K162631 · STERIS Corporation · JOJ · Jan 27, 2017 · General Hospital
Device Facts
Record ID
K162631
Device Name
VERIFY STEAM Integrating Indicator - Short
Applicant
STERIS Corporation
Product Code
JOJ · General Hospital
Decision Date
Jan 27, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The integrating indicator is designed to chemically react over time with the critical parameters of a steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: Steam Sterilization Cycles: - 250°F/121°C, 30 minutes Gravity - 270°F/132°C, 4 minutes Dynamic Air Removal - 270°F/132°C, 15 minutes Gravity - 275°F/135°C, 3 minutes dynamic air removal - 275°F/135°C, 10 minutes Gravity Steam Sterilization Cycles (IUSS): - 270°F/132°C, 4 minutes Dynamic Air Removal - 270°F/132°C, 3 minutes Gravity - 270°F/132°C, 10 minutes Gravity - 275°F/135°C, 3 minutes dynamic air removal - 275°F/135°C, 3 minutes Gravity - 275°F/135°C, 10 minutes Gravity
Device Story
Single-use chemical integrating indicator; monitors steam sterilization cycles. Placed inside sterilization containment devices (packs, trays, pouches). Principle of operation: temperature-sensitive chemical dye migrates within an embossed cavity upon exposure to critical steam sterilization parameters (time, temperature, saturated steam). Output: visual indicator; dark bar enters 'ACCEPT' window if sterilization criteria met; remains outside if criteria not met. Used by healthcare providers in clinical settings. Provides immediate visual confirmation of cycle success or failure; informs decision to use or re-process sterilized materials. Benefits patient by ensuring medical instruments meet required sterilization standards before use.
Clinical Evidence
Bench testing only. Testing conducted in ISO 18472-conforming steam resistometer and simulated use cycles. Results confirmed device passes at stated values (SV) and fails at 15% below SV or 1°C below SV. Confirmed no endpoint reached before biological indicator inactivation. All performance criteria met.
Technological Characteristics
Single-use chemical indicator. Components: backing material, embossed cavity with temperature-sensitive chemical/dye, wicking strip, laminated paper cover with viewing window. Principle: chemical migration. Standards: ANSI/AAMI/ISO 11140-1:2014. No energy source, software, or connectivity.
Indications for Use
Indicated for use by healthcare providers as an independent monitor of critical parameters in steam sterilization cycles (gravity and dynamic air removal) for packs, pouches, containers, and trays.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized graphic of three human profiles facing right, with flowing lines connecting them. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 27, 2017
Steris Corporation Tricia Cregger, Ph.D. Manager, Regulatory Affairs 5960 Heisley Road Mentor, Ohio 44060
Re: K162631
Trade/Device Name: Verify Steam Integrating Indicator - Short Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: December 19, 2016 Received: December 19, 2016
Dear Tricia Cregger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162631
Device Name VERIFY STEAM Integrating Indicator- Short
Indications for Use (Describe)
The integrating indicator is designed to chemically react over time with the critical parameters of a steam sterilization cycle within a specified tolerance. The integrating intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:
Steam Sterilization Cycles:
- · 250°F/121°C, 30 minutes Gravity
- · 270°F/132°C, 4 minutes Dynamic Air Removal
- · 270°F/132°C, 15 minutes Gravity
- · 275°F/135°C, 3 minutes dynamic air removal
- · 275°F/135°C, 10 minutes Gravity
Steam Sterilization Cycles (IUSS):
- · 270°F/132°C, 4 minutes Dynamic Air Removal
- · 270°F/132°C, 3 minutes Gravity
- · 270°F/132°C, 10 minutes Gravity
- · 275°F/135°C, 3 minutes dynamic air removal
- · 275°F/135°C, 3 minutes Gravity
- · 275°F/135°C, 10 minutes Gravity
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for STERIS. The logo consists of the word "STERIS" in a serif font, with the registered trademark symbol to the right of the word. Below the word is a graphic of several horizontal wavy lines, resembling water.
# 510(k) Summary For K162631 VERIFY® STEAM Integrating Indicator – Short
STERIS Corporation 11457 Melrose Street Franklin Park, IL 60131
Contact:
Tricia Cregger, Ph.D. Manager, Regulatory Affairs Telephone: (440) 392-7189 Fax No: (440) 357-9198 e-mail: Tricia_Cregger@steris.com
Submission Date: January 23, 2017
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{4}------------------------------------------------
#### 1. Device Name
| Trade Name: | VERIFY® STEAM Integrating Indicator – Short |
|------------------------|------------------------------------------------------------------------------|
| Device Classification: | Class II |
| Common/usual Name: | Indicator, physical/chemical sterilization process |
| Classification Name: | Indicator, physical/chemical sterilization process<br>(21 CFR 880.2800, JOJ) |
#### 2. Predicate Device
VERIFY® |STEAM Integrating Indicator (K152630)
#### 3. Description of Device
The VERIFY® STEAM Integrating Indicator – Short is a single use device used by healthcare providers to monitor steam sterilization cycles. The VERIFY® STEAM Integrating Indicator - Short is included in a pack or load in a steam sterilizer and the load is processed in accordance with the sterilizer manufacturer's directions. Prior to the use of the processed items, the integrator is observed. If the dark bar on the device enters the ACCEPT (OK) area of the viewing window, the integrator is read as a PASS to indicate that the steam sterilization criteria for the cycle have been met. If the dark bar on the device does not enter the ACCEPT (OK) area of the viewing window, the integrator is read as a FAIL, indicating that sufficient steam sterilization criteria has not been met and processed materials should be subjected to another steam sterilization cycle prior to use.
#### 4. Intended Use
The integrating indicator is designed to chemically react over time with the critical parameters of a steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:
Steam Sterilization Cycles:
- · 250°F/121°C, 30 minutes Gravity
- · 270°F/132°C, 4 minutes Dynamic Air Removal
- · 270°F/132°C, 15 minutes Gravity
- · 275°F/135°C, 3 minutes dynamic air removal
- · 275°F/135°C, 10 minutes Gravity
{5}------------------------------------------------
Steam Sterilization Cycles (IUSS):
- · 270°F/132°C. 4 minutes Dynamic Air Removal
- · 270°F/132°C, 3 minutes Gravity
- · 270°F/132°C, 10 minutes Gravity
- · 275°F/135°C, 3 minutes dynamic air removal
- · 275°F/135°C, 3 minutes Gravity
- · 275°F/135°C, 10 minutes Gravity
#### ട്. Description of Substantial Equivalence
The proposed and predicate devices are both single use integrating indicators for use in monitoring steam sterilization cycles. The differences between the proposed VERIFY® |STEAM Integrating Indicator - Short and the predicate device are limited to differences in the materials used and the overall length of the device. However, these differences do not affect the intended use or performance characteristics, nor do the differences raise any new concerns of equivalence when compared to the predicate device. Performance testing has been conducted for monitoring all claimed steam sterilization cycles in accordance with the Guidance for Industry and FDA Staff: Premarket [510(k)] Submissions for Chemical Indicators.
#### Technological Characteristics 6.
The proposed and predicate devices are single use integrating indicators for use in monitoring particular steam sterilization cycles. The device components, mechanism of action, and endpoint are the same as the predicate's and when exposed to the defined processing conditions, the proposed and predicate devices exhibit a visible migration of the included chemistry to indicate that stated values have been achieved.
Table 5-1 contains a comparison of technological characteristics and specifications of the proposed VERIFY® STEAM Integrating Indicator - Short to the predicate VERIFY® STEAM Integrating Indicator.
{6}------------------------------------------------
| | Proposed | K152630 | Comparison |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | VERIFY® STEAM | VERIFY® STEAM | |
| | Integrating Indicator - Short | Integrating Indicator | |
| Intended use | The integrating indicator is<br>designed to chemically react<br>over time with the critical<br>parameters of a steam<br>sterilization cycle within a<br>specified tolerance. The<br>integrating indicator strip is<br>intended to be placed in each<br>pack, pouch, container, tray or<br>other containment device to<br>function as an independent<br>monitor of critical parameters<br>for the following sterilization<br>cycles: | The integrating indicator is<br>designed to chemically react<br>over time with the critical<br>parameters of steam<br>sterilization cycle within a<br>specified tolerance. The<br>integrating indicator strip is<br>intended to be placed in each<br>pack, pouch, container, tray or<br>other containment device to<br>function as an independent<br>monitor of critical parameters<br>for the following sterilization<br>cycles: | The intended use<br>statement is<br>identical between<br>the proposed and<br>predicate devices. |
| | Steam Sterilization Cycles:<br>•250°F/121°C, 30 minutes<br>Gravity<br>•270°F/132°C, 4 minutes<br>Dynamic Air Removal<br>•270°F/132°C, 15 minutes<br>Gravity<br>•275°F/135°C, 3 minutes<br>dynamic air removal<br>•275°F/135°C, 10 minutes<br>Gravity | Steam Sterilization Cycles:<br>•250°F/121°C, 30 minutes<br>Gravity<br>•270°F/132°C, 4 minutes<br>Dynamic Air Removal<br>•270°F/132°C, 15 minutes<br>Gravity<br>•275°F/135°C, 3 minutes<br>dynamic air removal<br>•275°F/135°C, 10 minutes<br>Gravity | |
| | Steam Sterilization Cycles<br>(IUSS):<br>•270°F/132°C, 4 minutes<br>Dynamic Air Removal<br>•270°F/132°C, 3 minutes<br>Gravity<br>•270°F/132°C, 10 minutes<br>Gravity<br>•275°F/135°C, 3 minutes<br>dynamic air removal<br>•275°F/135°C, 3 minutes<br>Gravity<br>•275°F/135°C, 10 minutes<br>Gravity | Steam Sterilization Cycles<br>(IUSS):<br>•270°F/132°C, 4 minutes<br>Dynamic Air Removal<br>•270°F/132°C, 3 minutes<br>Gravity<br>•270°F/132°C, 10 minutes<br>Gravity<br>•275°F/135°C, 3 minutes<br>dynamic air removal<br>•275°F/135°C, 3 minutes<br>Gravity<br>•275°F/135°C, 10 minutes<br>Gravity | |
| Device<br>design -<br>components | Backing material with<br>embossed cavity containing<br>temperature sensitive chemical<br>and coloring dye, wicking strip,<br>covered with laminated paper<br>containing labeling and a<br>viewing window. | Backing material with<br>embossed cavity containing<br>temperature sensitive chemical<br>and coloring dye, wicking strip,<br>covered with laminated paper<br>containing labeling and viewing<br>windows. | Device design and<br>components are<br>similar and any new<br>questions of safety<br>and effectiveness will<br>be mitigated through<br>performance data. |
| Indicator<br>agent | Proprietary formulation | Proprietary formulation | The indicator agent<br>is identical to the<br>predicate. |
| Feature | Proposed | K152630 | Comparison |
| | VERIFY® STEAM<br>Integrating Indicator - Short | VERIFY® STEAM<br>Integrating Indicator | |
| Sterilization<br>method and<br>cycles | Steam Sterilization Cycles:<br>•250°F/121°C, 30 minutes<br>Gravity<br>•270°F/132°C, 4 minutes<br>Dynamic Air Removal<br>•270°F/132°C, 15 minutes<br>Gravity<br>•275°F/135°C, 3 minutes<br>dynamic air removal<br>•275°F/135°C, 10 minutes<br>Gravity<br>Steam Sterilization Cycles<br>(IUSS):<br>•270°F/132°C, 4 minutes<br>Dynamic Air Removal<br>•270°F/132°C, 3 minutes<br>Gravity<br>•270°F/132°C, 10 minutes<br>Gravity<br>•275°F/135°C, 3 minutes<br>dynamic air removal<br>•275°F/135°C, 3 minutes<br>Gravity<br>•275°F/135°C, 10 minutes<br>Gravity | Steam Sterilization Cycles:<br>•250°F/121°C, 30 minutes<br>Gravity<br>•270°F/132°C, 4 minutes<br>Dynamic Air Removal<br>•270°F/132°C, 15 minutes<br>Gravity<br>•275°F/135°C, 3 minutes<br>dynamic air removal<br>•275°F/135°C, 10 minutes<br>Gravity<br>Steam Sterilization Cycles<br>(IUSS):<br>•270°F/132°C, 4 minutes<br>Dynamic Air Removal<br>•270°F/132°C, 3 minutes<br>Gravity<br>•270°F/132°C, 10 minutes<br>Gravity<br>•275°F/135°C, 3 minutes<br>dynamic air removal<br>•275°F/135°C, 3 minutes<br>Gravity<br>•275°F/135°C, 10 minutes<br>Gravity | The intended cycles<br>are identical<br>between the<br>proposed and<br>predicate devices. |
| Mechanism<br>of action | Proprietary | Proprietary | Mechanism of<br>action is identical to<br>predicate. |
| Endpoint<br>specifications | The end point is determined<br>by the migration of the steam<br>sensitive dye to an area<br>marked ACCEPT (OK) on the<br>indicator. Endpoint is reached<br>at the stated value (SV) for<br>each claimed temperature.<br>Endpoint is not reached at the<br>stated value - 15% time and/or<br>1°C. | The end point is determined<br>by the migration of the steam<br>sensitive dye to an area<br>marked ACCEPT (OK) on the<br>indicator. Endpoint is reached<br>at the stated value (SV) for<br>each claimed temperature.<br>Endpoint is not reached at the<br>stated value - 15% time and/or<br>1°C. | The end point is the<br>same for proposed<br>and predicate<br>devices. |
| | Proposed | K152630 | Comparison |
| Feature | VERIFY® STEAM<br>Integrating Indicator - Short | VERIFY® STEAM<br>Integrating Indicator | |
| Comparison<br>of integrator<br>stated values<br>at biological<br>indicator<br>growth-<br>negative<br>cycle<br>conditions | Integrator does not reach<br>endpoint before the biological<br>indicator is inactivated. | Integrator does not reach<br>endpoint before the biological<br>indicator is inactivated | Same performance<br>observed. |
| Shelf-life | 5 years | 5 years | Formulation of ink<br>is identical to<br>predicate. |
| Standard /<br>Guidance | Conforms to the Guidance for<br>Industry and FDA Staff:<br>Premarket Notification<br>[510(k)] Submissions for<br>Chemical Indicators for<br>integrating indicators and<br>ANSI/AAMI/ISO 11140-<br>1:2014: Sterilization of Health<br>Care Products - Chemical<br>Indicators - Part 1: General<br>Requirements. | Conforms to the Guidance for<br>Industry and FDA Staff:<br>Premarket Notification<br>[510(k)] Submissions for<br>Chemical Indicators for<br>integrating indicators and<br>ANSI/AAMI/ISO 11140-<br>1:2014: Sterilization of Health<br>Care Products - Chemical<br>Indicators - Part 1: General<br>Requirements. | Both proposed and<br>predicate devices<br>meet the same<br>applicable standard. |
Table 5-1. Device Comparison Table
{7}------------------------------------------------
{8}------------------------------------------------
#### 7. Performance Testing
Performance testing was conducted to verify that the proposed VERIFY® STEAM Integrating Indicator - Short meets the requirements for integrating indicators in accordance with the Guidance for Industry and FDA Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators for integrating indicators as well as ANSI/AAMI/ISO 11140-1:2014.
Table 5-2 summarizes the verification activities that were performed, with their respective acceptance criteria and results, to demonstrate that the proposed VERIFY® |STEAM | Integrating Indicator - Short is as safe and is as effective as the predicate device. These studies confirm that the proposed device's performance meets the requirements of its pre-defined acceptance criteria and intended uses, and qualify the proposed device for use in the claimed steam sterilization cycles.
## Stated Values
The subject device (integrator indicator) testing was performed in an ISO 18472conforming steam resistometer using the following stated values shown below at the following temperatures (see Table 5-2). The integrator indicator showed a passing result when the chemicals in the subject device reacted to all critical parameters in the bier vessel.
{9}------------------------------------------------
| Test of 3 Lots | Acceptance Criteria | Study Result |
|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| BIER Vessel Testing | Pass result at SV for each temperature claimed<br>Stated values are:<br>• 26.5 min at 121°C<br>• 2.7 min at 132 °C<br>• 1.4 min at 135 °C | PASS |
| | Fail result at 15% below SV for each<br>temperature claimed | PASS |
| | Fail result at 1°C below SV for each<br>temperature claimed | PASS |
| Simulated Use Testing in<br>Claimed Sterilization<br>Cycles | 100% Pass result under pass conditions<br>100% Fail result under fail conditions | PASS |
| Dry Heat Testing | Fail result when exposed to 140°C dry heat<br>for 30 min | PASS |
| Verification of integrator<br>stated values at biological<br>indicator growth-negative<br>cycle conditions | Integrator does not reach endpoint before the<br>biological indicator is inactivated | PASS |
| Offset/ Transference | Integrator shall not transfer or offset | PASS |
| Table 5-2. Verification Results Summary | |
|-----------------------------------------|--|
| | |
The results of the VERIFY® |STEAM Integrating Indicator – Short performance testing demonstrate that it performs as intended in the claimed steam sterilization cycles and the proposed device is substantially equivalent to the predicate device.
#### 8. Conclusion
Based on the conclusions from the non-clinical tests, the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device (K152630), Class II (21 CFR 880.2800), product code JOJ.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.