K162743 · GE Medical Systems Ultrasound and Primary Care Diagnostics · LLZ · Nov 22, 2016 · Radiology
Device Facts
Record ID
K162743
Device Name
ViewPoint 6
Applicant
GE Medical Systems Ultrasound and Primary Care Diagnostics
Product Code
LLZ · Radiology
Decision Date
Nov 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
ViewPoint 6 is intended to be used in medical practices and in clinical departments and serves the purposes of diagnostic interpretation of images, electronic documentation of examinations in the form of text and images, and generation of medical reports primarily for diagnostic ultrasound. ViewPoint 6 provides the user the ability to include images, drawings, and charts into medical reports. ViewPoint 6 is designed to accept, transfer, display, calculate, store, and process medical images and data, and enables the user to measure and annotate the images. The medical images, which 6 displays to the user, can be used for diagnostic purposes. ViewPoint 6 is intended for professional use only. ViewPoint 6 is not intended to be used as an automated diagnosis system. ViewPoint 6 is not intended to operate medical devices in surgery related procedures.
Device Story
ViewPoint 6 is a PACS software for diagnostic interpretation of medical images and reporting. It accepts, transfers, displays, calculates, stores, and processes medical images and data. Users (clinicians) perform measurements and annotations on images to generate reports. It supports single-workstation or client-server network configurations. Features include exam-specific reporting forms, image review/comparison, clinical scheduling, resource management, and HL7 order management. It is not an automated diagnosis system and does not operate surgical devices. It is distributed via a cloud-based solution with smartcard authentication and encrypted database storage. It aids clinical decision-making by providing standardized terminology and organized documentation for diagnostic ultrasound.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including verification and validation of software modules, system integration, performance, and safety testing. Compliance with IEC 62366-1, IEC 62304, NEMA PS 3.1-3.20 (DICOM), and ISO 14971.
Technological Characteristics
Software-based PACS; runs on Microsoft Windows; supports single-workstation or client-server network architecture. Connectivity via DICOM and HL7 standards. Features include encrypted database, smartcard authentication, and cloud-based distribution. Complies with IEC 62304 (software lifecycle) and IEC 62366-1 (usability).
Indications for Use
Indicated for professional use in medical practices and clinical departments for diagnostic interpretation of images, electronic documentation, and generation of medical reports, primarily for diagnostic ultrasound.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 22, 2016
GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC % Ms. Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226
Re: K162743
Trade/Device Name: ViewPoint 6 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: September 28, 2016 Received: September 30, 2016
Dear Ms. Ortiz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' intertwined within a blue circle. The 'G' and 'E' are stylized and connected, creating a distinctive monogram. The blue color of the circle and letters is consistent, providing a clean and recognizable brand identity.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|------------------------------------------------------------------------------------------------|---------|---------------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | K162743 | |
Device Name ViewPoint 6
Indications for Use (Describe)
ViewPoint 6 is intended to be used in medical practices and in clinical departments and serves the purpostic interpretation of images, electronic documentations in the form of text and images, and generation of medical reports primarily for diagnostic ultrasound.
ViewPoint 6 provides the user the ability to include images, drawings, and charts into medical reports. ViewPoint 6 is designed to accept, transfer, display, calculate, store, and process medical images and data, and enables the user to measure and annotate the images. The medical images, which 6 displays to the user, can be used for diagnostic purposes.
ViewPoint 6 is intended for professional use only. ViewPoint 6 is not intended to be used as an automated diagnosis system.
ViewPoint 6 is not intended to operate medical devices in surgery related procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
EF PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. There are also some white swirls around the letters.
### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided: September 28, 2016 Date: GE Medical Systems Ultrasound and Primary Care Diagnostics, Submitter: 9900 Innovation Drive Wauwatosa, WI 53226
- Primary Contact Person: Tracey Ortiz Regulatory Affairs Director GE Healthcare T:(262)676-6120 F:(414)918-8275
- Secondary Contact Nadine Gebhard- Höflinger Requlatory Affairs Person: GE Healthcare Austria GmbH & Co OG
| Trade Name: | ViewPoint 6 |
|-----------------------|----------------------------------------------------|
| Common/Usual Name: | PACS / Picture archiving and communications system |
| Classification Names: | Class II |
- Product Code: Picture archiving and communications system, 21 CFR 892.2050, LLZ
- Primary Predicate ViewPoint 6 (K103458) Device(s):
- Reference Predicate none Device(s):
ViewPoint 6 is an image archiving and reporting software for Device Description: medical practices and clinical radiological departments. It is used for diagnostic interpretation of images and other data.
> ViewPoint 6 is for professional use only and enables quick diagnostic reporting with standardized terminology. It has an intuitive graphical user interface (GUI) and is based on Microsoft Windows® with defined hardware requirements for the user to install on their computer.
ViewPoint 6 provides exam type specific reporting forms for
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has decorative swirls or flourishes around the letters. The logo is presented in a light blue color.
various medical care areas. Forms are composed of different sections with data entry fields. The documentation can include measurements, exam findings, images, and graphs. All data is saved in the ViewPoint 6 database and can be compiled to a professional report. Images and image sequences can be reviewed in the ViewPoint 6 display area based on user preference.
ViewPoint 6 supports both a single workstation and a clientserver setup. The number of user licenses determines how many workstations in the network have concurrent access to the database. Access can be limited to read-only functionality.
#### Intended Use/ Indication for Use:
## Indications for Use of ViewPoint 6
ViewPoint 6 is intended to be used in medical practices and in clinical departments and serves the purposes of diagnostic interpretation of images, electronic documentation of examinations in the form of text and images, and generation of medical reports primarily for diagnostic ultrasound.
ViewPoint 6 provides the user the ability to include images. drawings, and charts into medical reports. ViewPoint 6 is designed to accept, transfer, display, calculate, store, and process medical images and data, and enables the user to measure and annotate the images. The medical images, which ViewPoint 6 displays to the user, can be used for diagnostic purposes.
ViewPoint 6 is intended for professional use only. ViewPoint 6 is not intended to be used as an automated diagnosis system. ViewPoint 6 is not intended to operate medical devices in surgery related procedures.
- Technology: ViewPoint 6 employs the same fundamental scientific technology as its predicate device.
#### Determination of Comparison to Predicates
Substantial Equivalence:
The proposed ViewPoint 6 is substantially equivalent to the predicate ViewPoint 6 with regards to intended use. capabilities, technological characteristics, safety and effectiveness. There was no change in the indication for use.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo, which is a stylized monogram of the letters 'g' and 'e' enclosed within a circular shape. The monogram is rendered in a cursive style, with the 'g' and 'e' intertwined. The logo is colored in a light blue hue, and the circular shape has a double-line border, also in the same blue color.
Following changes were done compared to the predicate device:
- Distribution changed from physical DVD to an access controlled cloud based solution (e-distribution)
- Labelling updates made as needed to alian with the o product changes
- Expanded capabilities for image review/compare, ● measure, annotate, and adjust
- o Added organization features – auto log off tracking, smartcard authentication, clinical scheduler, resource manager, spell check, code lists for billing, HL7 order management
- Added options for export of data and reports with an o encrypted database
- Added adaptable lighting environment setting ( lightson/lights-off mode)
- Added workflow features for exam type definition, ● exam auto lock, multi-monitor support and new report forms
# Summary of Non-Clinical Tests:
The ViewPoint 6 and its applications comply with voluntary standards:
- 1. IEC 62366-1:2015 Medical devices Application of usability engineering to medical devices
- 2. IEC 62304:2006, Medical device software Software life cycle process
- 3. NEMA PS 3.1 3.20 (2016), Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology)
- 4. ISO 14971:2012 Medical Devices Application of risk management to medical devices
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are decorative swirls around the outer edge of the circle, giving it a classic and recognizable appearance.
The following quality assurance measures are applied to the development of the system:
- Risk Analysis
- Requirements Reviews ●
- . Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification)
- Final Acceptance Testing (Design Validation and ● Service Validation)
- o Performance testing (Verification)
- Safety testing (Verification) ●
## Summary of Clinical Tests:
The subject of this premarket submission, ViewPoint 6, did not require clinical studies to support substantial equivalence.
- GE Healthcare considers the proposed ViewPoint 6 to be as Conclusion: safe, as effective, and performance is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.