K162767 · First Sense Medical, LLC · LHQ · Jan 9, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K162767
Device Name
Sentinel BreastScan II System
Applicant
First Sense Medical, LLC
Product Code
LHQ · Obstetrics/Gynecology
Decision Date
Jan 9, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.2980
Device Class
Class 1
Indications for Use
The Sentinel BreastScan II System is intended for viewing and recording heat patterns generated by the human body in the hospital, acute care settings, outpatient surgery, healthcare practitioner facilities or in an environment where patient care is provided by qualified healthcare personnel. The patient populations include adult. The device is for adjunctive diagnostic screening for detection of breast cancer and diseases affecting blood perfusion or reperfusion of tissue or organs. This device is intended for use by qualified healthcare personnel trained in its use.
Device Story
Portable telethermographic system; captures thermal infrared energy (heat) emitted from human body. Input: thermal images of breasts captured via FLIR A315 camera. Process: patient sits in chair with IR-reflecting mirrors; system blows cool air on breasts during test cycle; motorized vertical alignment adjusts camera/cooling unit to patient. Output: thermal data stored in cloud for review by healthcare professional or third-party lab. Used in hospitals, acute care, outpatient surgery, or practitioner facilities by trained personnel. No breast compression; no radiation; no patient contact. Benefits: non-invasive adjunctive screening for breast cancer and perfusion-related diseases.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing confirmed system performance, thermal camera accuracy, and software functionality met all requirements. Compliance with ANSI/AAMI ES60601-1 and IEC 60601-1-2 standards demonstrated.
Technological Characteristics
Telethermographic system; FLIR A315 thermal camera (45º lens); motorized vertical alignment; air cooling unit; IR-reflecting mirrors; Wi-Fi connectivity; cloud-based data storage. Operates via non-contact infrared sensing. Complies with ANSI/AAMI ES60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for adult patients as an adjunctive diagnostic screening tool for the detection of breast cancer and diseases affecting blood perfusion or reperfusion of tissue or organs.
Regulatory Classification
Identification
A telethermographic system for adjunctive diagnostic screening for detection of breast cancer or other uses is an electrically powered device with a detector that is intended to measure, without touching the patient's skin, the self-emanating infrared radiation that reveals the temperature variations of the surface of the body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, resembling a family or group of people.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 9, 2017
First Sense Medical, LLC % John J. Smith, M.D., J.D. Regulatory Counsel Hogan Lovells US LLP 555 13th Street, NW WASHINGTON DC 20004
Re: K162767
Trade/Device Name: Sentinel BreastScan II System Regulation Number: 21 CFR 884.2980 Regulation Name: Telethermographic system Regulatory Class: I Product Code: LHQ Dated: December 14, 2016 Received: December 14, 2016
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use Statement
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement on last page |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
**Indications for Use**
| 510(k) Number ( <i>if known</i> ) | K162767 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Sentinel BreastScan II System |
| Indications for Use (Describe) | The Sentinel BreastScan II System is intended for viewing and recording heat patterns generated by the human body in the hospital, acute care settings, outpatient surgery, healthcare practitioner facilities or in an environment where patient care is provided by qualified healthcare personnel. The patient populations include adult. The device is for adjunctive diagnostic screening for detection of breast cancer and diseases affecting blood perfusion or reperfusion of tissue or organs. This device is intended for use by qualified healthcare personnel trained in its use. |
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.235.235 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including
the time to review instructions, search existing data sources, gather and maintain the data needed and
complete and review the collection of information. Send comments regarding this burden estimate or any
other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human
Services Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA)
Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
#### 510(k) SUMMARY
First Sense Medical, LLC's Sentinel BreastScan II System
#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
First Sense Medical, LLC 2001 Centerpoint Parkway, Suite 110 Pontiac, MI 48341
Phone: 248-392-3006 Facsimile: 248-876-9278
> Contact Person: Alan Klevens, CEO
Date Prepared: September 30, 2016
#### Name of Device and Name/Address of Sponsor
Sentinel BreastScan II System
First Sense Medical, LLC 2001 Centerpoint Parkway, Suite 110 Pontiac, MI 48341
#### Common or Usual Name
Telethermographic System (Adjunctive Use)
#### Classification Name
Telethermographic System (Adjunctive Use) Product Code: LHQ 21 C.F.R. 884.2980
#### Predicate Devices
Infrared Sciences Corp., BreastScan IR (K032350)
#### Purpose of the Special 510(k) Notice
The Sentinel BreastScan II System is a modification to the Infrared Sciences Corp. BreastScan IR system.
{4}------------------------------------------------
### Device Description
The Sentinel BreastScan II System consists of a portable device that captures and records thermal infrared energy (heat) emitting from a person's body. There is no compression of the breast or patient contact with the device and the test emits no radiation to the patient. The device consists of a thermal camera, a metal enclosure to secure and protect components, a motor and electronics to raise and lower the tester to adjust the camera for various sized patients, a chair with attached armrests, adjustable special heat reflecting mirrors, a computer with a touchscreen monitor, data storage, software, and Wi-Fi communication. The tester includes an air cooling unit that blows cool air during part of the test cycle.
## Intended Use / Indications for use
The Sentinel BreastScan II System is intended for viewing and recording heat patterns generated by the human body in the hospital, acute care settings, outpatient surgery, healthcare practitioner facilities or in an environment where patient care is provided by qualified healthcare personnel. The patient populations include adult. The device is for adjunctive diagnostic screening for detection of breast cancer and diseases affecting blood perfusion of tissue or organs. This device is intended for use by qualified healthcare personnel trained in its use.
### Summary of Technological Characteristics
Telethermography is the technological principle for both the subject and predicate devices. Using an advanced non-contact infrared thermal camera, the Sentinel BreastScan II system records thermal data (emission of body temperature) from a patient's breasts. During the procedure, the patient sits facing a thermal camera, cool air is blown on the patient's breasts for part of the test and additional thermal data is recorded. The data is verified for errors, securely stored and made available to the healthcare professional or a third party reference lab for later processing and evaluation. The test procedure does not involve any compression or touching of the breasts and the patient has no contact with the tester.
At a high level, the subject and predicate devices are based on the following same technological elements:
- . Advanced digital Thermal Camera (FDA cleared).
- · The Sentinel BreastScan II uses a FLIR A315 camera with a 45º lens at 2.5 ft. from the patient. The BreastScan IR uses a FLIR S40 camera with a 25º lens at 5 ft. from the patient. The FOV is the same for both cameras at patient chest.
- Test Sequence is the same for capturing thermal data and cooling patient. .
- . Tester housing contains computer, monitor(s), thermal camera, software and air cooling system equipment.
- . Patient Chair with attached special (IR reflecting) mirrors to capture side images of breasts and custom armrests for proper patient positioning.
- Air Cooling System blows cool air onto patient breasts during test. .
{5}------------------------------------------------
The following technological differences exist between the subject and predicate devices:
- The Sentinel BreastScan II System Tester housing is custom designed and includes . motorized vertical travel for alignment of thermal camera and air cooling system with the patient. The BreastScan IR System Tester housing is a standard commercial cart containing the tester components at a fixed camera height. The vertical alignment of the patient is achieved by an adjustable chair.
- The Sentinel BreastScan II chair is a fixed height chair with vertically adjustable . mirrors and armrests. The BreastScan IR chair is vertically adjustable with fixed mirrors and armrests.
- . The Sentinel BreastScan II stores data securely in the cloud. The BreastScan IR stores data locally on the computer.
# Performance Data
The Sentinel BreastScan II System was tested to and complies with the following standards:
- ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 . or IEC 60601-1 ed. 3.1 2012-08.
- [IEC] [60601-1-2:2007] Medical Electrical Equipment Part 1-2 Requirements for Basic . Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements and Tests} [2007}
The performance testing conducted for the thermal camera, system performance monitoring, software and the device functionality as a whole demonstrates that the device meets all Sentinel BreastScan II requirements and specifications as provided in the verification and validation testing results reports. In all instances, the Sentinel BreastScan II functioned and the results observed and reported were as expected.
## Conclusions
The Sentinel BreastScan II System has the same intended use and indications, principles of operation, and technological characteristics as the Infrared Sciences Corp. BreastScan IR. The minor differences in the Sentinel BreastScan II System's technological characteristics do not raise any new questions of safety or effectiveness. Performance data demonstrates that the Sentinel BreastScan II System is as safe and effective as the Infrared Sciences Corp., BreastScan IR. Thus, the Sentinel BreastScan II System is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.