K162775 · Smith & Nephew, Inc. · JWH · Jan 13, 2017 · Orthopedic
Device Facts
Record ID
K162775
Device Name
Legion Cone System
Applicant
Smith & Nephew, Inc.
Product Code
JWH · Orthopedic
Decision Date
Jan 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Legion Cones are intended to be used with the Legion Revision and Legion Hinge knee systems. The Legion Cones are indicated for the following: 1. Rheumatoid arthritis. 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 3. Failed osteotomies, unicompartmental replacement, or total knee replacement. 4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact. 5. Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral collateral ligament) are incompetent. 6. Hinge knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral collateral ligament) are absent of incompetent. These are single use implants and are intended for use with and without bone cement.
Device Story
Legion Cone System consists of porous-coated titanium alloy (Ti-6Al-4V) tibial and femoral augments; designed to address cavitary bone defects in proximal tibia or distal femur during knee revision or primary surgery. Implants are conically shaped with central cannulation for fixation to Legion Revision and Legion Hinge knee system components (baseplates, stems, couplers). Used in clinical/OR settings by orthopedic surgeons. Implants provide structural support for prosthetic components in patients with significant bone loss; can be used with or without bone cement. Surgeon prepares bone site using provided instrumentation; implants are single-use.
Clinical Evidence
Bench testing only. Mechanical evaluation included fatigue strength, push-out strength, torsional shear, and tensile attachment strength. No clinical data was required to support safety and effectiveness.
Technological Characteristics
Substrate: Titanium alloy (Ti-6Al-4V) per ASTM F1472. Porous coating: C.P. Titanium per ASTM F67 (Stiktite). Interior surface: Grit-blasted for cement adhesion. Form factor: Conical, cannulated augments. Sterilization: Gamma radiation. Connectivity: None (mechanical implant).
Indications for Use
Indicated for patients requiring primary or revision knee surgery with rheumatoid, post-traumatic, osteoarthritis, or degenerative arthritis, or failed previous knee procedures (osteotomy, unicompartmental, or total knee replacement). Specific use depends on ligamentous stability: posterior stabilized (cruciates incompetent, collaterals intact), constrained (cruciate and one/both collaterals incompetent), or hinge (cruciate and one/both collaterals absent/incompetent).
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Zimmer Trabecular Metal Tibial and Femoral Cone Augments (K053340)
Smith & Nephew Genesis II Total Knee System (K951987)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles forming the staff and a flowing ribbon representing the snake.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 13, 2017
Smith & Nephew, Inc. Ms. Dongeun Kim Regulatory Affairs Specialist I 1450 Brooks Road Memphis. Tennessee 38116
Re: K162775 Trade/Device Name: Legion Cone System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, MBH Dated: September 7, 2016 Received: October 3, 2016
Dear Ms. Dongeun Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) WX K162775
Device Name Smith & Nephew Legion Cone System
Indications for Use (Describe)
The Legion Cones are intended to be used with the Legion Revision and Legion Hinge knee systems. The Legion Cones are indicated for the following:
1. Rheumatoid arthritis.
2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
5. Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral collateral ligament) are incompetent.
6. Hinge knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral collateral ligament) are absent of incompetent.
These are single use implants and are intended for use with and without bone cement.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image displays the Smith & Nephew logo. The logo features an orange flower-like symbol on the left, followed by the text "We are smith&nephew". The words "smith&nephew" are in orange, while "We are" is in gray.
| Submitted by: | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116 |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary: | September 30, 2016 |
| Contact Person and Address: | Ms. Dongeun Kim<br>Regulatory Affairs Specialist I<br>T 901-399-1151<br>F 901-566-7120 |
| Name of Device: | Smith & Nephew, Inc. Legion Cone System |
| Common Name: | Knee Prosthesis |
| Device Classification Name and<br>Reference: | 21 CFR 888.3560 Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis<br>21 CFR 888.3565 Knee joint patellofemorotibial<br>metal/polymer porous-coated uncemented prosthesis |
| Device Class: | Class II |
| Panel Code: | Orthopaedics/87 |
| Product Code: | JWH, MBH |
#### Device Description
Subject of this Traditional premarket notification is the Smith & Nephew Legion™ Cone System. The Smith & Nephew Legion Cone System consists of porous coated tibial and femoral augments to address cavitary defects in the proximal tibia or distal femur. The system will also include the necessary instrumentation for the surgeon to prepare the tibia and femur for implantation of the augments.
The components of the Smith & Nephew Legion Cone System are as follows:
- Titanium alloy (Ti-6Al-4V) tibial cones of 2 heights and 7 sizes .
- · Titanium alloy (Ti-6Al-4V) hand-specific femoral cones of 1 height and 7 sizes
The Leqion Tibial and Femoral Cones are a one piece, conically shaped device with cannulation all the way through the cone which allows fixation to the Legion Revision and Legion Hinge tibial baseplates, femoral components, offset couplers, augments/wedges, and stem constructs.
The Legion Tibial and Femoral Cones are similar to the primary predicate Zimmer Trabecular™ Metal Tibial and Femoral Cone Augments cleared via premarket notification K053340. The Legion Tibial Cones will be available in heights of 25mm (short) and 40mm (long) and inner diameters (ID) of 18mm-30mm, in 2mm increments. The Legion Femoral Cones will be available in a height of approximately 35mm and inner diameters of 18mm-30mm, in 2mm increments. The Legion Femoral Cones will be provided in left and right orientations. The subject tibial and femoral cones are manufactured from forged titanium alloy (Ti-6Al-4V) that conforms to ASTM F 1472. The subject tibial and femoral cones also have an external
{4}------------------------------------------------
asymmetric (Stiktite™) porous coating specification identical to the porous coating of the reference predicate Smith & Nephew SMF™ Hip Stems cleared via premarket notifications K080625. K103256, and K123012. The interior side of the subject cones has a grit-blasted surface specification for cement adhesion identical to the gritblasted surface of reference predicate Smith & Nephew Genesis II Total Knee System cement-on-augment devices cleared via premarket notification K951987.
#### Intended Use
The Legion Cones are intended to be used with the Legion Revision and Legion Hinge knee systems. The Legion Cones are indicated for the following:
- 1. Rheumatoid Arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
- 5. Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral ligament) are incompetent.
- 6. Hinge knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral ligament) are absent or incompetent.
These are single use implants and are intended for use with and without bone cement.
#### Technological Characteristics
Smith & Nephew has conducted a mechanical evaluation of the subject Legion Tibial and Femoral Cone Augments of the following:
- Fatigue Strength
- Push-Out Strength .
- Torsional Shear ●
- Tensile Attachment Strength o
A review of testing has demonstrated that there are no new issues related to the safety or effectiveness of the subject devices. Clinical data was not needed to support the safety and effectiveness of the subject devices.
#### Substantial Equivalence Information
The intended use, indications for use and technological characteristics of the subject Smith & Nephew Legion Tibial and Femoral Cones are substantially equivalent to the predicate devices listed in Table 5.1 below.
{5}------------------------------------------------
| Design Aspect<br>Reviewed | Legion Tibial and<br>Femoral Cones<br>(Subject Devices) | Zimmer Trabecular<br>Metal™ Tibial and<br>Femoral Cone<br>Augments (Primary<br>Predicate) | SMF Hip Stems with<br>Stiktite (Stiktite<br>Reference Predicate) | Genesis II Hemi<br>Stepped Augment<br>(Grit-Blast<br>Reference<br>Predicate) |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| 510(k) Number | None, Subject<br>Devices | K053340 | K080625, K103256,<br>K123012 | K951987 |
| Manufacturer | Smith & Nephew,<br>Inc. | Zimmer | Smith & Nephew, Inc. | Smith & Nephew, Inc. |
| Similar<br>Indications for<br>Use | Yes | Yes | Yes | Yes |
| Size Offering | Femoral Cones:<br>18mm, 20mm,<br>22mm, 24mm,<br>26mm, 28mm,<br>30mm (inner<br>diameters) in one<br>height<br>Tibial Cones: 18mm,<br>20mm, 22mm,<br>24mm, 26mm,<br>28mm, 30mm (inner<br>diameters) in<br>Short and Long | Femoral Cones: 30mm,<br>40mm, 50mm; Small,<br>Medium, and Large; M/L<br>or A/P<br>Tibial Cones: Full and<br>Stepped (15mm/30mm,<br>30mm/15mm); Sizes 48-<br>67 | -1, 0, 1, 2, 3, 4, 5, 6, 7, 8,<br>9 | 1-2, 3-4, 5-6, 7-8<br>Thickness: 10mm<br>and 15mm |
| Sterilization<br>Method | Gamma | Gamma | Gamma | Gamma |
| Material | Substrate - Titanium<br>Alloy (ASTM F1472)<br>Porous Coating -<br>C.P. Titanium<br>(ASTM F67) | Trabecular™ Metal | Substrate - Titanium Alloy<br>(ASTM F1472)<br>Porous Coating - C.P.<br>Titanium (ASTM F67) | Ti 6Al 4V<br>(ASTM F1472) |
| Grit-Blast<br>Surface Exists | Yes | No | Yes | Yes |
| Cemented or<br>Cementless | Cemented and<br>Cementless | Cemented and<br>Cementless | Cementless | Cemented |
# Table 5.1: Predicate Comparison Matrix – Legion Tibial and Femoral Cones
{6}------------------------------------------------
### Conclusion
This Traditional 510(k) Premarket Notification is being submitted to request clearance for the Smith & Nephew Legion Cone System. Based on the similarities to the predicate devices and a review of the mechanical testing performed, the subject devices are substantially equivalent to the above predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.