2% | 0.03 | 0.5% | 0.05 | 1.0% | 0.05 | 1.0% | 0.04 | 0.8% | 0.09 | 1.7% |
| Control 2, 9.1% | 0.02 | 0.3% | 0.05 | 0.5% | 0.05 | 0.5% | 0.03 | 0.4% | 0.08 | 0.8% |
| Control 3, 13.6% | 0.03 | 0.2% | 0.05 | 0.4% | 0.05 | 0.4% | 0.06 | 0.4% | 0.10 | 0.7% |
Hemolysate Application: ADAMS HA-8180V analyzer # 3 (NGSP)
| Mean % HbA1c | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient 1, 5.2% | 0.02 | 0.3% | 0.00 | 0.0% | 0.01 | 0.3% | 0.00 | 0.1% | 0.02 | 0.4% |
| Patient 2, 6.5% | 0.03 | 0.5% | 0.02 | 0.4% | 0.00 | 0.0% | 0.03 | 0.5% | 0.05 | 0.8% |
| Patient 3, 7.9% | 0.01 | 0.1% | 0.01 | 0.1% | 0.00 | 0.0% | 0.06 | 0.7% | 0.06 | 0.8% |
| Patient 4, 11.9% | 0.03 | 0.2% | 0.03 | 0.2% | 0.02 | 0.2% | 0.10 | 0.9% | 0.11 | 0.9% |
| Control 1, 5.2% | 0.02 | 0.4% | 0.04 | 0.7% | 0.04 | 0.8% | 0.03 | 0.5% | 0.07 | 1.2% |
| Control 2, 9.1% | 0.03 | 0.3% | 0.04 | 0.4% | 0.03 | 0.4% | 0.04 | 0.5% | 0.07 | 0.8% |
| Control 3, 13.6% | 0.04 | 0.3% | 0.05 | 0.4% | 0.04 | 0.3% | 0.08 | 0.6% | 0.11 | 0.8% |
{13}
Hemolysate Application: ADAMS HA-8180V analyzers combined (NGSP)
| Mean % HbA1c | Repeatability | | Between Run | | Between Day | | Between Lot | | Between Instrument | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Pt 1, 5.2% | 0.02 | 0.3% | 0.02 | 0.4% | 0.01 | 0.3% | 0.01 | 0.1% | 0.00 | 0.0% | 0.03 | 0.6% |
| Pt 2, 6.5% | 0.02 | 0.4% | 0.02 | 0.3% | 0.02 | 0.3% | 0.04 | 0.7% | 0.02 | 0.2% | 0.06 | 0.9% |
| Pt 3, 7.9% | 0.02 | 0.2% | 0.02 | 0.2% | 0.02 | 0.2% | 0.05 | 0.7% | 0.01 | 0.1% | 0.06 | 0.8% |
| Pt 4, 11.9% | 0.03 | 0.3% | 0.03 | 0.2% | 0.03 | 0.3% | 0.09 | 0.7% | 0.03 | 0.3% | 0.11 | 0.9% |
| Cntrl 1, 5.2% | 0.02 | 0.5% | 0.05 | 1.1% | 0.05 | 1.0% | 0.04 | 0.7% | 0.01 | 0.3% | 0.09 | 1.7% |
| Cntrl 2, 9.1% | 0.03 | 0.3% | 0.04 | 0.4% | 0.04 | 0.5% | 0.03 | 0.3% | 0.02 | 0.2% | 0.07 | 0.8% |
| Cntrl 3, 13.6% | 0.03 | 0.2% | 0.05 | 0.4% | 0.05 | 0.4% | 0.05 | 0.4% | 0.03 | 0.3% | 0.10 | 0.8% |
Hemolysate Application: ADAMS HA-8180V analyzer # 1 (IFCC)
| Mean mmol/mol | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient 1, 33 | 0.3 | 0.8% | 0.3 | 0.8% | 0.2 | 0.7% | 0.4 | 1.1% | 0.6 | 1.7% |
| Patient 2, 47 | 0.3 | 0.6% | 0.2 | 0.3% | 0.2 | 0.4% | 0.2 | 0.5% | 0.4 | 0.9% |
| Patient 3, 63 | 0.3 | 0.5% | 0.3 | 0.5% | 0.2 | 0.3% | 0.5 | 0.8% | 0.7 | 1.2% |
| Patient 4, 106 | 0.3 | 0.3% | 0.3 | 0.2% | 0.2 | 0.2% | 0.7 | 0.6% | 0.8 | 0.8% |
| Control 1, 33 | 0.2 | 0.6% | 0.7 | 2.0% | 0.7 | 2.0% | 0.3 | 1.0% | 1.0 | 3.0% |
| Control 2, 76 | 0.3 | 0.4% | 0.3 | 0.4% | 0.4 | 0.5% | 0.2 | 0.2% | 0.6 | 0.8% |
| Control 3, 125 | 0.3 | 0.3% | 0.6 | 0.5% | 0.5 | 0.4% | 0.4 | 0.3% | 0.9 | 0.7% |
{14}
Hemolysate Application: ADAMS HA-8180V analyzer # 2 (IFCC)
| Mean mmol/mol | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient 1, 33 | 0.2 | 0.7% | 0.2 | 0.5% | 0.2 | 0.5% | 0.4 | 1.1% | 0.5 | 1.5% |
| Patient 2, 47 | 0.3 | 0.5% | 0.2 | 0.5% | 0.2 | 0.4% | 0.3 | 0.7% | 0.5 | 1.1% |
| Patient 3, 63 | 0.2 | 0.4% | 0.3 | 0.4% | 0.2 | 0.3% | 0.6 | 1.0% | 0.7 | 1.2% |
| Patient 4, 107 | 0.4 | 0.3% | 0.4 | 0.4% | 0.2 | 0.2% | 1.0 | 0.9% | 1.2 | 1.1% |
| Control 1, 33 | 0.2 | 0.6% | 0.6 | 1.7% | 0.7 | 2.0% | 0.2 | 0.6% | 0.9 | 2.7% |
| Control 2, 76 | 0.3 | 0.4% | 0.6 | 0.8% | 0.5 | 0.7% | 0.4 | 0.6% | 0.9 | 1.2% |
| Control 3, 125 | 0.3 | 0.3% | 0.5 | 0.4% | 0.6 | 0.5% | 0.7 | 0.5% | 1.1 | 0.9% |
Hemolysate Application: ADAMS HA-8180V analyzer # 3 (IFCC)
| Mean mmol/mol | Repeatability | | Between Run | | Between Day | | Between Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Patient 1, 33 | 0.2 | 0.7% | 0.2 | 0.7% | 0.1 | 0.4% | 0.4 | 1.2% | 0.5 | 1.6% |
| Patient 2, 47 | 0.2 | 0.4% | 0.2 | 0.3% | 0.1 | 0.3% | 0.4 | 0.8% | 0.5 | 1.0% |
| Patient 3, 63 | 0.2 | 0.3% | 0.2 | 0.3% | 0.0 | 0.0% | 0.6 | 1.0% | 0.7 | 1.0% |
| Patient 4, 107 | 0.4 | 0.4% | 0.3 | 0.3% | 0.2 | 0.2% | 1.2 | 1.1% | 1.3 | 1.2% |
| Control 1, 33 | 0.2 | 0.7% | 0.5 | 1.5% | 0.4 | 1.2% | 0.2 | 0.7% | 0.7 | 2.2% |
| Control 2, 76 | 0.3 | 0.4% | 0.4 | 0.5% | 0.4 | 0.5% | 0.5 | 0.6% | 0.8 | 1.0% |
| Control 3, 125 | 0.4 | 0.3% | 0.5 | 0.4% | 0.5 | 0.4% | 0.9 | 0.7% | 1.2 | 0.9% |
{15}
Hemolysate Application: ADAMS HA-8180V analyzers combined (IFCC)
| Mean mmol/mol | Repeatability | | Between Run | | Between Day | | Between Lot | | Between Instrument | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Pt 1, 33 | 0.25 | 0.7% | 0.22 | 0.7% | 0.18 | 0.5% | 0.38 | 1.2% | 0.06 | 0.2% | 0.54 | 1.6% |
| Pt 2, 47 | 0.24 | 0.5% | 0.19 | 0.4% | 0.21 | 0.4% | 0.29 | 0.6% | 0.13 | 0.3% | 0.49 | 1.0% |
| Pt 3, 63 | 0.27 | 0.4% | 0.26 | 0.4% | 0.18 | 0.3% | 0.57 | 0.9% | 0.15 | 0.2% | 0.72 | 1.1% |
| Pt 4, 107 | 0.36 | 0.3% | 0.32 | 0.3% | 0.30 | 0.3% | 0.96 | 0.9% | 0.30 | 0.3% | 1.15 | 1.1% |
| Cntrl 1, 33 | 0.21 | 0.6% | 0.57 | 1.7% | 0.57 | 1.7% | 0.28 | 0.9% | 0.00 | 0.0% | 0.88 | 2.7% |
| Cntrl 2, 76 | 0.28 | 0.4% | 0.44 | 0.6% | 0.44 | 0.6% | 0.36 | 0.5% | 0.23 | 0.3% | 0.81 | 1.1% |
| Cntrl 3, 125 | 0.35 | 0.3% | 0.53 | 0.4% | 0.58 | 0.5% | 0.62 | 0.5% | 0.34 | 0.3% | 1.12 | 0.9% |
# b. Linearity/assay reportable range:
The linearity of the HA-8180V system was evaluated in accordance with CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. Linearity across the reportable range was performed using K2-EDTA whole blood patient samples ranging from $3\%$ to $19\%$ (9.3-184.2 mmol/mol) HbA1c. Intermediate HbA1c concentration samples were prepared by performing equal distance dilutions of the high HbA1c sample with the low HbA1c sample. The sample concentrations tested were as follows: 3.0, 4.6, 6.2, 7.8, 9.4, 10.9, 12.5, 14.1, 15.7, 17.37, and $19.0\%$ HbA1c (9.3, 26.8, 44.3, 61.8, 79.2, 96.7, 114.2, 131.7, 149.2, 166.7, and $184.2\mathrm{mmol / mol}$ ). The samples were tested in replicates of three.
The following linear regression equations were obtained:
NGSP: $y = 0.9958x + 0.0064$ , $R^2 = 0.999$
IFCC: $y = 0.9958x - 0.0275$ , $R^2 = 0.999$
The linearity study supports the measuring range for this device of $4.0 - 15.0\%$ HbA1c (20-140 mmol/mol HbA1c)
{16}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The ADAMS HbA1c test method is traceable to the Diabetes Control and Complications Trial Reference Method through The National Glycohemoglobin Standardization Program (NGSP) and has received NGSP certification. See the NGSP website for current certification at http://www.ngsp.org.
The International Federation of Clinical Chemistry (IFCC) units of mmol/mol are calculated using the Master Equation NGSP (%) = 0.09148 x IFCC (mmol/mol) + 2.52. HbA1c results are provided to the customers using two different units: NGSP equivalent units (%) and IFCC equivalent units (mmol/mol).
CALIBRATOR 80 Value Assignment:
The value assignment for the ADAMS H1c HA-8180V Calibrators is traceable to the IFCC reference method. Using an ADAMS A1c HA-8180V instrument calibrated with standard material JCCRM-411 (provided by ReCCS), CALIBRATOR 80 Low and High are value assigned by measuring each calibrator 9 times per day over 3 days for a total of 27 values. The mean generated over the 3 days is used as the reference value for each calibrator.
The HbA1c values for the CALIBRATOR 80 are lot specific and target the following % HbA1c concentrations:
CALIBRATOR 80 Low: 5.0-6.0% HbA1c
CALIBRATOR 80 High: 10.0-11.0% HbA1c
CALIBRATOR 80 Stability:
Shelf life: The shelf-life stability study protocol and acceptance criteria were reviewed and found to be acceptable to support the sponsor's shelf-life stability claim that un-opened CALIBRATOR 80 is stable for 18 months when stored at 2-8°C.
Open vial: The open-vial stability study protocol and acceptance criteria were reviewed and found to be acceptable to support the sponsor's open-vial stability claim that upon reconstitution, CALIBRATOR 80 may be stored at 2-8°C for 8 hours. The calibrators are intended for single use only.
d. Detection limit:
The claimed measuring range, 4.0% to 15.0% HbA1c (20 to 140 mmol/mol), was validated by the linearity study.
17
{17}
# e. Analytical specificity:
# i.) Interference
The effect of various potential interfering substances on the HA-8180V HbA1c results was evaluated in accordance with CLSI EP7-A2, Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition.
with HbA1c values of $6.5\%$ Human whole blood K2-EDTA samples HbA1c and $\sim 8\%$ HbA1c were analyzed by spiking the interfering substance into each of the two whole blood samples. Samples were analyzed in replicates of ten on the HA-8180V test system and the results were compared to the reference sample (sample containing no interferent). The sponsor defines significant interference as $\geq + / - 7\%$ change in $\% \mathrm{HbA1c}$ value from the control value. The following substances showed no significant interference at the concentrations described below:
| Substance | Highest concentration tested with no significant interference |
| --- | --- |
| Albumin | 20 g/dL |
| Bilirubin, conjugated | 100 mg/dL |
| Bilirubin, unconjugated | 100 mg/dL |
| Rheumatoid Factor | 750 IU/mL |
| Triglycerides | 2,000 mg/dL |
| Acetaminophen | 20 mg/dL |
| Acetylcysteine | 330 mg/dL |
| Acetylsalicylic Acid (Aspirin) | 65 mg/dL |
| Ampicillin-Na | 1,000 mg/dL |
| Ascorbic Acid | 200 mg/dL |
| Cefoxitin-Na | 2,500 mg/dL |
| Cyclosporine | 0.67 mg/dL |
| Doxycycline | 50 mg/dL |
| Ibuprofen | 50 mg/dL |
| Levodopa | 20 mg/dL |
| Metformin | 5 mg/dL |
| Methyldopa | 30 mg/dL |
| Metronidazole | 200 mg/dL |
| Rifampicin | 6.4 mg/dL |
| Salicylic Acid | 60 mg/dL |
| Theophylline | 10 mg/dL |
# ii.) Cross Reactivity:
Potential interferences from acetylated hemoglobin (Hb), carbamylated Hb, and labile HbA1c were evaluated using Human whole blood K2- EDTA samples with HbA1c values of $6.5\%$ ,and $8.0\%$ . Aliquots of the $6.5\%$ and $8.0\%$ HbA1c samples were spiked with 12.5, 25, 37.5, and $50~\mathrm{mg / dL}$ acetaldehyde solution evaluate the effects of
{18}
acetylated hemoglobin; 6.25, 12.5, 18.75, and $25\mathrm{mg / dL}$ of sodium cyanate solution was added to the A1c samples to evaluate the effects of carbamylated hemoglobin; and 500, 1000, 1500, and $2000\mathrm{mg / dL}$ glucose solution were added to the A1c samples to evaluate the effects of Labile A1c. The samples were allowed to incubate for two hours at $37^{\circ}\mathrm{C}$ prior to testing. The samples were run in replicates of ten on one HA-8180V analyzer and the results were compared to the reference sample (sample containing no added solutions). Significant interference was defined as $\%$ recovery $\geq + / - 7\%$ of the expected $100\%$ recovery.
The results were concluded as follows:
- Acetylated Hb (up to $50\mathrm{mg / dL}$ ) does not interfere with this assay.
- Carbamylated Hb (up to $25\mathrm{mg / dL}$ ) does not interfere with this assay.
- Labile HbA1c (up to $2000\mathrm{mg / dL}$ ) does not interfere with this assay.
# iii.) Hemoglobin Variant Interference:
The effect of various hemoglobin variants on the HA-8180V results was evaluated to determine if the variants interfere with HbA1c results. To assess the effects of hemoglobin variant interference, a total of 165 K2-EDTA whole blood samples known to contain Hemoglobin variants $\mathrm{A}_2$ , C, D, E, F and S were tested on the HA-8180V and results compared to a NGSP boronate affinity HPLC reference method that has been demonstrated to be free from interference by the hemoglobin variants (Trinity Biotech Ultra2 A1c for HbC, HbD, HbE, and Tosoh G8 HPLC for HbF and HbA2). A minimum of 10 samples were tested in duplicate at each of the two HbA1c concentrations of $\sim 6.5\%$ and $\sim 8.0\%$ containing the Hemoglobin variants A2, C, D, E, F, and S. The table below shows the number and concentrations of the variants and HbA1c concentrations that were tested.
| Hemoglobin Variant | Number of Samples | % Concentration of Variant in Sample | %HbA1c Concentration Range |
| --- | --- | --- | --- |
| HbA2 | 30 | 1.9-25.4% | 4.8- 8.9% |
| HbC | 26 | 26.9-39.0% | 4.6-9.2% |
| HbD | 22 | 31.6-36% | 5.7-10.3% |
| HbE | 22 | 18.0-30.0 % | 4.9-9.7% |
| HbF | 24 | 1.0-30.3% | 4.7-11.7% |
| HbS | 41 | 12.9-42.1% | 4.5-11.6% |
The results obtained by the HA-8180V were compared to the comparative method and the percent bias was calculated. The sponsor defines acceptable bias as less than or equal to $7\%$ . The result summary is presented below:
{19}
Hemoglobin Variant Results Summary
| Hemoglobin Variant | %Bias (Range of % Bias) for ~6.5% HbA1c | % Bias (95% CI) for ~8.0% HbA1c |
| --- | --- | --- |
| HbA2 | -1.6% (-6.7% to -1.5%) | 1.2% (-1.3%, 3.4%) |
| HbC | 0.1% (-5.6% to 4.6%) | -0.5% (-4.7% to 5.7%) |
| HbD | -1.7% (-6.8% to 1.7%) | -2.6% (-5.7% to 1.2%) |
| HbE | -0.1% (-3.0% to 4.8%) | -1.1% (-2.8% to 3.5%) |
| HbF | -0.3% (-6.5% to4.0%) | 0.5% (-2.7% to 6.9%) |
| HbS | 0.1% (-8.8% to 5.7%) | -0.4% (-4.7 to% 5.5%) |
The results show there is no significant interference for HbA2 concentrations of less than or equal to $16\%$ , HbC concentrations of less than or equal to $39\%$ , HbD concentrations of less than or equal to $36\%$ , HbE concentrations of less than or equal to $30\%$ , HbF concentrations of less than or equal to $30\%$ , and HbS concentrations of less than or equal to $40\%$ .
The labeling contains the above results summary in the limitations section and the summary and tables are included in the performance characteristic section of the labeling.
f. Assay cut-off:
Not applicable
# 2. Comparison studies:
a. Method comparison with predicate device:
A method comparison study was performed in which $\% \mathrm{HBA1c}$ was measured in 143 variant-free venous whole blood K2-EDTA samples and hemolysate samples on the candidate HA-8180V device. The candidate device HbA1c results were compared to those obtained from testing performed at secondary NGSP reference laboratory on the Tosoh G8 HPLC method. The distribution of samples spanned the measuring interval (with a higher percentage of samples around the clinical decision points) for both whole blood and hemolysate methods as follows:
| Hemoglobin A1c level | Number of samples | %samples tested |
| --- | --- | --- |
| ≤ 4.9% | 5 | 3% |
| 5 – 5.9% | 19 | 13% |
| 6 – 6.4% | 40 | 28% |
| 6.5 – 6.9% | 40 | 25% |
| 7 – 7.9% | 21 | 15% |
| 8 – 8.9% | 10 | 7% |
| > 9% | 12 | 89% |
| Total samples | 143 | 100% |
{20}
Deming (weighted) regression analysis was performed for the HA-8180V system versus the NGSP SRL reference method. A summary of the results for both the whole blood and the hemolysate methods are provided in the tables below.
| Whole Blood | Slope (95% CI) | Intercept (95% CI) | R² |
| --- | --- | --- | --- |
| Weighted Deming | 0.9864 (0.9626 to 1.010) | 0.09585 (-0.06895 to 0.2670) | 0.998 |
| Hemolysate | Slope (95% CI) | Intercept (95% CI) | R² |
| Weighted Deming | 0.9906 (0.9670-1.014) | 0.0847 (-0.08251 to 0.24347) | 0.998 |
Scatter plot using weighted Deming Fit, %HbA1c, Reference Method vs ADAMS A1c HA-8180V (Whole Blood Samples)

{21}

Scatter plot using weighted Deming Fit, %HbA1c, Reference Method vs ADAMS A1c HA-8180V (Hemolysate Blood Samples)
The following biases between the candidate HA-8180V system versus the NGSP Reference Method (Tosoh G8 HPLC analyzer) were observed:
Bias Estimation (whole blood)
| % HbA1c Level | Bias | %Bias |
| --- | --- | --- |
| 5.0 | 0.028 | 0.56% |
| 6.5 | 0.007 | 0.11% |
| 8.0 | -0.013 | -0.16% |
| 12.0 | -0.067 | -0.56% |
{22}
Bias Estimation (hemolysate)
| % HbA1c Level | Bias | %Bias |
| --- | --- | --- |
| 5.0 | 0.033 | 0.67% |
| 6.5 | 0.019 | 0.30% |
| 8.0 | 0.005 | 0.07% |
| 12.0 | -0.032 | -0.27% |
# Total Error Near Cutoff
Using the results of bias estimation (%Bias) in the method comparison study and precision estimated in the reproducibility study, Total Error (TE) at four concentrations: $(5.0\%, 6.5\%, 8.0\%$ and $12\%)$ was calculated as follows: $\% \mathrm{TE} = |\% \mathrm{Bias}| + 1.96 \times \% \mathrm{CV} \times (1 + (\% \mathrm{Bias} / 100)$ . The results are presented in the tables below:
Whole Blood
| Decision Level | %Bias | %CV | %TE |
| --- | --- | --- | --- |
| 5.0 | 0.56 | 0.5 | 1.5 |
| 6.5 | 0.11 | 0.9 | 1.9 |
| 8.0 | -0.16 | 1.0 | 2.1 |
| 12.0 | -0.56 | 1.0 | 2.4 |
Hemolysate
| Decision Level | %Bias | %CV | %TE |
| --- | --- | --- | --- |
| 5.0 | 0.67 | 0.6 | 1.8 |
| 6.5 | 0.31 | 0.9 | 2.1 |
| 8.0 | 0.07 | 0.8 | 1.6 |
| 12.0 | -0.27 | 0.9 | 2.0 |
# b. Matrix comparison:
A matrix comparison study was performed to evaluate the effects of the different anticoagulants, K2-EDTA and K3-EDTA, on $\% \mathrm{HbA1c}$ measurement by the ADAMS A1c HA-8180V system. Venous whole blood specimens with concentrations spanning 4.7 to $15.7\%$ HbA1c were collected in K2-EDTA (reference anticoagulant) and K3-EDTA blood tubes from a total of 40 donors. The samples were measured on the candidate device and the results were compared. The following regression equation was obtained :
$$
y = 1. 0 0 5 x - 0. 0 4 7, R ^ {2} = 1. 0 0 0
$$
The results of the study support the sponsor's claim that blood samples collected in blood collection tubes containing K2-EDTA and K3-EDTA anti-coagulants are
{23}
acceptable for use with this device.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Hemoglobin A1c expected values reference range cited below is from the American Diabetes Association (ADA), "Diagnosis and Classification of Diabetes Mellitus" Diabetes Care; 39 (Supplement 1): S13-S22; 2016.
| Suggested Diagnosis | HbA1c (%) | HbA1c (mmol/mol) |
| --- | --- | --- |
| Diabetic | ≥ 6.5 | ≥ 48 |
| Prediabetes | 5.7 – 6.4 | 39 – 47 |
| Normal | < 5.7 | < 39 |
N. Instrument Name:
ADAMS H1c HA-8180V analyzer
O. System Descriptions:
1. Modes of Operation:
Continuous testing mode
2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
{24}
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
Manual sample identification or internal barcode reader
4. Specimen Sampling and Handling:
Hemoglobin test specimens are prepared by automated dilution (1:101) of anticoagulated venous whole blood in hemolysis washing solution 80H. Alternately, hemolysates may be prepared "off-line" by manual dilution of anticoagulated venous whole blood with the diluent 80 provided in control dilution set 80. The samples are placed in plastic racks and placed on the analyzer.
5. Calibration:
Automatic calibration: The analyzer measures two HbA1c standard solutions (CALIBRATOR 80 Low and High solutions), and uses the results to automatically determine the calibration coefficients. Standard values of the standard solutions can be set by entering the numbers using numeric buttons, or reading the calibration information barcodes during measurement when using the internal barcode reader.
6. Quality Control:
Recommendations for quality control are described in the labeling. Additionally, users should follow local, state and federal regulations.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
25
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.