Medline ReNewal Reprocessed Ligasure Curved, Small Jaw Open Sealer/Divider
K162751 · Surgical Instrument Services and Savings, Inc. · NUJ · May 1, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162751
Device Name
Medline ReNewal Reprocessed Ligasure Curved, Small Jaw Open Sealer/Divider
Applicant
Surgical Instrument Services and Savings, Inc.
Product Code
NUJ · General, Plastic Surgery
Decision Date
May 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medline ReNewal Reprocessed LigaSure Curved Small Jaw, Open Sealer/Divide LF1212A is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic and reconstructive. Procedures may include, but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis. The instrument is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, parotidectomy, and tonsillectomy) for ligation and division of vessels, lymphatics and tissue bundles 2 - 3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands. The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
Device Story
Reprocessed single-use bipolar electrosurgical instrument; used in open surgical procedures to ligate and divide vessels, tissue bundles, and lymphatics. Surgeon grasps tissue with jaws; initiates RF energy via hand switch or footswitch; generator delivers energy for controlled tissue fusion; surgeon actuates blade to divide tissue along seal line. Used in OR/clinic settings by surgeons. Benefits include permanent tissue fusion with minimal thermal spread, sticking, or charring. Device is reprocessed version of original Covidien LigaSure LF1212A; cord/connector/generator not reprocessed.
Bipolar electrosurgical instrument; reprocessed single-use. Energy source: compatible electrosurgical generator (e.g., ForceTriad). Dimensions/form factor: curved, small jaw. Connectivity: cable connector to generator. Sterilization: validated re-sterilization process. No software algorithm internal to the instrument; relies on external generator software.
Indications for Use
Indicated for open surgical procedures requiring ligation and division of vessels (up to 7 mm), tissue bundles, and lymphatics. Used in general, urologic, thoracic, plastic, and reconstructive surgery. Also indicated for adult open ENT procedures (thyroidectomy, radical neck dissection, parotidectomy, tonsillectomy) requiring 2-3 mm clearance from thermally-sensitive structures. Contraindicated for tubal coagulation/sterilization.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
LigaSure Curved, Small Jaw, Open Sealer/Divider LF1212A (K152286)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus-like symbol with three intertwined figures, representing health and well-being. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 1, 2017
Surgical Instrument Services and Savings, Inc. Ms. Brandi Panteleon Director, Quality Assurance and Regulatory Affairs 2747 SW 6th St. Redmond, Oregon 97756
Re: K162751
Trade/Device Name: Medline ReNewal Reprocessed LigaSure Curved, Small Jaw Open Sealer/Divider LF1212A Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II
Product Code: NUJ Dated: April 19, 2017 Received: April 21, 2017
Dear Ms. Panteleon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely, Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Reprocessed Single-Use Device Models Subject to Clearance:
| Model<br>Number | Device Name/Description | Original Manufacturer |
|-----------------|-------------------------------------------------|-----------------------|
| LF1212A | LigaSure Curved, Small Jaw, Open Sealer/Divider | Covidien |
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162751
Device Name
Medline ReNewal Reprocessed LigaSure Curved, Small Jaw Open Sealer/Divider LF1212A
### Indications for Use (Describe)
The Medline ReNewal Reprocessed LigaSure Curved Small Jaw, Open Sealer/Divide LF1212A is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgical specialties as urologic, thoracic, plastic and reconstructive. Procedures may include, but are not limited to, bowel resections, gall bladder procedures. Nissen fundoplication, and adhesiolysis.
The instrument is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, parotidectory, and tonsillectorny) for ligation and division of vessels, lymphatiss and tissue bundles 2 - 3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands.
The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the Medline Renewal logo. The word "MEDLINE" is in blue, with a blue star-like symbol to the right. To the right of that is a vertical line, followed by the word "Renewal" in green and blue. Underneath "Renewal" is the phrase "Full Circle Reprocessing" in a smaller font.
# K162751 Summary
| Submitter/<br>Owner | Medline ReNewal<br>2747 SW 6th St.<br>Redmond, OR 97756 |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name | Brandi Panteleon<br>Director, Regulatory Affairs<br>P: 541-516-4180<br>F: 541-923-3375<br>E: bpanteleon@medline.com |
| Prepared by | Stephanie Boyle Mays<br>Requlatory Affairs Specialist, Requlatory Affairs<br>P: 541-516-4205<br>F: 541-923-3375<br>E: smays@medline.com |
| Date Prepared | September 29, 2016 |
| Device Names | Proprietary Name: Medline ReNewal Reprocessed LigaSure Curved, Small<br>Jaw Open Sealer/Divider LF1212A<br>Common Name: Bipolar electrosurgical instrument |
| Classification | Electrosurgical cutting and coagulation device and accessories<br>reprocessed<br>Product code: NUJ<br>Class: Class II, non-exempt |
| Predicate<br>Device | K152286 LigaSure Curved, Small Jaw, Open Sealer/Divider LF1212A |
| Device<br>Description | The Medline ReNewal Reprocessed LigaSure Curved, Small Jaw, Open<br>Sealer/Divider LF1212A is a sterile, single-use, hand-held bipolar<br>electrosurgical instrument designed for use with compatible Covidien<br>generators that include vessel sealing capabilities to ligate (seal) and divide<br>(cut) vessels, tissue bundles and lymphatics during open general surgical<br>procedures. Covidien electrosurgical generators that include vessel sealing<br>capabilities deliver precise energy through the device to tissue for a<br>controlled tissue response to achieve complete and permanent tissue<br>fusion while producing minimal sticking, charring, and thermal spread to<br>adjacent tissue.<br>The surgeon first grasps the vessel or tissue with the jaws of the<br>instrument. Once the surgeon is confident that the instrument is placed<br>correctly, he or she initiates a seal by pressing a hand switch or footswitch.<br>The instrument creates a seal by application of RF electrosurgical energy to<br>vascular structures (vessels and lymph) or tissue bundles interposed<br>between the jaws of the instrument. Once the seal cycle is complete, the<br>surgeon actuates a blade within the instrument to divide tissue along the<br>seal line.<br>The 10-foot cord, cable connector, and generator are not reprocessed by<br>Medline ReNewal. |
| Intended Use | The Medline ReNewal Reprocessed LigaSure Curved Small Jaw, Open<br>Sealer/Divide LF1212A is a bipolar electrosurgical instrument intended for<br>use in open surgical procedures where ligation and division of vessels,<br>tissue bundles, and lymphatics is desired. The LigaSure Sealer/Divider can<br>be used on vessels (arteries and veins) up to and including 7 mm. It is<br>indicated for use in general surgery and in such surgical specialties as<br>urologic, thoracic, plastic and reconstructive. Procedures may include, but<br>are not limited to, bowel resections, gall bladder procedures, Nissen<br>fundoplication, and adhesiolysis.<br>The instrument is also indicated for open ENT procedures in adults<br>(thyroidectomy, radical neck dissection, parotidectomy, and tonsillectomy)<br>for ligation and division of vessels, lymphatics and tissue bundles 2 - 3 mm<br>away from unintended thermally-sensitive structures such as nerves and<br>parathyroid glands.<br>The LigaSure system has not been shown to be effective for tubal<br>sterilization or tubal coagulation for sterilization procedures. Do not use the<br>LigaSure system for these procedures. |
| Technological<br>Characteristics | The technological characteristics and the fundamental scientific technology<br>of the subject devices are identical to the predicate device. The proposed<br>devices are a reprocessed version of the predicate devices. K152286<br>LigaSure Curved, Small Jaw, Open Sealer/Divider LF1212A was used as<br>the primary predicate to support intended use, technological characteristics,<br>and functional performance specifications. |
| Performance<br>Testing | The functional characteristics of the proposed devices have been evaluated<br>and found to be equivalent to the predicate devices based on the following<br>tests:<br>electrical safety and electromagnetic compatibility in accordance<br>with IEC 60601-2-2 and 60601-1-2; simulated use; device integrity; blade trigger advance/return; activation button device recognition; thermal analysis characterization; tissue sticking; blade sharpness; transient cooling analysis burst pressure; histopathology; seal quality; Cleaning: protein, carbohydrates, and endotoxin; Biocompatibility: cytotoxicity sensitization, irritation; pyrogenicity, and acute systemic toxicity; performance qualification. |
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image contains the Medline Renewal logo. The Medline logo is on the left, and the Renewal logo is on the right. The Renewal logo is green and blue, and the text "Full Circle Reprocessing" is below the Renewal logo.
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the Medline Renewal logo. The Medline logo is on the left, and the Renewal logo is on the right. The Renewal logo is green and blue, and the words "Full Circle Reprocessing" are below it.
sterilization validation; and ● ● product stability. Based on comparisons of the indications for use, intended use, technological characteristics, and performance data to the predicate Conclusion devices, Medline ReNewal reprocessed Medline ReNewal LigaSure Impact, Curved, Large Jaw, Open Sealer/Divider (LF1212A) is substantially equivalent to the predicate devices.
{7}------------------------------------------------
K162751
Image /page/7/Picture/2 description: The image features the word "ReNewal" in a vertically stacked format, with the "Re" in green and the "Newal" in blue. To the right of "ReNewal" is the phrase "Full Circle Reprocessing" in a smaller, vertically stacked font. Below "ReNewal" is the word "MEDLINE" in blue, accompanied by a stylized logo. The logo consists of a blue star-like shape with an arrow pointing to the right, suggesting a sense of direction or progress.
# Predicate and Medline ReNewal Reprocessed LigaSure Curved, Small Jaw Open Sealer/Divider (LF1212A) device
comparison chart. Table 1:
| Device<br>Characteristics | Predicate | Proposed | Comparison |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Predicate 510(k) | K152286 | K162751 | N/A |
| Model Numbers | LF1212A | LF1212A | N/A |
| Intended<br>Use/Indications for<br>Usea | The Covidien LigaSure Curved Small Jaw, Open<br>Sealer/Divide model LF1212A is a bipolar<br>electrosurgical instrument intended for use in<br>open surgical procedures where ligation and<br>division of vessels, tissue bundles, and lymphatics<br>is desired. The LigaSure Sealer/Divider can be<br>used on vessels (arteries and veins) up to and<br>including 7 mm. It is indicated for use in general<br>surgery and in such surgical specialties as<br>urologic, thoracic, plastic and reconstructive.<br>Procedures may include, but are not limited to,<br>bowel resections, gall bladder procedures, Nissen<br>fundoplication, and adhesiolysis.<br>The instrument is also indicated for open ENT<br>procedures in adults (thyroidectomy, radical neck<br>dissection, parotidectomy, and tonsillectomy) for<br>ligation and division of vessels, lymphatics and<br>tissue bundles 2 – 3 mm away from unintended<br>thermally-sensitive structures such as nerves and<br>parathyroid glands. | The Medline ReNewal Reprocessed LigaSure<br>Curved Small Jaw, Open Sealer/Divide model<br>LF1212A is a bipolar electrosurgical instrument<br>intended for use in open surgical procedures<br>where ligation and division of vessels, tissue<br>bundles, and lymphatics is desired. The LigaSure<br>Sealer/Divider can be used on vessels (arteries<br>and veins) up to and including 7 mm. It is<br>indicated for use in general surgery and in such<br>surgical specialties as urologic, thoracic, plastic<br>and reconstructive. Procedures may include, but<br>are not limited to, bowel resections, gall bladder<br>procedures, Nissen fundoplication, and<br>adhesiolysis.<br>The instrument is also indicated for open ENT<br>procedures in adults (thyroidectomy, radical neck<br>dissection, parotidectomy, and tonsillectomy) for<br>ligation and division of vessels, lymphatics and<br>tissue bundles 2 - 3 mm away from unintended<br>thermally-sensitive structures such as nerves and<br>parathyroid glands. | Same device;<br>different<br>manufacturer |
| Device<br>Characteristics | Predicate<br>Covidien LigaSure Impact Curved, Large Jaw<br>Sealer/Divider | Proposed<br>Medline ReNewal Reprocessed LigaSure<br>Impact Curved, Large Jaw Sealer/Divider | Comparison |
| Intended<br>Use/Indications for<br>Use (concluded)a | The LigaSure system has not been shown to be<br>effective for tubal sterilization or tubal<br>coagulation for sterilization procedures. Do not<br>use the LigaSure system for these procedures. | The LigaSure system has not been shown to be<br>effective for tubal sterilization or tubal<br>coagulation for sterilization procedures. Do not<br>use the LigaSure system for these procedures.<br>parathyroid glands. | Same device;<br>different<br>manufacturer |
| Power Platform | ForceTriad Electrosurgical Generator running<br>software version 3.40 or higher.b | ForceTriad Electrosurgical Generator running<br>software version 3.40 or higher.b | Same |
| Technological<br>Characteristics | The LigaSure Impact Curved Large Jaw, Open<br>Sealer/Divider uses bipolar energy to seal and<br>divide vessels up to and including 7 mm in<br>diameter. | The LigaSure Impact Curved Large Jaw, Open<br>Sealer/Divider uses bipolar energy to seal and<br>divide vessels up to and including 7 mm in<br>diameter. | Same |
| | a Intended Use and Indications for Use were the same category in the K1522864 predicate.<br>b Force Triad generator was cleared under K070162: It will not be reprocessed and it is not part of this submission | | |
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the Medline Renewal logo. The word "Renewal" is written vertically in a combination of green and blue colors. Below the word "Renewal" is the phrase "Full Circle Reprocessing" in a smaller, blue font. To the left of the word "Renewal" is the Medline logo.
- Predicate and: Medline ReNewal Reprocessed LigaSure Impact Curved, Large Jaw Sealer/Divider (LF1212A) device compariso
chart (concluded). Table 1:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.