The RehaStim 2 stimulator is intended for general rehabilitation 1. Relaxation of muscle spasms 2. Prevention or retardation of disuse atrophy 3. Increasing local blood circulation 4. Maintaining or increasing range of motion The RehaStim 2 is an External Functional Neuromuscular Stimulator, designed for use as stand-alone device as well as in combination with exercise equipment for walking and cycling applications, that has been validated for compatibility by HASOMED GmbH and should be used under the supervision of a clinician.
Device Story
RehaStim 2 is an 8-channel external functional neuromuscular stimulator (FES). It functions as a stand-alone device or integrates with external motion trainers (cycling/walking) via a serial interface (USB/FTDI). The device receives motion data (gait cycle percentage) from the exerciser every 10ms; it uses this data to synchronize electrical stimulation pulses to specific muscle groups based on the movement phase. Stimulation parameters are configured within the stimulator. Operated under clinician supervision in clinical settings. Output consists of charge-balanced rectangular electrical impulses (0-130mA, max 500µs pulse width, max 50Hz frequency) delivered via skin electrodes. Benefits include muscle spasm reduction, atrophy prevention, improved circulation, and range of motion maintenance. The device uses a custom controller running embedded Linux.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, verification, and validation of hardware and software, and comparison of technical specifications against predicate devices.
Technological Characteristics
8-channel FES; 0-130mA output; max 50Hz; max 500µs pulse width; Type BF applied part. Power: AC (Cincon TR30M, EN60601-1) or Li-Ion battery (BMZ18650). Connectivity: Serial interface (USB/FTDI) for motion data. Controller: Embedded Linux. Electrodes: HASOMED RehaTrodes. Sterilization: Not specified.
Indications for Use
Indicated for general rehabilitation in patients requiring relaxation of muscle spasms, prevention/retardation of disuse atrophy, increased local blood circulation, or maintenance/increase of range of motion. Used under clinician supervision.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular arrangement of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a symbol of three stylized human profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 3, 2017
HASOMED Gmbh Matthias Ueltzen Quality Management & Regulatory Affairs Paul-Ecke-Str. 1 Magdeburg, SA 39114 Germany
Re: K162683
Trade/Device Name: RehaStim 2 Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI Dated: December 22, 2016 Received: December 27, 2016
Dear Mr. Ueltzen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162683
Device Name RehaStim 2
Indications for Use (Describe) The RehaStim 2 stimulator is intended for general rehabilitation
- 1. Relaxation of muscle spasms
- 2. Prevention or retardation of disuse atrophy
- 3. Increasing local blood circulation
- 4. Maintaining or increasing range of motion
The RehaStim 2 is an External Functional Neuromuscular Stimulator, designed for use as stand-alone device as well as in combination with exercise equipment for walking and cycling applications, that has been validated for compatibility by HASOMED GmbH and should be used under the supervision of a clinician.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### l. SUBMITTER
| 510k Submitter: | Hasomed GmbH |
|-----------------|---------------------------------------------------|
| Contact person: | Matthias Weber |
| Address: | Paul-Ecke-Strasse 1<br>39114 Magdeburg<br>Germany |
| Phone: | +493916230112 |
| Fax: | +493916230113 |
Prepared 2016-09-22, revised 2016-11-22, 2017-01-20, 2017-01-26, and 2017-02-02.
#### II. DEVICE
## Modification to an existing device
| Proprietary name: | RehaStim 2 |
|----------------------|---------------------------|
| Common Name: | Powered Muscle Stimulator |
| Classification Name: | Powered Muscle Stimulator |
| Device Class: | class 2 device |
| Classification: | Neurology |
| Product code: | GZI |
#### III. PREDICATE DEVICE
## Device 1 – primary predicate
RehaStim 2; Proprietary name:
{4}------------------------------------------------
| | RehaMove 2 (RehaStim 2 with movement exerciser) | |
|-----------------|-------------------------------------------------|--|
| 510(k): | K112844 | |
| Device Class: | class 2 device | |
| Classification: | Neurology | |
| Product code: | GZI | |
| Common Name: | Powered Muscle Stimulator | |
## Device 2
| Proprietary name: | RehaStim 2; |
|-------------------|---------------------------|
| | ErigoPro (FES) |
| 510(k): | K132416 |
| Device Class: | class 2 device |
| Classification: | Neurology |
| Product code: | GZI, BXB, INQ |
| Common Name: | Powered Muscle Stimulator |
#### IV. DEVICE DESCRIPTION
RehaStim 2 is a 8 channel External Functional Neuromuscular Stimulator. RehaStim 2 can be used as a stand-alone device or in combination with an external motion trainer for cycling or walking applications, that has been validated for compatibility by HASOMED GmbH and should be used under the supervision of a clinician.
A connection cable enables the RehaStim 2 to receive motion data from the external movement exerciser device for walking or cycling applications in order to apply electrical stimulation synchronized with the movement. The stimulation is performed according to the motion data. The stimulation parameters can be configured inside the stimulator.
Electrode lead wires are used to connect RehaStim2 output and electrodes on skin.
{5}------------------------------------------------
#### V. INDICATIONS FOR USE
The RehaStim 2 stimulator is intended for general rehabilitation for:
- 1. Relaxation of muscle spasms
- 2. Prevention or retardation of disuse atrophy
- 3. Increasing local blood circulation
- 4. Maintaining or increasing range of motion
The RehaStim 2 is an External Functional Neuromuscular Stimulator, designed for use as stand-alone device as well as in combination with exercise equipment for walking and cycling applications, that has been validated for compatibility by HASOMED GmbH and should be used under the supervision of a clinician.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The functions of the RehaStim 2 (2016) are the same as the predicate devices ErigoPro (FES), RehaMove 2, and RehaStim 2 remained the same, however there are certain technological similarities and differences as described below:
| Technology | RehaStim 2<br>(2016) | RehaStim 2 /<br>ErigoPro (FES) | RehaStim 2 /<br>RehaMove 2 |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | K132416 | K112844<br>primary predicate |
| Power Source | AC and/or storage<br>battery:<br>Power supply:<br>cincon tr30m ac-<br>cording to<br>EN60601-1<br>Battery:<br>BMZ18650 V, Li-<br>Ion Mangan cells,<br>2S1P, C= 1600<br>mAh, | AC:<br>ErigoPro:<br>Total System: 110-<br>230V 800VA<br>Includes<br>RehaStim2 power<br>supply: cincon<br>tr30m according to<br>EN60601-1 | AC and/or storage<br>battery:<br>Power supply:<br>cincon tr30m ac-<br>cording to<br>EN60601-1<br>Battery:<br>BMZ18650 V, Li-<br>Ion Mangan cells,<br>2S1P, C= 1600<br>mAh, |
| | | RehaStim2 (only):<br>AC: power supply:<br>cincon tr30m ac-<br>cording to<br>EN60601-1<br>Battery:<br>BMZ18650 V, Li-<br>Ion Mangan cells,<br>2S1P, C= 1600<br>mAh, | |
| Power input<br>(charging) | 9V, 3A, max. 30 W | RehaStim2: 9V,<br>3A, max. 30 W | 9V, 3A, max. 30 W |
| Operation time | ca. 90 min | ErigoPro: No<br>limitation by<br>battery<br>RehaStim2: ca. 90<br>min | ca. 90 min |
| Charging time | ca. 180 min | ErigoPro: N/A<br>RehaStim2: ca.<br>180 min | ca. 180 min |
| Controller | uses custom<br>controller<br>using embedded<br>Linux as basis<br>operating system,<br>running custom<br>software | uses custom<br>controller<br>using embedded<br>Linux as basis<br>operating system,<br>running custom<br>software<br>ErigoPro:<br>RehaStim2 is<br>controlled by | uses custom<br>controller<br>using embedded<br>Linux as basis<br>operating system,<br>running custom<br>software |
| | | ErigoPro Control<br>Panel | |
| Stimulator<br>(energy delivered) | 0-130mA charge<br>balanced<br>stimulator with<br>rectangular<br>impulses / same<br>hardware | 0-130mA charge<br>balanced<br>stimulator with<br>rectangular<br>impulses / same<br>hardware | 0-130mA charge<br>balanced<br>stimulator with<br>rectangular<br>impulses / same<br>hardware |
| Pulse width | max. 500 µs | max. 500 µs | max. 500 µs |
| Stimulation<br>frequency | max. 50 Hz | max. 50 Hz | max. 50 Hz |
| Number of<br>channels | 8 | 8 | 8 |
| Electrodes | HASOMED's<br>RehaTrodes only | HASOMED's<br>RehaTrodes only<br>ErigoPro:<br>HASOMED's<br>electrodes with<br>"distributed by<br>Hocoma" label | HASOMED's<br>RehaTrodes only |
| FES00200<br>RehaTrode | 2.0" x 3.5",<br>rectangle | 2.0" x 3.5",<br>rectangle | 2.0" x 3.5",<br>rectangle |
| FES00201<br>RehaTrode | 3.0" x 5.0",<br>rectangle | 3.0" x 5.0",<br>rectangle | 3.0" x 5.0",<br>rectangle |
| FES00202<br>RehaTrode | 1.5" x 2.5" oval | 1.5" x 2.5" oval | 1.5" x 2.5" oval |
| Applied part | Type BF | Type BF | Type BF |
| Movement | RehaStim 2 (2016) | ErigoPro: uses | RehaMove: Uses |
| exerciser | is intended for<br>combination with<br>different movement<br>exercisers for<br>walking and<br>cycling<br>applications | motor to robotic leg<br>movement and<br>cyclic leg loading<br>for early functional<br>mobilization | motor to create<br>flywheel effect with<br>reduced weight<br>and space |
| Communication<br>interface | FES 2 + FES 3;<br>FES 3-<br>bidirectional<br>communication,<br>controlling of<br>movement<br>exerciser<br>functionality<br>interface allows for<br>combination with<br>different devices | RehaStim2:<br>FES 2 + FES 3;<br>FES 3-<br>bidirectional<br>communication,<br>controlling of<br>movement<br>exerciser<br>functionality<br>ErigoPro: RS2-<br>ErigoMode -<br>bidirectional<br>communication,<br>controlling of FES,<br>synchronized with<br>the ErigoPro<br>device | FES 2 + FES 3;<br>FES 3-<br>bidirectional<br>communication,<br>controlling of<br>movement<br>exerciser<br>functionality |
| Control philosophy<br>of matching the<br>combined devices | Complete<br>stimulation control<br>implemented in<br>RehaStim 2,<br>movement<br>exerciser delivers<br>motion data for<br>synchronization of<br>stimulation and<br>movement | Parts of stimulation<br>control<br>implemented in<br>Erigo device | Complete<br>stimulation control<br>implemented in<br>RehaStim 2,<br>movement<br>exerciser delivers<br>motion data for<br>synchronization of<br>stimulation and<br>movement |
| Patient position | As usual in the<br>movement<br>exerciser device<br>combined with | Verticalization<br>table for early<br>rehabilitation | Allows user to<br>remain in their own<br>seating, e.g.<br>wheelchair<br>eliminating the<br>need for transfer |
| Passive cycling | Exercise<br>equipment: Utilizes<br>motor to provide<br>assistance during<br>passive walking or<br>cycling | ErigoPro (FES):<br>Utilizes motor for<br>robotic leg<br>movement and<br>cyclic leg loading | RehaMove:<br>Utilizes motor to<br>provide assistance<br>during passive<br>cycling |
| Database interface | Utilizes database<br>interface for<br>storage and<br>retrieval of patient<br>therapy settings<br>and storage of<br>session logs | Utilizes database<br>interface for<br>storage and<br>retrieval of patient<br>therapy settings<br>and storage of<br>session logs | Utilizes database<br>interface for<br>storage and<br>retrieval of patient<br>therapy settings<br>and storage of<br>session logs |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
#### Table 1 Technological similarities and differences
The difference is in the control philosophy of the device combinations: In RehaStim 2 / RehaMove 2 the complete stimulation know-how is in the controller of the stimulation device RehaStim 2. In ErigoPro (FES) some of the stimulation control is implemented in the Erigo control. In the new device, stimulation control is completely implemented in the stimulation device and an interface is used for input of motion data into the stimulator. The stimulation is performed according to the motion data. The motion data are delivered via a serial interface (USB, FTDI) as percentage of full gait cycle at a rate of about 10 ms from the exerciser to the stimulator. A proprietary protocol is used. Based on this gait information, stimulation is activated at pre-defined values to the muscles needed according to the actual phase of gait. A proprietary testing protocol conducted by Hasomed GmbH is being used for validation and verification of RehaStim 2 for use with an exercise equipment. Hasomed will conduct risk hazard analysis to determine if additional testing (e.g. electrical safety, EMC, verification and validation) is necessary to ensure safety and effectiveness when using the RehaStim 2 with an exercise equipment.
{10}------------------------------------------------
#### VII. PERFORMANCE DATA
The RehaStim 2 (2016) and both predicate devices (ErigoPro (FES) and RehaStim 2 / RehaMove 2) have the same intended use and the same output characteristics. The different technological characteristics do not raise new questions of safety and effectiveness.
The safety and effectiveness of the single stimulator device RehaStim 2 is extensively demonstrated by using the device for many years.
The safety and effectiveness of combining the electrostimulation device RehaStim 2 with movement exercisers or with devices for early functional mobilization of bedridden patients have been extensively demonstrated by the ongoing use of the predicate devices RehaStim 2 / ErigoPro (FES) and RehaStim 2 / RehaMove 2, respectively.
The safety and effectiveness of the controller has been demonstrated over the development period of the RehaStim 2 (2016). For combining the device with exercise equipment for walking and cycling applications, clear assessment instructions for verification and validation ensure, that no new questions of safety and effectiveness can be introduced by the combination of the products.
| Test or procedure | Task |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Review of user documentation for<br>predicate device | It was reviewed that equivalent<br>functionality was implemented. |
| Review of 510(k) submissions for<br>predicate device | Confirm technical specifications for<br>completion of predicate details in<br>comparison tables (see above) |
| Output characteristic measurement of<br>new device | The RehaStim 2 (2016) device was<br>tested and technically compared with<br>the predicate devices. |
| Control of system testing | The system testing was aligned to<br>verify performance to specification. |
## Determination of substantial equivalence
#### Table 2 Performance data
It can be matched that there is a substantial equivalence in all important technical and medical characteristics to the predicate devices.
VIII. CONCLUSIONS
{11}------------------------------------------------
### 510(k) RehaStim 2 (2016)
Since the intended use of the predicate devices is the same and the therapeutic parameters did not change, no clinical testing was required. The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that the device will perform as intended in the specified use conditions. Hasomed concludes that the device performs comparably to the predicate devices that are currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.