The BLUEPRINT™ Glenoid Guides are intended to be used as surgical instruments to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed shoulder arthroplasty procedures using anatomic landmarks that are identifiable on patient-specific preoperative CT scans. The BLUEPRINT 3D Planning Software is intended to be used as a medical software to assist in pre-surgical planning for shoulder orthopedic surgery.
Device Story
System comprises patient-specific glenoid drill guides and 3D planning software. Software imports patient CT scans (DICOM) to allow surgeons to visualize, measure, reconstruct, and annotate anatomic data; position/select glenoid and humeral implants; display bone density/reaming surfaces; and simulate prosthetic range of motion. Software generates a surgical report and 3D file for guide fabrication. Guides are patient-specific, single-use, non-sterile, and manufactured from polyamide or titanium. Used in clinical settings by orthopedic surgeons for preoperative planning and intraoperative guidance. Guides assist in positioning glenoid components by providing a drill path for K-wires based on preoperative plans. Benefits include improved surgical planning and accurate intraoperative component placement.
Clinical Evidence
Bench testing only. Validation of software and hardware design, manufacturing, biocompatibility, sterility, and dimensional accuracy was previously established in predicate submissions (K143374, K160555). No clinical data presented.
Technological Characteristics
Materials: Medical grade polyamide (PA 2200) or titanium (Ti6Al4V). Software: 3D planning application for Windows/Mac OS. Connectivity: DICOM standard for CT image import; Online Management System (OMS) integration. Sterilization: Delivered non-sterile. Function: Patient-specific drill guides for K-wire placement; software-based visualization, measurement, and 3D modeling.
Indications for Use
Indicated for adult patients undergoing total anatomic or reversed shoulder arthroplasty. Used by surgeons to assist in preoperative planning and intraoperative positioning of glenoid components based on patient-specific CT scans.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
BLUEPRINT™ Patient Specific Instrumentation (K160555)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, suggesting a focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 22, 2017
TORNIER SAS Aymen Azaiez Regulatory Affairs Specialist 161 rue Lavoisier 38330 Montbonnot Saint Martin France
Re: K162800
Trade/Device Name: BLUEPRINT Patient Specific Instrumentation Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: KWS Dated: September 30, 2016 Received: October 5, 2016
Dear Aymen Azaiez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K162800
Device Name
BLUEPRINT Patient Specific Instrumentation
Indications for Use (Describe)
Hardware
The BLUEPRINT™ Glenoid Guides are patient-specific drill guides. They have been specially designed to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed shoulder arthroplasty procedures using anatomic landmarks that are identifiable on patient-specific preoperative CT scans.
Software
BLUEPRINT™ 3D Planning Software is a medical device for surgeon composed of one software component. It is intended to be used as a pre-surgical planner for shoulder orthopedic surgery.
BLUEPRINT™ 3D Planning Software runs on standard personal and business computers running Microsoft Windows or Mac OS operating systems.
The software supports DICOM standard to import the CT scan (Computed Tomography) images of the patient. Only CT scan modality can be loaded with BLUEPRINT™ 3D Planning Software.
BLUEPRINT™ 3D Planning Software allows surgeon to visualize, measure, reconstruct, annotate and edit anatomic data. It allows surgeon to design glenoid patient-specific guides based on the pre-surgical plan.
The software leads to the generation of a surgery report along with a 3D file of the glenoid patient-specific guide.
BLUEPRINT™ 3D Planning Software does not include any system to manufacture the glenoid patient-specific guide.
BLUEPRINT™ 3D Planning Software is to be used for adult patients only and should not be used for diagnostic purpose.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Tornier Implants Chirurgicaux. The word "TORNIER" is in large, bold, blue font. Below that, the words "Implants Chirurgicaux" are in a smaller, bold, blue font.
### 510k Summary
Traditional 510(k) Premarket – BLUEPRINT™ Patient Specific Instrumentation
Regulatory authority: Safe Medical Devices Act of 1990, 21 CRF 807.92
#### 1) Device name
| Trade name: | BLUEPRINT™ Patient Specific Instrumentation |
|----------------------|--------------------------------------------------------------------------------|
| Common name: | Patient Specific Instrumentation + 3D Planning Software |
| Classification name: | Prosthesis, Shoulder, Semi-constrained, Metal/Polymer Cemented<br>(\$888.3660) |
#### 2) Submitter
TORNIER SAS 161 rue Lavoisier 38330 Montbonnot Saint Martin- France Registration Number: 3000931034
#### 3) Company contact
TORNIER SAS Mr Aymen AZAIEZ Regulatory Affairs Specialist 161 rue Lavoisier 38334 Montbonnot Tel: 00 33 4 76 61 35 00 Fax: 00 33 4 76 61 35 59 Email: aymen.azaiez(@wright.com
#### 4) Classification
Device class: Class II Classification panel: Orthopedic Product code: KWS
#### 5) Primary Predicate Device and Reference devices
| Trade name | 510(k) Number | Decision date | Applicant |
|-------------------------------------------------------------|---------------|--------------------|---------------------|
| BLUEPRINT™ Patient Specific<br>Instrumentation<br>(PRIMARY) | K160555 | 10th June 2016 | TORNIER SAS |
| Materialise Glenoid Positioning System<br>(Reference) | K153602 | 26th April 2016 | MATERIALISE NV |
| TraumaCad Mobile Release 2.0 (Reference) | K160001 | 12th February 2016 | VOYANT HEALTH LTD |
| DYONICS PLAN HIP IMPINGEMENT<br>PLANNING SYSTEM (Reference) | K132636 | 17th October 2016 | SMITH & NEPHEW, INC |
| OSIRIX MD (Reference) | K101342 | 20th August 2010 | PIXMEO SARL |
Image /page/4/Picture/14 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is rendered in a light blue color and appears to be constructed from geometric shapes, giving it a modern and abstract look. The hexagon provides a border around the 'T', further defining the logo's shape and structure. The overall design is simple, clean, and easily recognizable.
TORNIER S.A.S. 161, rue Lavoisier 38334 MONTBONNOT CEDEX FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33 S.A.S. au capital de 35 043 008 € SIRET : 070 501 275 000 21 R.C.S. : 070 501 275 CODE APE : 3250 A
{5}------------------------------------------------
#### 6) Device description
BLUEPRINT™ Patient Specific Instrumentation is composed of two components: BLUEPRINT™ Glenoid Guides (hardware) and BLUEPRINT™ 3D Planning Software (software).
#### Hardware
The BLUEPRINT™ Glenoid Guides are patient-specially designed to facilitate the implantation of WRIGHT-TORNIER glenoid prostheses.
The BLUEPRINT™ Glenoid Guides are designed and manufactured based on a pre-operative plan generated by the BLUEPRINTTM 3D Planning Software. All BLUEPRINT™ Glenoid Guides are patient-specific, single use and delivered non-sterile.
#### Software
BLUEPRINT™ 3D Planning Software is composed of one software component connected to an Online Management System (OMS). The software installed on a computer is intended to be used by orthopedic surgeons, as a preoperative planning software for shoulder arthroplasty surgery (anatomic and reversed).
It is intended to help plan an operation by allowing surgeons to:
- Position and select glenoid implant,
- Position and select humeral implant,
- Display bone density and reaming surface, ●
- Simulate the prosthetic range of motion, 0
- Design a patient specific guide for the glenoid component. o
#### 7) Materials
The commercially available BLUEPRINT™ Glenoid Guides are manufactured from medical grade polyamide (PA 2200) or titanium (Ti6Al4V).
#### 8) Intended Use
#### Hardware
The BLUEPRINT™ Glenoid Guides are intended to be used as surgical instruments to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed shoulder arthroplasty procedures using anatomic landmarks that are identifiable on patient-specific preoperative CT scans.
#### Software
The BLUEPRINT 3D Planning Software is intended to be used as a medical software to assist in pre-operative surgical planning for shoulder surgery.
Image /page/5/Picture/21 description: The image shows a logo with a stylized letter "T" inside a hexagon. The "T" is formed by two overlapping shapes, creating a three-dimensional effect. The color of the logo is a light shade of purple or blue. The logo appears to be simple and modern.
TORNIER S.A.S. 161. rue Lavoisier 38334 MONTBONNOT CEDEX FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33 S.A.S. au capital de 35 043 008 € SIRET : 070 501 275 000 21 R.C.S. : 070 501 275 CODE APE : 3250 A
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Tornier Implants Chirurgicaux. The word "TORNIER" is in large, bold, dark blue font. Below that, the words "Implants Chirurgicaux" are in a smaller, dark blue font.
#### 9) Indications
#### Hardware
The BLUEPRINT™ Glenoid Guides are patient-specific drill guides. They have been specially designed to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed shoulder arthroplasty procedures using anatomic landmarks that are identifiable on patient-specific preoperative CT scans.
#### Software
BLUEPRINT™ 3D Planning Software is a medical device for surgeon composed of one software component. It is intended to be used as a pre-surgical planner for shoulder orthopedic surgery.
BLUEPRINT™ 3D Planning Software runs on standard personal and business computers running Microsoft Windows or Mac OS operating systems.
The software supports DICOM standard to import the CT scan (Computed Tomography) images of the patient. Only CT scan modality can be loaded with BLUEPRINT™ 3D Planning Software.
BLUEPRINT™ 3D Planning Software allows surgeon to visualize, measure, reconstruct, annotate and edit anatomic data. It allows surgeon to design glenoid patient-specific guides based on the pre-surgical plan.
The software leads to the generation of a surgery report along with a 3D file of the glenoid patient-specific guide.
BLUEPRINT™ 3D Planning Software does not include any system to manufacture the glenoid patient-specific guide.
BLUEPRINT™ 3D Planning Software is to be used for adult patients only and should not be used for diagnostic purpose.
#### 10) Summary of technological characteristics
BLUEPRINT™ Patient Specific Instrumentation (Subject Device System) has an equivalent intended use and the same fundamental scientific technology as the primary predicate and reference devices. The subject device is intended for total shoulder arthroplasty, (same as the predicate device) and for reverse shoulder arthroplasty (same as the reference device Materialise Glenoid Positioning System (K153602)). The subject device is equivalent to the predicate device for the main technological characteristics. Additionally, the subject device is equivalent to reference devices:
- Voyant Health, Ltd. TRAUMACAD VERSION 2.0 (K160001) for the humeral planning functionality -
- The Smith & Nephew DYONICS PLAN Hip Impingement Planning System (K132636) for the range of motion functionality
- -PIXMEO SARL OsiriX MD (K101342) for the bone density functionality
The design differences have been demonstrated to not affectiveness or raise new issues of safety or effectiveness.
Image /page/6/Picture/18 description: The image shows a logo with a stylized letter "T" inside a hexagon. The "T" is composed of two overlapping shapes, creating a layered effect. The color of the logo is a light shade of purple or blue. The overall design is simple and modern.
TORNIER S.A.S. 161. rue Lavoisier 38334 MONTBONNOT CEDEX FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33 S.A.S. au capital de 35 043 008 € SIRET : 070 501 275 000 21 R.C.S. : 070 501 275 CODE APE : 3250 A
{7}------------------------------------------------
# Implants Chirurgicau
#### 11) Performance data
BLUEPRINT™ Patient Specific Instrumentation (Subject Device System) was validated through nonclinical testing for the BLUEPRINT 3D Planning Software.
The validation of BLUEPRINT™ Patient Specific Instrumentation (Subject Device System) was already proven via BLUEPRINT™ Patient Specific Instrumentation (K143374) and BLUEPRINT™ Patient Specific Instrumentation (K160555), including polyamide and titanium materials, when used for helping the positioning of the Aequalis PerFORM (K111902) implant. The performed testing for validation of the design, manufacturing, biocompatibility, sterility, dimensions and the accuracy of the guide are applicable to the subject device. Technical, clinical and biological equivalences have been demonstrated in order to use BLUEPRINT™ Glenoid Guides with additional WRIGHT-TORNIER implants such as Aequalis PerFORM+ (K160975) and Aequalis Reversed II (K151293).
All BLUEPRINT™ Glenoid Guides aim to drill a K-wire guiding the main reaming, regardless of the implant planned. All BLUEPRINT™ Glenoid Guides have exactly the same design, except BLUEPRINT™ Glenoid Guides in titanium dedicated to be used with Aequalis PerFORM (K111902) and Aequalis PerFORM+ (K160975) implants which include an orientation hole for controlling rotation of the implant (cleared via BLUEPRINTTM Patient Specific Instrumentation (K160555)).
#### 12) Conclusion
BLUEPRINT™ Patient Specific Instrumentation (Subject Device System) described in this section has an equivalent intended use and the same fundamental scientific technology as the cleared BLUEPRINT™ Patient Specific Instrumentation (K160555) and the reference devices.
Based on the performance data presented for the design differences between the subject and predicate device, Tornier concludes that subject device is substantially equivalent to the predicate device.
Image /page/7/Picture/8 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is rendered in a light purple or blue color and appears to be constructed from geometric shapes, giving it a modern and somewhat abstract look. The hexagon provides a clear border around the 'T', making it stand out against the white background. The logo is simple, clean, and likely represents a company or organization whose name begins with the letter 'T'.
TORNIER S.A.S. 161. rue Lavoisier 38334 MONTBONNOT CEDEX FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33
S.A.S. au capital de 35 043 008 € SIRET : 070 501 275 000 21 R.C.S. : 070 501 275 CODE APE : 3250 A
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.