K162624 · International Medical Devices, Inc. · MIB · Feb 1, 2017 · Ear, Nose, Throat
Device Facts
Record ID
K162624
Device Name
Pre-Formed Penile Silicone Block
Applicant
International Medical Devices, Inc.
Product Code
MIB · Ear, Nose, Throat
Decision Date
Feb 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3620
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K162624 · Feb 1, 2017
Pre-Formed Penile Silicone Block
International Medical Devices, Inc.
Clinical patient follow-up records
Clinical data from 100 patients were reviewed to assess post-operative outcomes, including pain scores, erosion, migration, and infection rates, to support the safety profile of the device.
Pain (0-10 scale), erosion, migration, and infection
Indications for Use
The Pre-Formed Penile Silicone Block is intended for use in the cosmetic correction of soft tissue deformities, and is contoured at the surgeon's discretion to create a custom implant.
Device Story
Pre-Formed Penile Silicone Block is a medical-grade silicone implant with embedded polyester mesh; available in three sizes (L, XL, XXL) and one durometer (Soft). Used by surgeons for cosmetic correction of penile soft tissue deformities; implant is trimmed and contoured at surgeon's discretion to create custom fit. Device provides structural support to soft tissue. Clinical benefits include cosmetic improvement; risks include potential migration, erosion, pain, and infection. Healthcare providers use the device during surgical procedures; patient outcomes depend on adherence to post-operative instructions, specifically avoiding external devices for skin stretching/expansion.
Clinical Evidence
Clinical study of 100 patients evaluated risks of migration, erosion, and pain. Follow-up occurred at 3 days, 2 weeks, 1 month, 3 months, 6 months, and 12 months. Results: weighted average pain rating 3.2 (0-10 scale); pain relief averaged 7.2 days. Adverse events: 3 cases of erosion (avg 8 months), 4 cases of migration (avg 1.5 months), 3 cases of infection (avg 3.5 months). Most erosion/migration cases attributed to patient non-compliance with post-operative instructions.
Technological Characteristics
Medical grade silicone with embedded polyester mesh. Available in three sizes (L: 12cm x 0.5cm x 2cm; XL: 15cm x 0.8cm x 3cm; XXL: 18cm x 1.1cm x 3.5cm) and one durometer (Soft). Implantable device.
Indications for Use
Indicated for cosmetic correction of soft tissue deformities in the penis. Intended for use by surgeons to create custom implants.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
Predicate Devices
Silicone Block / National Medical Devices, Inc. (K042380)
AART Calf Implant / Aesthetic and Reconstructive Technologies, Inc. (K022511)
AART Gluteal Implant / Aesthetic and Reconstructive Technologies, Inc. (K021839)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized representation of a human figure in profile, with three overlapping faces suggesting community and support. The figure is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 1, 2017
International Medical Devices, Inc. % Allison Komiyama Principal Consultant Acknowledge Regulatory Strategies 2834 Hawthorn St. San Diego, CA 92104
Re: K162624
> Trade/Device Name: Pre-Formed Penile Silicone Block Regulation Number: 21 CFR 874.3620 Regulation Name: Ear, Nose, and Throat Synthetic Polymer Material Regulatory Class: Class II Product Code: MIB Dated: December 20, 2016 Received: December 22, 2016
Dear Allison Komiyama,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K162624
Device Name Pre-Formed Penile Silicone Block
Indications for Use (Describe)
The Pre-Formed Penile Silicone Block is intended for use in the cosmetic correction of soft tissue deformities, and is contoured at the surgeon's discretion to create a custom implant.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for HMD International Medical Devices Inc. The logo consists of the letters "HMD" in bold black font, with a blue plus sign to the left of the "H". To the right of "HMD" are the words "International Medical Devices Inc." in a smaller, blue font. The logo is simple and professional, and the use of blue and black gives it a clean and modern look.
# 510(k) Summary K162624
#### DATE PREPARED
January 27, 2016
#### MANUFACTURER AND 510(k) OWNER
International Medical Devices, Inc. 717 North Maple Drive, Beverly Hills, CA 90210, USA Telephone: (310) 652-2600 Fax: (310) 657-0500 Official Contact: James Elist, MD
#### REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. AcKnowledge Regulatory Strategies Telephone: +1 (619) 208-7888 Email: akomiyama@acknowledge-rs.com
#### PROPRIETARY NAME OF SUBJECT DEVICE
Pre-Formed Penile Silicone Block
COMMON NAME Elastomer, Silicone Block
### DEVICE CLASSIFICATION
21 CFR 874.3620, Product Code MIB, Class II
#### INDICATIONS FOR USE
The Pre-Formed Penile Silicone Block is intended for use in the cosmetic correction of soft tissue deformities, and is contoured at the surgeon's discretion to create a custom implant.
#### DEVICE DESCRIPTION
The Pre-Formed Penile Silicone Block is made from medical grade silicone with an embedded polyester mesh. The device comes in variations that include three sizes (L, XL, and XXL) and one durometer ("Soft"). Size "L" is 12 cm in length with a maximum thickness of 0.5 cm and a height of 2 cm. Size "XL" is 15 cm in length with a maximum thickness of 0.8 cm and a height of 3 cm. Size "XXL" is 18 cm in length with a maximum thickness of 1.1 cm and a height of 3.5 cm. The device is used in the cosmetic correction of soft tissue deformities in the penis, and may be trimmed to allow the surgeon to tailor the device to the needs of a specific patient.
{4}------------------------------------------------
### PREDICATE DEVICE IDENTIFICATION
The Pre-Formed Penile Silicone Block is substantially equivalent to the following predicates:
| 510(k) | Predicate Device Name / Manufacturer | Primary |
|---------|------------------------------------------------------------------------|-----------|
| Number | | Predicate |
| K042380 | Silicone Block / National Medical Devices, Inc.* | ✓ |
| K040042 | Medisil Silicone Sheeting / Medisil Corporation | |
| K022511 | AART Calf Implant / Aesthetic and Reconstructive Technologies, Inc. | |
| K021839 | AART Gluteal Implant / Aesthetic and Reconstructive Technologies, Inc. | |
* National Medical Devices, Inc. is now International Medical Devices, Inc.
#### SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the Pre-Formed Penile Silicone Block. No additional non-clinical testing was provided in this submission in order to demonstrate substantial equivalence.
#### SUMMARY OF CLINICAL TESTING
Clinical evidence on 100 patients was collected to evaluate the risks of migration, erosion, and pain. After the surgical procedure, follow-up data for all patients was collected and evaluated at the following time points: the three successive days following the procedure, and at two weeks, one month, three months, six months, and twelve months following the procedure.
Pain was assessed post-operatively using a validated 0 to 10 Comparative Pain Scale. In 100 patients the weighted average pain rating across the patient population reviewed was 3.2. Pain relief was experienced on average in 7.2 days. In 100 patients there were 3 cases of erosion. This adverse event was observed by the clinic on average 8 months after the patient's procedure with a minimum of 6 months and a maximum of 10 months. In 100 patients there were 4 cases of migration. This adverse event was observed by the clinic on average 1.5 months after the patient's procedure with a minimum of 1 month and a maximum of 2 months. In 100 patients there were 3 cases of infection. This adverse event was observed by the clinic on average 3.5 months after the patient's procedure with a minimum of 2.5 months and a maximum of 4.5 months. 3 out of the 3 cases of erosion and 3 out of the 4 cases of migration were determined to be caused by failure of the patient to follow post-operative instructions (e.g., avoid use of external devices for the stretching/expansion of penile skin). Labeling to mitigate these risks has been added to the package insert in order to more directly address post-operative care.
IMD believes that the unique anatomy, physiology, and function of the penis does not increase the overall potential risks compared to the predicate devices. Clinical evidence demonstrates that rates of pain, erosion, migration, and infection are low compared to reports of other silicone implants on the market.
{5}------------------------------------------------
## EQUIVALENCE TO PREDICATE DEVICES
International Medical Devices, Inc. believes that the Pre-Formed Penile Silicone Block is substantially equivalent to the predicate devices based on the information summarized here:
The subject device has a similar design and dimensions, has an equivalent intended use, and uses identical materials as the device cleared in K042380. The subject device has similar or identical technological characteristics to the devices cleared in K042380, K040042, K022511 and K021839.
## CONCLUSION
The Pre-Formed Penile Silicone Block is considered substantially equivalent to the predicate devices based on the similar indications for use, clinical testing, and similar or identical technological characteristics. Based on this comparison, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.