An AC-powered slit-lamp biomicroscope and accessories intended for use in the examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or traumas which affect the structural properties of the anterior eye segment.
Device Story
AC-powered slit-lamp biomicroscope; projects thin, intense beam of light into patient's eye via control diaphragm. Components include light source, slit, collimation/imaging optics, infrared/ultraviolet filters, and dielectric mirror. Patient positioned with chin rest and forehead band; clinician operates joystick to focus slit on cornea. Magnification, slit width, and illumination angle are adjustable. Used in clinical settings by eye care professionals to visualize anterior eye segment structures. Output is magnified optical image observed through microscope; aids diagnosis of ocular diseases or trauma. No automated processing or AI/ML components.
Clinical Evidence
Bench testing only. Compliance with ISO 15004-2:2007 (radiation hazards), ISO 10939:2007, IEC 60601-1 (electrical safety), and IEC 60601-1-2 (EMC) demonstrated.
Technological Characteristics
AC-powered slit-lamp biomicroscope; LED light source; optical magnification system; mechanical joystick control for positioning; adjustable slit width (0-14mm) and length (1-14mm). Standards: ISO 15004-2:2007, ISO 10939:2007, IEC 60601-1, IEC 60601-1-2. No software or connectivity.
Indications for Use
Indicated for patients requiring examination of the anterior eye segment (cornea epithelium to posterior capsule) to aid in the diagnosis of diseases or traumas affecting structural properties.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2017
Takagi Seiko Co., Ltd. Hagiwara Toru Official Correspondent 330-2 Iwafune, Nakano-Shi, Nagano-Ken Nakano, 383-8585 JP
Re: K162636
Trade/Device Name: Marco Ultra M3, Marco Ultra M4 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: HJO Dated: March 2, 2017 Received: March 7, 2017
Dear Hagiwara Toru:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting
{1}------------------------------------------------
(reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation
(21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/10 description: The image contains the text "Bradley S. Cunningham -S". The text is arranged in two lines, with "Bradley S." on the first line and "Cunningham -S" on the second line. The text is in a sans-serif font and is black in color. There is a logo in the background.
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162636
Device Name Marco Ultra M3, Marco Ultra M4
Indications for Use (Describe)
An AC-powered slit-lamp biomicroscope and accessories intended for use in the examination of the anterior eye segment, from the cornea epithelium to the posterior capsule.
It is used to aid in the diagnosis of diseases or traumas which affect the structural properties of the anterior eye segment.
Type of Use (Select one or both, as applicable):
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
B-TQMS4101Letterhead Rev. 1
Image /page/3/Picture/2 description: The image shows the logo of Takagi Seiko Co., Ltd. The logo consists of a blue abstract symbol on the left and the company name in bold, black letters on the right. A black line is present under the company name. The text "510(k) Summary" is present below the black line.
330-2 IWAFUNE, NAKANO-SHI, NAGANO-KEN, +81-269-22-4512 FAX +81-269-2 http://www.takagi-i.com E-mail: in
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c).
1. Submitter of this pre-market notification:
TAKAGI SEIKO CO., LTD. 330-2 Iwafune, Nakano-shi, Nagano-ken, Japan 383-8585
Contact Person: Toru Hagiwara Phone: +80-269-22-4511 Fax : +81-269-26-6321 Email: hagiwara.Toru@takagi-j.com
Date prepared: This summary was prepared on April 10, 2017
2.Trade names of the devices: MARCO ULTRA M3 MARCO ULTRA M4
3.Common / Usual name: AC-Powered Slit-lamp Biomicroscope
# 4.Classification Information:
- Classification name: CFR title: Product code: Device Class: Classification Panel:
Biomicroscope, Slit-Lamp, AC-Powered 21 CFR 886.1850 HJO Class II Ophthalmic Panel
{4}------------------------------------------------
5. Predicate Devices:
We claim substantial equivalence to the following devices Manufacture: Haag-Streit AG Device name: Slit Lamp BM900 510(k) Premarket Notification No: K100202 Classification name: Biomicroscope, Slit-Lamp, AC-Powered CFR title: 21 CFR 886.1850 Product code: HJO Device Class: Class II Classification Panel: Ophthalmic Panel Manufacture: TAKAGI SEIKO CO., LTD Device name: Z2 Slit lamp microscope 510(k) Premarket Notification No: K152535 Biomicroscope, Slit-Lamp, AC-Powered Classification name: CFR title:
Product code: Device Class: Classification Panel: 21 CFR 886.1850 HJO Class II Ophthalmic Panel
#### 6. General Device Description:
An AC-powered slit lamp biomicroscope is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment.
An AC-Powered slit lamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
The slit lamp illumination is composed of the light source, the slit, collimation and imaging optics, and infrared and ultra violet filters and a dielectric mirror. The slit lamp have the option to combine a background illumination together with the slit illumination.
The patient sits in front of the slit lamp with his chin rest and his forehead against the forehead band. The chin rest is adjusted in height until the eyes of the patient are level with the black mark of the headrest column. The light is switched on and the brightness is controlled with a knob on the power supply. With the joystick control lever the instrument can be moved back and forward until the slit appears in focus on the cornea. The image can be observed through the microscope. Various magnifications can be selected on the microscope. For different observations the slit width can be changed, the slit can be tilted horizontally and the angle between the illumination unit and the microscope can also be varied horizontally.
{5}------------------------------------------------
# 7. Indication for use:
An AC-powered slit-lamp biomicroscope and accessories intended for use in the examination of the anterior eye segment, from the cornea epithelium to the posteriorcapsule. It is used to aid in the diagnosis of diseases or traumas which affect the structural properties of the anterior eye segment.
#### 8. Comparison with predicate devices:
The MARCO ULTRA M3 is substantially equivalent to the predicate device Slit Lamp BM900 because they use similar technology and perform similar functions to provide the physician with the necessary information to aid in diagnosis.
The MARCO ULTRA M4 is substantially equivalent to the predicate device Z2 Slit lamp microscope because they use similar technology and perform similar functions to provide the physician with the necessary information to aid in diagnosis.
| | Predicate Device Slit lamps | TAKAGI Slit lamps |
|------------------------|-------------------------------------------|--------------------------------------------------|
| | Slit Lamp BM900((K100202) | MARCO ULTRA M3(K162636) |
| | Z2 Slit lamp microscope(K152535) | MARCO ULTRA M4(K162636) |
| Brightness<br>Controls | For BM900 (K100202) | For ULTRA M3(K162636) |
| | Variable control by potentiometer | Control by Light intensity control knob |
| | Maximum brightness | Maximum brightness |
| | approx. 450'000 Lux | approx. 240'000 Lux |
| | For Z2(K152535) | For ULTRA M4(K162636) |
| | Control by Light intensity control knob | Control by Light intensity control knob |
| | Maximum brightness<br>approx. 150'000 Lux | Maximum brightness<br>approx. 150'000 Lux |
| Slit image<br>width | For BM 900(K100202)<br>0-8mm continuous | For all Slit lamps(K162636)<br>0-14mm continuous |
| | For Z2(K152535) | |
| | 0-14mm | |
| Slit image<br>length | For BM 900 (K100202)<br>1-8mm continuous | For all Slit lamps(K162636)<br>1-14mm continuous |
| | For Z2(K152535) | |
| | 1-14mm continuous | |
Major different technological characteristics are as follows:
{6}------------------------------------------------
| | Predicate Device Slit lamps | TAKAGI Slit lamp |
|--------------------------------------------------------------------------------------------------|----------------------------------------|----------------------------------------|
| | Slit Lamp BM900 (K100202) | MARCO ULTRA M3(K162636) |
| | Z2 Slit lamp microscope(K152535) | MARCO ULTRA M4(K162636) |
| Illumination<br>field<br>Diameter | For BM900(K100202) | For M3(K162636) |
| | φ8,φ5,φ3,φ2,φ1,φ0.2 mm | φ10,φ5, φ3, φ2, φ1, φ0.2 mm |
| | For Z2(K152535) | For M4(K162636) |
| | φ14,φ8,φ5, φ3,φ0.3 mm | φ14,φ8,φ5, φ3,φ0.3 mm |
| Radial<br>movement of<br>the slit light<br>illumination<br>relative to the<br>microscope<br>axis | For BM900(K100202) | For M3(K162636) |
| | Horizontal ±90° | Horizontal±90° |
| | Vertical 0 - 20° | Vertical 0°,5°,10°,15°,20° |
| | For Z2(K152535) | For M4(K162636) |
| | Horizontal±90°<br>Vertical don't apply | Horizontal±90°<br>Vertical don't apply |
| Light source | For BM900(K100202) and Z2(K152535) | For all Slit lamps(K162636) |
| | LED | LED |
| Background<br>illumination | For BM900(K100202) | For M3(K162636) |
| | Option | None |
| | For Z2(K152535) | For M4(K162636) |
| | None | Mounted in Slit lamp unit |
# 9. Performance, safety and EMC Data:
The slit lamps M3 and M4 were tested according to ISO 15004-2:2007 and ISO10939:2007 for radiation hazards, to IEC60601-1 for electrical safety and IEC-60601-1-2 for electromagnetic compatibility. In all tests, the slit lamps were in compliance with these FDA recognized standards.
# 10.Conclusions:
In accordance to 21 CFR 807.92(d) and based on the technical characteristics and the results of the performance tests we conclude that the slit lamps M3 and M4 are safe and effective compared to the predicate device slit lamps BM 900 and Z2.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.