K162785 · Pari Respiratory Equipment, Inc. · CAF · Apr 14, 2017 · Anesthesiology
Device Facts
Record ID
K162785
Device Name
Velox Nebulizer System
Applicant
Pari Respiratory Equipment, Inc.
Product Code
CAF · Anesthesiology
Decision Date
Apr 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Velox Nebulizer System is a handheld nebulizer that is to be used with patients for whom doctors have prescribed medication for nebulization. It is intended for adult and pediatric patients 4 years and older who can coordinate breathing through a mouthpiece, and may be used in hospital-type facilities, nursing homes, sub-acute institutions and home environments.
Device Story
Handheld, portable, reusable, single-patient electronic nebulizer; uses piezo-driven, micro-perforated vibrating membrane to aerosolize liquid medication. Input: liquid medication in reservoir. Operation: user presses on/off switch; controller sends electrical charge to aerosol head; piezo-actuator vibrates; fluid ejected through membrane as droplets; aerosol deposited into chamber. Breath-enhanced two-way valve system allows inhalation of aerosolized medication. Used in hospitals, nursing homes, sub-acute institutions, and homes. Operated by patients or caregivers. Output: aerosolized medication for inhalation. Provides optical/acoustical feedback on status and errors. Benefits: efficient delivery of prescribed respiratory medication. No pause function, breath trigger, or interrupter; continuous nebulization until reservoir empty or Tmax reached.
Clinical Evidence
Bench testing only. Aerosol characterization (next generation impaction) showed MMAD, GSD, RF, and RD comparable to predicate. Biocompatibility (ISO 10993-1), gas path analysis, electrical safety (IEC 60601-1 series), EMC (IEC 60601-1-2), and simulated use (365 nebulizations/52 disinfection cycles) confirmed device performance and safety over one-year lifetime.
Technological Characteristics
Materials: polypropylene, thermoplastic elastomer, stainless steel, piezo-ceramic, polyamide film, medical adhesives. Principle: piezo-driven micro-perforated vibrating membrane. Energy: AC power or 3x AA batteries. Form factor: handheld, integrated controller/nebulizer. Connectivity: standalone. Sterilization: non-sterile. Software: moderate level of concern, embedded control logic.
Indications for Use
Indicated for adult and pediatric patients 4 years and older who can coordinate breathing through a mouthpiece, for whom doctors have prescribed medication for nebulization.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three figures in profile facing right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2017
Pari Respiratory Equipment, Inc. Michael Judge VP, Operations And Regulatory Affairs 2412 Pari Way Midlothian, Virginia 23112
Re: K162785
Trade/Device Name: Velox Nebulizer System Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: March 8, 2017 Received: March 8, 2017
Dear Mr. Judge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K162785
Device Name Velox Nebulizer System
#### Indications for Use (Describe)
The Velox Nebulizer System is a handheld nebulizer that is to be used with patients for whom doctors have prescribed medication for nebulization. It is intended for adult and pediatric patients 4 years and older who can coordinate breathing through a mouthpiece, and may be used in hospital-type facilities, nursing homes, sub-acute institutions and home environments.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# SECTION 6: 510(K) SUMMARY
#### 1. Submitter
PARI Respiratory Equipment, Inc. 2412 PARI Way Midlothian, Virginia 23112
Phone: 804-253-7274 804-253-0260 Fax:
Contact Person: Mike Judge x 269 Date Prepared: September 15, 2016
#### 2. Device Name
| Name of Device: | Velox™ Nebulizer System |
|-----------------------|-------------------------------------------------------|
| Common or Usual Name: | Nebulizer |
| Classification Name: | Nebulizer, Direct Patient Interface (21 CFR 868.5630) |
| Regulatory Class: | II |
| Product Code: | CAF |
There are no other variants or model numbers of this device.
#### 3. Legally Marketed Predicate Device
The following predicate device has not been subject to a design-related recall.
### a. Primary Predicate Device
eRapid® Nebulizer System, K112859
Additionally, the Philips Respironics I-Neb AAD System, cleared under K042991, is used in this 510(k) submission as a reference device.
#### Device Description 4.
The Velox Nebulizer System is a portable, reusable, single-patient use, handheld electronic nebulizer that uses a piezo-driven, micro-perforated vibrating membrane technology to aerosolize liquid medications for the treatment or prophylaxis of respiratory diseases. It is to be used with adult and pediatric patients for whom doctors have prescribed medication, i.e. it is for prescription use only. It may be used in hospital-type facilities, nursing homes, sub-acute institutions and home environments. It is provided non-sterile.
The device consists of two major components. One of these components is the nebulizer itself, which has the following materials in its construction: (1) polypropylene: (2); thermoplastic elastomer; (3) stainless steel; (4) piezo-ceramic; (5) capton; and, (6) medical grade adhesives. It contains, as subcomponents, the following: (1) a medication cap, (2) a medication reservoir, (3) an aerosol head, (4) an aerosol chamber; and, (5) a mouthpiece.
The second major component is a Controller that is attached to the Nebulizer for its operation. This component, the body of which is constructed of acrylonitrile butadiene styrene and thermoplastic elastomer, has the following principal internal subcomponents: (1)
{4}------------------------------------------------
a circuit board: (2) a USB connector: (3) a patient interface (on/off switch); and (4) AA battery housing.
The controller's functions are to; (1) conduct power pre-processing; (2) supervise, through incorporated software, the nebulizer's aerosol generation; and, (3) generate optical and acoustical feedback to the user on nebulization status and out-of-range parameters.
The incorporated software is device-specific and: (1) controls the power to the aerosol head, i.e., either to have the aerosol head active and generate aerosol, or shut-off; (2) controls optical and acoustical feedback to the user; and; (3) is not dependent for its operation on any external devices, such as a monitor, printer, keyboard or mouse.
Energy sources for the device are provided by: (1) connection to a mains power source, via an AC Power Supply; or. (2) disposable/rechargeable AA batteries.
The device has two associated accessories: (1) a cleaning aid, branded VELOXcare, that assists in debris removal from the membrane prior to cleaning; and, as heretofore noted. (2) an AC Power Supply to provide mains power to the device when desired or required.
#### Mechanism of Action 5.
The Velox Nebulizer System's mode of action is permanent, i.e., once the on/off switch is on the controller pressed, permanent nebulization occurs until the medication is nebulized. There is no pause function. Further, there is no breath trigger or interrupter.
Nebulization is accomplished as follows. When the on/off button is pressed the controller sends an electrical charge to the aerosol head in the nebulizer. The charge alternation causes the piezo-actuator to vibrate. Vibration motion then builds up a positive pressure in the boundary layer of the liquid column, ejecting the fluid through the holes as droplets and creating the aerosol on the front side of the membrane. This aerosol is deposited vertically into the aerosol chamber. Once in the chamber it is available for inhalation by the patient.
The inhalation is via a breath-enhanced two-way valve system. During inhalation the inhalation valves open to allow fresh air to enter the aerosol chamber. This inhaled fresh air stream, takes the aerosolized medication, with minimum depositional loss, through the mouthpiece into the patient's respiratory tract. Note that, during inhalation, the exhalation valve on the mouthpiece is closed, preventing any aerosol dilution on its way to the patient. During exhalation, the inhalation valves close and the exhalation valve opens, driving the expired air out of the mouthpiece.
#### 6. Indications for Use
The Velox Nebulizer System is a handheld nebulizer that is to be used with patients for whom doctors have prescribed medication for nebulization. It is intended for adult and pediatric patients 4 years and older who can coordinate breathing through a mouthpiece, and may be used in hospitals, hospital-type facilities, nursing homes, sub-acute institutions and home environments.
The foregoing statement for the subject device is identical to that of the predicate eRapid® device (with the exception of the "4 years and older who can coordinate breathing through a mouthpiece" restriction. The indications are substantially equivalent; the subject device indication statement simply provides a quantification of patient age, and clarification of the patient's breathing
{5}------------------------------------------------
coordination ability at that age). Further, both the subject and predicate device have the same intended use.
#### Comparison of Technological Characteristics with the Predicate Device 7.
# Table 1. Substantial Equivalence Comparison
| PRODUCT | VELOX NEBULIZER<br>SYSTEM | ERAPID NEBULIZER<br>SYSTEM |
|----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DEVICE | CAF, Nebulizer (direct patient interface), | CAF, Nebulizer (direct patient interface), |
| CLASSIFICATION | 21 CFR 868.5630 | 21 CFR 868.5630 |
| 510K NO. | n/a | K112859 |
| MANUFACTURER<br>(REG. NO.) | PARI Respiratory Equipment, Inc. (2954963) | PARI Respiratory Equipment, Inc. (2954963) |
| INTENDED USE | The Velox is a handheld nebulizer that will be<br>used with patients for whom doctors have<br>prescribed medication for nebulization. | The eRapid is a handheld nebulizer that will be used<br>with patients for whom doctors have prescribed<br>medication for nebulization. |
| Environment of use | Home care, nursing home, sub-acute institution, or<br>hospital | Home care, nursing home, sub-acute institution, or<br>hospital |
| Target Population | Adult and pediatric patients 4 years and older who<br>can coordinate breathing through a mouthpiece | Adult and pediatric |
| Prescription Use | Rx Only | Rx Only |
| METHOD OF OPERATION | | |
| Technology Used | Micro-perforated vibrating membrane technology:<br>active, vibrating membrane | Micro-perforated vibrating membrane technology:<br>active, vibrating membrane |
| Aerosol<br>Direction/Output | Vertically sprayed into aerosol chamber, cleared<br>by inspiratory flow | Horizontally sprayed into aerosol chamber, cleared<br>by inspiratory flow |
| Breath Enhanced/<br>Controlled/<br>Triggered | Breath enhanced two-way valve system.<br>Permanent nebulization - no breath trigger or<br>control. | Breath enhanced two-way valve system. Permanent<br>nebulization - no breath trigger or control. |
| Fluid Delivery -<br>Generator | Direct contact between aerosol head and fluid in<br>sealed chamber (medication reservoir). | Direct contact between aerosol head and fluid in<br>sealed chamber (medication reservoir). |
| Configuration | Controller component is connected directly to the<br>nebulizer component. No connection cord for<br>operation. | Nebulizer and controller are separate components,<br>joined by a connection cord for operation. |
| Automatic Shut-off | Yes. Automatic shut-off when: (1) medication<br>reservoir is empty; (2) out-of-range parameters<br>are detected; or (3) programmed maximum<br>operating time) is reached. | Yes. Automatic shut-off when: (1) medication<br>reservoir is empty: (2) out-of-range parameters are<br>detected; or (3) programmed maximum operating<br>time) is reached. |
| PRODUCT | VELOX NEBULIZER<br>SYSTEM | ERAPID NEBULIZER<br>SYSTEM |
| DESIGN CAPACITIES | | |
| Medication Reservoir | 2.0 mL | 2.0 mL |
| Min. Fill | 6.0 mL | 6.0 mL |
| Max. Fill | Approx. 1.0 ml residue (depending on filled | Approx. 1.0 ml residue (depending on filled volume) |
| Residue | volume) | |
| PERFORMANCE | | |
| MMAD | Mass Median Aerodynamic Diameter of subject device is slightly lower than that of eRapid. | |
| GSD | Geometric Standard Deviation of subject device is substantially equivalent to that of eRapid | |
| RF | Respirable Fraction of subject device is slightly higher than that of eRapid | |
| RD | Respirable Dose of subject device is substantially equivalent to that of eRapid | |
| SOFTWARE | | |
| Level of Concern | Moderate | Moderate |
| Device-Specific | Yes | Yes |
| Dependent on<br>External Devices | No | No |
| Function | By continuous loop: (1) conducts power pre-<br>processing; (2) supervises the nebulizer's aerosol<br>generation; and (3) communicates to the user on<br>the operational status of the device and out-of-<br>range parameters. These functions are performed<br>continuously until the medication reservoir is<br>empty, out-of-range parameters are detected, or<br>programmed maximum operating time (Tmax) is<br>reached. | By continuous loop: (1) conducts power pre-<br>processing; (2) supervises the nebulizer's aerosol<br>generation; and (3) communicates to the user on the<br>operational status of the device and out-of-range<br>parameters. These functions are performed<br>continuously until the medication reservoir is empty,<br>out-of-range parameters are detected, or<br>programmed maximum operating time (Tmax) is<br>reached. |
| | It should be noted however, that, unlike the<br>predicate device's software, the subject device's<br>software has: (1) no integral pause function that<br>enables treatment interruption for user<br>convenience; and, (2) no separate cleaning aid<br>mode that allows the user to operate a cleaning<br>aid. | It should be noted however, that the predicate<br>device's software has: (1) an integral pause function<br>that enables treatment interruption for user<br>convenience; and, (2) a separate cleaning aid mode<br>that allows the user to operate a cleaning aid. |
| Audiovisual Signals | LED and tone sounds concerning battery level,<br>operational status and failure mode. | LED, tone sounds and illuminated display showing<br>graphical symbols concerning battery level,<br>operational status and failure mode. |
| ELECTRICAL CONNECTION | | |
| PRODUCT | VELOX NEBULIZER<br>SYSTEM | ERAPID NEBULIZER<br>SYSTEM |
| Power Wattage | < 2.0 W under normal load | < 2.0 W under normal load |
| AC Power Supply | | |
| Input | 100-240 VAC / 50-60 Hz | 100-240 VAC / 50-60 Hz |
| Output | 5 V | 12 V |
| Battery Operation | | |
| AA Disposable | 3 x 1.5 V | 4 x 1.5 V |
| AA Rechargeable | 3 x 1.2 V | 4 x 1.2 V |
| Classification<br>According To IEC<br>60601 - 1 | | |
| Type of electric shock<br>protection (AC Power<br>Supply) | Protection Class II | Protection Class II |
| Degree of protection<br>from electric shock of<br>the part used<br>(nebulizer) | Type BF | Type BF |
| Degree of protection<br>against water ingress<br>in accordance with EN<br>60529 (IP rating) | IP-22 | No protection |
| Degree of protection<br>when used in the<br>presence of flammable<br>mixtures of<br>anesthetics with air,<br>with oxygen, or with<br>nitrous oxide | No protection | No protection |
| Operating Mode | Continuous operation | Continuous operation |
| NEBULIZER MATERIALS | | |
| Medication Cap | PP, TPE | PP, POM, TPE |
| Medication Reservoir | PP, TPE | – |
| Fluid Feed | N/A | PP, TPE |
| Headmount | N/A | PP, stainless steel |
| Housing | PP, TPE | N/A |
| Aerosol Head | Stainless steel, piezoceramic,<br>polyamide film, adhesive | Stainless steel, piezoceramic, polyamide film,<br>adhesive |
| PRODUCT | VELOX NEBULIZER<br>SYSTEM | ERAPID NEBULIZER<br>SYSTEM |
| Inhalation Valve | TPE | Silicone |
| Aerosol chamber | PP | PP |
| Mouthpiece | PP | PP |
| Exhalation Valve | TPE | Silicone |
| MECHANICAL | | |
| Dimensions | | |
| Nebulizer (L /W /H) | 145 x 61 x 65 mm | 145 x 50 x 63 mm |
| Controller (H /Ø) | 70 x 61 x 92 mm | 40 mm x Ø 116 mm |
| Weight (approx.) | | |
| Handset w/o med. | 45g | 55g |
| System w/batt. | 285g | 300g |
| CLEANING / DISINFECTION |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.