K162601 · Shinva Ande Healthcare Apparatus Co., Ltd. · FPA · May 31, 2017 · General Hospital
Device Facts
Record ID
K162601
Device Name
Infusion Set for Single Use
Applicant
Shinva Ande Healthcare Apparatus Co., Ltd.
Product Code
FPA · General Hospital
Decision Date
May 31, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is indicated for the delivery of fluids from a container to a patient's vascular system.
Device Story
Gravity-fed infusion set; delivers fluids from container to patient vascular system. Components include protective cap, spike, air filter, drip chamber, precision filter, flexible tube, injection site, roller clamp, flow regulator, pinch clamp, and luer lock connector. Used in clinical settings for fluid administration. Operated by healthcare professionals. Provides controlled fluid delivery path. Benefits patient by enabling sterile, regulated intravenous therapy.
Clinical Evidence
No clinical data; bench testing only. Compliance demonstrated via ISO 8536-4 (gravity feed), ISO 8536-12 (check valves), ISO 594-1/2 (luer fittings), ISO 10993 (biocompatibility/sterilization), ASTM F88 (seal strength), ASTM F838 (bacterial retention), and internal specifications for flow regulation and filtration.
Technological Characteristics
Gravity-fed infusion set. Components: spike, air/fluid filters, drip chamber, flexible tubing, clamps, injection sites, luer lock. Materials evaluated per ISO 10993 (cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis, pyrogenicity). Sterilization: Ethylene Oxide (SAL 10^-6). Standards: ISO 8536-4, ISO 8536-12, ISO 594-1/2, ASTM F88, ASTM F756, ASTM F838.
Indications for Use
Indicated for delivery of fluids from a container to a patient's vascular system via IV catheter or infusion needle.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, connected by flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 31, 2017
SHINVA ANDE Healthcare Apparatus Co., Ltd. c/o Ms. Diana Hong Mid-Link Consulting Co., Ltd. P.O. Box 120-119 Shanghai 200120 CHINA
Re: K162601
Trade/Device Name: Infusion Set for Single Use Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: April 26, 2017 Received: April 28, 2017
Dear Ms. Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature and the name "Tina Kiang". The signature is a complex, looping design to the left of the name. The name "Tina Kiang" is written in a clear, legible font, positioned to the right of the signature.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Change Control Table, Change History
# Change Control Table
| Version | Document Author | Document Approver | Date Approved |
|---------|---------------------|---------------------|---------------|
| 1.00 | Name, Title, Office | Name, Title, Office | MM/DD/YYYY |
Complete Change Control Table (all versions) retained in SWIFT Docs.
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162601
Device Name Infusion Set for Single Use
Indications for Use (Describe)
The device is indicated for the delivery of fluids from a container to a patient's vascular system.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K162601
- Date of Preparation: 05/25/2017 1.
- 2. Sponsor Identification
## SHINVA ANDE Healthcare Apparatus Co., Ltd.
No.77 Development Zone North Road, Zibo, Shandong, 255086 China
Contact Person: Xiaolei Tian Position: R&D Director Tel: +86-533-3917821 Fax: +86-533- 3918218 Email: tianxiaolei@andemed.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
## Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
- 4. Identification of Proposed Device
Trade Name: Infusion Set for Single Use
Regulatory Information Regulation Name: Intravascular administration set Classification: II Product Code: FPA Regulation Number: 21CFR 880.5440 Review Panel: General Hospital
{5}------------------------------------------------
#### Indications for Use:
The device is indicated for the delivery of fluids from a container to a patient's vascular system.
### Device Description
The proposed devices are indicated for the gravity infusion of fluids from a container to a patient's vascular system through an IV catheter or infusion needle. . There are ten different models, each a different configuration comprised of various components which may include: protective cap, spike, air filter, drip chamber, precision filter, flexible tube, injection site, roller clamp, flow regulator, pinch clamp and luer lock connector. The devices are provided sterile and are single use.
- 5. Identification of Predicate Device
510(k) Number: K121803 Product Name: Intravascular Administration Set 510k Holder: Acta Medical
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to determine that the proposed device met all design specifications to claim Substantially Equivalence (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > ISO 8536-4:2010 AMD 1 2013 Infusion equipment for medical use, Part 4: Infusion sets for single use, gravity feed
- > ISO 8536-12:2007 AMD 1 2012 Infusion equipment for medical use- Part 12: Check valves.
- > ISO 594-1:1986 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment- Part 1: General requirements
- > ISO 594-2:1998 Conical fittings with 6% (Luer) taper for syringes, needles and certain other medical equipment- Part 2: Lock fittings
- > ISO 10993-7:2008 Biological evaluation of medical devices- Part 7: Ethylene oxide sterilization residuals:
- > ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materials;
- > USP38-NF33 <85> Bacterial Endotoxins Test.
- > ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity.
- > ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
- > ISO 10993-11:2006 Biological evaluation of medical devices- Part 11: Tests for systemic toxicity
- > ASTM F 756-13 Standard practice for assessment of hemolytic properties of materials
- A ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process
- A USP 38-NF 33<151> Pyrogen Test
- > Microbial Ingress Testing (Needleless Valve-Y Injection Site)
{6}------------------------------------------------
- > ASTM F838-15-Standard Test for Determining Bacterial Retention of Membrane Filter (0.2 Micron Filter test)
- > Flow Regulator Performance (Internal Specification)
- > 1.2 Micron Filter Test (Internal Specification)
- Clinical Test Conclusion 7.
No clinical study is included in this submission.
- 8. Substantially Equivalent (SE) Comparison
| | Table 1 Comparison of Technology Characteristics | |
|--|--------------------------------------------------|--|
| | | |
| | | |
| Item | Proposed Device | Predicate Device | |
|--------------------|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Product Code | FPA | Same | |
| Regulation Number | 21 CFR 880.5440 | Same | |
| Indication for Use | The device is indicated for the<br>delivery of fluids from a container<br>to a patient's vascular system. | Acta Medical Intravascular administration set<br>intended use is to deliver sterile, infusion fluid<br>from a container to the patient with or without<br>flow control features. Acta Medical infusion<br>tubing may act as an extension of other infusion<br>tubing in delivering intravenous fluids from a<br>container to patient. | |
| | Configuration | Protector Cap of Spike<br>Spike<br>Air Vent<br>Air Filter<br>Drip Chamber<br>Fluid Filter<br>Flexible Tube<br>Check Valve<br>Needle Free Y Injection Site<br>Roller Clamp<br>Y Injection Site<br>Precision Filter<br>Pinch Clamp<br>Luer Lock Connector<br>Protector Cap of Luer Lock<br>Connector | Similar, except for the additional model<br>Minibore extension set |
| | | Sterile | EO sterilized |
| SAL 10-6 | | | Same |
| Single Use | | Single Use | Same |
{7}------------------------------------------------
| Biocompatibility | | |
|---------------------------|-------------------------------------|-------------------------------------|
| Cytotoxicity | Conform with ISO 10993 requirements | Conform with ISO 10993 requirements |
| Intracutaneous Reactivity | | |
| Skin Sensitization | | |
| Acute Systemic Toxicity | | |
| Hemolysis Test | | |
| Pyrogen Test | | |
Device Comparison: The predicate device and subject device are similar in model configurations except the predicate device has an additional model, an extension set. The indications for use for the subject device and predicate device are similar except the predicate device has additional language to include the extension set and some models will come with or without a flow control feature. These minor differences in device design and indications for use do not raise different questions of safety or effectiveness.
#### 9. Substantially Equivalent (SE) Conclusion
Based on the non clinical performance testing, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.