K162648 · Tianjin Uwell Medical Device Manufacturing Co., Ltd. · HIF · Dec 12, 2016 · Obstetrics/Gynecology
Device Facts
Record ID
K162648
Device Name
U-Blade Veress Needle
Applicant
Tianjin Uwell Medical Device Manufacturing Co., Ltd.
Product Code
HIF · Obstetrics/Gynecology
Decision Date
Dec 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1730
Device Class
Class 2
Indications for Use
The U-Blade Veress Needle is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.
Device Story
The U-Blade Veress Needle is a sterile, single-use surgical instrument used to establish pneumoperitoneum. It consists of an external stainless steel needle tube with a sharp point for tissue puncture and an internal spring-loaded, round-tipped obturator. The obturator extends automatically when the needle tip enters the peritoneal cavity, protecting internal organs from the sharp point. The handle includes a red indicator to verify obturator retraction and a valve connector for attachment to an insufflation pump. Used by surgeons in clinical settings during laparoscopic procedures, the device facilitates the introduction of carbon dioxide into the peritoneal cavity. The device is available in 120mm and 150mm lengths with either U-shaped or bugle-shaped handles.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1, -5, and -10 (cytotoxicity, irritation, sensitization). Performance testing included tip pull, switch operation, spring obturator operation, and needle puncture force tests. Sterilization validated per ISO 11137-1 and -2. Package integrity and shelf-life testing were also conducted.
Technological Characteristics
Materials: PC resin, SUS 304/316 stainless steel. Dimensions: 120-150 mm length. Principle: Mechanical spring-loaded obturator for organ protection. Connectivity: Valve connector for insufflation pump. Sterilization: Gamma radiation. Standards: ISO 10993 (biocompatibility), ISO 11137 (sterilization).
Indications for Use
Indicated for patients undergoing laparoscopic procedures requiring percutaneous insertion into the peritoneal cavity for CO2 insufflation to establish pneumoperitoneum.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 12, 2016
Tianjin UWell Medical Device Manufacturing Co. Ltd. Tao Fan General Manager A02, Plant B. No. 278, Hangkong Rd Tianjin Free Trade Zone Tianjin, CN 300308
Re: K162648 Trade/Device Name: U-Blade Veress Needle Regulation Number: 21 CFR 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: Class II Product Code: HIF, FHO Dated: September 2, 2016 Received: September 22, 2016
Dear Tao Fan,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
For Division
Douglas Silverstein -S 2016.12.12 12:31:16 -05'00'
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162648
Device Name U-Blade Veress Needle
Indications for Use (Describe)
The U-Blade Veress Needle is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for U-Well. The logo consists of a blue and green circular design with a plus sign in the center, followed by the text "U-Well" in blue and green. To the right of "U-Well" is Chinese text in green.
TIANJIN UWELL MEDICAL DEVICE MANUFACTURING CO., LTD
# 510(k) Summary K162648
| Submitter: | Tianjin UWell Medical Device Manufacturing Co., Ltd.<br>A02, Plant B, No. 278, Hangkong Road, Tianjin Free Trade Zone<br>(Airport Industrial Park), Tianjin, 300308 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Mr. Fan, Tao General Manager<br>Phone: (86)18616806984<br>Email: fant01@uwellmed.com |
| Date Prepared: | December 9, 2016 |
| Device Trade Name: | U-Blade Veress Needle |
| Device Common Name: | Veress Needle |
| Classification Name: | Laparoscopic insufflator |
| Class: | II |
| Regulation Number: | 884.1730 |
| Panel: | Obstetrical and Gynecological Devices |
| Product Code: | HIF (Insufflator. Laparoscopic). FHQ (Pneumoperitoneum needle |
# 1. Predicate Device:
| Device | Company | Product Code | 510(k) Number |
|-----------------------------------------------------------------------|----------------------------|--------------|---------------|
| Unimax Veress Needle | Unimax Medical Systems Inc | HIF<br>FHO | K111441 |
| The Predicate Device has not been subject to a design-related recall. | | | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for U-Well. The logo consists of a blue and green circular design with a plus sign in the center, followed by the text "U-Well" in blue and green. To the right of the text is Chinese characters in green.
## 2. Indications for Use:
The U-Blade Veress Needle is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.
# 3. Device Description:
The U-Blade Veress Needle is a slender, sharply-pointed metal tube designed to introduce or remove gas from the peritoneal cavity as a surgical/radiological procedural method. It is used for percutaneous insertion into the peritoneal cavity for the purpose of insufflation [e.g., with carbon dioxide (CO2)] to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.
The device consists of an external stainless steel needle tube with a sharp point to puncture tissue. It is also equipped with an internal spring-loaded round-tipped obturator that extends from the tube when the device does not encounter resistance (i.e., when the cavity has been reached) to prevent damage to internal organs from the sharp point. The subject device also has a handle with a red indicator mark to verify when the spring-loaded obturator is retracted and a valve connector which connects to an insufflation pump.
As described in the following table, the U-Blade Veress Needle is available in 120mm and 150mm needle lengths and two types of handles with different shapes (U shape and Bugle).
| Model Number | Length (mm) | Comments |
|--------------|-------------|--------------------|
| UV120 | 120 mm | U shape handle |
| UV150 | 150 mm | U shape handle |
| BV120 | 120 mm | Bugle shape handle |
| BV150 | 150 mm | Bugle shape handle |
This is a sterile and single-use device.
## 4. Non-Clinical Testing
A series of tests were performed to assess the safety and effectiveness of the U-Blade Veress Needle. These tests were conducted in accordance with ISO 10993-1:2009. ISO 10993-5:2009, ISO 10993-10:2010, ISO 10993-12:2012, ISO 11137-1:2006, and ISO 11137-2:2012.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for U-Well. The logo consists of a green and blue symbol that looks like three people holding hands in a circle with a plus sign in the middle. To the right of the symbol is the text "U-Well" in blue, followed by some Chinese characters in green.
Per ISO 10993-1:2009, the subject device is an external communicating device (tissue/bone/dentin) with limited contact duration (<24 h); and therefore, the following biocompatibility tests were completed and demonstrated biocompatibility:
- Cytotoxicity Test (ISO 10993-5:2009)
- Intracutaneous Irritation Test (ISO 10993-10:2010) ●
- . Skin Sensitization Test (the Guinea Pig maximization test) (ISO 10993-10:2010)
The performance testing conducted on subject device and predicate device are listed below:
- Tip Pull Test
- Switch Operation Test
- Spring Obturator Operation Test ●
- Needle Puncture Force Test ●
The performance test results demonstrate that the U-Blade Veress Needle meet its acceptance criteria.
Sterilization validation was conducted in accordance with ISO 11137-1:2006 and ISO 11137-2:2012.
Package integrity and functional performance testing were completed on the subject device following aging to support the proposed shelf life.
## 5. Substantial Equivalence Determination
The U-Blade Veress Needle and the predicate device have the same intended use but different technological characteristics. The differences in technological characteristics between the devices do not raise different questions of safety or effectiveness.
The comparison to predicate device is described in the table below.
| | Proposed Device<br>(U-Blade Veress Needle) | Predicate Device<br>(Unimax Veress Needle, K111441) |
|---------------------------------|--------------------------------------------------------|-----------------------------------------------------|
| Trade Name | U-Blade Veress Needle | Unimax Veress Needle |
| 510(k) Submitter | Tianjin UWell Medical Device<br>Manufacturing Co., Ltd | Unimax Medical Systems Inc. |
| 510(k) Number | K162648 | K111441 |
| Common Name | Veress Needle | Veress Needle |
| Classification | CFR 884.1730 | 21 CFR 884.1730 |
| Classification and Product Code | Class II,<br>HIF, FHO | Class II,<br>HIF, FHO |
| Device Classification Name | Laparoscopic insufflator | Laparoscopic insufflator |
| Models | UV120/150, BV 120/150 | xVN Series |
## Comparison to Predicate Device
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for U-Well. The logo consists of a blue and green graphic to the left, followed by the text "U-Well" in blue and the text "优外" in green. The graphic appears to be a stylized representation of a medical symbol.
| | TIANJIN UWELL MEDICAL DEVICE MANUFACTURING CO., LTD | |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comprised Elements | Needle tube, Inner Blunder<br>stylet, Handle, and Valve<br>connector | Needle tube, Inner Blunder<br>stylet, Handle, and Valve<br>connector |
| Materials: | PC resin, SUS 304/316 | PC resin, SUS 304,<br>Acrylonitrile styrene |
| Dimension | Length:<br>120-150 mm | Length:<br>120-150 mm |
| Indications for Use | Intended for percutaneous<br>insertion into the peritoneal<br>cavity for the purpose of<br>insufflation with carbon<br>dioxide to establish<br>pneumoperitoneum prior to the<br>placement of trocars during<br>laparoscopic procedures. | Intended for percutaneous<br>insertion into the peritoneal<br>cavity for the purpose of<br>insufflation with carbon<br>dioxide to establish<br>pneumoperitoneum prior to the<br>placement of trocars during<br>laparoscopic procedures. |
| Sterilization method | Gamma sterilization | EO sterilization |
| Disposable for Single<br>Use | Yes | Yes |
# 6. Conclusion
Based upon the performance tests, it can be concluded that the UWell Veress Needle is as safe and effective as the predicate device. Therefore, UWell Veress Needle is substantially equivalent to the Unimax Veress Needle.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.