K162811 · Essex Industries, Inc. · CAN · Mar 2, 2017 · Anesthesiology
Device Facts
Record ID
K162811
Device Name
MR Conditional CGA 870
Applicant
Essex Industries, Inc.
Product Code
CAN · Anesthesiology
Decision Date
Mar 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2700
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Essex MR Conditional CGA 870 pressure regulators are used with a portable oxygen delivery system intended to provide supplemental oxygen to adults in hospital, sub-acute care, and pre-hospital / ground transport settings. It is offered in models that are MR-conditional (per ASTM F2052-15), and may be used during MR imaging for static magnetic fields of 3.0 T or less.
Device Story
The MR Conditional CGA 870 is a mechanical oxygen pressure regulator designed for attachment to medical CGA 870 post-valve cylinders. It regulates high-pressure oxygen (500–2,000 psig) down to a nominal 50 psig for delivery via an integrated flow selector valve (0–25 L/min). Operated by clinicians in hospitals, outpatient centers, and pre-hospital transport, the device allows for supplemental oxygen delivery during MRI procedures due to its MR-conditional material composition. It functions as a single-stage piston regulator. The device provides a visual contents gauge for monitoring. By enabling oxygen therapy within the MRI suite, it ensures patient safety and continuity of care during imaging. It does not require external power or software.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including performance testing per CGA E-4, CGA E-7, ISO 10524, and EN 738-1, as well as MR-conditional testing per ASTM F2052-15 and biocompatibility evaluations (VOC and PM2.5).
Technological Characteristics
Single-stage piston pressure regulator; yoke-style inlet (CGA 870); flow selector (0-25 L/min); output pressure ~50 psig. Materials meet ASTM F2052-15 for MR-conditional use (3.0 T or less). Standards: CGA E-4, CGA E-7, ISO 10524-1, EN 738-1. Mechanical operation; no energy source or software.
Indications for Use
Indicated for adult patients requiring supplemental oxygen in hospital, sub-acute care, and pre-hospital/ground transport settings, including during MR imaging (up to 3.0 T).
Regulatory Classification
Identification
A pressure regulator is a device, often called a pressure-reducing valve, that is intended for medical purposes and that is used to convert a medical gas pressure from a high variable pressure to a lower, more constant working pressure. This device includes mechanical oxygen regulators.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized symbol featuring three human profiles facing to the right, layered on top of each other to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 2, 2017
Essex Industries, Inc. % Paul Dryden Consultant 7700 Gravois St. Louis, Missouri 63123
Re: K162811
Trade/Device Name: MR Conditional CGA 870 Regulation Number: 21 CFR 868.2700 Regulation Name: Pressure Regulator Regulatory Class: Class I Product Code: CAN Dated: January 27, 2017 Received: January 31, 2017
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Lori A. Wiggins -S
for
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162811
Device Name MR Conditional CGA 870
#### Indications for Use (Describe)
The Essex MR Conditional CGA 870 pressure regulators are used with a portable oxygen delivery system intended to provide supplemental oxygen to adults in hospital / ground transport settings. It is offered in models that are MR-conditional (per ASTM F2052-15), and may be used during MR imaging for static magnetic fields of 3.0 T or less.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CER 801 Subpart D) |
|------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary Page 1 of 6 3/2/2017
Essex Industries, Inc. 7700 Gravois St. Louis, MO 63123
| Official Contact: | Russ Jacobsmeyer - Vice President of Engineering<br>Tel – 314.338.8723 |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Proprietary or Trade Name: | Essex MR Conditional CGA 870 |
| Common/Usual Name: | Cylinder, Compressed Gas, and Valve |
| Classification Name/Code: | CAN - Cylinder, Compressed Gas, and Valve<br>Class I, 21 CFR 868.2700 |
| Predicate Device:<br>Reference Device: | Essex CGA 870 oxygen pressure regulator (Class 1, exempt)<br>K101792 – Linde – LIV Portable Oxygen System |
#### Device Description:
The Essex MR Conditional CGA 870 pressure regulator is designed to be installed on a medical CGA 870 post valve cylinder, regulate high pressure oxygen from 500 to 2,000 psig nominal, deliver a specific amount of oxygen to an attached flow selector, and be ignition resistant.
The device consists of:
- A yoke style inlet fitting per CGA 870. .
- The regulator can be installed on any cylinder with a CGA 870 connection. O
- A pressure regulator section is to reduce the pressure from 500-2,000 psig to 50 psig nominal. ●
- A flow selector valve to control the flow at the regulated pressure between 0 and 25 L/min. ●
- Made of materials which meet the MR Conditional requirements of ASTM F2052-15.
This is identical to our Class I, exempt model CGA 870 except for different materials to allow for the device to meet the ASTM F2052-15 - Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment.
#### Indications for Use:
The Essex MR Conditional CGA 870 pressure regulators are used with a portable oxygen delivery system intended to provide supplemental oxygen to adults in hospital, sub-acute care, and pre-hospital / ground transport settings. It is offered in models that are MR-conditional (per ASTM F2052-15), and may be used during MR imaging for static magnetic fields of 3.0 T or less.
Environments of Use: Their environments of use include all healthcare settings including:
- Hospital, including intra-hospital transport ●
- Outpatient
- Imaging center (MRI Suite) ●
- Pre-hospital ●
#### Contraindications: None.
#### Summary of substantial equivalence
{4}------------------------------------------------
#### 510(k) Summary Page 2 of 6 3/2/2017
The proposed device has been compared to the predicate Essex CGA 870 oxygen pressure regulator (Class 1, exempt) for the technical characteristics, see Table 1 and the reference device for environment of use, MR Conditional, and see Table 2.
# Device Comparison
| Description | Predicate | Subject Device |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | CGA 870 Device, Class I, exempt | MR Conditional CGA 870 |
| Device Name | CGA 870 | MR Conditional CGA 870 |
| Device Classification<br>& Product code | Class I / CAN | Class I / CAN |
| Prescription Device | Yes | Yes |
| Indications for Use | The Essex CGA 870 pressure regulator<br>is a portable oxygen delivery system<br>intended to provide supplemental<br>oxygen too pediatrics and adults in<br>hospital, sub-acute care, and pre-hospital<br>/ ground transport settings. | The Essex MR Conditional CGA 870<br>pressure regulators are used with a<br>portable oxygen delivery system<br>intended to provide supplemental<br>oxygen to adults in hospital, sub-acute<br>care, and pre-hospital / ground<br>transport settings. It is offered in<br>models that are MR-conditional (per<br>ASTM F2052-15), and may be used<br>during MR imaging for static magnetic<br>fields of 3.0 T or less. |
| Flow selector and<br>Flow Outlet | Yes | Yes |
| Fixed pressure<br>outlet | Yes ~ 50 psig | Yes ~ 50 psig |
| Cylinder On/Off | Yes | Yes |
| Filling Port | No<br>When cylinder is empty the device is<br>attached to a new cylinder | No<br>When cylinder is empty the device is<br>attached to a new cylinder |
| Contents Gauge | Yes / active | Yes / active |
| Excess Flow | No | No |
| Burst Disk | Burst Disk is located on the cylinder<br>post valve. | Burst Disk is located on the cylinder<br>post valve. |
| Single stage piston<br>type | Yes | Yes |
| MR Conditional | N/A | 3.0 Tesla or less |
| Pressure Inlet | 2000 psig | 2000 psig |
| Flow rates | 0-25 L/min | 0-25 L/min |
| Operating Pressure<br>range | 500-2000 psig | 500-2000 psig |
| Operating<br>Temperature Range | -20°F to 130°F | -20°F to 130°F |
#### Table 1 - Comparison of Predicate Essex CGA 870 to the Subject Device -MR Conditional CGA 870
{5}------------------------------------------------
# 510(k) Summary Page 3 of 6 3/2/2017
| Description | Predicate | Subject Device |
|------------------|-----------------------------------|-----------------------------------|
| Standards | CGA 870 Device, Class I, exempt | MR Conditional CGA 870 |
| | ISO 10524-1 | ISO 10524-1 |
| | CGA E-4 | CGA E-4 |
| | CGA E-7 | CGA E-7 |
| | EN 738-1 | EN 738-1 |
| | | ASTM F2052-15 |
| Biocompatibility | Externally communicating, Tissue, | Externally communicating, Tissue, |
| | Permanent duration | Permanent duration |
| | VOC | VOC |
| | PM2.5 | PM2.5 |
# Table 2 - Comparison of Reference Linde LIV (K101792) to the Subject Device -MR Conditional CG 870
| Description | Reference Device | Subject Device |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Linde LIV - K101792 | MR Conditional CGA 870 |
| Device Name | LIV | MR Conditional CGA 870 |
| Device Classification<br>& Product code | Unclassified/ECX | Class I / CAN |
| Prescription Device | Yes | Yes |
| Indications for Use | The LIV is an integrated portable oxygen<br>delivery system intended to provide<br>supplemental oxygen to pediatrics and<br>adults. The device is MR-conditional (per<br>ASTM F2503-05*), and intended for use<br>during MR imaging for MRI systems up to<br>3.0T.<br><br>Compressed gas cylinders in service or in<br>storage shall be stabilized or otherwise<br>secured to prevent falling and rolling. | The Essex MR Conditional CGA 870<br>pressure regulators are used with a<br>portable oxygen delivery system intended<br>to provide supplemental oxygen to adults<br>in hospital, sub-acute care, and pre-<br>hospital / ground transport settings. It is<br>offered in models that are MR-conditional<br>(per ASTM F2052-15), and may be used<br>during MR imaging for static magnetic<br>fields of 3.0 T or less. |
| Flow selector and<br>Flow Outlet | Yes | Yes |
| Fixed pressure<br>outlet | Yes ~ 50 psig | Yes ~ 50 psig |
| Cylinder On/Off | Yes | Yes |
| Filling Port | Yes | When cylinder is empty the device is<br>attached to a new cylinder |
| Contents Gauge | Yes / active | Yes / active |
| Excess Flow | No | No |
| Burst Disk | Yes integral to the system | Burst Disk is located on the cylinder post<br>valve. |
| Single stage piston<br>type | Yes | Yes |
| Hand grip | Yes | No |
| Access ports | Yes | No |
| MR Conditional | 3.0 Tesla | 3.0 Tesla or less |
| Pressure Inlet | 2000 psig | 2000 psig |
{6}------------------------------------------------
#### 510(k) Summary Page 4 of 6 3/2/2017
| Description | Reference Device | Subject Device |
|-----------------------------|---------------------------------------------------------------------|-------------------------------------------------------------------------|
| | Linde LIV - K101792 | MR Conditional CGA 870 |
| DISS 1240 fittings | Yes | No<br>Not applicable |
| Flow rates | 0-25 L/min | 0-25 L/min |
| Operating Pressure range | Up to 3000 psig | 500-2000 psig |
| Operating Temperature Range | | -20°F to 130°F |
| | ISO 10524-3<br>ASTM G173-03<br>CGA E-7<br>CGA 540<br>ASTM F2503-05* | ISO 10524-1<br>CGA E-4<br>CGA E-7<br>EN 738-1<br>ASTM F2052-15* |
| Standards | | |
| Biocompatibility | Externally communicating, Tissue,<br>Permanent duration | Externally communicating, Tissue,<br>Permanent duration<br>VOC<br>PM2.5 |
*Note ASTM F2503-05 is a device marking and labeling standard whereas ASTM F2052-15 is the Testing and Performance standard.
The proposed MR Conditional CGA 870 is viewed as substantially equivalent to the predicate device for the performance, intended use and technical characteristics and the reference device for the addition of the MR conditional environment of use. In the tables above, we have outlined the similarities and differences, if any, between the subject device, the predicate and reference devices.
#### Indications -
- The MR Conditional CGA 870 is a standalone oxygen pressure regulator intended to provide supplemental oxygen.
- Discussion - The indications for the predicate and reference are similar with the predicate. One can find the indications for use to be similar and the differences do not raise any concerns of safety or effectiveness compared to the predicate or reference devices.
#### Patient Population -
- It is intended for adult patients requiring supplemental oxygen ●
- Discussion The patient population is similar to the predicate and reference device. There are ● no differences that raise any concerns of safety or effectiveness compared to the predicate or reference devices.
#### Environment of Use -
- For use in all healthcare settings hospitals, ambulatory, imaging centers, pre-hospital. ●
- Discussion The environments of use are similar to the predicate and reference devices. The . addition of MR Conditional use has been demonstrated to meet the standard which would not raise any new concerns of safety and effectiveness.
#### Technology -
- The principle of pressure regulating is based upon a single stage piston style with adjustable flow rates, flow gauge, burst disk and use of a separate cylinder.
{7}------------------------------------------------
#### 510(k) Summary Page 5 of 6 3/2/2017
- . Discussion - This technology is identical to the predicate device and similar to the reference.
#### Differences
The only difference between the reference and the subject device is the change in materials to be MR Conditional. The difference between the reference and the subject device is that the predicate is an integrated regulator and the subject device is not. Discussion - These differences do not raise any new concerns of safety or effectiveness.
#### Non-clinical Testing Summary -
We have performed a number of tests appropriate for the proposed device. These tests include:
- CGA E-4: Standard for Gas Pressure Regulators ●
- o par. 4.1 Operating Temperature Range
- par. 4.2.2.3 Resistance to Ignition (Oxygen Service) O
- par. 4.4 Gas Tightness O
- par. 4.5 Mechanical Resistance O
- par. 4.6 Endurance O
- par. 5.4 Pressure Relief Devices O
- par. 6.1 Pressure Regulation Coefficient, i O
- par. 6.2 Static Increment, SI O
- par. 6.3 Flow Regulation O
- par. 6.4 Flow Capacity O
- CGA E-7: American National Standard for Medical Gas Regulators and Flowmeters ●
- par. 5.4 Relief Valves റ
- par. 5.6.1 Minimum Burst Strength O
- par. 5.6.2 Leakage o
- o par. 5.7 Temperature for Storage and Operation
- ISO 10524: Pressure Regulators and Pressure Regulators with Flow-Metering Devices for Medical Gas Systems
- o par. 7.3 Pressure relief valve
- par. 7.5 Resistance to ignition O
- par. 7.8 Gas tightness O
- par. 7.9.3 Mechanical resistance (high pressure section burst) O
- par. 7.9.4 Mechanical resistance (low pressure section burst) O
- par. 8.0 Environmental temperatures O
- EN 738-1: Pressure Regulators for use with Medical Gases ●
- par. 5.4.2.7 Performance, functional, and flow characteristics o
- O par. 5.4.2.8 Relief valve
- par. 5.4.2.9 Leakage o
- par. 5.4.2.10 Mechanical Strength O
- par. 5.4.2.11 Resistance to ignition O
- MR Conditional Testing per ASTM F2052-15 (Note testing was done according to ASTM ● F2052-06 but there are no testing differences between the 2006 version and the 2015 version.
- Biocompatibility (VOC and PM2.5) ●
#### Bench Testing -
- The performance testing is based upon standards for material, construction, ignition, MRI ● compatibility, and accuracy.
{8}------------------------------------------------
### 510(k) Summary Page 6 of 6 3/2/2017
- . Discussion - Based upon the testing and comparison to the standards, we can find the subject device and the predicate and reference devices to be substantially equivalent.
# Biocompatibility of Materials -
- ISO 10993-1 would consider the materials in the gas pathway as - Externally communicating, Tissue / Dentin / Bone contact, Permanent Duration.
- The materials were evaluated per CGA E-4 for ignition and for biocompatibility via VOC and . PM23.5 The testing demonstrated that the materials were biocompatible for their intended use.
- Discussion – The materials were found to be compatible for their intended use and identical to the reference.
#### Animal Testing:
There was no animal testing.
#### Clinical Testing:
There was no clinical testing.
#### Substantial Equivalence Conclusion:
The sponsor has demonstrated through performance testing and non-clinical testing that the proposed device is substantially equivalent to the predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.