K162783 · Bioptigen, Inc. · OBO · Dec 9, 2016 · Ophthalmic
Device Facts
Record ID
K162783
Device Name
EnFocus 2300, EnFocus 4400
Applicant
Bioptigen, Inc.
Product Code
OBO · Ophthalmic
Decision Date
Dec 9, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.1570
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Bioptigen EnFocus™ device is intended to acquire, process, display and save depth resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SDOCT). The EnFocus™ is indicated for use as an aid in the visualization of physiologic and pathologic conditions of the eye through non-contact optical imaging. It is indicated for use on patient populations from premature and neonatal infants to adult. The system is indicated for use in supine imaging, mounted to a surgical microscope, with cooperative patients or patients under anesthesia.
Device Story
EnFocus 2300/4400 is a non-contact, noninvasive ophthalmic imaging device using Spectral Domain Optical Coherence Tomography (SD-OCT) and near-infrared light to capture ocular tissue microstructure images. The system includes an OCT engine, scan head, and computer with InVivoVue software. It mounts to a surgical microscope for intraoperative imaging without interrupting procedures. The device provides high-speed volume data acquisition; clinicians view images on-screen to visualize ocular structures, aiding surgical decision-making. The system is operated by surgical staff in an OR setting. Recent modifications include repackaging optical and computing subsystems into a single integrated chassis for improved logistics.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser safety (IEC 60825-1), light hazard protection (ISO 15004-2), and optical performance/image quality validation compared to the predicate.
Technological Characteristics
Spectral Domain OCT; near-infrared light source (800 nm SLD); Class 1 Laser Product. Includes OCT engine, scan head, and integrated computer chassis. Compatible with 175mm/200mm objective lenses. Connectivity: digital video output. Software: InVivoVue 2.10. Standards: AAMI/ANSI 60601-1, IEC 60601-1-2, ISO 62304, ISO 14971, IEC 60825-1, ISO 15004-2.
Indications for Use
Indicated for visualization of physiologic and pathologic ocular conditions in patients ranging from premature/neonatal infants to adults. Used in supine position, mounted to surgical microscope, with cooperative patients or patients under anesthesia.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing the feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 9. 2016
Bioptigen, Inc. Ravi Kommineni Director of Quality and Regulatory Affairs 633 Davis Drive, Suite 480 Morrisville, NC 27560
Re: K162783
Trade/Device Name: EnFocus 2300, EnFocus 4400 Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO Dated: November 16, 2016 Received: November 17, 2016
Dear Mr. Kommineni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name EnFocusTM (EnFocus 2300, EnFocus 4400)
Indications for Use (Describe)
The Bioptigen EnFocus™ device is intended to acquire, process, display and save depth resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SDOCT).
The EnFocus™ is indicated for use as an aid in the visualization of physiologic and pathologic conditions of the eye through non-contact optical imaging. It is indicated for use on patient populations from premature and neonatal infants to adult. The system is indicated for use in suping, mounted to a surgical microscope, with cooperative patients or patients under anesthesia.
| Concession of the program of a concession of a concession of concession of concession a promote of consisted on annual consistence of consisted of consistences of consistence | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| | Type of Use (Select one or both, as applicable) |
| | |
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for bioptigen. The logo consists of a stylized symbol on the left, followed by the word "bioptigen" in a bold, sans-serif font. The symbol is made up of three stacked, horizontally oriented shapes in different shades of blue and green.
# 510(k) Summary
This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safety Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92
#### l. General Information
| Manufacturer: | Bioptigen, Inc.<br>A Leica Microsystems Company<br>633 Davis Drive, Suite 480<br>Morrisville, NC 27560 |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ravi Kommineni<br>Director for Quality and Regulatory Affairs<br>Bioptigen, Inc.<br>Phone: 919-348-6439<br>Email: ravi.kommineni@leica-microsystems.com |
- Date Prepared: 12/5/2016
#### II. Device Information
| Trade Name: | EnFocus 2300, EnFocus 4400 |
|-----------------------|----------------------------------------------|
| Model Number(s): | 2300 (70-10091) |
| | 4400 (70-10094) |
| Common Name: | Spectral Domain Optical Coherence Tomography |
| Classification Name: | Tomography, Optical Coherence |
| Regulation Number: | 21 CFR § 886.1570 |
| Product Code: | OBO |
| Classification Panel: | Ophthalmology |
| Class: | II |
#### lll. Predicate Device
EnFocus (EnFocus 2300, EnFocus 4400) - K150722
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "bioptigen". The logo consists of a stylized graphic to the left of the word "bioptigen". The graphic is made up of three stacked, rounded rectangles, with the top one being blue, the middle one being a darker shade of green, and the bottom one being a lighter shade of green. The word "bioptigen" is written in a bold, sans-serif font and is black.
#### Purpose of Special 510(k) IV.
The purpose of the special 510(k) submission is to gain clearance for the changes made to the EnFocus™ (K150722) cleared device. The EnFocus™ optical and computing subsystems are now repackaged into a single enclosure for improved manufacturing and servicing logistics. The associated hardware and software changes do not affect the intended use, functionality of the system nor alter the principle of operation of the EnFocus™ device.
#### > Indications for Use
The Bioptigen EnFocus™ device is intended to acquire, process, display and save depth resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SDOCT).
The EnFocus™ is indicated for use as an aid in the visualization of physiologic and pathologic conditions of the eye through non-contact optical imaging. It is indicated for use on patient populations from premature and neonatal infants to adult. The system is indicated for use in supine imaging, mounted to a surgical microscope, with cooperative patients under anesthesia.
#### VI. Device Description
The EnFocus™ is a non-contact, noninvasive ophthalmic imaging device that includes an OCT engine, a scan head and a system computer with system software. The EnFocus™ uses Spectral Domain Optical Coherence Tomography (SD-OCT) and a near infrared light source to image ocular tissue microstructures.
The EnFocus™ is coupled to a surgical microscope for OCT imaging during ophthalmic surgical procedures. The software, InVivoVue™, works with the hardware and the controller to offer intuitive, flexible system control for high-speed volume data acquisition and imaging.
The EnFocus™ system includes two OCT-compatible objective lenses for use with the surgical microscope: a 175mm lens and 200mm lens. The system also offers a choice of accessory masks that may be deployed to manage illumination glare artifacts when necessary.
Using the EnFocus™, OCT imaging may be acquired during the surgical procedure, without stopping a procedure or repositioning the surgical microscope. The surgical microscope position is stationary relative to the surgical procedure, and the surgical view is unaltered by the scanning of the OCT beam.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "bioptigen". The logo consists of a stylized graphic to the left of the word "bioptigen". The graphic is made up of three stacked, rounded shapes in blue, teal, and green. The word "bioptigen" is written in a bold, sans-serif font in black.
#### VII. Substantial Equivalence
The modified EnFocus™ device has the same intended use, indications for use, technological characteristics and principle of operation as the cleared EnFocus™ device (K150722) to which it is a modification.
| Category | Predicate (EnFocus™ - K150722) | Subject (Modified EnFocus™) |
|--------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Indications for Use | The Bioptigen EnFocus™ device is intended to acquire,<br>process, display and save depth resolved images of<br>ocular tissue microstructure using Spectral Domain<br>Optical Coherence Tomography (SDOCT).<br>The EnFocus™ is indicated for use as an aid in the<br>visualization of physiologic and pathologic conditions of<br>the eye through noncontact optical imaging. It is<br>indicated for use on patient populations from premature<br>and neonatal infants to adult. The system is indicated<br>for use in supine imaging, mounted to a surgical<br>microscope, with cooperative patients or patients under<br>anesthesia. | Same as predicate |
| Target Population | • Premature and neonatal infants to adult<br>• Supine<br>• Cooperative patients or patients under anesthesia | Same as predicate |
| Anatomical Sites | Anterior and Posterior Segment of Eye | Same as predicate |
| Method of Operation | SD-OCT | Same as predicate |
| Principle of Operation | OCT dichroically coupled into collimated space of<br>common main objective surgical microscope. | Same as predicate |
| Energy Used and/or<br>Delivered | Near Infrared Optical Emissions | Same as predicate |
| Light Source | 800 nm band SLD:<br>High Resolution (HR)<br>Very High Resolution (VHR) | Same as predicate |
| Light Source Classification | Class 1 Laser Product | Same as predicate |
| Optical Power | < 750 µW | Same as predicate |
| PERFORMANCE | | |
| Resolution, Lateral | < 31.0 µm (175 mm Obj., low NA)<br>< 15.1 µm (175 mm Obj., high NA)<br>< 35.4 µm (200 mm Obj., low NA)<br>< 17.3 µm (200 mm Obj., high NA) | Same as predicate |
| Field of View, Lateral | ≥ 20 mm | Same as predicate |
| Resolution, Axial | Model 2300 VHR: < 4 µm in tissue<br>Model 4400: < 9 µm in tissue | Same as predicate |
| Field of View,<br>Longitudinal<br>(Depth Range)<br>(in tissue/air) | Model 2300: 2.5 / 3.4 mm<br>Model 4400: 11.1 / 15.3 mm | Same as predicate |
| Category | Predicate (EnFocus™ - K150722) | Subject (Modified EnFocus™) |
| Scanner Type | Galvanometric mirror pair | Same as predicate |
| Scan Patterns Available | Line, rectangular volume, circle, concentric rings<br>(annular volume), radial lines (radial volume) | Same as predicate |
| Scan Pixels | Axial (depth) :<br>Model 2300: 1024<br>Model 4400: 2048<br>Lateral: User Selectable<br>Maximum A-scans / B-scan: 2000<br>Maximum total A-scans/volume: 1,000,000 | Same as predicate |
| Scan Rate | Model 2300: 32,000 A-scans/sec<br>Model 4400: 18,000 A-scans/sec | Same as predicate |
| DIMENSIONS | | |
| System Footprint | Mobile Cart: 24" x 22" x 37.5" | Same as predicate |
| Subsystem Dimensions | Computer: 13" (w), 15.5" (l), 5.75" (h)<br>Interface Box: 13" (w), 15.5" (l), 5.75" (h)<br>OCT Engine: 17.75" (w), 17" (l), 7.5" (h) | Integrated Chassis:<br>19" (w) x 10.5" (l) x 8.25" (h) |
| Scanner | Scan head: 3.2" (h) x 3.2" (od) | Same as predicate |
| Dimensions | Relay Arm: 10.9" (1) x 1.7" (od) Scanner Assembly: 2.2"<br>(I) x 5.5" (w) x 8.0" (h)<br>Mounting Adapter: 2.2" (l) x 2.8" (w) x 1.8" (h) | Same as predicate |
| Scanner<br>Ergonomics | Surgical Microscope Mounted | Same as predicate |
| SOFTWARE | | |
| Software | InVivoVue™ (IVV) 2.6 | InVivoVue™ (IVV) 2.10 |
| Scan Setting Procedures | Protocol Pre-Sets, Scan Pre-Sets, Custom Scan<br>definitions | Same as predicate |
| Measurements | Manual placement of on-screen calipers | Same as predicate |
| Doppler | Qualitative blood flow visualization with color Doppler<br>OCT | Same as predicate |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "bioptigen". The logo consists of a stylized "B" made up of three overlapping shapes in blue and green, followed by the word "bioptigen" in black sans-serif font. The logo is simple and modern, with a focus on the company name.
#### VIII. Performance
The modified EnFocus™ device is in compliance with the following recognized consensus standards:
| Standard | Version | Title |
|----------------------|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| AAMI/ANSI<br>60601-1 | 2005/(R)2012 | Medical electrical equipment - Part 1: General requirements for basic safety<br>and essential performance |
| IEC 60601-1-2 | Edition 4.0 2014-02 | Medical electrical equipment - Part 1-2: General requirements for safety –<br>Collateral Standard: Electromagnetic compatibility – Requirements and tests |
| ISO 62304 | 2015 | Medical device software - Software life-cycle processes |
| ISO 14971 | Second Edition<br>2007-03-01 | Medical Devices - Application of Risk Management to Medical Device |
| IEC 60825-1 | Edition 2.0 2007-03 | Safety of laser products - Part 1:Equipment classification and requirements |
| ISO 15004-2 | 2007 | Ophthalmic Instruments - Fundamental Requirements And Test Methods<br>Part 2: Light Hazard Protection |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Bioptigen. The logo consists of a stylized "B" made up of three horizontal bars in different shades of blue and green. To the right of the "B" is the word "bioptigen" in a sans-serif font. The word is in black.
#### IX. Non-Clinical Testing
## Safety
Electrical safety and EMC testing were conducted on the modified EnFocus™ device and demonstrate that the system complies with the IEC 60601-1 and IEC 60601-1-2 recognized standards.
IEC 60825-1 Class 1 laser product safety testing and ISO 15004-2 Group 2 light hazard test results together confirm the system to be eye safe when used as labeled.
## Performance
Optical performance testing demonstrates that the modified EnFocus™ maintains optical performance attributes equivalent to the original device.
Image quality testing validates that the OCT image quality of the modified EnFocus™ device is statistically equivalent to the original device.
Video compatibility testing confirms that the modified EnFocus™ hardware remains capable of displaying digital video inputs to primary and secondary monitors.
Fundus viewing system testing validates that the modified EnFocus™ is compatible with qualified fundus viewing systems.
# Software
InVivoVue™ software was verified and validated in accordance with the software lifecycle requirements of ISO 62304.
#### X. Clinical Testing
No clinical information is included in this submission.
#### XI. Conclusion
The results of verification and validation demonstrate that the modified device is as safe and effective as the predicate device and does not introduce new issues of safety and effectiveness. Therefore, the modified EnFocus™ is substantially equivalent to its cleared predicate device, EnFocus™ (K150722).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.