The Wrist Type Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the Wrist. It can be used at medical facilities or at home. The cuff size is 310 mm x 75 mm, and the intended wrist circumference is 13.5 - 19.5 cm.
Device Story
Battery-driven automatic non-invasive blood pressure monitor; uses oscillometric technique via inflatable wrist cuff. Device automatically manages inflation, deflation, and measurement of systolic/diastolic blood pressure and pulse rate. Features include user ID settings for data storage, time/date configuration, optional voice broadcast of results, unit selection (mmHg/KPa), and low battery indicator. Used in medical facilities or home environments by adults. Healthcare providers or patients review stored data (systolic/diastolic pressure, pulse rate, measurement time) to monitor cardiovascular status. Benefits include convenient, automated blood pressure tracking.
Clinical Evidence
Clinical trial performed per ISO 81060-2:2013 with 150 subjects. Auscultation using a calibrated mercurial sphygmomanometer served as the gold standard control. Results demonstrated accuracy within ±5 mmHg, meeting ISO 81060-2:2013 requirements.
Technological Characteristics
Battery-powered (2x AAA, 3V) automated oscillometric sphygmomanometer. Components: LCD, keypad, inflatable cuff (310x75mm), MCU, pump. Connectivity: None. Biocompatibility: ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization). Safety/EMC: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-30. Software level of concern: Moderate.
Indications for Use
Indicated for adult patients requiring non-invasive measurement of systolic/diastolic blood pressure and pulse rate via oscillometric wrist cuff. Intended for use in medical facilities or home settings. Wrist circumference must be 13.5 - 19.5 cm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
PG 800A Series Electronic Blood Pressure Monitor (K131569)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 9, 2018
Xuzhou Yongkang Electronic Science Technology Co., Ltd. % Ray Wang General Manager Beijing Believe Technology Service Co., Ltd. 5-1206, Build 332, DaFangJu, No.25 BanBiDian Rd. LiYuan Town, TongZhou District, Beijing, 101121 China
Re: K162616
Trade/Device Name: Wrist Type Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN Dated: January 21, 2018 Received: January 25, 2018
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M.A. Willemsen
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K162616
Device Name Wrist Type Blood Pressure Monitor
#### Indications for Use (Describe)
The Wrist Type Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the Wrist. It can be used at medical facilities or at home.
The cuff size is 310 mm x 75 mm, and the intended wrist circumference is 13.5 - 19.5 cm.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
EF
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
#### Tab #3 510(k) Summary
This 510(k) Summary of 510(k) substantial equivalence information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K162616
- 1. Date of Preparation
01/21/2018
- Sponsor 2.
## Xuzhou Yongkang Electronic Science Technology Co., Ltd
4F Building C8, 40 Jingshan Road, Econonic and Technolgical Development Zone, Xuzhou, Jiangsu, China Establishment Registration Number: Not yet registered or the Number Contact Person: YanLi Li Position: General Manager Tel: 86-516- 87892766-631 Fax:86- 516-87892766 -606 Email: lyl(@yonker.cn
- Submission Correspondent 3.
Mr. Ray Wang
#### Beijing Believe Technology Service Co., Ltd.
5-1206, Build 332, DaFangJu, No.25 BanBiDian Rd., LiYuan Town, TongZhou District, Beijing, 101121, China Tel: +86-21-50313932 Fax: +86-21-68093116 Email: ray.wang@believe-med.com
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Wrist Type Blood Pressure Monitor Common Name: Blood Pressure Monitor Model(s): YK-BPW1
#### Regulatory Information:
Classification Name: System, measurement, blood-pressure, non-invasive; Classification: II; Product Code: DXN; Regulation Number: 21 CFR 870.1130; Review Panel: Cardiovascular;
#### Indication For Use Statement:
The Wrist Type Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the Wrist. It can be used at medical facilities or at home. The cuff size is 310 mm x 75 mm, and the intended wrist circumference is 13.5 - 19.5 cm.
#### Device Description 5.
The proposed device, Wrist Type Blood Pressure Monitor, is a battery driven automatic non-invasive blood pressure monitor.
The Wrist Type Blood Pressure Monitor can automatically complete the inflation, deflation and measurement of systolic and diastolic blood pressure as well as the pulse rate of adult person. User can set different user ID for data storage, which can be set as two independent groups of measured blood pressure data;
User can set the time as year, month, day, hour and minute;
User can enable or disable the function of voice broadcast about measurement result;
User can select the unit of the measurement: mmHg or KPa;
The device has the data storage function for data reviewing, including the data of systolic pressure. diastolic pressure, pulse rate and measurement time.
The device has low power indicator to indicate the states of low batteries power.
{5}------------------------------------------------
- 6. Identification of Predicate Device
Predicate #
510(k) Number: K131569
Product Name: PG 800A Series Electronic Blood Pressure Monitor Manufacturer: Shenzhen Pango Electronic Co., Ltd.
- 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1:2012, Medical electrical equipment-Part 1: General requirements for basic safety, and essential performance.
IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
ISO 80601-2-30:2013, Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Non-Invasive Sphygmomanometers.
ISO 10993-5: 2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity.
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
- 8. Clinical Test Conclusion
The clinical trials for YK-BPW1 were performed according to ISO 81060-2:2013, Non-Invasive Sphygmomanometers - Part 2: Clinical Validation of Automated Measurement Type, and relevant volunteers were collected to conduct actual clinical trial of blood pressure measurement.
There were 150 subjects been selected to participate in the trial, and Auscultation was applied as gold standard with the qualified calibrated mercurial sphygmomanometer used as control group for comparison with the proposed device.
The results shown that the accuracy of proposed device meet the requirements of ISO 81060-2:2013 within the ±5mmHg.
{6}------------------------------------------------
# 9. Substantially Equivalent (SE) Comparison
Table 1 General Comparison
| ITEM | Proposed Device | Predicate Device<br>K131569 | Remark |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code | DXN | DXN | SE |
| Regulation No. | 21 CFR 870.1130 | 21 CFR 870. 1130 | SE |
| Class | 2 | 2 | SE |
| Intended Use | The Wrist Type Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the Wrist. It can be used at medical facilities or at home. The cuff size is 310 mm x 75 mm, and the intended wrist circumference is 13.5 - 19.5 cm. | The PG-800A Series Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist. It can be used at medical facilities or at home. The intended wrist circumference is 13.5-19.5 cm. | SE |
| Application Site | Wrist | Wrist | SE |
| Patient<br>Population | Adult | Adult | SE |
| Measurements<br>Item | SYS,DYS,Pulse | SYS,DYS,Pulse | SE |
| Principle | Oscillometric | Oscillometric | SE |
| Main<br>Component | LCD / Key / Cuff / MCU / Pump / Batteries | LCD / Key / Cuff / MCU / Pump / Batteries | SE |
{7}------------------------------------------------
| ITEM | Proposed Device | Predicate Device<br>K131569 | Remark |
|-------------------------|-----------------------------------------------------------------|------------------------------------------------|----------|
| BP Range | 20-280 mmHg | 30-280 mmHg | SE |
| BP Accuracy | $ \pm $ 3 mmHg | $ \pm $ 3 mmHg | SE |
| PR Range | 40-199 | 40-199 bpm | SE |
| Operation<br>Condition | 5-40 °C<br>15%-80% RH<br>50kPa~105kPa | 5-40 °C<br>30%-80% RH<br>Atmosphere Pressure | SE |
| Storage Condition | -20-55 °C<br>$ \le $ 90% RH<br>50kPa~105kPa | -20-55 °C<br>10%-90% RH<br>Atmosphere Pressure | SE |
| Data Storage | SYS, DIA, PR, Measurement Time, No. | SYS, DIA, PR, Measurement Time, No. | SE |
| Performance<br>Standard | Comply with ISO 80601-2-30 | Comply with AAMI SP10 | SE |
| Cuff Size | 310 x 75 mm | 30.8 cm (length) x 8 cm (width) | Analysis |
| Power Supply | 2 x AAA | Two AAA or LR03 Batteries | SE |
| Voltage | 3V | 3V | SE |
Table 2 Performance Comparison
## Analysis
The proposed device is substantially equivalent to the predicate device. Based on the nonclinical and clinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device.
{8}------------------------------------------------
| ITEM | Proposed Device | Predicate Device | Remark |
|-------------------------------------|-----------------------------------------|--------------------------------------------|--------|
| | | K131569 | |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 | SE |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SE |
| Biocompatibility | Comply with ISO 10993-1 | Comply with ISO 10993-1 | SE |
| Label and<br>Labeling | Conforms to FDA Regulatory Requirements | Conforms to FDA Regulatory<br>Requirements | SE |
| Level of Concern<br>of the Software | Moderate | Moderate | SE |
Table 3Safety Comparison
# 10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.