K162608 · Arrinex, Inc. · GEH · Feb 14, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162608
Device Name
ClariFix Device
Applicant
Arrinex, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Feb 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K162608 · Feb 14, 2017
ClariFix Device
Arrinex, Inc.
Published clinical literature (11 clinical studies)
The sponsor performed a literature review of 11 clinical studies involving various cryoprobes used in the nasal passageway to demonstrate that the safety and effectiveness profile of the ClariFix device is comparable to historical clinical outcomes for similar cryosurgical procedures.
Literature review; Chronic rhinitis; Cryosurgery; Historical clinical outcomes
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature review of cryosurgery in the nasal passageway; Systematic review of published clinical literature; Follow-up/Duration: Four decades of published data; Study Period: Last four decades
Patients with nasal obstruction or symptoms of rhinitis; Sample Size: 11 clinical studies
Not applicable for this study
Clinical outcomes after cryosurgery, safety profile, adverse events
Indications for Use
The ClariFix Device is intended to be used as a cryosurgical tool for the destruction of unwanted tissue during surgical procedures, including in adults with chronic rhinitis.
Device Story
Handheld, single-use, disposable cryosurgical device; consists of handle, cannula, and distal cryoprobe. Operates by inserting nitrous oxide canister into handle; mechanical valve controls cryogen flow to cryoprobe. Cryoprobe cools to < -80°C via partial evaporation of nitrous oxide; freezing zone destroys target tissue. Used by physicians in clinical settings; requires direct visualization of target tissue. Physician controls treatment time and duration via valve. No power source, electronics, or software required. Benefits include targeted tissue ablation for chronic rhinitis symptoms.
Clinical Evidence
Prospective, multicenter clinical study (n=27) evaluated safety and effectiveness in adults with chronic rhinitis. Primary safety endpoint: zero device/procedure-related serious adverse events (SAEs) at 90 days. Primary effectiveness: rTNSS improved 56% (6.2 to 2.7) and VAS improved 53% (7.6 to 3.7) at 90 days. Secondary: 89% of investigators rated ease of use as easy/moderately easy. Literature review of 11 studies supported safety and effectiveness of nasal cryosurgery.
Technological Characteristics
Handheld, portable, non-powered cryosurgical device. Uses nitrous oxide cryogen in single-use cartridges. Cryoprobe cooling principle: partial evaporation of cryogen. No electronics, software, or external power source. Sterile, disposable design.
Indications for Use
Indicated for destruction of unwanted tissue during surgical procedures in adults with chronic rhinitis.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, with flowing lines above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 14, 2017
Arrinex, Inc. Vahid Saadat CEO 1755 East Bayshore RD., Suite 26 Redwood City, California 94063
Re: K162608
Trade/Device Name: ClariFix Device Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH Dated: January 11, 2017 Received: January 12, 2017
Dear Vahid Saadat:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Image /page/1/Picture/8 description: The image shows the text "Jennifer R. Stevenson -S" in a large, sans-serif font. The text is black, and the background is white. The letters are well-defined and easy to read. The text appears to be a name, possibly of a person or organization.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162608
Device Name ClariFix Device
Indications for Use (Describe)
The ClariFix Device is intended to be used as a cryosurgical tool for the destruction of unwanted tissue during surgical procedures, including in adults with chronic rhinitis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "Arrinex, Inc.". The logo features a stylized letter "A" with a red accent on the left side. The text "rrinex, Inc." is written in a cursive font to the right of the "A".
# 510(k) Summary (K162608)
| I. Submitter Information | |
|----------------------------|----------------------------------------------------------------------|
| Submitter name: | Arrinex, Inc<br>1755 Bayshore Rd, Suite 26<br>Redwood City, CA 94063 |
| Contact person: | Vahid Saadat<br>CEO<br>408.802.2052 |
| Date Prepared: | 13 February 2017 |
| II. Product Classification | |
| Device Name: | ClariFix Device |
| Common Name: | Cryosurgical Unit and accessories |
| CFR Classification: | 21 CRF 878.4350 |
| Device Class: | II |
III. Predicate Device Arrinex, Inc., ClariFix Device (K160669) Predicate: This predicate has not been subject to any recalls.
GEH
### IV. Device Description
Product Code:
The ClariFix Device is a handheld, single patient-use, disposable cryosurgical device used for the destruction of tissue during surgical procedures. The device consists of a Handle attached to a Cannula with a Cryoprobe at the distal end. The ClariFix Device is provided sterile to the user. The user can select from two nitrous oxide Canister (cryogen Canister) sizes.
To perform cryosurgery, the ClariFix Device is removed from the sterile packaging and the desired cryogen Canister is inserted into the Handle. A Canister Cap is then tightened onto the Handle, which pierces the Canister. The cryogen is held in the Handle until the user opens a mechanical valve that allows cryogen to flow down the Cannula into the Cryoprobe.
The Cryoprobe is placed into contact with the target tissue via direct visualization. The Cannula can be rotated to ensure proper positioning. Once the Cryoprobe is in the desired position, cryosurgery can start.
The Cryoprobe is held in contact with the target location during cryosurgery. As cryogen flows into the Cryoprobe, the liquid partially evaporates and the inside of the Cryoprobe cools to < -80°C and a freezing zone forms in the adjacent tissue, destroying the unwanted tissue. Nitrous oxide is fully contained within the Cryoprobe and does not contact the tissue. Cryosurgery is stopped by the user closing the Valve. Once the Cryoprobe has thawed, it can be safely removed. Additional cycles can be initiated per the physician's discretion.
#### V. Indications for Use
The ClariFix Device is intended to be used as a cryosurgical tool for the destruction of unwanted tissue during surgical procedures, including in adults with chronic rhinitis.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "Arrinex, Inc.". The logo features a stylized letter "A" with a red accent on the left side. The text "rrinex, Inc." is written in a cursive font to the right of the "A" symbol.
## VI. Comparison of Intended Use, Indications for Use and Technological Characteristics with the Predicate Device
Cryoablation is the fundamental technological principle for the ClariFix subject device and the predicate, the ClariFix Device (K160669). The intended use for the ClariFix Device, destruction of tissue with extreme cold, remains unchanged for the subject device.
The only modification to the subject device is to update the indications for use to include in adults with chronic rhinitis. Results from a clinical study combined with published literature demonstrated that the reference to adults with chronic rhinitis does not impact safety and effectiveness.
# Comparison of Indications for Use
| ClariFix Device (Subject) | The ClariFix Device is intended to be used as a cryosurgical tool for the destruction of unwanted tissue during surgical procedures, including in adults with chronic rhinitis |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ClariFix Device (Predicate, K160669) | The ClariFix Device is intended to be used as a cryosurgical tool for the destruction of unwanted tissue during surgical procedures |
The subject and the legally marketed predicate device share identical technological elements:
- Hand-held, portable device containing a single use nitrous oxide cryogen cartridge ●
- Application of cryogen to ablate (freeze) unwanted tissue
- Use of a cryo balloon as the mechanism for applying cold treatment and containment and ● exhaust of the cryogen
- . User controls treatment time
- No power source, electronics or software required to use the device.
- Use of direct visualization to the target location ●
As such, the intended use, design, materials and function of the Arrinex ClariFix Device are substantially equivalent to the predicate device.
# VII. Performance Data
Because there have been no technological changes to the subject device, no additional bench testing, preclinical testing, sterilization or biocompatibility testing was performed,
A prospective, multicenter clinical study was performed in the United States to support the modification of the indications for use statement to include in adults with chronic rhinitis. Twenty seven (27) subjects were enrolled and cryosurgery was successfully performed with the ClariFix device bilaterally (n=54 treatments) while the subjects were awake using local anesthesia. Cryosurgery was well-tolerated and subjects reported an average pain rating of 1.19 on the Wong-Baker FACES pain scale (0 = minimum score, 5 = maximum score).
The primary safety endpoint was the frequency of device and procedure-related serious adverse events (SAEs). No device or procedure-related SAEs were reported throughout the 90-day follow-up duration. The primary effectiveness endpoints were subject-reported outcomes of nasal symptoms using reflective Total Nasal Symptom Score (rTNSS) and Visual Analog Scale (VAS). Three subjects did not return for the 90-day follow-up visit. Of the remaining 24 subjects, the average rTNSS improved 56% from 6.2 at baseline to 2.7 at 90 days (0 = minimum score, 12 = maximum score). The average VAS score improved 53% from 7.6 at baseline to 3.7 at 90 days (0= minimum score, 10 = maximum score).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "Arrinex, Inc.". The logo features a stylized letter "A" in red and gray, followed by the company name in a cursive font. The text is in a dark color, possibly black or dark gray.
The secondary safety endpoint was the frequency of device and procedure-related adverse events (AEs). One subject reported moderate nasal bleeding at Day 27 that was controlled with standard nasal packing and cautery. The investigator deemed it to be remotely related to the device as the target treatment location was completely healed during endoscopic examination at Day 7. Other adverse events commonly associated with healing after cryosurgery in the nasal passageways (pain/discomfort, headache, facial pain, bleeding, dry nose and ear blockage) were observed and by day 90, they had either self-resolved or the remaining events rated as mild with a probable cause relating to pre-existing conditions. The secondary performance endpoint was investigator evaluation of the ease of use of the ClariFix device. Investigators rated cryosurgery using the ClariFix device as "easy" to "moderately easy" in 89% (24/27) of the subjects and "moderately difficult" in 11% (3/27) of the subjects.
Subjects were asked to rate their experience using a subject satisfaction questionnaire. A total of 17 of the 27 subjects responded and fifty nine percent (59%) of the subjects reported cryosurgery to be somewhat (53%) or very (6%) uncomfortable, 41% of the patients felt that it was somewhat (29%) or very (12%) comfortable. Additionally, endoscopic evaluation was performed by the investigators at the follow-up visits, and the average nasal congestion score improved 67% from 1.5 at baseline to 0.5 at 90 days (0 = minimum score, 3 = maximum score). On average, the investigators rated bleeding, swelling and crusting of the treatment site as 0 at 90 days (0 = minimum score, 3 = maximum score) and the average Lund-Kennedy assessment of the entire nasal passageway was 0.21 at 90 days (0 = minimum score, 10 = maximum score) indicating the treatment site was well healed.
A review of published literature over the last four decades was performed to determine clinical outcomes after cryosurgery in the nasal passageway. Eleven clinical studies examined use of various cryoprobes to deliver extreme cold to the nasal mucosal surface in order to treat nasal obstruction or symptoms of rhinitis. Cryosurgical ablation of tissue in the nasal passageways was shown to result in a reduction in rhinitis symptoms with rare reports of major adverse or permanent complications. Minor complications were transient in nature and resolved. There is paucity of studies concerning the vidian neurectomy by means of cryosurgery. Nevertheless, clinical study results demonstrated that both the safety profile and effectiveness results achieved by the ClariFix device in adults with chronic rhinitis are by and large comparable to those published in scientific literature.
The clinical study results demonstrated that when the subject device is used in the nasal passageway according to its Instructions for Use, it is a safe and effective tool for use in destroying unwanted tissue in adults with chronic rhinitis.
## VIII. Conclusions
The ClariFix device is substantially equivalent to the predicate device. Clinical study results demonstrated that both the safety profile and effectiveness results achieved by the ClariFix device in adults with chronic rhinitis are comparable to that reported in published literature on cryosurgery in the nasal passageway.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.