K162781 · Formatk Systems , Ltd. · GEX · Nov 9, 2016 · General, Plastic Surgery
Device Facts
Record ID
K162781
Device Name
Magma Spark
Applicant
Formatk Systems , Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 9, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Magma Spark device is indicated for use in aesthetic and cosmetic applications in dermatology. Magma Spark device has connection capability with the following available treatment hand-pieces (applicators), for multi-application treatment options. All handpieces are designed for aesthetic and cosmetic dermatologic procedure applications. Magma Spark device with ALD and LLD diode laser applicators iis indicated for hair removal, permanent hair reduction in skin types I-VI and for the treatment of pseudofolliculitis barbae (PFB). Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after completions of treatment regime
Device Story
Magma Spark is a tabletop, multi-application laser platform for aesthetic dermatology. It utilizes 808 nm diode laser applicators (ALD and LLD) to deliver optical energy for hair removal and PFB treatment. The system includes a melanin meter, water-based cooling, and a CPU-controlled user interface with a touchscreen. Operators (clinicians) select and monitor treatment modes and intensity via the digital control panel. The device transforms electrical input into controlled laser output based on patient skin and hair type parameters. It is used in clinical settings to provide permanent hair reduction, benefiting patients through long-term hair regrowth stability.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including comprehensive software-system verification covering firmware functionality, hardware interfaces, performance, load testing, integrity, recovery, and usability.
Technological Characteristics
Tabletop laser platform; 808 nm diode laser (ALD/LLD); water-based cooling; CPU-controlled; touchscreen interface. Complies with AAMI/ANSI 60601-1, IEC 60601-1-2, IEC 60825-1, and IEC 60601-2-22. Dimensions: 26.0"W x 21.7"D x 23.6"H; Weight: 50.7 lb. Input: 100-240 VAC, 50-60 Hz.
Indications for Use
Indicated for aesthetic/cosmetic dermatologic procedures, specifically hair removal and permanent hair reduction in skin types I-VI, and treatment of pseudofolliculitis barbae (PFB).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 9, 2016
Formatk Systems Ltd. % Ahava Stein Regulatory Manager A. Stein - Regulatory Affairs Consulting Ltd. 20 Hata'as Str., Suite 102 Kfar Saba, 4442520 IL
Re: K162781
Trade/Device Name: Magma Spark Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: September 29, 2016 Received: October 3, 2016
Dear Ahava Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely, Jennifer R. Stevenson -A For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices
Center for Devices and Radiological Health
Office of Device Evaluation
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K162781 Device Name Magma Spark
#### Indications for Use (Describe)
Magma Spark device is indicated for use in aesthetic applications in dermatology.
Magma Spark device has connection capability with the follo\ving available treatment hand-pieces (applicators), for multiapplication treatment options. All hand-pieces are designed for aesthetic demlatologic procedure applications. Magma Spark device \vith ALD or LLD diode laser applicators is indicated for hair removal, pennanent hair reduction* in skin types I-VI and for the treatment of pseudofolliculitis barbae (PFB).
(*Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after completion of treatment regime)
Type of Use (Select one or both, as applicable)
|ZJ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section 6: Special 510(k) Summary
### 510(K) SUMMARY
# MAGMA SPARK DEVICE
510(k) Number K162781
| Applicant Name: | Formatk Systems Ltd |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Company Name: | Tavor Building, POB 533<br>Yokneam Iillit 2069206 Israel<br>Tel: +972-4-909-7440<br>Fax: +972-4-909-7471<br>E-mail: ahava@asteinrac.com |
| Contact Person:<br>Official Correspondent: | Ahava Stein |
| Company Name: | A. Stein - Regulatory Affairs Consulting Ltd. |
| Address: | 20 Hata'as Str., Suite 102<br>Kfar Saba 4442520 Israel<br>Tel: + 972-9-7670002<br>Fax: +972-9-7668534<br>E-mail: ahava@asteinrac.com |
| Date Prepared: | September 29, 2016 |
| Trade Name: | Magma Spark |
| Classification Name: | CFR Classification section 878.4810 (Product code GEX) |
| Classification: | Class II Medical Device |
Predicate Device:
Magma Spark device is substantially equivalent to the previously cleared, Magma device, also manufactured by Formatk Systems Ltd.
| Device | Manufacturer | 510(k) No. |
|--------|----------------------|------------|
| Magma | Formatk Systems Ltd. | K153566 |
{4}------------------------------------------------
Device Description:
The Magma Spark device is a computerized multi-application platform intended for non-invasive aesthetic applications utilizing laser optical power.
Magma Spark is a relatively small tabletop device that provided with melanin meter and two diode laser applicators (ALD and LLD) at nominal 808 nm wavelength. The system platform includes AC/DC power supply unit, water based cooling system, CPU main card and user interface including an LCD display and touch screen module. The operator of Magma Spark chooses and monitors the mode and intensity of the treatment from a digital control panel on the front of the device.
| Main Line Frequency (nominal) | 50 - 60 Hz | |
|---------------------------------------|--------------------------|--|
| Input Voltage (nominal) | 100 - 240 VAC | |
| Input Current (A) | 6.0 max | |
| Platform dimensions (inch) | 26.0"W x 21.7"D x 23.6"H | |
| Platform weight (lb.) | 50.7 | |
| ALD & LLD Diode laser wavelength (nm) | 808 | |
| Maximal Fluence (J/cm2): | | |
| ALD | Up to 75 | |
| LLD | Up to 35 | |
Device Specifications:
Intended Use/Indication for Use:
Magma Spark device is indicated for use in aesthetic and cosmetic applications in dermatology.
Magma Spark device has connection capability with the following available treatment hand-pieces (applicators), for multi-application treatment options. All handpieces are designed for aesthetic and cosmetic dermatologic procedure applications. Magma Spark device with ALD and LLD diode laser applicators iis indicated for hair removal, permanent hair reduction in skin types I-VI and for the treatment of pseudofolliculitis barbae (PFB).
Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after completions of treatment regime
{5}------------------------------------------------
Section 6: Special 510(k) Summary
Performance Standards:
Magma Spark complies with the following recognized consensus standards:
- . AAMI/ANSI 60601-1 (2012). Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, Mod).
- . IEC 60601-1-2 (Edition 3.0, 2007), Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance -Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.
- . IEC 60825-1 (Edition 2.0, 2007), Safety Of Laser Products - Part 1: Equipment Classification, And Requirements JIncluding: Technical Corrigendum 1 (2008), Interpretation Sheet 1 (2007), Interpretation Sheet 2 (2007)].
- . IEC 60601-2-22 (Edition 3.0, 2007 & Edition 3.1 2012), Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment.
Non-Clinical (Bench) Performance Data:
The underlying technology and principle of operation of Magma Spark is essential the same as the predicate device. Therefore, results were gathered to support the safety of the modified device and thorough test of the software was executed. Comprehensive software-system test plan and results report covers the full system test of the Magma Spark. It includes operation and user procedures, as well as programs. In addition to comprehensively testing firmware functionality, hardware interfaces, performance, load test, download procedure, integrity, recovery and usability.
Pre-Clinical (Animal Study) Performance Data:
Not Applicable
Clinical Performance Data: Not Applicable
## Substantial Equivalence:
The components of modified Magma Spark device, similarly to predicate Magma device, generate its mechanism of operation using the same underlying technology. Delivery of the laser energy through each specific applicator (either ALD or LLD) is monitored and controlled by the device's CPU. The user interface control panel
{6}------------------------------------------------
provides the user with the optimal treatment settings taking into consideration the patient skin type and hair type (for hair removal treatments). Since Magma Spark device is limited to operate with diode laser applicators only, the intended use of Magma Spark device is same as predicate Magma device in relation to the diode laser-relevant applications only. Using both modified and cleared devices the device user can decide on the optimal treatment settings and adjust these treatment settings through the control panel. Furthermore, Magma Spark device, as the cleared Magma device, introduces similar safety features and comply with same relevant consensus standards.
### Conclusions:
Based on the comparison to predicate device, Magma Spark device is substantially equivalent to the predicate Magma device for the mentioned indication for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.